Power Supply · SCPI command reference

Fluke PM 2832 SCPI commands

The Fluke PM 2832 power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 61 commands for it, 34 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the PM 2832 on a real bench.

Commands
61
Queries
34
Subsystems
13
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Fluke PM 2832 and writes the report from the measured results.

Get started

Control the Fluke PM 2832 from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Fluke PM 2832's VISA address

# Paste the address of your Fluke PM 2832 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("OUTP:STAT ON")  # Enable Output
print(inst.query("MEAS:VOLT?"))  # Measure DC Voltage

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Fluke PM 2832 and writes the report from the measured results.

Get started

Run a PM 2832 test without writing code

Describe the test in a sentence, for example “enable output on the PM 2832, then measure dc voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

61 commands in 13 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

LIST subsystem (14)

CommandWhat it doesArguments
LIST:IND <address>Set Recall Memory Address. Select the actual memory address index.Number: memory address 1 to 999
LIST:IND?Query Recall Memory Address. Read the selected memory address index.None
LIST:VOLT <voltage>Store Recall Voltage. Store a voltage in the selected recall memory address.Number: voltage in volts; several values separated by commas for consecutive addresses
LIST:CURR <current>Store Recall Current. Store a current in the selected recall memory address.Number: current in amperes; several values separated by commas for consecutive addresses
LIST:VOLT:POIN?Query Voltage List Length. Read the number of voltage points in the sequence list.None
LIST:CURR:POIN?Query Current List Length. Read the number of current points in the sequence list.None
LIST:SEQ:STAR <address>Set Sequence Start Address. Set the start address of the sequence list.Number: memory address 1 to 999
LIST:SEQ:STAR?Query Sequence Start Address. Read the start address of the sequence list.None
LIST:SEQ:STOP <address>Set Sequence Stop Address. Set the stop address of the sequence list.Number: memory address 1 to 999
LIST:SEQ:STOP?Query Sequence Stop Address. Read the stop address of the sequence list.None
LIST:TIM <seconds>Set Step Interval Time. Set the interval time between automatic steps.Number: interval time in seconds, e.g. 5.0
LIST:TIM?Query Step Interval Time. Read the step interval time.None
LIST:SOUR <source>Set Step Trigger Source. Set the source of the next step in the sequence list.Text: TIM, EXT or BUS
LIST:SOUR?Query Step Trigger Source. Read the source of the next step in the sequence list.None

Current (CURRent) (10)

CommandWhat it doesArguments
CURR <current>Set Current. Program the output current setting Iset.Number: current setpoint in amperes, within the channel rating
CURR?Query Current Setting. Read the programmed current setting Iset.None
CURR:PROT:STAT ONEnable Overcurrent Protection. Enable overcurrent protection.None
CURR:PROT:STAT OFFDisable Overcurrent Protection. Disable overcurrent protection.None
CURR:PROT:STAT?Query Overcurrent Protection State. Read whether overcurrent protection is enabled.None
CURR:PROT:DEL <seconds>Set Overcurrent Protection Delay. Program the overcurrent protection delay time.Number: delay in seconds, within the instrument range
CURR:PROT:DEL?Query Overcurrent Protection Delay. Read the overcurrent protection delay time.None
CURR:PROT:TRIP?Query Overcurrent Tripped. Read whether the overcurrent protection has tripped.None
CURR:LIM:HIGH?Query Current High Limit. Read the output specific maximum current.None
CURR:LIM:LOW?Query Current Low Limit. Read the output specific minimum current.None

Output (OUTPut) (8)

CommandWhat it doesArguments
OUTP:STAT ONEnable Output. Enable the output of the selected channel.None
OUTP:STAT OFFDisable Output. Disable the output of the selected channel.None
OUTP:STAT?Query Output State. Read whether the output is enabled.None
OUTP:PROT:COUP:STAT <state>Set Coupled Protection. Enable or disable coupled protection so one channel's trip shuts down all channels.Text: ON or OFF
OUTP:PROT:COUP:STAT?Query Coupled Protection. Read whether coupled protection is enabled.None
OUTP:PROT:TRIP?Query Output Protection Tripped. Read whether an output protection circuit has tripped.None
OUTP:PROT:CLEClear Protection Trip. Reset an overcurrent or overvoltage trip.None
:OUTP:STAT ONEnable Output Channel. Enables the selected output channel so it can deliver power in OPERATE mode.None

Voltage (VOLTage) (7)

CommandWhat it doesArguments
VOLT <voltage>Set Voltage. Program the output voltage setting Vset.Number: voltage setpoint in volts, within the channel rating
VOLT?Query Voltage Setting. Read the programmed voltage setting Vset.None
VOLT:PROT <voltage>Set Overvoltage Protection Level. Program the overvoltage protection trip level OVPset.Number: OVP trip level in volts
VOLT:PROT?Query Overvoltage Protection Level. Read the overvoltage protection trip level.None
VOLT:PROT:TRIP?Query Overvoltage Tripped. Read whether the overvoltage protection has tripped.None
VOLT:LIM:HIGH?Query Voltage High Limit. Read the output specific maximum voltage.None
VOLT:LIM:LOW?Query Voltage Low Limit. Read the output specific minimum voltage.None

Instrument (INSTrument) (5)

CommandWhat it doesArguments
INST:NSEL <channel>Select Output Channel. Select the output channel that subsequent commands address.Number: output channel number, 1 to the number of channels on the unit
INST:NSEL?Query Selected Output Channel. Read the currently selected output channel number.None
INST:STAT ONSet Operate Mode. Make OPERATE mode active so the selected channel can deliver power.None
INST:STAT OFFSet Standby Mode. Set STANDBY mode so the output channels are disabled.None
INST:STAT?Query Operate State. Read whether OPERATE mode is active.None

Initiate (INITiate) (3)

CommandWhat it doesArguments
INITInitiate Sequence. Start execution of the sequence list from start to stop address.None
INIT:CONT <state>Set Continuous Stepping. Set the sequence list to run once or repeatedly.Text: ON or OFF
INIT:CONT?Query Continuous Stepping. Read whether continuous stepping is enabled.None

Measurement (MEASure) (2)

CommandWhat it doesArguments
MEAS:VOLT?Measure DC Voltage. Read back the actual output voltage.None
MEAS:CURR?Measure DC Current. Read back the actual output current.None

TEST subsystem (2)

CommandWhat it doesArguments
:TEST:SYSTEM?Run System Self-Test. Runs the power supply self-test and returns 0 or the hardware error number.None
:TEST:INSTRUMENT?Run Output Channel Self-Test. Runs the self-test of the currently selected output channel and returns 0 or the hardware error number.None

POW subsystem (1)

CommandWhat it doesArguments
POW:LIM:HIGH?Query Power High Limit. Read the output specific maximum power.None

Function (FUNCtion) (1)

CommandWhat it doesArguments
FUNC:MODE?Query Operating Mode. Read whether the channel is in constant voltage or constant current mode.None

ABORT subsystem (1)

CommandWhat it doesArguments
ABORTAbort Sequence. Stop execution of the sequence list; the last recalled setting stays active.None

System (SYSTem) (1)

CommandWhat it doesArguments
SYST:ERR?Query Last Error. Read the next error from the error queue.None

IEEE 488.2 common commands (6)

CommandWhat it doesArguments
*IDN?Identify. Read the instrument identification string.None
*RSTReset. Reset the instrument to its reset state.None
*CLSClear Status. Clear the status byte and event registers.None
*TRGTrigger Step. Send the bus trigger that performs the next step when the trigger source is BUS.None
*ESR?Query Standard Event Register. Read and clear the standard event status register.None
*TST?Run Power-On Self-Test Result Check. Runs the power-on test and reads its result, where 0 means the test passed.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Fluke's programming documentation for the PM 2832 and checked when the agent runs it. The library marks this instrument as supporting *OPC? for operation-complete synchronisation. It reads errors back with SYSTem:ERRor?. Ranges, timing, and option-dependent commands are defined by Fluke's programming manual: check it before you rely on an edge case. Last updated October 29, 2026.

Frequently asked questions

Does the Fluke PM 2832 support SCPI remote control?

Yes. The Fluke PM 2832 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 61 SCPI commands for it: 34 queries that read values or state, and 27 settings and actions.

How do I control the Fluke PM 2832 from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTP:STAT ON and MEAS:VOLT?.

Which command reads a value from the Fluke PM 2832?

MEAS:VOLT? (measure dc voltage). Read back the actual output voltage. The full list of queries is in the command reference on this page, grouped by subsystem.