The Fluke PM 2832 power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 61 commands for it, 34 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the PM 2832 on a real bench.
Commands
61
Queries
34
Subsystems
13
Models covered
1
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Fluke PM 2832 and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Fluke PM 2832's VISA address
# Paste the address of your Fluke PM 2832 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("OUTP:STAT ON") # Enable Output
print(inst.query("MEAS:VOLT?")) # Measure DC Voltage
inst.close()
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Fluke PM 2832 and writes the report from the measured results.
Describe the test in a sentence, for example “enable output on the PM 2832, then measure dc voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
Query Overcurrent Protection State. Read whether overcurrent protection is enabled.
None
CURR:PROT:DEL <seconds>
Set Overcurrent Protection Delay. Program the overcurrent protection delay time.
Number: delay in seconds, within the instrument range
CURR:PROT:DEL?
Query Overcurrent Protection Delay. Read the overcurrent protection delay time.
None
CURR:PROT:TRIP?
Query Overcurrent Tripped. Read whether the overcurrent protection has tripped.
None
CURR:LIM:HIGH?
Query Current High Limit. Read the output specific maximum current.
None
CURR:LIM:LOW?
Query Current Low Limit. Read the output specific minimum current.
None
Output (OUTPut) (8)
Command
What it does
Arguments
OUTP:STAT ON
Enable Output. Enable the output of the selected channel.
None
OUTP:STAT OFF
Disable Output. Disable the output of the selected channel.
None
OUTP:STAT?
Query Output State. Read whether the output is enabled.
None
OUTP:PROT:COUP:STAT <state>
Set Coupled Protection. Enable or disable coupled protection so one channel's trip shuts down all channels.
Text: ON or OFF
OUTP:PROT:COUP:STAT?
Query Coupled Protection. Read whether coupled protection is enabled.
None
OUTP:PROT:TRIP?
Query Output Protection Tripped. Read whether an output protection circuit has tripped.
None
OUTP:PROT:CLE
Clear Protection Trip. Reset an overcurrent or overvoltage trip.
None
:OUTP:STAT ON
Enable Output Channel. Enables the selected output channel so it can deliver power in OPERATE mode.
None
Voltage (VOLTage) (7)
Command
What it does
Arguments
VOLT <voltage>
Set Voltage. Program the output voltage setting Vset.
Number: voltage setpoint in volts, within the channel rating
VOLT?
Query Voltage Setting. Read the programmed voltage setting Vset.
None
VOLT:PROT <voltage>
Set Overvoltage Protection Level. Program the overvoltage protection trip level OVPset.
Number: OVP trip level in volts
VOLT:PROT?
Query Overvoltage Protection Level. Read the overvoltage protection trip level.
None
VOLT:PROT:TRIP?
Query Overvoltage Tripped. Read whether the overvoltage protection has tripped.
None
VOLT:LIM:HIGH?
Query Voltage High Limit. Read the output specific maximum voltage.
None
VOLT:LIM:LOW?
Query Voltage Low Limit. Read the output specific minimum voltage.
None
Instrument (INSTrument) (5)
Command
What it does
Arguments
INST:NSEL <channel>
Select Output Channel. Select the output channel that subsequent commands address.
Number: output channel number, 1 to the number of channels on the unit
INST:NSEL?
Query Selected Output Channel. Read the currently selected output channel number.
None
INST:STAT ON
Set Operate Mode. Make OPERATE mode active so the selected channel can deliver power.
None
INST:STAT OFF
Set Standby Mode. Set STANDBY mode so the output channels are disabled.
None
INST:STAT?
Query Operate State. Read whether OPERATE mode is active.
None
Initiate (INITiate) (3)
Command
What it does
Arguments
INIT
Initiate Sequence. Start execution of the sequence list from start to stop address.
None
INIT:CONT <state>
Set Continuous Stepping. Set the sequence list to run once or repeatedly.
Text: ON or OFF
INIT:CONT?
Query Continuous Stepping. Read whether continuous stepping is enabled.
None
Measurement (MEASure) (2)
Command
What it does
Arguments
MEAS:VOLT?
Measure DC Voltage. Read back the actual output voltage.
None
MEAS:CURR?
Measure DC Current. Read back the actual output current.
None
TEST subsystem (2)
Command
What it does
Arguments
:TEST:SYSTEM?
Run System Self-Test. Runs the power supply self-test and returns 0 or the hardware error number.
None
:TEST:INSTRUMENT?
Run Output Channel Self-Test. Runs the self-test of the currently selected output channel and returns 0 or the hardware error number.
None
POW subsystem (1)
Command
What it does
Arguments
POW:LIM:HIGH?
Query Power High Limit. Read the output specific maximum power.
None
Function (FUNCtion) (1)
Command
What it does
Arguments
FUNC:MODE?
Query Operating Mode. Read whether the channel is in constant voltage or constant current mode.
None
ABORT subsystem (1)
Command
What it does
Arguments
ABORT
Abort Sequence. Stop execution of the sequence list; the last recalled setting stays active.
None
System (SYSTem) (1)
Command
What it does
Arguments
SYST:ERR?
Query Last Error. Read the next error from the error queue.
None
IEEE 488.2 common commands (6)
Command
What it does
Arguments
*IDN?
Identify. Read the instrument identification string.
None
*RST
Reset. Reset the instrument to its reset state.
None
*CLS
Clear Status. Clear the status byte and event registers.
None
*TRG
Trigger Step. Send the bus trigger that performs the next step when the trigger source is BUS.
None
*ESR?
Query Standard Event Register. Read and clear the standard event status register.
None
*TST?
Run Power-On Self-Test Result Check. Runs the power-on test and reads its result, where 0 means the test passed.
None
Where these commands come from
This reference is generated from TestFlow's instrument library, which is drafted from Fluke's programming documentation for the PM 2832 and checked when the agent runs it. The library marks this instrument as supporting *OPC? for operation-complete synchronisation. It reads errors back with SYSTem:ERRor?. Ranges, timing, and option-dependent commands are defined by Fluke's programming manual: check it before you rely on an edge case. Last updated October 29, 2026.
Frequently asked questions
Does the Fluke PM 2832 support SCPI remote control?
Yes. The Fluke PM 2832 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 61 SCPI commands for it: 34 queries that read values or state, and 27 settings and actions.
How do I control the Fluke PM 2832 from Python?
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTP:STAT ON and MEAS:VOLT?.
Which command reads a value from the Fluke PM 2832?
MEAS:VOLT? (measure dc voltage). Read back the actual output voltage. The full list of queries is in the command reference on this page, grouped by subsystem.
Can I automate the Fluke PM 2832 without writing code?
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Fluke PM 2832 through the TestFlow desktop app, and writes the report from the measured results.
Your next test, automated today
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.