The Fluke 8558A multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 97 commands for it, 29 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 8558A on a real bench.
Commands
97
Queries
29
Subsystems
18
Models covered
1
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Fluke 8558A and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
Describe the test in a sentence, for example “func dcv on the 8558A, then measure dcv and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
Res 4W Autorange On. Enables autoranging for 4-wire resistance.
None
FRESistance:RANGe:AUTO OFF
Res 4W Autorange Off. Disables autoranging for 4-wire resistance.
None
FRESistance:NPLC <NPLC>
Set 4W Res NPLC. Sets the integration time for 4-wire resistance in number of power line cycles.
Number: 0.02,0.2,1,10,100
System (SYSTem) (4)
Command
What it does
Arguments
SYSTem:ERRor?
Read Error. Reads and clears the oldest entry from the error queue (returns 0,"No error" if empty).
None
SYSTem:VERSion?
Read SCPI Version. Returns the SCPI version that the meter complies with.
None
SYSTem:LOCal
Local. Places the meter in local mode and enables the front-panel keys.
None
SYSTem:REMote
Remote. Places the meter in remote mode and disables the front-panel keys.
None
Display (DISPlay) (2)
Command
What it does
Arguments
DISPlay ON
Display On. Turns the front-panel display on.
None
DISPlay OFF
Display Off. Turns the front-panel display off to speed up measurements.
None
Read (READ) (1)
Command
What it does
Arguments
READ?
Read Reading. Starts a new measurement and returns the reading using the current configuration.
None
Fetch (FETCh) (1)
Command
What it does
Arguments
FETCh?
Fetch Reading. Returns readings already stored in memory from the current trigger without starting a new measurement.
None
Initiate (INITiate) (1)
Command
What it does
Arguments
INITiate
Initiate. Changes the trigger state from idle to wait-for-trigger and stores readings in memory.
None
Other subsystem (1)
Command
What it does
Arguments
<Output Variable>=<Equation>
Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)
Number; Text
WAIT subsystem (1)
Command
What it does
Arguments
wait(<ms>)
Wait. Waits for number of milliseconds
Number
SAVE subsystem (1)
Command
What it does
Arguments
save2var <Variable_name>
Save2Variable. Saves last measurement to a variable
Text
IEEE 488.2 common commands (7)
Command
What it does
Arguments
*IDN?
Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).
None
*RST
Reset. Resets the instrument to its default state.
None
*CLS
Clear Status. Clears all event registers and the error queue.
None
*TST?
Self Test. Runs the self-test and returns 0 for pass or 1 for failure.
None
*OPC?
Operation Complete. Returns 1 when all pending operations are complete.
None
*TRG
Trigger Bus. Generates a bus trigger when the trigger source is set to BUS.
None
*WAI
Wait To Continue. Waits until all pending operations are complete before continuing.
None
Where these commands come from
This reference is generated from TestFlow's instrument library, which is drafted from Fluke's programming documentation for the 8558A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Fluke's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.
Frequently asked questions
Does the Fluke 8558A support SCPI remote control?
Yes. The Fluke 8558A accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 97 SCPI commands for it: 29 queries that read values or state, and 68 settings and actions.
How do I control the Fluke 8558A from Python?
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, FUNCtion "VOLTage:DC" and MEASure:VOLTage:DC?.
Which command reads a value from the Fluke 8558A?
MEASure:VOLTage:DC? (measure dcv). Configures DC voltage and returns one measured value in volts. The full list of queries is in the command reference on this page, grouped by subsystem.
Can I automate the Fluke 8558A without writing code?
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Fluke 8558A through the TestFlow desktop app, and writes the report from the measured results.
Your next test, automated today
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.