The GW Instek GDM-9060 and GDM-9061 multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 66 commands for it, 21 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the GDM-9060 on a real bench.
Commands
66
Queries
21
Subsystems
15
Models covered
2
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your GW Instek GDM-9060 or GDM-9061 and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the GW Instek GDM-9060's VISA address
# Paste the address of your GW Instek GDM-9060 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("CONFigure:VOLTage:DC") # Conf DCV
print(inst.query("MEASure:VOLTage:DC?")) # Measure DCV
inst.close()
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your GW Instek GDM-9060 or GDM-9061 and writes the report from the measured results.
Describe the test in a sentence, for example “conf dcv on the GDM-9060, then measure dcv and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
66 commands in 15 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The GDM-9060 and GDM-9061 share this command set in TestFlow's library.
Sense (SENSe) (14)
Command
What it does
Arguments
SENSe:VOLTage:DC:RANGe <Range>
Set DCV Range. Sets the DC voltage measurement range in volts.
Number
SENSe:VOLTage:AC:RANGe <Range>
Set ACV Range. Sets the AC voltage measurement range in volts.
Number
SENSe:CURRent:DC:RANGe <Range>
Set DCI Range. Sets the DC current measurement range in amps.
Number
SENSe:RESistance:RANGe <Range>
Set Resistance Range. Sets the 2-wire resistance measurement range in ohms.
Number
SENSe:VOLTage:DC:RANGe:AUTO ON
DCV Autorange On. Enables autorange for DC voltage.
None
SENSe:VOLTage:DC:RANGe:AUTO OFF
DCV Autorange Off. Disables autorange for DC voltage.
None
SENSe:RESistance:RANGe:AUTO ON
Resistance Autorange On. Enables autorange for resistance.
None
SENSe:RESistance:RANGe:AUTO OFF
Resistance Autorange Off. Disables autorange for resistance.
None
SENSe:VOLTage:DC:NPLCycles <NPLC>
Set DCV NPLC. Sets the DC voltage integration time in power line cycles.
Number: 0.025,0.1,1,10,100
SENSe:VOLTage:DC:ZERO:AUTO ON
Set Autozero On. Enables autozero.
None
SENSe:VOLTage:DC:ZERO:AUTO OFF
Set Autozero Off. Disables autozero.
None
SENSe:VOLTage:DC:NULL:STATe ON
Rel On. Enables the relative (null) operation.
None
SENSe:VOLTage:DC:NULL:STATe OFF
Rel Off. Disables the relative (null) operation.
None
SENSe:FUNCtion2 "<Function>"
Enable Second Display. Enables the secondary measurement display.
Text: VOLT:DC,VOLT:AC,CURR:DC,CURR:AC,FREQ,PER
Configure (CONFigure) (13)
Command
What it does
Arguments
CONFigure:VOLTage:DC
Conf DCV. Configures the meter to measure DC voltage on the primary display.
None
CONFigure:VOLTage:AC
Conf ACV. Configures the meter to measure AC voltage on the primary display.
None
CONFigure:CURRent:DC
Conf DCI. Configures the meter to measure DC current on the primary display.
None
CONFigure:CURRent:AC
Conf ACI. Configures the meter to measure AC current on the primary display.
None
CONFigure:RESistance
Conf 2W Resistance. Configures the meter to measure 2-wire resistance.
None
CONFigure:FRESistance
Conf 4W Resistance. Configures the meter to measure 4-wire resistance.
None
CONFigure:FREQuency
Conf Frequency. Configures the meter to measure frequency.
None
CONFigure:PERiod
Conf Period. Configures the meter to measure period.
None
CONFigure:CAPacitance
Conf Capacitance. Configures the meter to measure capacitance.
None
CONFigure:TEMPerature
Conf Temperature. Configures the meter to measure temperature.
None
CONFigure:CONTinuity
Conf Continuity. Configures the meter for continuity testing.
None
CONFigure:DIODe
Conf Diode. Configures the meter for diode testing.
None
CONFigure2:OFF
Disable Second Display. Disables the secondary measurement display.
None
Measurement (MEASure) (10)
Command
What it does
Arguments
MEASure:VOLTage:DC?
Measure DCV. Configures DC voltage and returns one measured value in volts.
None
MEASure:VOLTage:AC?
Measure ACV. Configures AC voltage and returns one measured value in volts.
None
MEASure:CURRent:DC?
Measure DCI. Configures DC current and returns one measured value in amps.
None
MEASure:CURRent:AC?
Measure ACI. Configures AC current and returns one measured value in amps.
None
MEASure:RESistance?
Measure 2W Resistance. Configures 2-wire resistance and returns one measured value in ohms.
None
MEASure:FRESistance?
Measure 4W Resistance. Configures 4-wire resistance and returns one measured value in ohms.
None
MEASure:FREQuency?
Measure Frequency. Configures frequency and returns one measured value in Hz.
None
MEASure:PERiod?
Measure Period. Configures period and returns one measured value in seconds.
None
MEASure:CAPacitance?
Measure Capacitance. Configures capacitance and returns one measured value in farads.
None
MEASure:TEMPerature?
Measure Temperature. Configures temperature and returns one measured value in degrees.
None
Calculate (CALCulate) (9)
Command
What it does
Arguments
CALCulate:FUNCtion DB ; CALCulate:STATe ON
DB On. Enables the dB math operation.
None
CALCulate:STATe OFF
DB Off. Disables the dB math operation.
None
CALCulate:FUNCtion DBM ; CALCulate:STATe ON
DBm On. Enables the dBm math operation.
None
CALCulate:DBM:REFerence <Reference>
Set dBm Reference. Sets the dBm reference resistance in ohms.
Number
CALCulate:AVERage:STATe ON
Statistics On. Enables the min/max/average statistics operation.
None
CALCulate:AVERage:STATe OFF
Statistics Off. Disables the min/max/average statistics operation.
None
CALCulate:AVERage:MINimum?
Read Min. Returns the minimum of the collected statistics.
None
CALCulate:AVERage:MAXimum?
Read Max. Returns the maximum of the collected statistics.
None
CALCulate:AVERage:AVERage?
Read Average. Returns the average of the collected statistics.
None
Fetch (FETCh) (2)
Command
What it does
Arguments
FETCh?
Fetch Value. Returns the last measurement without re-triggering.
None
FETCh2?
Read Second Value. Returns the secondary-display measured value.
None
Trigger (TRIGger) (2)
Command
What it does
Arguments
TRIGger:SOURce <Source>
Set Trigger Source. Sets the trigger source.
Text: IMMediate,EXTernal,BUS
TRIGger:COUNt <Count>
Set Trigger Count. Sets the number of triggers accepted before returning to idle.
Number
System (SYSTem) (2)
Command
What it does
Arguments
SYSTem:SCPi:MODE <Mode>
Set SCPI Mode. Selects the SCPI command-compatibility mode.
Text: NOR,GDM,COMP
SYSTem:ERRor?
Read Error. Reads the first message from the error queue.
None
Function (FUNCtion) (1)
Command
What it does
Arguments
FUNCtion?
Read Function. Queries the present primary measurement function.
None
Read (READ) (1)
Command
What it does
Arguments
READ?
Read Value. Triggers and returns one new measurement of the active function.
None
SAMP subsystem (1)
Command
What it does
Arguments
SAMPle:COUNt <Count>
Set Sample Count. Sets the number of samples taken per trigger.
Number
Initiate (INITiate) (1)
Command
What it does
Arguments
INITiate
Init Trigger. Places the meter in the wait-for-trigger state.
None
Other subsystem (1)
Command
What it does
Arguments
<Output Variable>=<Equation>
Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)
Number; Text
WAIT subsystem (1)
Command
What it does
Arguments
wait(<ms>)
Wait. Waits for number of milliseconds
Number
SAVE subsystem (1)
Command
What it does
Arguments
save2var <Variable_name>
Save2Variable. Saves last measurement to a variable
Text
IEEE 488.2 common commands (7)
Command
What it does
Arguments
*IDN?
Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).
None
*RST
Reset. Resets the instrument to its default state.
None
*CLS
Clear Status. Clears all event registers and the error queue.
None
*TST?
Self Test. Runs the self-test and returns 0 for pass or 1 for failure.
None
*OPC?
Operation Complete. Returns 1 when all pending operations are complete.
None
*WAI
Wait To Continue. Waits until all pending operations are complete before continuing.
None
*TRG
Trigger Bus. Generates a bus trigger when the instrument is in wait-for-trigger state.
None
Where these commands come from
This reference is generated from TestFlow's instrument library, which is drafted from GW Instek's programming documentation for the GDM-9060 and GDM-9061 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by GW Instek's programming manual: check it before you rely on an edge case. Last updated November 12, 2026.
Frequently asked questions
Does the GW Instek GDM-9060 support SCPI remote control?
Yes. The GW Instek GDM-9060 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 66 SCPI commands for it: 21 queries that read values or state, and 45 settings and actions.
How do I control the GW Instek GDM-9060 from Python?
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, CONFigure:VOLTage:DC and MEASure:VOLTage:DC?.
Which command reads a value from the GW Instek GDM-9060?
MEASure:VOLTage:DC? (measure dcv). Configures DC voltage and returns one measured value in volts. The full list of queries is in the command reference on this page, grouped by subsystem.
Can I automate the GW Instek GDM-9060 without writing code?
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the GW Instek GDM-9060 through the TestFlow desktop app, and writes the report from the measured results.
Your next test, automated today
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.