Function Generator · SCPI command reference

GW Instek AFG-2225 SCPI commands

The GW Instek AFG-2225 function generator is controlled remotely with SCPI commands. TestFlow's instrument library holds 65 commands for it, 10 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the AFG-2225 on a real bench.

Commands
65
Queries
10
Subsystems
8
Models covered
1

Control the GW Instek AFG-2225 from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the GW Instek AFG-2225's VISA address

# Paste the address of your GW Instek AFG-2225 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("OUTPut1 ON")  # Turn On Output
print(inst.query("COUNter:VALue?"))  # Read Counter Frequency

inst.close()

Run a AFG-2225 test without writing code

Describe the test in a sentence, for example “turn on output on the AFG-2225, then read counter frequency and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

65 commands in 8 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Source (SOURce) (46)

CommandWhat it doesArguments
SOURce<Channel>:FUNCtion <Shape>Set Function. Sets the output waveform shape for the selected channel.Number: 1,2; Text: SINusoid,SQUare,RAMP,PULSe,NOISe,USER
SOURce<Channel>:FUNCtion?Read Function. Reads the output waveform shape for the selected channel.Number: 1,2
SOURce<Channel>:FREQuency <Freq>Set Frequency. Sets the output frequency in Hz for the selected channel.Number: 1,2; Number
SOURce<Channel>:FREQuency?Read Frequency. Reads the output frequency for the selected channel.Number: 1,2
SOURce<Channel>:AMPlitude <Amp>Set Amplitude. Sets the output amplitude (Vpp) for the selected channel.Number: 1,2; Number
SOURce<Channel>:AMPlitude?Read Amplitude. Reads the output amplitude for the selected channel.Number: 1,2
SOURce<Channel>:DCOffset <Volt>Set Voltage Offset. Sets the DC offset voltage for the selected channel.Number: 1,2; Number
SOURce<Channel>:DCOffset?Read Voltage Offset. Reads the DC offset voltage for the selected channel.Number: 1,2
SOURce<Channel>:VOLTage:UNIT <Unit>Set Voltage Unit. Sets the amplitude unit for the selected channel.Number: 1,2; Text: VPP,VRMS,DBM
SOURce<Channel>:PHASe <Deg>Set Phase. Sets the output phase in degrees for the selected channel.Number: 1,2; Number
SOURce<Channel>:RAMP:SYMMetry <Pct>Set Ramp Symmetry. Sets the ramp symmetry percentage for the selected channel.Number: 1,2; Number: 0-100
SOURce<Channel>:SQUare:DCYCle <Pct>Set Square Duty Cycle. Sets the square-wave duty cycle for the selected channel.Number: 1,2; Number
SOURce<Channel>:PULSe:WIDTh <Sec>Set Pulse Width. Sets the pulse width in seconds for the selected channel.Number: 1,2; Number
SOURce<Channel>:FUNCtion USERSet Arb Waveform. Selects the arbitrary waveform for the selected channel.Number: 1,2
SOURce<Channel>:AM:STATe ONEnable AM. Enables AM modulation on the selected channel.Number: 1,2
SOURce<Channel>:AM:STATe OFFDisable AM. Disables AM modulation on the selected channel.Number: 1,2
SOURce<Channel>:AM:DEPTh <Pct>Set AM Deviation. Sets the AM depth percentage for the selected channel.Number: 1,2; Number
SOURce<Channel>:AM:SOURce <Src>Set AM Source. Sets the AM modulation source for the selected channel.Number: 1,2; Text: INTernal,EXTernal
SOURce<Channel>:AM:INTernal:FREQuency <Freq>Set AM Int Freq. Sets the internal AM modulating frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:FM:STATe ONEnable FM. Enables FM modulation on the selected channel.Number: 1,2
SOURce<Channel>:FM:STATe OFFDisable FM. Disables FM modulation on the selected channel.Number: 1,2
SOURce<Channel>:FM:DEViation <Freq>Set FM Deviation. Sets the FM frequency deviation for the selected channel.Number: 1,2; Number
SOURce<Channel>:FM:SOURce <Src>Set FM Source. Sets the FM modulation source for the selected channel.Number: 1,2; Text: INTernal,EXTernal
SOURce<Channel>:FM:INTernal:FREQuency <Freq>Set FM Int Freq. Sets the internal FM modulating frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:PM:STATe ONEnable PM. Enables PM modulation on the selected channel.Number: 1,2
SOURce<Channel>:PM:STATe OFFDisable PM. Disables PM modulation on the selected channel.Number: 1,2
SOURce<Channel>:PM:DEViation <Deg>Set PM Deviation. Sets the PM phase deviation for the selected channel.Number: 1,2; Number
SOURce<Channel>:PM:SOURce <Src>Set PM Source. Sets the PM modulation source for the selected channel.Number: 1,2; Text: INTernal,EXTernal
SOURce<Channel>:PM:INTernal:FREQuency <Freq>Set PM Int Freq. Sets the internal PM modulating frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:FSKey:STATe ONEnable FSK. Enables FSK modulation on the selected channel.Number: 1,2
SOURce<Channel>:FSKey:STATe OFFDisable FSK. Disables FSK modulation on the selected channel.Number: 1,2
SOURce<Channel>:FSKey:FREQuency <Freq>Set FSK Hop Freq. Sets the FSK hop frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:SWEep:STATe ONEnable Sweep. Enables the frequency sweep on the selected channel.Number: 1,2
SOURce<Channel>:SWEep:STATe OFFDisable Sweep. Disables the frequency sweep on the selected channel.Number: 1,2
SOURce<Channel>:SWEep:TIME <Sec>Set Sweep Time. Sets the sweep time for the selected channel.Number: 1,2; Number
SOURce<Channel>:SWEep:SPACing <Type>Set Sweep Spacing. Sets the sweep spacing for the selected channel.Number: 1,2; Text: LINear,LOGarithmic
SOURce<Channel>:FREQuency:STARt <Freq>Set Sweep Start Freq. Sets the sweep start frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:FREQuency:STOP <Freq>Set Sweep Stop Freq. Sets the sweep stop frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:BURSt:STATe ONEnable Burst. Enables burst mode on the selected channel.Number: 1,2
SOURce<Channel>:BURSt:STATe OFFDisable Burst. Disables burst mode on the selected channel.Number: 1,2
SOURce<Channel>:BURSt:MODE <Mode>Set Burst Mode. Sets the burst mode for the selected channel.Number: 1,2; Text: TRIGgered,GATed
SOURce<Channel>:BURSt:NCYCles <N>Set Burst Cycles. Sets the number of cycles per burst for the selected channel.Number: 1,2; Number
SOURce<Channel>:BURSt:INTernal:PERiod <Sec>Set Burst Period. Sets the internal burst period for the selected channel.Number: 1,2; Number
SOURce<Channel>:BURSt:PHASe <Deg>Set Burst Phase. Sets the burst start phase for the selected channel.Number: 1,2; Number
SOURce<Channel>:BURSt:TRIGger:SOURce <Src>Set Trigger Source. Sets the trigger source.Number: 1,2; Text: IMMediate,EXTernal,MANual
SOURce<Channel>:BURSt:TRIGger:MANualTrigger Now. Issues an immediate trigger.Number: 1,2

Output (OUTPut) (4)

CommandWhat it doesArguments
OUTPut<Channel> ONTurn On Output. Enables the selected channel output.Number: 1,2
OUTPut<Channel> OFFTurn Off Output. Disables the selected channel output.Number: 1,2
OUTPut<Channel>?Read Output State. Reads the selected channel output state.Number: 1,2
OUTPut<Channel>:LOAD <Load>Set Output Load. Sets the selected channel output load in ohms; use INFinity for high-Z.Number: 1,2; Text: DEFault,INFinity

COUN subsystem (3)

CommandWhat it doesArguments
COUNter:STATe ONEnable Counter. Enables the frequency counter.None
COUNter:STATe OFFDisable Counter. Disables the frequency counter.None
COUNter:VALue?Read Counter Frequency. Reads the measured frequency.None

System (SYSTem) (1)

CommandWhat it doesArguments
SYSTem:ERRor?Read Error. Reads and clears the next error-queue entry.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (8)

CommandWhat it doesArguments
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 for pass.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete.None
*SAV <Slot>Save State Slot. Stores the current state to a memory slot.Number
*RCL <Slot>Recall State Slot. Recalls a state from a memory slot.Number

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from GW Instek's programming documentation for the AFG-2225 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by GW Instek's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the GW Instek AFG-2225 support SCPI remote control?

Yes. The GW Instek AFG-2225 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 65 SCPI commands for it: 10 queries that read values or state, and 55 settings and actions.

How do I control the GW Instek AFG-2225 from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTPut1 ON and COUNter:VALue?.

Which command reads a value from the GW Instek AFG-2225?

COUNter:VALue? (read counter frequency). Reads the measured frequency. The full list of queries is in the command reference on this page, grouped by subsystem.