The Elektro-Automatik BT 20010-400 Triple dynamic load is controlled remotely with SCPI commands. TestFlow's instrument library holds 96 commands for it, 29 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the BT 20010-400 Triple on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Elektro-Automatik BT 20010-400 Triple and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Elektro-Automatik BT 20010-400 Triple's VISA address
# Paste the address of your Elektro-Automatik BT 20010-400 Triple (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("OUTPut ON,(@1)") # Turn On
print(inst.query("MEASure:VOLTage? (@1)")) # Measure Voltage
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Elektro-Automatik BT 20010-400 Triple and writes the report from the measured results.
Describe the test in a sentence, for example “turn on on the BT 20010-400 Triple, then measure voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
96 commands in 15 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| BATTery:MODe <Mode>,(@<Channel>) | Set Battery Mode. Selects the battery test mode for the selected channel. IDLE leaves the battery test function. | Text: IDLE,STATic,CHARge,DYNamic,COMBined; Number: 1,2,3 |
| BATTery:CURRent <Current>,(@<Channel>) | Set Battery Discharge Current. Sets the constant discharge current for the static discharge battery test on the selected channel. | Number; Number: 1,2,3 |
| BATTery:POWer <Power>,(@<Channel>) | Set Battery Discharge Power. Sets the maximum discharge power for static and dynamic discharge modes on the selected channel. | Number; Number: 1,2,3 |
| BATTery:RESistance <Resistance>,(@<Channel>) | Set Battery Discharge Resistance. Sets the discharge resistance in Ohm for static/dynamic discharge on the selected channel (0 disables resistance mode). | Number; Number: 1,2,3 |
| BATTery:DISCharge:VOLTage <Voltage>,(@<Channel>) | Set Battery Discharge Stop Voltage. Sets the discharge end voltage of the selected channel. The test stops when the battery voltage reaches this level. | Number; Number: 1,2,3 |
| BATTery:DISCharge:CAPacity <Capacity>,(@<Channel>) | Set Battery Discharge Stop Capacity. Sets the maximum capacity in Ah to discharge on the selected channel before the capacity action is triggered. | Number; Number: 1,2,3 |
| BATTery:DISCharge:TIME <Time>,(@<Channel>) | Set Battery Discharge Stop Time. Sets the maximum discharge test time of the selected channel in HH:MM:SS format (up to 10:00:00). | Text; Number: 1,2,3 |
| BATTery:ACTion:CAPacity <Action>,(@<Channel>) | Set Battery Capacity Action. Defines the action when the capacity limit is reached on the selected channel (NONE ignore SIGNal continue END stop). | Text: NONE,SIGNal,END; Number: 1,2,3 |
| BATTery:ACTion:TIME <Action>,(@<Channel>) | Set Battery Time Action. Defines the action when the maximum test time is reached on the selected channel (NONE ignore SIGNal continue END stop). | Text: NONE,SIGNal,END; Number: 1,2,3 |
| BATTery:CHARge:VOLTage <Voltage>,(@<Channel>) | Set Battery Charge Voltage. Sets the charging voltage for the static charge battery test on the selected channel. | Number; Number: 1,2,3 |
| BATTery:CHARge:CURRent <Current>,(@<Channel>) | Set Battery Charge Current. Sets the charging current for the static charge battery test on the selected channel. | Number; Number: 1,2,3 |
| BATTery:CHARge:STOP:CURRent <Current>,(@<Channel>) | Set Battery Charge Stop Current. Sets the charge end current on the selected channel. The test stops once the charge current falls to this threshold. | Number; Number: 1,2,3 |
| BATTery:CHARge:STOP:CAPacity <Capacity>,(@<Channel>) | Set Battery Charge Stop Capacity. Sets the maximum capacity in Ah to charge on the selected channel before the capacity action is triggered. | Number; Number: 1,2,3 |
| BATTery:CHARge:STOP:TIME <Time>,(@<Channel>) | Set Battery Charge Stop Time. Sets the maximum charge test time of the selected channel in HH:MM:SS format (up to 10:00:00). | Text; Number: 1,2,3 |
| BATTery:INDex <Index>,(@<Channel>) | Set Battery Pulse Index. Selects the dynamic discharge pulse parameter to edit on the selected channel (0 level1 1 level2 2 time1 3 time2). | Number: 0,1,2,3; Number: 1,2,3 |
| BATTery:PULSe <Value>,(@<Channel>) | Set Battery Pulse Value. Sets the value for the pulse parameter selected with Set_Battery_Pulse_Index on the selected channel (amplitude in A or time in s). | Number; Number: 1,2,3 |
| BATTery:COMBined:CURRent <Current>,(@<Channel>) | Set Battery Combined Current. Sets the discharge current for the discharge phase of the combined dynamic test on the selected channel. | Number; Number: 1,2,3 |
| BATTery:COMBined:STOP:CAPacity <Capacity>,(@<Channel>) | Set Battery Combined Stop Capacity. Sets the global maximum capacity in Ah for the combined dynamic test on the selected channel (both phases). | Number; Number: 1,2,3 |
| BATTery:COMBined:STOP:TIME <Time>,(@<Channel>) | Set Battery Combined Stop Time. Sets the global maximum time in HH:MM:SS for the combined dynamic test on the selected channel (both phases). | Text; Number: 1,2,3 |
| BATTery:COMBined:TIME <Time>,(@<Channel>) | Set Battery Combined Rest Time. Sets the rest time in seconds between the charge and discharge phases of the combined test on the selected channel. | Number; Number: 1,2,3 |
| BATTery:COMBined:STARt <Phase>,(@<Channel>) | Set Battery Combined Start Phase. Defines which phase the combined dynamic test starts with on the selected channel. | Text: CHARge,DISCharge; Number: 1,2,3 |
| BATTery:COMBined:CYCLes <Cycles>,(@<Channel>) | Set Battery Combined Cycles. Sets how many times the combined dynamic test runs through on the selected channel (0 infinite or 1 to 999). | Number; Number: 1,2,3 |
| BATTery:STATe RUN,(@<Channel>) | Battery Test Start. Starts the configured battery test on the selected channel. | Number: 1,2,3 |
| BATTery:STATe STOP,(@<Channel>) | Battery Test Stop. Stops the running battery test on the selected channel immediately. | Number: 1,2,3 |
| BATTery:CONDition? (@<Channel>) | Read Battery Condition. Queries the battery test condition of the selected channel (IDLE RUN FINISHED ERROR or phase/limit states). | Number: 1,2,3 |
| BATTery:TEST? (@<Channel>) | Read Battery Test Result. Returns the battery test results of the selected channel (capacity in Ah energy in Wh and elapsed time). | Number: 1,2,3 |
| BATTery:TEST RESet,(@<Channel>) | Reset Battery Test Result. Resets all battery test result values (capacity energy and time) of the selected channel to zero. | Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| SYSTem:LOCK ON | Remote On. Activates remote control of the device. Remote control is global for all three channels. Must be sent before any setting or status changing command. | None |
| SYSTem:LOCK OFF | Remote Off. Deactivates remote control and returns the device to manual operation (global for all channels). | None |
| SYSTem:LOCK:OWNer? | Read Remote Owner. Queries the remote control owner (returns REMOTE NONE or LOCAL). Global for all channels. | None |
| SYSTem:CONFig:MODe <Mode>,(@<Channel>) | Set Regulation Mode. Selects the DC power stage regulation mode of the selected channel between U/I/P and U/I/R (resistance) mode. | Text: UIP,UIR; Number: 1,2,3 |
| SYSTem:CONFig:OUTPut:RESTore <State>,(@<Channel>) | Set DC State After Power On. Defines the DC terminal state of the selected channel after the device is powered on. | Text: AUTO,OFF; Number: 1,2,3 |
| SYSTem:CONFig:CONTroller:SPEed <Speed>,(@<Channel>) | Set Controller Speed. Sets the voltage regulator/controller speed of the selected channel. | Text: FAST,SLOW,NORMal; Number: 1,2,3 |
| SYSTem:ALARm:ACTion:PFAil <Action>,(@<Channel>) | Set PF Alarm Action. Defines the DC terminal state of the selected channel after a power fail (PF) alarm has cleared. | Text: AUTO,OFF; Number: 1,2,3 |
| SYSTem:ALARm:ACTion:OTEMperature <Action>,(@<Channel>) | Set OT Alarm Action. Defines the DC terminal state of the selected channel after an overtemperature (OT) alarm has cleared. | Text: AUTO,OFF; Number: 1,2,3 |
| SYSTem:NOMinal:VOLTage? (@<Channel>) | Read Nominal Voltage. Queries the rated (nominal) voltage of the selected channel. | Number: 1,2,3 |
| SYSTem:NOMinal:CURRent? (@<Channel>) | Read Nominal Current. Queries the rated (nominal) current of the selected channel. | Number: 1,2,3 |
| SYSTem:NOMinal:POWer? (@<Channel>) | Read Nominal Power. Queries the rated (nominal) power of the selected channel. | Number: 1,2,3 |
| SYSTem:MA:ENABle <State> | Set MA Enable. Enables or disables master-auxiliary (MA) operation mode (system level). | Text: ON,OFF |
| SYSTem:MA:LINK <Role> | Set MA Link. Defines the role of the device in the master-auxiliary system. | Text: MASTER,AUX |
| SYSTem:MA:INITialisation | MA Initialize. Starts the master-auxiliary system initialization with the configured settings. | None |
| SYSTem:MA:CONDition? | Read MA Condition. Queries the result of the master-auxiliary initialization (INIT or NO INIT). | None |
| SYSTem:MA:UNITs? | Read MA Units. Queries the number of units successfully initialized in the master-auxiliary system. | None |
| SYSTem:ERRor? | Read Error. Reads the next error from the SCPI error queue and acknowledges device alarms. The return includes a channel marker. | None |
| SYSTem:ERRor:ALL? | Read All Errors. Reads all errors (up to 5) from the SCPI error queue in one string. Each return includes a channel marker. | None |
| Command | What it does | Arguments |
|---|---|---|
| SINK:CURRent <Current>,(@<Channel>) | Set Sink Current. Sets the sink mode current (discharge current) of the selected channel. | Number; Number: 1,2,3 |
| SINK:POWer <Power>,(@<Channel>) | Set Sink Power. Sets the sink mode power limit of the selected channel. | Number; Number: 1,2,3 |
| SINK:RESistance <Resistance>,(@<Channel>) | Set Sink Resistance. Sets the sink mode internal resistance value in Ohm for the selected channel. | Number; Number: 1,2,3 |
| SINK:CURRent:PROTection <Current>,(@<Channel>) | Set Sink OCP. Sets the sink mode overcurrent protection (OCP) threshold of the selected channel. | Number; Number: 1,2,3 |
| SINK:POWer:PROTection <Power>,(@<Channel>) | Set Sink OPP. Sets the sink mode overpower protection (OPP) threshold of the selected channel. | Number; Number: 1,2,3 |
| SINK:CURRent:LIMit:LOW <Current>,(@<Channel>) | Set Sink Current Limit Low. Sets the lower adjustment limit for the sink current set value of the selected channel. | Number; Number: 1,2,3 |
| SINK:CURRent:LIMit:HIGH <Current>,(@<Channel>) | Set Sink Current Limit High. Sets the upper adjustment limit for the sink current set value of the selected channel. | Number; Number: 1,2,3 |
| SINK:POWer:LIMit:HIGH <Power>,(@<Channel>) | Set Sink Power Limit High. Sets the upper adjustment limit for the sink power set value of the selected channel. | Number; Number: 1,2,3 |
| SINK:RESistance:LIMit:HIGH <Resistance>,(@<Channel>) | Set Sink Resistance Limit High. Sets the upper adjustment limit for the sink resistance set value of the selected channel. | Number; Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| VOLTage <Voltage>,(@<Channel>) | Set Voltage. Sets the source mode output voltage of the selected channel. | Number; Number: 1,2,3 |
| VOLTage? (@<Channel>) | Read Voltage Setting. Queries the currently set source voltage value of the selected channel. | Number: 1,2,3 |
| VOLTage:PROTection <Voltage>,(@<Channel>) | Set OVP. Sets the overvoltage protection (OVP) threshold of the selected channel (0 to 110 percent of rated voltage). | Number; Number: 1,2,3 |
| VOLTage:PROTection? (@<Channel>) | Read OVP. Queries the overvoltage protection threshold of the selected channel. | Number: 1,2,3 |
| VOLTage:LIMit:LOW <Voltage>,(@<Channel>) | Set Voltage Limit Low. Sets the lower adjustment limit (U-min) for the voltage set value of the selected channel. | Number; Number: 1,2,3 |
| VOLTage:LIMit:HIGH <Voltage>,(@<Channel>) | Set Voltage Limit High. Sets the upper adjustment limit (U-max) for the voltage set value of the selected channel. | Number; Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| CURRent <Current>,(@<Channel>) | Set Current. Sets the source mode output current (charge current) of the selected channel. | Number; Number: 1,2,3 |
| CURRent? (@<Channel>) | Read Current Setting. Queries the currently set source current value of the selected channel. | Number: 1,2,3 |
| CURRent:PROTection <Current>,(@<Channel>) | Set OCP. Sets the source mode overcurrent protection (OCP) threshold of the selected channel. | Number; Number: 1,2,3 |
| CURRent:PROTection? (@<Channel>) | Read OCP. Queries the source mode overcurrent protection threshold of the selected channel. | Number: 1,2,3 |
| CURRent:LIMit:LOW <Current>,(@<Channel>) | Set Current Limit Low. Sets the lower adjustment limit (I-min) for the source current set value of the selected channel. | Number; Number: 1,2,3 |
| CURRent:LIMit:HIGH <Current>,(@<Channel>) | Set Current Limit High. Sets the upper adjustment limit (I-max) for the source current set value of the selected channel. | Number; Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| POWer <Power>,(@<Channel>) | Set Power. Sets the source mode power limit of the selected channel. | Number; Number: 1,2,3 |
| POWer:PROTection <Power>,(@<Channel>) | Set OPP. Sets the source mode overpower protection (OPP) threshold of the selected channel. | Number; Number: 1,2,3 |
| POWer:PROTection? (@<Channel>) | Read OPP. Queries the source mode overpower protection threshold of the selected channel. | Number: 1,2,3 |
| POWer:LIMit:HIGH <Power>,(@<Channel>) | Set Power Limit High. Sets the upper adjustment limit (P-max) for the source power set value of the selected channel. | Number; Number: 1,2,3 |
| POWer:STAGe:AFTer:REMote <State>,(@<Channel>) | Set DC State After Remote. Defines the DC terminal state of the selected channel after leaving remote control. | Text: AUTO,OFF; Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| MEASure:VOLTage? (@<Channel>) | Measure Voltage. Measures and returns the actual voltage on the DC terminal of the selected channel in Volt. | Number: 1,2,3 |
| MEASure:CURRent? (@<Channel>) | Measure Current. Measures and returns the actual current of the selected channel in Ampere (negative value means sink mode). | Number: 1,2,3 |
| MEASure:POWer? (@<Channel>) | Measure Power. Returns the actual power of the selected channel in Watt (negative value means sink mode). | Number: 1,2,3 |
| MEASure:ARRay? (@<Channel>) | Measure Array. Returns the actual voltage current and power of the selected channel in one query separated by commas. | Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| FETCh:AHOur? (@<Channel>) | Read Amp Hours. Queries the ampere hours supplied during source mode on time of the selected channel in Ah. | Number: 1,2,3 |
| FETCh:WHOur? (@<Channel>) | Read Watt Hours. Queries the watt hours supplied during source mode on time of the selected channel in kWh. | Number: 1,2,3 |
| FETCh:SINK:AHOur? (@<Channel>) | Read Sink Amp Hours. Queries the ampere hours consumed during sink mode on time of the selected channel in Ah. | Number: 1,2,3 |
| FETCh:SINK:WHOur? (@<Channel>) | Read Sink Watt Hours. Queries the watt hours consumed during sink mode on time of the selected channel in kWh. | Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| OUTPut ON,(@<Channel>) | Turn On. Switches the DC terminal of the selected channel on. | Number: 1,2,3 |
| OUTPut OFF,(@<Channel>) | Turn Off. Switches the DC terminal of the selected channel off. | Number: 1,2,3 |
| OUTPut? (@<Channel>) | Read Output State. Queries whether the DC terminal of the selected channel is ON or OFF. | Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| RESistance <Resistance>,(@<Channel>) | Set Resistance. Sets the source mode internal resistance value in Ohm for the selected channel (requires U/I/R mode). | Number; Number: 1,2,3 |
| RESistance:LIMit:HIGH <Resistance>,(@<Channel>) | Set Resistance Limit High. Sets the upper adjustment limit (R-max) for the source resistance set value of the selected channel. | Number; Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| STATus:QUEStionable? (@<Channel>) | Read Questionable Status. Reads the Questionable status register of the selected channel (device alarms and conditions as a bit value). | Number: 1,2,3 |
| STATus:OPERation? (@<Channel>) | Read Operation Status. Reads the Operation status register of the selected channel (regulation mode and operation state as a bit value). | Number: 1,2,3 |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Returns the instrument identification string (manufacturer model serial number and firmware). Global command. | None |
| *RST | Reset. Resets the device to a defined state switches all DC terminals off and clears the alarm buffer. Global command. | None |
| *CLS | Clear Status. Clears the error queue the status byte (STB) and the Event Status Register (ESR). Global command. | None |
| *ESE <Value> | Set Event Status Enable. Sets the Event Status Enable register a signal filter for the ESR. Global command. | Number |
| *ESR? | Read Event Status. Reads the Event Status Register related to command transfer and execution. Global command. | None |
| *STB? | Read Status Byte. Reads the Status Byte register. Global command. | None |
| *SRE <Value> | Set Service Request Enable. Sets the Service Request Enable register. Global command. | Number |
This reference is generated from TestFlow's instrument library, which is drafted from Elektro-Automatik's programming documentation for the BT 20010-400 Triple and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Elektro-Automatik's programming manual: check it before you rely on an edge case. Last updated June 11, 2026.
Yes. The Elektro-Automatik BT 20010-400 Triple accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 96 SCPI commands for it: 29 queries that read values or state, and 67 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTPut ON,(@1) and MEASure:VOLTage? (@1).
MEASure:VOLTage? (@1) (measure voltage). Measures and returns the actual voltage on the DC terminal of the selected channel in Volt. The full list of queries is in the command reference on this page, grouped by subsystem.