Oscilloscope · SCPI command reference

Teledyne LeCroy SDA 8080 SCPI commands

The Teledyne LeCroy SDA 8080 oscilloscope is controlled remotely with SCPI commands. TestFlow's instrument library holds 30 commands for it, 9 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the SDA 8080 on a real bench.

Commands
30
Queries
9
Subsystems
15
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Teledyne LeCroy SDA 8080 and writes the report from the measured results.

Get started

Control the Teledyne LeCroy SDA 8080 from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Teledyne LeCroy SDA 8080's VISA address

# Paste the address of your Teledyne LeCroy SDA 8080 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("TRMD AUTO")  # Run Continuous
print(inst.query("TDIV?"))  # Read Timebase Scale

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Teledyne LeCroy SDA 8080 and writes the report from the measured results.

Get started

Run a SDA 8080 test without writing code

Describe the test in a sentence, for example “run continuous on the SDA 8080, then read timebase scale and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

30 commands in 15 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

C subsystem (9)

CommandWhat it doesArguments
C<n>:TRA ONChannel On. Number: 1,2,3,4
C<n>:TRA OFFChannel Off. Number: 1,2,3,4
C<n>:VDIV <VoltDiv>Channel Scale. Number: 1,2,3,4; Number
C<n>:VDIV?Read Channel Scale. Number: 1,2,3,4
C<n>:OFST <Volt>Channel Offset. Number: 1,2,3,4; Number
C<n>:OFST?Read Channel Offset. Number: 1,2,3,4
C<n>:CPL <Mode>Channel Coupling. Number: 1,2,3,4; Text: A1M,D1M,D50,GND
C<n>:WF? DAT1Read Waveform Data. Number: 1,2,3,4
C<n>:WF? DESCRead Waveform Preamble. Number: 1,2,3,4

TRMD subsystem (3)

CommandWhat it doesArguments
TRMD SINGLESingle Acquisition. None
TRMD AUTORun Continuous. None
TRMD STOPStop Acquisition. None

TDIV subsystem (2)

CommandWhat it doesArguments
TDIV <SecDiv>Timebase Scale. Number
TDIV?Read Timebase Scale. None

TRDL subsystem (2)

CommandWhat it doesArguments
TRDL <Sec>Timebase Offset. Number
TRDL?Read Timebase Offset. None

TRLV subsystem (2)

CommandWhat it doesArguments
TRLV <Volt>Trigger Level. Number
TRLV?Read Trigger Level. None

Other subsystem (2)

CommandWhat it doesArguments
<Source>:PAVA? <Type>Read Measurement. Text: C1,C2,C3,C4,F1,F2,F3,F4; Text: AMPL,AREA,BASE,DLY,DUTY,FALL,FREQ,MAX,MEAN,MIN,OVSN,OVSP,PHASE,PKPK,PER,RISE,RMS,SDEV,TOP,WID,WIDN
<Output Variable>=<Equation>Math. Number; Text

ASET subsystem (1)

CommandWhat it doesArguments
ASETAuto Setup. None

TRSE subsystem (1)

CommandWhat it doesArguments
TRSE EDGE,SR,<Source>Trigger Source. Text: C1,C2,C3,C4,EX,LINE

TRSL subsystem (1)

CommandWhat it doesArguments
TRSL <Slope>Trigger Slope. Text: POS,NEG

PACU subsystem (1)

CommandWhat it doesArguments
PACU <n>,<Type>,<Source>Measurement. Number: 1,2,3,4,5,6,7,8; Text: AMPL,AREA,BASE,DLY,DUTY,FALL,FREQ,MAX,MEAN,MIN,OVSN,OVSP,PHASE,PKPK,PER,RISE,RMS,SDEV,TOP,WID,WIDN; Text: C1,C2,C3,C4,F1,F2,F3,F4

PADL subsystem (1)

CommandWhat it doesArguments
PADL <n>Clear Measurement. Number: 1,2,3,4,5,6,7,8

HCSU subsystem (1)

CommandWhat it doesArguments
HCSU DEV,PNG,FORMAT,LANDSCAPE,BCKG,WHITE,DEST,REMOTE,AREA,GRIDAREAONLYCapture. None

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Number

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Select_var>Save2Variable. Text

IEEE 488.2 common commands (2)

CommandWhat it doesArguments
*IDN?Identify. None
*RSTReset. None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Teledyne LeCroy's programming documentation for the SDA 8080 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Teledyne LeCroy's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the Teledyne LeCroy SDA 8080 support SCPI remote control?

Yes. The Teledyne LeCroy SDA 8080 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 30 SCPI commands for it: 9 queries that read values or state, and 21 settings and actions.

How do I control the Teledyne LeCroy SDA 8080 from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, TRMD AUTO and TDIV?.

Which command reads a value from the Teledyne LeCroy SDA 8080?

TDIV? (read timebase scale). The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Teledyne LeCroy SDA 8080 without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Teledyne LeCroy SDA 8080 through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today