Power Supply · SCPI command reference

Sorensen SGA50-200 SCPI commands

The Sorensen SGA50-200 power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 37 commands for it, 14 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the SGA50-200 on a real bench.

Commands
37
Queries
14
Subsystems
8
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Sorensen SGA50-200 and writes the report from the measured results.

Get started

Control the Sorensen SGA50-200 from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Sorensen SGA50-200's VISA address

# Paste the address of your Sorensen SGA50-200 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write(":OUTPut:STATe ON")  # Turn On
print(inst.query(":MEASure:VOLTage?"))  # Measure Voltage

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Sorensen SGA50-200 and writes the report from the measured results.

Get started

Run a SGA50-200 test without writing code

Describe the test in a sentence, for example “turn on on the SGA50-200, then measure voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

37 commands in 8 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Output (OUTPut) (13)

CommandWhat it doesArguments
:OUTPut:STATe ONTurn On. Enables the power supply output.None
:OUTPut:STATe OFFTurn Off. Disables the power supply output.None
:OUTPut:STATe?Read Output State. Reads whether the output is ON or OFF.None
:OUTPut:PROTection:FOLD <Mode>Set Foldback. Sets foldback mode: 0 off, 1 program down on CV, 2 program down on CC.Number: 0,1,2
:OUTPut:PROTection:FOLD?Read Foldback. Reads the foldback (program-down) mode.None
:OUTPut:PROTection:DELay <Sec>Set Protection Delay. Sets the protection reporting delay in seconds.Number
:OUTPut:ISOLation ONEnable Isolation. Closes the rear-panel isolation relay.None
:OUTPut:ISOLation OFFDisable Isolation. Opens the rear-panel isolation relay.None
:OUTPut:POLarity NORMSet Polarity Normal. Sets the output polarity relay to normal.None
:OUTPut:POLarity INVSet Polarity Inverted. Sets the output polarity relay to inverted.None
:OUTPut:SENSe ONEnable Remote Sense. Closes the remote-sense relay (4-wire sensing).None
:OUTPut:SENSe OFFDisable Remote Sense. Opens the remote-sense relay (local sensing).None
:OUTPut:TRIPped?Read Output Tripped. Returns 1 if the output is tripped, 0 otherwise.None

Source (SOURce) (10)

CommandWhat it doesArguments
:SOURce:VOLTage <Volt>Set Voltage. Sets the output voltage in volts.Number
:SOURce:VOLTage?Read Voltage Setting. Reads the programmed output voltage.None
:SOURce:VOLTage:LIMit <Volt>Set Voltage Limit. Sets the soft upper voltage limit in volts.Number
:SOURce:CURRent <Amp>Set Current. Sets the output current in amps.Number
:SOURce:CURRent?Read Current Setting. Reads the programmed output current.None
:SOURce:CURRent:LIMit <Amp>Set Current Limit. Sets the soft upper current limit in amps.Number
:SOURce:VOLTage:PROTection <Volt>Set OVP. Sets the over-voltage protection trip point in volts.Number
:SOURce:VOLTage:PROTection?Read OVP. Reads the over-voltage protection trip point.None
:SOURce:VOLTage:PROTection:TRIPped?Read OVP Tripped. Returns 1 if OVP has tripped, 0 otherwise.None
:SOURce:VOLTage:PROTection:CLEarClear OVP. Clears a latched over-voltage protection trip.None

Measurement (MEASure) (2)

CommandWhat it doesArguments
:MEASure:VOLTage?Measure Voltage. Measures the actual output voltage in volts.None
:MEASure:CURRent?Measure Current. Measures the actual output current in amps.None

System (SYSTem) (2)

CommandWhat it doesArguments
:SYSTem:FAULt?Read Fault. Returns the next fault from the fault queue.None
:SYSTem:ERRor?Read Error. Returns the next message from the error queue.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (7)

CommandWhat it doesArguments
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 for pass or 1 for failure.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None
*TRGTrigger Bus. Generates a bus trigger when the instrument is in wait-for-trigger state.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Sorensen's programming documentation for the SGA50-200 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Sorensen's programming manual: check it before you rely on an edge case. Last updated June 30, 2026.

Frequently asked questions

Does the Sorensen SGA50-200 support SCPI remote control?

Yes. The Sorensen SGA50-200 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 37 SCPI commands for it: 14 queries that read values or state, and 23 settings and actions.

How do I control the Sorensen SGA50-200 from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, :OUTPut:STATe ON and :MEASure:VOLTage?.

Which command reads a value from the Sorensen SGA50-200?

:MEASure:VOLTage? (measure voltage). Measures the actual output voltage in volts. The full list of queries is in the command reference on this page, grouped by subsystem.