Power Supply · SCPI command reference

Rohde & Schwarz NGE103B SCPI commands

The Rohde & Schwarz NGE103B power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 28 commands for it, 9 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the NGE103B on a real bench.

Commands
28
Queries
9
Subsystems
6
Models covered
1

Control the Rohde & Schwarz NGE103B from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Rohde & Schwarz NGE103B's VISA address

# Paste the address of your Rohde & Schwarz NGE103B (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
print(inst.query("SYSTem:ERRor?"))  # Read Error

inst.close()

Run a NGE103B test without writing code

Describe the test in a sentence, for example “configure the NGE103B, then read error and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

28 commands in 6 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Instrument (INSTrument) (16)

CommandWhat it doesArguments
INSTrument:NSELect 1 ; OUTPut:STATe ONTurn On CH1. Enables the output of channel 1.None
INSTrument:NSELect 1 ; OUTPut:STATe OFFTurn Off CH1. Disables the output of channel 1.None
INSTrument:NSELect 2 ; OUTPut:STATe ONTurn On CH2. Enables the output of channel 2.None
INSTrument:NSELect 2 ; OUTPut:STATe OFFTurn Off CH2. Disables the output of channel 2.None
INSTrument:NSELect 3 ; OUTPut:STATe ONTurn On CH3. Enables the output of channel 3.None
INSTrument:NSELect 3 ; OUTPut:STATe OFFTurn Off CH3. Disables the output of channel 3.None
INSTrument:NSELect <Channel> ; OUTPut:STATe?Read Output State. Reads whether the selected channel output is ON or OFF.Number: 1,2,3
INSTrument:NSELect <Channel> ; SOURce:VOLTage <Volt>Set Voltage. Sets the output voltage of the selected channel in volts.Number: 1,2,3; Number
INSTrument:NSELect <Channel> ; SOURce:VOLTage?Read Voltage Setting. Reads the set output voltage of the selected channel.Number: 1,2,3
INSTrument:NSELect <Channel> ; SOURce:CURRent <Amp>Set Current. Sets the output current limit of the selected channel in amps.Number: 1,2,3; Number
INSTrument:NSELect <Channel> ; SOURce:CURRent?Read Current Setting. Reads the set output current limit of the selected channel.Number: 1,2,3
INSTrument:NSELect <Channel> ; SOURce:VOLTage:PROTection:LEVel <Volt>Set OVP. Sets the over-voltage protection level of the selected channel.Number: 1,2,3; Number: 0-32.2
INSTrument:NSELect <Channel> ; SOURce:VOLTage:PROTection:STATe ONEnable OVP. Enables over-voltage protection on the selected channel.Number: 1,2,3
INSTrument:NSELect <Channel> ; SOURce:VOLTage:PROTection:STATe OFFDisable OVP. Disables over-voltage protection on the selected channel.Number: 1,2,3
INSTrument:NSELect <Channel> ; MEASure:VOLTage?Measure Voltage. Measures the output voltage of the selected channel in volts.Number: 1,2,3
INSTrument:NSELect <Channel> ; MEASure:CURRent?Measure Current. Measures the output current of the selected channel in amps.Number: 1,2,3

System (SYSTem) (1)

CommandWhat it doesArguments
SYSTem:ERRor?Read Error. Reads the first message from the error queue.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (8)

CommandWhat it doesArguments
*SAV <Slot>Save State. Saves the current instrument state to the specified storage slot.Number
*RCL <Slot>Recall State. Recalls an instrument state from the specified storage slot.Number
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 for pass or 1 for failure.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Rohde & Schwarz's programming documentation for the NGE103B and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Rohde & Schwarz's programming manual: check it before you rely on an edge case. Last updated November 12, 2026.

Frequently asked questions

Does the Rohde & Schwarz NGE103B support SCPI remote control?

Yes. The Rohde & Schwarz NGE103B accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 28 SCPI commands for it: 9 queries that read values or state, and 19 settings and actions.

How do I control the Rohde & Schwarz NGE103B from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN? and SYSTem:ERRor?.

Which command reads a value from the Rohde & Schwarz NGE103B?

SYSTem:ERRor? (read error). Reads the first message from the error queue. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Rohde & Schwarz NGE103B without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Rohde & Schwarz NGE103B through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.