The Rohde & Schwarz HMP4040 power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 28 commands for it, 9 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the HMP4040 on a real bench.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Rohde & Schwarz HMP4040's VISA address
# Paste the address of your Rohde & Schwarz HMP4040 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
print(inst.query("SYSTem:ERRor?")) # Read Error
inst.close()Describe the test in a sentence, for example “configure the HMP4040, then read error and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
28 commands in 6 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| INSTrument:NSELect 1 ; OUTPut:STATe ON | Turn On CH1. Enables the output of channel 1. | None |
| INSTrument:NSELect 1 ; OUTPut:STATe OFF | Turn Off CH1. Disables the output of channel 1. | None |
| INSTrument:NSELect 2 ; OUTPut:STATe ON | Turn On CH2. Enables the output of channel 2. | None |
| INSTrument:NSELect 2 ; OUTPut:STATe OFF | Turn Off CH2. Disables the output of channel 2. | None |
| INSTrument:NSELect 3 ; OUTPut:STATe ON | Turn On CH3. Enables the output of channel 3. | None |
| INSTrument:NSELect 3 ; OUTPut:STATe OFF | Turn Off CH3. Disables the output of channel 3. | None |
| INSTrument:NSELect 4 ; OUTPut:STATe ON | Turn On CH4. Enables the output of channel 4. | None |
| INSTrument:NSELect 4 ; OUTPut:STATe OFF | Turn Off CH4. Disables the output of channel 4. | None |
| INSTrument:NSELect <Channel> ; OUTPut:STATe? | Read Output State. Reads whether the selected channel output is ON or OFF. | Number: 1,2,3,4 |
| INSTrument:NSELect <Channel> ; SOURce:VOLTage <Volt> | Set Voltage. Sets the output voltage of the selected channel in volts. | Number: 1,2,3,4; Number |
| INSTrument:NSELect <Channel> ; SOURce:VOLTage? | Read Voltage Setting. Reads the set output voltage of the selected channel. | Number: 1,2,3,4 |
| INSTrument:NSELect <Channel> ; SOURce:CURRent <Amp> | Set Current. Sets the output current limit of the selected channel in amps. | Number: 1,2,3,4; Number |
| INSTrument:NSELect <Channel> ; SOURce:CURRent? | Read Current Setting. Reads the set output current limit of the selected channel. | Number: 1,2,3,4 |
| INSTrument:NSELect <Channel> ; SOURce:VOLTage:PROTection:LEVel <Volt> | Set OVP. Sets the over-voltage protection level of the selected channel. | Number: 1,2,3,4; Number: 0.1-32.5 |
| INSTrument:NSELect <Channel> ; MEASure:VOLTage? | Measure Voltage. Measures the output voltage of the selected channel in volts. | Number: 1,2,3,4 |
| INSTrument:NSELect <Channel> ; MEASure:CURRent? | Measure Current. Measures the output current of the selected channel in amps. | Number: 1,2,3,4 |
| Command | What it does | Arguments |
|---|---|---|
| SYSTem:ERRor? | Read Error. Reads the first message from the error queue. | None |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *SAV <Slot> | Save State. Saves the current instrument state to the specified storage slot. | Number |
| *RCL <Slot> | Recall State. Recalls an instrument state from the specified storage slot. | Number |
| *IDN? | Identify. Returns the instrument identification string (manufacturer, model, serial, firmware). | None |
| *RST | Reset. Resets the instrument to its default state. | None |
| *CLS | Clear Status. Clears all event registers and the error queue. | None |
| *TST? | Self Test. Runs the self-test and returns 0 for pass or 1 for failure. | None |
| *OPC? | Operation Complete. Returns 1 when all pending operations are complete. | None |
| *WAI | Wait To Continue. Waits until all pending operations are complete before continuing. | None |
This reference is generated from TestFlow's instrument library, which is drafted from Rohde & Schwarz's programming documentation for the HMP4040 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Rohde & Schwarz's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.
Yes. The Rohde & Schwarz HMP4040 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 28 SCPI commands for it: 9 queries that read values or state, and 19 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN? and SYSTem:ERRor?.
SYSTem:ERRor? (read error). Reads the first message from the error queue. The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Rohde & Schwarz HMP4040 through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.