Keysight N5745A SCPI commands

The Keysight N5745A power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 61 commands for it, 27 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the N5745A on a real bench.

Commands
61
Queries
27
Subsystems
12
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight N5745A and writes the report from the measured results.

Get started

Control the Keysight N5745A from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keysight N5745A's VISA address

# Paste the address of your Keysight N5745A (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("OUTPut:STATe ON")  # Turn On
print(inst.query("MEASure:VOLTage:DC?"))  # Measure Voltage

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight N5745A and writes the report from the measured results.

Get started

Run a N5745A test without writing code

Describe the test in a sentence, for example “turn on on the N5745A, then measure voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

61 commands in 12 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Source (SOURce) (15)

CommandWhat it doesArguments
SOURce:VOLTage <Volt>Set Voltage. Sets the output voltage in volts (0 to 31.5 V; limited by OVP/1.05 and UVL/0.95).Number: 0,31.5
SOURce:VOLTage?Read Voltage Setting. Reads the programmed output voltage in volts.None
SOURce:VOLTage:TRIGgered <Volt>Set Triggered Voltage. Sets the voltage that is transferred to the output when a trigger occurs.Number: 0,31.5
SOURce:VOLTage:TRIGgered?Read Triggered Voltage. Reads the programmed triggered output voltage in volts.None
SOURce:VOLTage:LIMit:LOW <Volt>Set Low Voltage Limit. Sets the under-voltage limit (UVL) below which voltage programming is ignored (0 to 28.5 V).Number: 0,28.5
SOURce:VOLTage:LIMit:LOW?Read Low Voltage Limit. Reads the low-voltage limit setting in volts.None
SOURce:VOLTage:PROTection:LEVel <Volt>Set OVP Level. Sets the over-voltage protection trip level in volts (2 to 36 V; at least 1.05 x the voltage setting).Number: 2,36
SOURce:VOLTage:PROTection:LEVel?Read OVP Level. Reads the over-voltage protection trip level in volts.None
SOURce:CURRent <Amp>Set Current. Sets the output current limit in amps (0 to 25 A).Number: 0,25
SOURce:CURRent?Read Current Setting. Reads the programmed output current limit in amps.None
SOURce:CURRent:TRIGgered <Amp>Set Triggered Current. Sets the current limit that is transferred to the output when a trigger occurs.Number: 0,25
SOURce:CURRent:TRIGgered?Read Triggered Current. Reads the programmed triggered output current in amps.None
SOURce:CURRent:PROTection:STATe ONEnable OCP. Enables over-current protection so the output latches off when it enters constant-current mode.None
SOURce:CURRent:PROTection:STATe OFFDisable OCP. Disables over-current protection.None
SOURce:CURRent:PROTection:STATe?Read OCP State. Reads whether over-current protection is ON (1) or OFF (0).None

Output (OUTPut) (7)

CommandWhat it doesArguments
OUTPut:STATe ONTurn On. Enables the DC output of the power supply.None
OUTPut:STATe OFFTurn Off. Disables the DC output (zero output voltage and zero source current).None
OUTPut:STATe?Read Output State. Reads whether the output is ON (1) or OFF (0).None
OUTPut:PON:STATe RSTSet Power On State Reset. Sets the unit to power on in the reset (*RST) state.None
OUTPut:PON:STATe AUTOSet Power On State Auto. Sets the unit to power on with the settings it had when it was turned off.None
OUTPut:PON:STATe?Read Power On State. Reads the power-on state setting (RST or AUTO).None
OUTPut:PROTection:CLEarClear Protection. Clears latched protection faults (OV/OC/OT/AC fail/inhibit) and restores the output to its previous state.None

Status (STATus) (7)

CommandWhat it doesArguments
STATus:OPERation:CONDition?Read Operation Condition. Reads the operation condition register (CV=256 / CC=1024 / waiting-for-trigger=32).None
STATus:OPERation:EVENt?Read Operation Event. Reads and clears the operation event register.None
STATus:OPERation:ENABle <Value>Set Operation Enable. Sets the operation enable register mask that drives the OPER bit of the status byte.Number: 0,32767
STATus:QUEStionable:CONDition?Read Questionable Condition. Reads the questionable condition register (OV=1 / OC=2 / PF=4 / OT=16 / INH=512 / UNR=1024).None
STATus:QUEStionable:EVENt?Read Questionable Event. Reads and clears the questionable event register.None
STATus:QUEStionable:ENABle <Value>Set Questionable Enable. Sets the questionable enable register mask that drives the QUES bit of the status byte.Number: 0,32767
STATus:PRESetStatus Preset. Presets the operation and questionable enable and transition registers to their power-on values.None

System (SYSTem) (5)

CommandWhat it doesArguments
SYSTem:COMMunicate:RLSTate <State>Set Remote Local State. Sets the remote/local state (RWLock disables the front panel keys).Text: LOCal,REMote,RWLock
SYSTem:COMMunicate:RLSTate?Read Remote Local State. Reads the remote/local state of the instrument.None
SYSTem:COMMunicate:TCPip:CONTrol?Read TCP Control Port. Returns the control socket port number used for LAN socket connections.None
SYSTem:ERRor?Read Error. Reads and removes the next error number and message from the error queue.None
SYSTem:VERSion?Read SCPI Version. Returns the SCPI version number that the instrument complies with.None

Initiate (INITiate) (4)

CommandWhat it doesArguments
INITiate:IMMediate:TRANsientInitiate Trigger. Initiates the trigger system so that the next trigger applies the triggered voltage and current.None
INITiate:CONTinuous:TRANsient ONEnable Continuous Trigger. Enables continuous initiation of the output trigger system.None
INITiate:CONTinuous:TRANsient OFFDisable Continuous Trigger. Disables continuous initiation so the trigger system must be initiated before each trigger.None
INITiate:CONTinuous:TRANsient?Read Continuous Trigger State. Reads whether continuous triggering is ON (1) or OFF (0).None

Trigger (TRIGger) (3)

CommandWhat it doesArguments
TRIGger:SOURce BUSSet Trigger Source Bus. Selects BUS as the trigger source for the output trigger system (the only available source).None
TRIGger:SOURce?Read Trigger Source. Reads the selected trigger source.None
TRIGger:TRANsient:IMMediateTrigger Immediate. Generates an immediate output trigger if the trigger system has been initiated.None

Measurement (MEASure) (2)

CommandWhat it doesArguments
MEASure:VOLTage:DC?Measure Voltage. Takes a new measurement and returns the DC output voltage in volts.None
MEASure:CURRent:DC?Measure Current. Takes a new measurement and returns the DC output current in amps.None

Abort (1)

CommandWhat it doesArguments
ABORtAbort Trigger. Cancels any trigger action in progress and returns the trigger system to the idle state.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (14)

CommandWhat it doesArguments
*STB?Read Status Byte. Reads the status byte register without clearing it.None
*ESE <Value>Set Event Status Enable. Sets the standard event status enable register mask.Number: 0,255
*ESR?Read Event Status Register. Reads and clears the standard event status register.None
*SRE <Value>Set Service Request Enable. Sets the service request enable register mask.Number: 0,255
*SAV <Slot>Save State. Saves the present instrument state to memory location 0 to 15 (lost at power-off).Number: 0,15
*RCL <Slot>Recall State. Recalls an instrument state previously saved in memory location 0 to 15.Number: 0,15
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*OPT?Read Options. Returns the installed options (0 means no options installed).None
*RSTReset. Resets the instrument to its default state (output off, V=0, I=0, OVP=max) and aborts triggers.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns the result (always returns 0 on this model).None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None
*TRGTrigger Bus. Generates a bus trigger when the trigger source is set to BUS.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the N5745A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated October 13, 2026.

Frequently asked questions

Does the Keysight N5745A support SCPI remote control?

Yes. The Keysight N5745A accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 61 SCPI commands for it: 27 queries that read values or state, and 34 settings and actions.

How do I control the Keysight N5745A from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTPut:STATe ON and MEASure:VOLTage:DC?.

Which command reads a value from the Keysight N5745A?

MEASure:VOLTage:DC? (measure voltage). Takes a new measurement and returns the DC output voltage in volts. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Keysight N5745A without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight N5745A through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.