Function Generator · SCPI command reference

Keysight EDU33212A SCPI commands

The Keysight EDU33212A function generator is controlled remotely with SCPI commands. TestFlow's instrument library holds 124 commands for it, 18 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the EDU33212A on a real bench.

Commands
124
Queries
18
Subsystems
11
Models covered
1

Control the Keysight EDU33212A from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keysight EDU33212A's VISA address

# Paste the address of your Keysight EDU33212A (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("OUTPut:SYNC:STATe ON")  # Turn On Sync CH1
print(inst.query("SYSTem:ERRor?"))  # Read Error

inst.close()

Run a EDU33212A test without writing code

Describe the test in a sentence, for example “turn on sync ch1 on the EDU33212A, then read error and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

124 commands in 11 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Source (SOURce) (84)

CommandWhat it doesArguments
SOURce<Channel>:FUNCtion <Shape>Set Function. Sets the output waveform shape for the selected channel.Number: 1,2; Text: SINusoid,SQUare,RAMP,PULSe,ARB,TRIangle,NOISe,PRBS,DC
SOURce<Channel>:FUNCtion?Read Function. Reads the output waveform shape for the selected channel.Number: 1,2
SOURce<Channel>:FREQuency <Freq>Set Frequency. Sets the output frequency in Hz for the selected channel.Number: 1,2; Number
SOURce<Channel>:FREQuency?Read Frequency. Reads the output frequency for the selected channel.Number: 1,2
SOURce<Channel>:FREQuency:MODE CWSet Frequency Mode Fixed. Sets the selected channel to fixed (continuous-wave) frequency mode.Number: 1,2
SOURce<Channel>:FUNCtion:PULSe:PERiod <Sec>Set Period. Sets the pulse period in seconds for the selected channel.Number: 1,2; Number
SOURce<Channel>:VOLTage <Amp>Set Amplitude. Sets the output amplitude (Vpp) for the selected channel.Number: 1,2; Number
SOURce<Channel>:VOLTage?Read Amplitude. Reads the output amplitude for the selected channel.Number: 1,2
SOURce<Channel>:VOLTage:HIGH <Volt>Set Voltage High. Sets the high voltage level for the selected channel.Number: 1,2; Number
SOURce<Channel>:VOLTage:HIGH?Read Voltage High. Reads the high voltage level for the selected channel.Number: 1,2
SOURce<Channel>:VOLTage:LOW <Volt>Set Voltage Low. Sets the low voltage level for the selected channel.Number: 1,2; Number
SOURce<Channel>:VOLTage:LOW?Read Voltage Low. Reads the low voltage level for the selected channel.Number: 1,2
SOURce<Channel>:VOLTage:OFFSet <Volt>Set Voltage Offset. Sets the DC offset voltage for the selected channel.Number: 1,2; Number
SOURce<Channel>:VOLTage:OFFSet?Read Voltage Offset. Reads the DC offset voltage for the selected channel.Number: 1,2
SOURce<Channel>:VOLTage:UNIT <Unit>Set Voltage Unit. Sets the amplitude unit for the selected channel.Number: 1,2; Text: VPP,VRMS,DBM
SOURce<Channel>:PHASe <Deg>Set Phase. Sets the output phase in degrees for the selected channel.Number: 1,2; Number
SOURce<Channel>:PHASe?Read Phase. Reads the output phase for the selected channel.Number: 1,2
SOURce<Channel>:PHASe:SYNChronizeSync Phase. Resets the phase reference between channels.Number: 1,2
SOURce1:APPLy:SINusoid <Freq>,<Amp>,<Offset>Apply Sine CH1. Sets channel 1 to sine with the given frequency, amplitude and offset.Number; Number; Number
SOURce2:APPLy:SINusoid <Freq>,<Amp>,<Offset>Apply Sine CH2. Sets channel 2 to sine with the given frequency, amplitude and offset.Number; Number; Number
SOURce1:APPLy:SQUare <Freq>,<Amp>,<Offset>Apply Square CH1. Sets channel 1 to square with the given frequency, amplitude and offset.Number; Number; Number
SOURce2:APPLy:SQUare <Freq>,<Amp>,<Offset>Apply Square CH2. Sets channel 2 to square with the given frequency, amplitude and offset.Number; Number; Number
SOURce1:APPLy:RAMP <Freq>,<Amp>,<Offset>Apply Ramp CH1. Sets channel 1 to ramp with the given frequency, amplitude and offset.Number; Number; Number
SOURce2:APPLy:RAMP <Freq>,<Amp>,<Offset>Apply Ramp CH2. Sets channel 2 to ramp with the given frequency, amplitude and offset.Number; Number; Number
SOURce1:APPLy:PULSe <Freq>,<Amp>,<Offset>Apply Pulse CH1. Sets channel 1 to pulse with the given frequency, amplitude and offset.Number; Number; Number
SOURce2:APPLy:PULSe <Freq>,<Amp>,<Offset>Apply Pulse CH2. Sets channel 2 to pulse with the given frequency, amplitude and offset.Number; Number; Number
SOURce<Channel>:APPLy:NOISe DEF,<Amp>,<Offset>Apply Noise. Sets the selected channel to noise with the given amplitude and offset.Number: 1,2; Number; Number
SOURce<Channel>:APPLy:DC DEF,DEF,<Offset>Apply DC. Sets the selected channel to DC with the given offset voltage.Number: 1,2; Number
SOURce1:APPLy:ARBitrary <Rate>,<Amp>,<Offset>Apply Arb CH1. Sets channel 1 to the arbitrary waveform with the given sample rate, amplitude and offset.Number; Number; Number
SOURce2:APPLy:ARBitrary <Rate>,<Amp>,<Offset>Apply Arb CH2. Sets channel 2 to the arbitrary waveform with the given sample rate, amplitude and offset.Number; Number; Number
SOURce<Channel>:APPLy?Read Apply. Reads the current function, frequency, amplitude and offset for the selected channel.Number: 1,2
SOURce<Channel>:FUNCtion:RAMP:SYMMetry <Pct>Set Ramp Symmetry. Sets the ramp symmetry percentage (0 to 100) for the selected channel.Number: 1,2; Number
SOURce<Channel>:FUNCtion:SQUare:DCYCle <Pct>Set Square Duty Cycle. Sets the square-wave duty cycle percentage for the selected channel.Number: 1,2; Number
SOURce<Channel>:FUNCtion:PULSe:WIDTh <Sec>Set Pulse Width. Sets the pulse width in seconds for the selected channel.Number: 1,2; Number
SOURce<Channel>:FUNCtion:PULSe:DCYCle <Pct>Set Pulse Duty Cycle. Sets the pulse duty cycle percentage for the selected channel.Number: 1,2; Number
SOURce<Channel>:FUNCtion:PULSe:TRANsition <Sec>Set Pulse Edge Both. Sets both pulse edge transition times for the selected channel.Number: 1,2; Number
SOURce<Channel>:FUNCtion:PULSe:TRANsition:LEADing <Sec>Set Pulse Lead Edge. Sets the pulse leading-edge transition time for the selected channel.Number: 1,2; Number
SOURce<Channel>:FUNCtion:PULSe:TRANsition:TRAiling <Sec>Set Pulse Trail Edge. Sets the pulse trailing-edge transition time for the selected channel.Number: 1,2; Number
SOURce<Channel>:FUNCtion:ARBitrary <Name>Set Arb Waveform. Selects the active arbitrary waveform by name for the selected channel.Number: 1,2; Text
SOURce<Channel>:FUNCtion:ARBitrary:SRATe <Rate>Set Arb Sample Rate. Sets the arbitrary waveform sample rate (Sa/s) for the selected channel.Number: 1,2; Number
SOURce<Channel>:AM:STATe ONEnable AM. Enables AM modulation on the selected channel.Number: 1,2
SOURce<Channel>:AM:STATe OFFDisable AM. Disables AM modulation on the selected channel.Number: 1,2
SOURce<Channel>:AM:DEPTh <Pct>Set AM Deviation. Sets the AM modulation depth percentage for the selected channel.Number: 1,2; Number
SOURce<Channel>:AM:SOURce <Src>Set AM Source. Sets the AM modulation source for the selected channel.Number: 1,2; Text: INTernal,CH1,CH2
SOURce<Channel>:AM:INTernal:FREQuency <Freq>Set AM Int Freq. Sets the internal AM modulating frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:AM:INTernal:FUNCtion <Shape>Set AM Int Function. Sets the internal AM modulating waveform for the selected channel.Number: 1,2; Text: SINusoid,SQUare,RAMP,NRAMp,TRIangle,NOISe
SOURce<Channel>:FM:STATe ONEnable FM. Enables FM modulation on the selected channel.Number: 1,2
SOURce<Channel>:FM:STATe OFFDisable FM. Disables FM modulation on the selected channel.Number: 1,2
SOURce<Channel>:FM:DEViation <Freq>Set FM Deviation. Sets the FM frequency deviation in Hz for the selected channel.Number: 1,2; Number
SOURce<Channel>:FM:SOURce <Src>Set FM Source. Sets the FM modulation source for the selected channel.Number: 1,2; Text: INTernal,CH1,CH2
SOURce<Channel>:FM:INTernal:FREQuency <Freq>Set FM Int Freq. Sets the internal FM modulating frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:FM:INTernal:FUNCtion <Shape>Set FM Int Function. Sets the internal FM modulating waveform for the selected channel.Number: 1,2; Text: SINusoid,SQUare,RAMP,NRAMp,TRIangle,NOISe
SOURce<Channel>:PM:STATe ONEnable PM. Enables PM modulation on the selected channel.Number: 1,2
SOURce<Channel>:PM:STATe OFFDisable PM. Disables PM modulation on the selected channel.Number: 1,2
SOURce<Channel>:PM:DEViation <Deg>Set PM Deviation. Sets the PM phase deviation in degrees for the selected channel.Number: 1,2; Number
SOURce<Channel>:PM:SOURce <Src>Set PM Source. Sets the PM modulation source for the selected channel.Number: 1,2; Text: INTernal,CH1,CH2
SOURce<Channel>:PM:INTernal:FREQuency <Freq>Set PM Int Freq. Sets the internal PM modulating frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:PM:INTernal:FUNCtion <Shape>Set PM Int Function. Sets the internal PM modulating waveform for the selected channel.Number: 1,2; Text: SINusoid,SQUare,RAMP,NRAMp,TRIangle,NOISe
SOURce<Channel>:FSKey:STATe ONEnable FSK. Enables FSK modulation on the selected channel.Number: 1,2
SOURce<Channel>:FSKey:STATe OFFDisable FSK. Disables FSK modulation on the selected channel.Number: 1,2
SOURce<Channel>:FSKey:FREQuency <Freq>Set FSK Hop Freq. Sets the FSK hop frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:FSKey:INTernal:RATE <Rate>Set FSK Rate. Sets the FSK internal keying rate for the selected channel.Number: 1,2; Number
SOURce<Channel>:FSKey:SOURce <Src>Set FSK Source. Sets the FSK modulation source for the selected channel.Number: 1,2; Text: INTernal,EXTernal
SOURce<Channel>:PWM:STATe ONEnable PWM. Enables PWM modulation (pulse carrier) on the selected channel.Number: 1,2
SOURce<Channel>:PWM:STATe OFFDisable PWM. Disables PWM modulation on the selected channel.Number: 1,2
SOURce<Channel>:PWM:DEViation <Sec>Set PWM Deviation. Sets the PWM width deviation for the selected channel.Number: 1,2; Number
SOURce<Channel>:PWM:SOURce <Src>Set PWM Source. Sets the PWM modulation source for the selected channel.Number: 1,2; Text: INTernal,CH1,CH2
SOURce<Channel>:SWEep:STATe ONEnable Sweep. Enables the frequency sweep on the selected channel.Number: 1,2
SOURce<Channel>:SWEep:STATe OFFDisable Sweep. Disables the frequency sweep on the selected channel.Number: 1,2
SOURce<Channel>:SWEep:SPACing <Type>Set Sweep Spacing. Sets the sweep spacing (LINear/LOGarithmic) for the selected channel.Number: 1,2; Text: LINear,LOGarithmic
SOURce<Channel>:SWEep:TIME <Sec>Set Sweep Time. Sets the sweep time in seconds for the selected channel.Number: 1,2; Number
SOURce<Channel>:SWEep:HTIMe <Sec>Set Sweep Hold Time. Sets the sweep hold time for the selected channel.Number: 1,2; Number
SOURce<Channel>:SWEep:RTIMe <Sec>Set Sweep Return Time. Sets the sweep return time for the selected channel.Number: 1,2; Number
SOURce<Channel>:FREQuency:STARt <Freq>Set Sweep Start Freq. Sets the sweep start frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:FREQuency:STOP <Freq>Set Sweep Stop Freq. Sets the sweep stop frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:FREQuency:CENTer <Freq>Set Sweep Center Freq. Sets the sweep center frequency for the selected channel.Number: 1,2; Number
SOURce<Channel>:FREQuency:SPAN <Freq>Set Sweep Span. Sets the sweep frequency span for the selected channel.Number: 1,2; Number
SOURce<Channel>:BURSt:STATe ONEnable Burst. Enables burst mode on the selected channel.Number: 1,2
SOURce<Channel>:BURSt:STATe OFFDisable Burst. Disables burst mode on the selected channel.Number: 1,2
SOURce<Channel>:BURSt:MODE <Mode>Set Burst Mode. Sets the burst mode for the selected channel.Number: 1,2; Text: TRIGgered,GATed
SOURce<Channel>:BURSt:NCYCles <N>Set Burst Cycles. Sets the number of cycles per burst for the selected channel.Number: 1,2; Number
SOURce<Channel>:BURSt:INTernal:PERiod <Sec>Set Burst Period. Sets the internal burst period for the selected channel.Number: 1,2; Number
SOURce<Channel>:BURSt:PHASe <Deg>Set Burst Phase. Sets the burst start phase for the selected channel.Number: 1,2; Number
SOURce<Channel>:BURSt:GATE:POLarity <Pol>Set Burst Gate Polarity. Sets the burst gate polarity for the selected channel.Number: 1,2; Text: NORMal,INVerted

Output (OUTPut) (12)

CommandWhat it doesArguments
OUTPut<Channel>:STATe ONTurn On Output. Enables the selected channel output.Number: 1,2
OUTPut<Channel>:STATe OFFTurn Off Output. Disables the selected channel output.Number: 1,2
OUTPut<Channel>:STATe?Read Output State. Reads whether the selected channel output is on.Number: 1,2
OUTPut<Channel>:LOAD <Ohm>Set Output Load. Sets the expected output load in ohms (1 to 10k) for the selected channel; use INF for high-Z.Number: 1,2; Number
OUTPut<Channel>:LOAD?Read Output Load. Reads the output load setting for the selected channel.Number: 1,2
OUTPut<Channel>:POLarity NORMalSet Output Polarity Normal. Sets the selected channel output polarity to normal.Number: 1,2
OUTPut<Channel>:POLarity INVertedSet Output Polarity Inverted. Sets the selected channel output polarity to inverted.Number: 1,2
OUTPut<Channel>:POLarity?Read Output Polarity. Reads the selected channel output polarity.Number: 1,2
OUTPut:SYNC:STATe ONTurn On Sync CH1. Enables the front-panel Sync output.None
OUTPut:SYNC:STATe OFFTurn Off Sync CH1. Disables the Sync output.None
OUTPut:SYNC:SOURce CH1Set Sync Source CH1. Sets the Sync output source to channel 1.None
OUTPut:SYNC:SOURce CH2Set Sync Source CH2. Sets the Sync output source to channel 2.None

System (SYSTem) (5)

CommandWhat it doesArguments
SYSTem:BEEPer:STATe ONEnable Beeper. Enables the beeper.None
SYSTem:BEEPer:STATe OFFDisable Beeper. Disables the beeper.None
SYSTem:BEEPerBeep Now. Sounds the beeper once.None
SYSTem:ERRor?Read Error. Reads and clears the next error-queue entry.None
SYSTem:VERSion?Read SCPI Version. Returns the SCPI version.None

Trigger (TRIGger) (3)

CommandWhat it doesArguments
TRIGger:SOURce <Src>Set Trigger Source. Sets the trigger source.Text: IMMediate,EXTernal,TIMer,BUS
TRIGger:SLOPe <Slope>Set Trigger Slope. Sets the external trigger slope.Text: POSitive,NEGative
TRIGgerTrigger Now. Issues an immediate trigger.None

Display (DISPlay) (2)

CommandWhat it doesArguments
DISPlay ONEnable Display. Turns the front-panel display on.None
DISPlay OFFDisable Display. Turns the front-panel display off.None

ROSC subsystem (2)

CommandWhat it doesArguments
ROSCillator:SOURce INTernalSet Clock Internal. Uses the internal 10 MHz reference clock.None
ROSCillator:SOURce EXTernalSet Clock External. Uses the external 10 MHz reference clock.None

Mass memory (MMEMory) (2)

CommandWhat it doesArguments
MMEMory:STORe:STATe <File>Save State. Stores the instrument state to a file.Text
MMEMory:LOAD:STATe <File>Load State. Loads an instrument state file.Text

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (11)

CommandWhat it doesArguments
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 for pass.None
*CAL?Self Calibration. Runs internal self-calibration and returns 0 on success.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete.None
*TRGTrigger Bus. Generates a bus trigger.None
*SAV <Slot>Save State Slot. Stores the current state to an internal memory slot (0 to 4).Number
*RCL <Slot>Recall State Slot. Recalls a state from an internal memory slot (0 to 4).Number
*OPT?Read Options. Queries the installed options.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the EDU33212A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the Keysight EDU33212A support SCPI remote control?

Yes. The Keysight EDU33212A accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 124 SCPI commands for it: 18 queries that read values or state, and 106 settings and actions.

How do I control the Keysight EDU33212A from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTPut:SYNC:STATe ON and SYSTem:ERRor?.

Which command reads a value from the Keysight EDU33212A?

SYSTem:ERRor? (read error). Reads and clears the next error-queue entry. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Keysight EDU33212A without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight EDU33212A through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.