The Keysight E36103B and E3634A power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 24 commands for it, 9 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the E36103B on a real bench.

Commands
24
Queries
9
Subsystems
8
Models covered
2

Control the Keysight E36103B from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keysight E36103B's VISA address

# Paste the address of your Keysight E36103B (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("OUTPut ON")  # Turn On CH1
print(inst.query("MEASure:VOLTage?"))  # Measure Voltage

inst.close()

Run a E36103B test without writing code

Describe the test in a sentence, for example “turn on ch1 on the E36103B, then measure voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

24 commands in 8 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The E36103B and E3634A share this command set in TestFlow's library.

Source (SOURce) (7)

CommandWhat it doesArguments
SOURce:VOLTage <Volt>Set Voltage. Sets the output voltage of the selected channel in volts.Number
SOURce:VOLTage?Read Voltage Setting. Reads the set output voltage of the selected channel.None
SOURce:CURRent <Amp>Set Current. Sets the output current limit of the selected channel in amps.Number
SOURce:CURRent?Read Current Setting. Reads the set output current limit of the selected channel.None
SOURce:VOLTage:PROTection:LEVel <Volt>Set OVP. Sets the over-voltage protection level of the selected channel.Number
SOURce:VOLTage:PROTection:STATe ONEnable OVP. Enables over-voltage protection on the selected channel.None
SOURce:VOLTage:PROTection:STATe OFFDisable OVP. Disables over-voltage protection on the selected channel.None

Output (OUTPut) (3)

CommandWhat it doesArguments
OUTPut ONTurn On CH1. Enables the output of channel 1.None
OUTPut OFFTurn Off CH1. Disables the output of channel 1.None
OUTPut?Read Output State. Reads whether the selected channel output is ON or OFF.None

Measurement (MEASure) (2)

CommandWhat it doesArguments
MEASure:VOLTage?Measure Voltage. Measures the output voltage of the selected channel in volts.None
MEASure:CURRent?Measure Current. Measures the output current of the selected channel in amps.None

System (SYSTem) (1)

CommandWhat it doesArguments
SYSTem:ERRor?Read Error. Reads the first message from the error queue.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (8)

CommandWhat it doesArguments
*SAV <Slot>Save State. Saves the current instrument state to the specified storage slot.Number
*RCL <Slot>Recall State. Recalls an instrument state from the specified storage slot.Number
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 for pass or 1 for failure.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the E36103B and E3634A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the Keysight E36103B support SCPI remote control?

Yes. The Keysight E36103B accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 24 SCPI commands for it: 9 queries that read values or state, and 15 settings and actions.

How do I control the Keysight E36103B from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTPut ON and MEASure:VOLTage?.

Which command reads a value from the Keysight E36103B?

MEASure:VOLTage? (measure voltage). Measures the output voltage of the selected channel in volts. The full list of queries is in the command reference on this page, grouped by subsystem.