Multimeter · SCPI command reference

GW Instek GDM-8351 SCPI commands

The GW Instek GDM-8351 multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 55 commands for it, 19 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the GDM-8351 on a real bench.

Commands
55
Queries
19
Subsystems
14
Models covered
1

Control the GW Instek GDM-8351 from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the GW Instek GDM-8351's VISA address

# Paste the address of your GW Instek GDM-8351 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("CONFigure:VOLTage:DC")  # Conf DCV
print(inst.query("MEASure:VOLTage:DC?"))  # Measure DCV

inst.close()

Run a GDM-8351 test without writing code

Describe the test in a sentence, for example “conf dcv on the GDM-8351, then measure dcv and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

55 commands in 14 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Configure (CONFigure) (19)

CommandWhat it doesArguments
CONFigure:VOLTage:DCConf DCV. Configures the meter to measure DC voltage on the primary display.None
CONFigure:VOLTage:ACConf ACV. Configures the meter to measure AC voltage on the primary display.None
CONFigure:CURRent:DCConf DCI. Configures the meter to measure DC current on the primary display.None
CONFigure:CURRent:ACConf ACI. Configures the meter to measure AC current on the primary display.None
CONFigure:RESistanceConf 2W Resistance. Configures the meter to measure 2-wire resistance.None
CONFigure:FRESistanceConf 4W Resistance. Configures the meter to measure 4-wire resistance.None
CONFigure:FREQuencyConf Frequency. Configures the meter to measure frequency.None
CONFigure:PERiodConf Period. Configures the meter to measure period.None
CONFigure:CAPacitanceConf Capacitance. Configures the meter to measure capacitance.None
CONFigure:TEMPerature:TCOupleConf Temperature. Configures the meter to measure temperature.None
CONFigure:CONTinuityConf Continuity. Configures the meter for continuity testing.None
CONFigure:DIODeConf Diode. Configures the meter for diode testing.None
CONFigure:VOLTage:DC <Range>Set DCV Range. Sets the DC voltage measurement range in volts.Number
CONFigure:VOLTage:AC <Range>Set ACV Range. Sets the AC voltage measurement range in volts.Number
CONFigure:CURRent:DC <Range>Set DCI Range. Sets the DC current measurement range in amps.Number
CONFigure:RESistance <Range>Set Resistance Range. Sets the 2-wire resistance measurement range in ohms.Number
CONFigure:AUTO ONDCV Autorange On. Enables autorange for DC voltage.None
CONFigure:AUTO OFFDCV Autorange Off. Disables autorange for DC voltage.None
CONFigure2:OFFDisable Second Display. Disables the secondary measurement display.None

Measurement (MEASure) (10)

CommandWhat it doesArguments
MEASure:VOLTage:DC?Measure DCV. Configures DC voltage and returns one measured value in volts.None
MEASure:VOLTage:AC?Measure ACV. Configures AC voltage and returns one measured value in volts.None
MEASure:CURRent:DC?Measure DCI. Configures DC current and returns one measured value in amps.None
MEASure:CURRent:AC?Measure ACI. Configures AC current and returns one measured value in amps.None
MEASure:RESistance?Measure 2W Resistance. Configures 2-wire resistance and returns one measured value in ohms.None
MEASure:FRESistance?Measure 4W Resistance. Configures 4-wire resistance and returns one measured value in ohms.None
MEASure:FREQuency?Measure Frequency. Configures frequency and returns one measured value in Hz.None
MEASure:PERiod?Measure Period. Configures period and returns one measured value in seconds.None
MEASure:CAPacitance?Measure Capacitance. Configures capacitance and returns one measured value in farads.None
MEASure:TEMPerature:TCOuple?Measure Temperature. Configures temperature and returns one measured value in degrees.None

Calculate (CALCulate) (8)

CommandWhat it doesArguments
CALCulate:FUNCtion REL ; CALCulate:STATe ONRel On. Enables the relative (null) operation.None
CALCulate:STATe OFFRel Off. Disables the relative (null) operation.None
CALCulate:FUNCtion DB ; CALCulate:STATe ONDB On. Enables the dB math operation.None
CALCulate:FUNCtion DBM ; CALCulate:STATe ONDBm On. Enables the dBm math operation.None
CALCulate:DBM:REFerence <Reference>Set dBm Reference. Sets the dBm reference resistance in ohms.Number
CALCulate:FUNCtion MAX ; CALCulate:STATe ONStatistics On. Enables the min/max/average statistics operation.None
CALCulate:MINimum?Read Min. Returns the minimum of the collected statistics.None
CALCulate:MAXimum?Read Max. Returns the maximum of the collected statistics.None

Trigger (TRIGger) (2)

CommandWhat it doesArguments
TRIGger:SOURce <Source>Set Trigger Source. Sets the trigger source.Text: INT,EXT
TRIGger:COUNt <Count>Set Trigger Count. Sets the number of triggers accepted before returning to idle.Number

Function (FUNCtion) (1)

CommandWhat it doesArguments
FUNCtion?Read Function. Queries the present primary measurement function.None

Read (READ) (1)

CommandWhat it doesArguments
READ?Read Value. Triggers and returns one new measurement of the active function.None

SAMP subsystem (1)

CommandWhat it doesArguments
SAMPle:COUNt <Count>Set Sample Count. Sets the number of samples taken per trigger.Number

Sense (SENSe) (1)

CommandWhat it doesArguments
SENSe:FUNCtion2 "<Function>"Enable Second Display. Enables the secondary measurement display.Text: VOLT:DC,VOLT:AC,CURR:DC,CURR:AC,RES,FRES,FREQ,PER

VAL subsystem (1)

CommandWhat it doesArguments
VAL2?Read Second Value. Returns the secondary-display measured value.None

System (SYSTem) (1)

CommandWhat it doesArguments
SYSTem:ERRor?Read Error. Reads the first message from the error queue.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (7)

CommandWhat it doesArguments
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 for pass or 1 for failure.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None
*TRGTrigger Bus. Generates a bus trigger when the instrument is in wait-for-trigger state.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from GW Instek's programming documentation for the GDM-8351 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by GW Instek's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the GW Instek GDM-8351 support SCPI remote control?

Yes. The GW Instek GDM-8351 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 55 SCPI commands for it: 19 queries that read values or state, and 36 settings and actions.

How do I control the GW Instek GDM-8351 from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, CONFigure:VOLTage:DC and MEASure:VOLTage:DC?.

Which command reads a value from the GW Instek GDM-8351?

MEASure:VOLTage:DC? (measure dcv). Configures DC voltage and returns one measured value in volts. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the GW Instek GDM-8351 without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the GW Instek GDM-8351 through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.