The Chroma 63101A dynamic load is controlled remotely with SCPI commands. TestFlow's instrument library holds 31 commands for it, 2 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 63101A on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Chroma 63101A and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Chroma 63101A's VISA address
# Paste the address of your Chroma 63101A (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write(":CONFigure:REMote ON") # Remote control
print(inst.query("MEAS:VOLT?")) # Measure voltage
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Chroma 63101A and writes the report from the measured results.
Describe the test in a sentence, for example “remote control on the 63101A, then measure voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
31 commands in 11 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| CURR:STAT:L1 <Amp> | Static Current low. Sets the static current lower value | Number |
| CURR:STAT:L2 <Amp> | Static Current High. Sets the static current higher value (Always set low and high) | Number |
| CURR:STAT:RISE <Slew> | Current rise slew. Sets the current rising slew rate | Number |
| CURR:STAT:FALL <Slew> | Current fall slew. Sets the current falling slew rate | Number |
| CURR:DYN:L1 <Amp> | DYN Current low. Sets the dynamic current lower value | Number |
| CURR:DYN:L2 <Amp> | DYN Current high. Sets the dynamic current higher value | Number |
| CURR:DYN:T1 <Time> | DYN Low time. Sets the period of low value | Number |
| CURR:DYN:T2 <Time> | DYN High time. Sets the period of high value | Number |
| CURR:DYN:RISE <Slew> | DYN Current rise. Sets the rise time for dynamic mode | Number |
| CURR:DYN:FALL <Slew> | DYN Current fall. Sets the falling time for dynamic mode | Number |
| Command | What it does | Arguments |
|---|---|---|
| VOLT:L1 <Volt>V | Static voltage low. Sets the lower value of CV mode | Number |
| VOLT:L2 <Volt>V | Static voltave high. Sets the higher value of the CV mode | Number |
| VOLTage:CURRent <Amp> | Static voltage current limit. Sets the current limit | Number |
| VOLT: MODE FAST | Voltage fast. Sets voltage fast mode | None |
| VOLT:MODE SLOW | Voltage slow. Sets voltage slow mode | None |
| Command | What it does | Arguments |
|---|---|---|
| RES:L1 <OHM> OHM | Resistance low. Sets the lower value of dynamic resistive load | Number |
| RES:L2 <OHM> OHM | Resistance high. Sets the higher value of dynamic resistive load (Set low and high for constant) | Number |
| RES:RISE <Slew> | Resistance rise. Sets resistive load rise | Number |
| RES:FALL <Slew> | Resistance fall. Sets resistive load fall | Number |
| Command | What it does | Arguments |
|---|---|---|
| :CONFigure:REMote ON | Remote control. Sets control to remote | None |
| CONF:VOLT:LATC ON | Voltage latch ON. Enables voltage latch | None |
| CONF:VOLT:LATC OFF | Voltage latch OFF. Disables voltage latch | None |
| Command | What it does | Arguments |
|---|---|---|
| LOAD 1 | Turn On. Turns the load on | None |
| LOAD 0 | Turn Off. Turns the load off | None |
| Command | What it does | Arguments |
|---|---|---|
| MEAS:VOLT? | Measure voltage. Measures the voltage | None |
| MEAS:CURR? | Measure current. Measures the current | None |
| Command | What it does | Arguments |
|---|---|---|
| MODE <Mode> | Mode. Sets the load mode | Text: CRH,CRL,CCH,CCL,CCDL,CCDH,CV |
| Command | What it does | Arguments |
|---|---|---|
| CHANnel <Channel> | Channel select. Selects the channel | Number: 1,2,3,4,5,6,7,8 |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last mearurment to a variable | Text |
This reference is generated from TestFlow's instrument library, which is drafted from Chroma's programming documentation for the 63101A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Chroma's programming manual: check it before you rely on an edge case. Last updated November 12, 2026.
Yes. The Chroma 63101A accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 31 SCPI commands for it: 2 queries that read values or state, and 29 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, :CONFigure:REMote ON and MEAS:VOLT?.
MEAS:VOLT? (measure voltage). Measures the voltage The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Chroma 63101A through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.