The Chroma 62006P-100-25 power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 55 commands for it, 26 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 62006P-100-25 on a real bench.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Chroma 62006P-100-25's VISA address
# Paste the address of your Chroma 62006P-100-25 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("CONFigure:OUTPut ON") # Turn On
print(inst.query("MEASure:VOLTage?")) # Measure Voltage
inst.close()Describe the test in a sentence, for example “turn on on the 62006P-100-25, then measure voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
55 commands in 11 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| PROGram:SELected <Program Index> | Program Select. Selects which program number to edit/use | Number |
| PROGram:SELected? | Read Program Selected. Reads the currently selected program number | None |
| PROGram:CLEAR | Program Clear. Clears the currently selected program data | None |
| PROGram:SAVE | Program Save. Saves the currently selected program to memory | None |
| PROGram:RUN <Program State> | Set Program Run. Starts or stops program execution | Text: ON,OFF |
| PROGram:RUN? | Read Program Run. Reads whether program execution is running | None |
| PROGram:COUNT <Count> | Set Program Count. Sets number of repeats (often 0 = continuous, if supported) | Number |
| PROGram:COUNT? | Read Program Count. Reads the repeat count setting | None |
| PROGram:LINK <Link> | Set Program Link. Sets program link mode/value (model-dependent) | Number |
| PROGram:LINK? | Read Program Link. Reads program link mode/value | None |
| PROGram:SEQuence:SELected <Step Index> | Step Select. Selects which sequence step to edit within the program | Number |
| PROGram:SEQuence:SELected? | Read Step Selected. Reads the currently selected sequence step index | None |
| PROGram:SEQuence:TYPE <Step Type> | Set Step Type. Sets step execution mode (AUTO/MANUAL/TTL trigger) | Text: AUTO,MANUAL,TRI |
| PROGram:SEQuence:TYPE? | Read Step Type. Reads the step execution mode | None |
| PROGram:SEQuence:VOLTage <Step Voltage> | Set Step Voltage. Sets the step voltage setpoint | Number |
| PROGram:SEQuence:VOLTage? | Read Step Voltage. Reads the step voltage setpoint | None |
| PROGram:SEQuence:CURRent <Step Current> | Set Step Current. Sets the step current setpoint/limit | Number |
| PROGram:SEQuence:CURRent? | Read Step Current. Reads the step current setpoint/limit | None |
| PROGram:SEQuence:VOLTage:SLEW <Step V Slew> | Set Step Voltage Slew. Sets the step voltage slew rate | Number |
| PROGram:SEQuence:VOLTage:SLEW? | Read Step Voltage Slew. Reads the step voltage slew rate | None |
| PROGram:SEQuence:CURRent:SLEW <Step Curr Slew> | Set Step Current Slew. Sets the step current slew rate | Number |
| PROGram:SEQuence:CURRent:SLEW? | Read Step Current Slew. Reads the step current slew rate | None |
| PROGram:SEQuence:TIME <Seconds> | Set Step Time. Sets the step duration (seconds) | Number |
| PROGram:SEQuence:TIME? | Read Step Time. Reads the step duration | None |
| PROGram:SEQuence:TTLport <Port> | Set Step TTLport. Assigns TTL port/line used by the step (model-dependent) | Number |
| PROGram:SEQuence:TTLport? | Read Step TTLport. Reads the TTL port/line assigned to the step | None |
| Command | What it does | Arguments |
|---|---|---|
| SOURce:VOLTage <Voltage> | Set Voltage. Sets output voltage | Number |
| SOURce:VOLTage? | Read Voltage Set. Reads programmed voltage | None |
| SOURce:CURRent <Current> | Set Current. Sets output current | Number |
| SOURce:CURRent? | Read Current Set. Reads programmed current | None |
| SOURce:VOLTage:SLEW <Voltage Slew> | Set Voltage Slew. Sets voltage slew rate | Number |
| SOURce:VOLTage:SLEW? | Read Voltage Slew. Reads voltage slew rate | None |
| SOURce:CURRent:SLEW <Current Slew> | Set Current Slew. Sets current slew rate | Number |
| SOURce:CURRent:SLEW? | Read Current Slew. Reads current slew rate | None |
| Command | What it does | Arguments |
|---|---|---|
| CONFigure:OUTPut ON | Turn On. Turns output ON | None |
| CONFigure:OUTPut OFF | Turn Off. Turns output OFF | None |
| CONFigure:OUTPut? | Read Output State. Reads output ON/OFF state | None |
| CONFigure:REMote ON | Set Remote Mode. Enables remote control | Text: ON,OFF |
| CONFigure:BEEPer <State> | Beeper Control. Enables or disables beeper | Text: ON,OFF |
| CONFigure:TTL <Value> | Set TTL Function. Configures TTL I/O behavior/mapping (model-dependent) | Number |
| CONFigure:TTL? | Read TTL Function. Reads TTL I/O behavior/mapping | None |
| Command | What it does | Arguments |
|---|---|---|
| MEASure:VOLTage? | Measure Voltage. Measures actual output voltage | None |
| MEASure:CURRent? | Measure Current. Measures actual output current | None |
| MEASure:POWer? | Measure Power. Measures output power | None |
| Command | What it does | Arguments |
|---|---|---|
| FETCh:VOLTage? | Fetch Voltage. Fetches last voltage reading | None |
| FETCh:CURRent? | Fetch Current. Fetches last current reading | None |
| FETCh:POWer? | Fetch Power. Fetches last power reading | None |
| Command | What it does | Arguments |
|---|---|---|
| SYSTem:ERRor? | Read Error. Reads system error queue | None |
| Command | What it does | Arguments |
|---|---|---|
| ABORt | Abort Operation. Aborts active program or operation | None |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last mearurment to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Reads instrument identification | None |
| *RST | Reset. Resets the power supply to default state | None |
| *CLS | Clear Status. Clears status and error registers | None |
This reference is generated from TestFlow's instrument library, which is drafted from Chroma's programming documentation for the 62006P-100-25 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Chroma's programming manual: check it before you rely on an edge case. Last updated February 10, 2026.
Yes. The Chroma 62006P-100-25 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 55 SCPI commands for it: 26 queries that read values or state, and 29 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, CONFigure:OUTPut ON and MEASure:VOLTage?.
MEASure:VOLTage? (measure voltage). Measures actual output voltage The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Chroma 62006P-100-25 through the TestFlow desktop app, and writes the report from the measured results.