Multimeter · SCPI command reference

B&K Precision 5493C SCPI commands

The B&K Precision 5493C multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 54 commands for it, 17 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 5493C on a real bench.

Commands
54
Queries
17
Subsystems
12
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your B&K Precision 5493C and writes the report from the measured results.

Get started

Control the B&K Precision 5493C from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the B&K Precision 5493C's VISA address

# Paste the address of your B&K Precision 5493C (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("CONFigure:VOLTage:DC")  # Func DCV
print(inst.query("MEASure:VOLTage:DC?"))  # Measure DCV

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your B&K Precision 5493C and writes the report from the measured results.

Get started

Run a 5493C test without writing code

Describe the test in a sentence, for example “func dcv on the 5493C, then measure dcv and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

54 commands in 12 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Configure (CONFigure) (11)

CommandWhat it doesArguments
CONFigure:VOLTage:DCFunc DCV. Selects DC voltage as the active measurement function.None
CONFigure:VOLTage:ACFunc ACV. Selects AC voltage as the active measurement function.None
CONFigure:CURRent:DCFunc DCI. Selects DC current as the active measurement function.None
CONFigure:CURRent:ACFunc ACI. Selects AC current as the active measurement function.None
CONFigure:RESistanceFunc 2W Resistance. Selects 2-wire resistance as the active measurement function.None
CONFigure:FRESistanceFunc 4W Resistance. Selects 4-wire resistance as the active measurement function.None
CONFigure:FREQuencyFunc Frequency. Selects frequency as the active measurement function.None
CONFigure:PERiodFunc Period. Selects period as the active measurement function.None
CONFigure:CONTinuityFunc Continuity. Selects continuity as the active measurement function.None
CONFigure:DIODeFunc Diode. Selects diode test as the active measurement function.None
CONFigure?Read Function. Queries the present measurement function.None

Sense (SENSe) (10)

CommandWhat it doesArguments
:SENSe:VOLTage:DC:RANGe <Range>Set DCV Range. Sets the DC voltage measurement range in volts.Number
:SENSe:VOLTage:AC:RANGe <Range>Set ACV Range. Sets the AC voltage measurement range in volts.Number
:SENSe:CURRent:DC:RANGe <Range>Set DCI Range. Sets the DC current measurement range in amps.Number
:SENSe:CURRent:AC:RANGe <Range>Set ACI Range. Sets the AC current measurement range in amps.Number
:SENSe:RESistance:RANGe <Range>Set Resistance Range. Sets the 2-wire resistance measurement range in ohms.Number
:SENSe:VOLTage:DC:RANGe:AUTO ONDCV Autorange On. Enables autorange for DC voltage.None
:SENSe:VOLTage:DC:RANGe:AUTO OFFDCV Autorange Off. Disables autorange for DC voltage.None
:SENSe:RESistance:RANGe:AUTO ONResistance Autorange On. Enables autorange for resistance.None
:SENSe:RESistance:RANGe:AUTO OFFResistance Autorange Off. Disables autorange for resistance.None
SENSe:<Function>:NPLC <NPLC>Set Rate. Sets the measurement integration rate (reading speed).Text: VOLTage:DC,CURRent:DC,RESistance,FRESistance,TEMPerature; Number: 0.02,0.2,1,10,100

Measurement (MEASure) (9)

CommandWhat it doesArguments
MEASure:VOLTage:DC?Measure DCV. Configures DC voltage and returns one measured value in volts.None
MEASure:VOLTage:AC?Measure ACV. Configures AC voltage and returns one measured value in volts.None
MEASure:CURRent:DC?Measure DCI. Configures DC current and returns one measured value in amps.None
MEASure:CURRent:AC?Measure ACI. Configures AC current and returns one measured value in amps.None
MEASure:RESistance?Measure 2W Resistance. Configures 2-wire resistance and returns one measured value in ohms.None
MEASure:FRESistance?Measure 4W Resistance. Configures 4-wire resistance and returns one measured value in ohms.None
MEASure:FREQuency?Measure Frequency. Configures frequency and returns one measured value in Hz.None
MEASure:PERiod?Measure Period. Configures period and returns one measured value in seconds.None
MEASure:DIODe?Measure Diode. Configures diode test and returns the forward voltage in volts.None

Calculate (CALCulate) (9)

CommandWhat it doesArguments
CALCulate:SCALe:FUNCtion DB;:CALCulate:SCALe:STATe ONDB On. Enables the dB math operation.None
CALCulate:SCALe:STATe OFFDB Off. Disables the dB math operation.None
CALCulate:SCALe:FUNCtion DBM;:CALCulate:SCALe:STATe ONDBm On. Enables the dBm math operation.None
CALCulate:SCALe:DBM:REFerence <Ohms>Set dBm Reference. Sets the dBm reference resistance in ohms.Number
:CALCulate:AVERage:STATe ONMinMax On. Enables the min/max/average statistics operation.None
:CALCulate:AVERage:STATe OFFMinMax Off. Disables the min/max/average statistics operation.None
:CALCulate:AVERage:MINimum?Read Min. Returns the minimum of the collected statistics.None
:CALCulate:AVERage:MAXimum?Read Max. Returns the maximum of the collected statistics.None
:CALCulate:AVERage:AVERage?Read Average. Returns the average of the collected statistics.None

Other subsystem (4)

CommandWhat it doesArguments
<Function>:NULL:STATe ONRel On. Enables the relative (null) operation.Text: VOLTage:DC,VOLTage:AC,CURRent:DC,CURRent:AC,RESistance,FRESistance,FREQuency,PERiod,TEMPerature,CAPacitance
<Function>:NULL:STATe OFFRel Off. Disables the relative (null) operation.Text: VOLTage:DC,VOLTage:AC,CURRent:DC,CURRent:AC,RESistance,FRESistance,FREQuency,PERiod,TEMPerature,CAPacitance
<Function>:NULL:VALue <Offset>Set Rel Offset. Sets the reference value used by the relative (null) operation.Text: VOLTage:DC,VOLTage:AC,CURRent:DC,CURRent:AC,RESistance,FRESistance,FREQuency,PERiod,TEMPerature,CAPacitance; Number
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

Trigger (TRIGger) (2)

CommandWhat it doesArguments
:TRIGger:SOURce <Source>Set Trigger Source. Sets the trigger source.Text: IMMediate,EXTernal,BUS
:TRIGger:COUNt <Count>Set Trigger Count. Sets the number of triggers accepted before returning to idle.Number

Read (READ) (1)

CommandWhat it doesArguments
READ?Read Value. Triggers and returns one new measurement of the active function.None

Fetch (FETCh) (1)

CommandWhat it doesArguments
FETCh?Fetch Value. Returns the last measurement without re-triggering.None

Initiate (INITiate) (1)

CommandWhat it doesArguments
:INITiateInit Trigger. Places the meter in the wait-for-trigger state.None

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (4)

CommandWhat it doesArguments
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*TRGTrigger Bus. Generates a bus trigger when the instrument is in wait-for-trigger state.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from B&K Precision's programming documentation for the 5493C and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by B&K Precision's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the B&K Precision 5493C support SCPI remote control?

Yes. The B&K Precision 5493C accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 54 SCPI commands for it: 17 queries that read values or state, and 37 settings and actions.

How do I control the B&K Precision 5493C from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, CONFigure:VOLTage:DC and MEASure:VOLTage:DC?.

Which command reads a value from the B&K Precision 5493C?

MEASure:VOLTage:DC? (measure dcv). Configures DC voltage and returns one measured value in volts. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the B&K Precision 5493C without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the B&K Precision 5493C through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today