Aim-TTi PL068-P, PL155-P, PL303-P, and PL601-P SCPI commands
The Aim-TTi PL068-P, PL155-P, PL303-P, and PL601-P power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 49 commands for it, 24 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the PL068-P on a real bench.
Commands
49
Queries
24
Subsystems
26
Models covered
4
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Aim-TTi PL068-P, PL155-P, PL303-P, or PL601-P and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Aim-TTi PL068-P's VISA address
# Paste the address of your Aim-TTi PL068-P (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("V1 volts (e.g. 5.0)") # Set Voltage
print(inst.query("V1O?")) # Measure Voltage
inst.close()
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Aim-TTi PL068-P, PL155-P, PL303-P, or PL601-P and writes the report from the measured results.
Describe the test in a sentence, for example “set voltage on the PL068-P, then measure voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
49 commands in 26 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The PL068-P, PL155-P, PL303-P, and PL601-P share this command set in TestFlow's library.
V subsystem (4)
Command
What it does
Arguments
V1 <Volt>
Set Voltage. Set the output voltage in volts, e.g. V1 5.0 (reference section 1).
Number: volts (e.g. 5.0)
V1V <Volt>
Set Voltage Verify. Set the output voltage in volts and wait until it has settled (reference section 1).
Number: volts (e.g. 5.0)
V1O?
Measure Voltage. Read the actual output voltage; the reply carries its unit, e.g. 5.002V (reference section 2).
None
V1?
Read Voltage Setting. Read the programmed voltage; the reply is V1 then the volts, e.g. V1 5.000 (reference section 2).
None
I subsystem (3)
Command
What it does
Arguments
I1 <Amp>
Set Current. Set the output current limit in amps, e.g. I1 1.0 (reference section 1).
Number: amperes (e.g. 1.0)
I1O?
Measure Current. Read the actual output current; the reply carries its unit, e.g. 0.854A (reference section 2).
None
I1?
Read Current Setting. Read the programmed current limit; the reply is I1 then the amps, e.g. I1 1.000 (reference section 2).
None
OP subsystem (3)
Command
What it does
Arguments
OP1 1
Turn On. Turn the output on (reference section 1).
None
OP1 0
Turn Off. Turn the output off (reference section 1).
None
OP1?
Read Output State. Read whether the output is on: 1 on, 0 off (reference section 2).
None
OVP subsystem (2)
Command
What it does
Arguments
OVP1 <Volt>
Set OVP. Set the over-voltage protection trip point in volts, e.g. OVP1 6.0 (reference section 1).
Number: volts (e.g. 6.0)
OVP1?
Read OVP. Read the over-voltage trip setting; the reply is VP1 then the volts, e.g. VP1 6.00 (reference section 2).
None
OCP subsystem (2)
Command
What it does
Arguments
OCP1 <Amp>
Set OCP. Set the over-current protection trip point in amps, e.g. OCP1 5.2 (reference section 1).
Number: amperes (e.g. 5.2)
OCP1?
Read OCP. Read the over-current trip setting; the reply is CP1 then the amps, e.g. CP1 5.000 (reference section 2).
None
DELTAV subsystem (2)
Command
What it does
Arguments
DELTAV1 <Step>
Set Voltage Step. Set the voltage step size used by Increment_Voltage and Decrement_Voltage, in volts (reference section 3).
Number: volts (e.g. 0.5)
DELTAV1?
Read Voltage Step. Read the voltage step size; the reply is DELTAV1 then the volts, e.g. DELTAV1 0.500 (Aim-TTi manual Iss 17 p.35).
None
DELTAI subsystem (2)
Command
What it does
Arguments
DELTAI1 <Step>
Set Current Step. Set the current step size used by Increment_Current and Decrement_Current, in amps (reference section 3).
Number: amperes (e.g. 0.1)
DELTAI1?
Read Current Step. Read the current step size; the reply is DELTAI1 then the amps, e.g. DELTAI1 0.100 (Aim-TTi manual Iss 17 p.35).
None
INCV subsystem (2)
Command
What it does
Arguments
INCV1
Increment Voltage. Raise the output voltage by one voltage step (reference section 3).
None
INCV1V
Increment Voltage Verify. Raise the output voltage by one voltage step and wait until it has settled (reference section 3).
None
DECV subsystem (2)
Command
What it does
Arguments
DECV1
Decrement Voltage. Lower the output voltage by one voltage step (reference section 3).
None
DECV1V
Decrement Voltage Verify. Lower the output voltage by one voltage step and wait until it has settled (reference section 3).
None
IRANGE subsystem (2)
Command
What it does
Arguments
IRANGE1 <Range>
Set Current Range. Set the current meter range: 1 low (800 mA max), 2 high (8 A max). Turn the output off first (reference section 4; the PL068-P's ranges from Aim-TTi manual Iss 17 p.3).
Text: 1,2
IRANGE1?
Read Current Range. Read the current range: 1 low (800 mA), 2 high (8 A) (reference section 4; Aim-TTi manual Iss 17 p.3).
None
IFLOCK subsystem (2)
Command
What it does
Arguments
IFLOCK
Lock Interface. Take exclusive control of the supply for this interface; replies 1 when granted, -1 when another interface holds the lock. Pressing Local also releases it (Aim-TTi manual Iss 17 p.37).
None
IFLOCK?
Read Interface Lock. Read the interface lock: 1 this interface holds it, 0 no lock, -1 another interface holds it (Aim-TTi manual Iss 17 p.37).
None
LSE subsystem (2)
Command
What it does
Arguments
LSE1 <Mask>
Set Limit Status Enable. Set the Limit Status Enable mask, 0 to 255 (reference section 6).
Number: 0 to 255
LSE1?
Read Limit Status Enable. Read the Limit Status Enable mask (reference section 6).
None
TRIPRST subsystem (1)
Command
What it does
Arguments
TRIPRST
Clear Trip. Attempt to clear every trip condition: OVP, OCP, sense and over-temperature (reference section 1).
None
INCI subsystem (1)
Command
What it does
Arguments
INCI1
Increment Current. Raise the current limit by one current step (reference section 3).
None
DECI subsystem (1)
Command
What it does
Arguments
DECI1
Decrement Current. Lower the current limit by one current step (reference section 3).
None
Save (SAVE) (1)
Command
What it does
Arguments
SAV1 <Slot>
Save State. Save the present settings (voltage, current, OVP, OCP, step sizes) to a store, 0 to 9 (reference section 4).
Number: store 0 to 9
RCL subsystem (1)
Command
What it does
Arguments
RCL1 <Slot>
Recall State. Recall the settings saved in a store, 0 to 9 (reference section 4).
Number: store 0 to 9
LOCAL subsystem (1)
Command
What it does
Arguments
LOCAL
Return To Local. Return the instrument to front-panel control (reference section 5).
None
IFUNLOCK subsystem (1)
Command
What it does
Arguments
IFUNLOCK
Unlock Interface. Release this interface's lock; replies 0 when released, -1 when this interface does not hold it (Aim-TTi manual Iss 17 p.37).
None
LSR subsystem (1)
Command
What it does
Arguments
LSR1?
Read Limit Status. Read and clear the Limit Event Status Register: 1 CV mode, 2 CC mode, 4 over-voltage trip, 8 over-current trip, 64 a trip only the front panel or a power cycle clears (Aim-TTi manual Iss 17 p.29).
None
EER subsystem (1)
Command
What it does
Arguments
EER?
Read Error. Read and clear the Execution Error Register: 0 no error, 100 value out of range, 101 store holds corrupt data, 102 store is empty, 103 no second output, 104 not valid with the output on (IRANGE), 200 this interface has no write access (Aim-TTi manual Iss 17 p.29).
None
QER subsystem (1)
Command
What it does
Arguments
QER?
Read Query Error. Read and clear the Query Error Register (reference section 6).
None
ADDRESS subsystem (1)
Command
What it does
Arguments
ADDRESS?
Read Bus Address. Read the GPIB bus address (reference section 6).
None
IPADDR subsystem (1)
Command
What it does
Arguments
IPADDR?
Read IP Address. Read the present LAN IP address (reference section 6).
None
NETMASK subsystem (1)
Command
What it does
Arguments
NETMASK?
Read Netmask. Read the present LAN netmask (reference section 6).
None
IEEE 488.2 common commands (8)
Command
What it does
Arguments
*IDN?
Identify. Read the identification: THURLBY THANDAR, PL068-P, serial number, firmware version (reference section 5).
None
*RST
Reset. Reset to the remote-operation defaults: 0.1 V, 100 mA, steps 10 mV and 1 mA, low current range off, OVP and OCP 5% above the range maximums; interface settings and stored set-ups are kept (Aim-TTi manual Iss 17 p.32).
None
*CLS
Clear Status. Clear all status registers (reference section 5).
None
*OPC?
Operation Complete. Return 1 once every pending operation has finished (reference section 5).
None
*OPC
Set Operation Complete. Set the Operation Complete bit in the Standard Event Status Register once every pending operation has finished (reference section 5).
None
*WAI
Wait To Continue. Hold further commands until every pending operation has finished (reference section 5).
None
*ESR?
Read Event Status. Read and clear the Standard Event Status Register (reference section 6).
None
*STB?
Read Status Byte. Read the Status Byte Register (reference section 6).
None
Where these commands come from
This reference is generated from TestFlow's instrument library, which is drafted from Aim-TTi's programming documentation for the PL068-P, PL155-P, PL303-P, and PL601-P and checked when the agent runs it. The library marks this instrument as supporting *OPC? for operation-complete synchronisation. Ranges, timing, and option-dependent commands are defined by Aim-TTi's programming manual: check it before you rely on an edge case. Last updated October 29, 2026.
Frequently asked questions
Does the Aim-TTi PL068-P support SCPI remote control?
Yes. The Aim-TTi PL068-P accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 49 SCPI commands for it: 24 queries that read values or state, and 25 settings and actions.
How do I control the Aim-TTi PL068-P from Python?
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, V1 volts (e.g. 5.0) and V1O?.
Which command reads a value from the Aim-TTi PL068-P?
V1O? (measure voltage). Read the actual output voltage; the reply carries its unit, e.g. 5.002V (reference section 2). The full list of queries is in the command reference on this page, grouped by subsystem.