Power Supply · SCPI command reference

TDK-Lambda Z36-6 SCPI commands

The TDK-Lambda Z36-6 power supply is controlled remotely with SCPI commands. TestFlow's instrument library holds 20 commands for it, 8 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the Z36-6 on a real bench.

Commands
20
Queries
8
Subsystems
8
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your TDK-Lambda Z36-6 and writes the report from the measured results.

Get started

Control the TDK-Lambda Z36-6 from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the TDK-Lambda Z36-6's VISA address

# Paste the address of your TDK-Lambda Z36-6 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("OUTPut:STATe ON")  # Turn On CH1
print(inst.query("MEASure:VOLTage?"))  # Measure Voltage

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your TDK-Lambda Z36-6 and writes the report from the measured results.

Get started

Run a Z36-6 test without writing code

Describe the test in a sentence, for example “turn on ch1 on the Z36-6, then measure voltage and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

20 commands in 8 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Source (SOURce) (5)

CommandWhat it doesArguments
SOURce:VOLTage <Volt>Set Voltage. Sets the output voltage of the selected channel in volts.Number: 0-36
SOURce:VOLTage?Read Voltage Setting. Reads the set output voltage of the selected channel.None
SOURce:CURRent <Amp>Set Current. Sets the output current limit of the selected channel in amps.Number: 0-6
SOURce:CURRent?Read Current Setting. Reads the set output current limit of the selected channel.None
SOURce:VOLTage:PROTection:LEVel <Volt>Set OVP. Sets the over-voltage protection level of the selected channel.Number: 2-40

Output (OUTPut) (3)

CommandWhat it doesArguments
OUTPut:STATe ONTurn On CH1. Enables the output of channel 1.None
OUTPut:STATe OFFTurn Off CH1. Disables the output of channel 1.None
OUTPut:STATe?Read Output State. Reads whether the selected channel output is ON or OFF.None

Measurement (MEASure) (2)

CommandWhat it doesArguments
MEASure:VOLTage?Measure Voltage. Measures the output voltage of the selected channel in volts.None
MEASure:CURRent?Measure Current. Measures the output current of the selected channel in amps.None

System (SYSTem) (1)

CommandWhat it doesArguments
SYSTem:ERRor?Read Error. Reads the first message from the error queue.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (6)

CommandWhat it doesArguments
*SAV <Slot>Save State. Saves the current instrument state to the specified storage slot.Number
*RCL <Slot>Recall State. Recalls an instrument state from the specified storage slot.Number
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from TDK-Lambda's programming documentation for the Z36-6 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by TDK-Lambda's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the TDK-Lambda Z36-6 support SCPI remote control?

Yes. The TDK-Lambda Z36-6 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 20 SCPI commands for it: 8 queries that read values or state, and 12 settings and actions.

How do I control the TDK-Lambda Z36-6 from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTPut:STATe ON and MEASure:VOLTage?.

Which command reads a value from the TDK-Lambda Z36-6?

MEASure:VOLTage? (measure voltage). Measures the output voltage of the selected channel in volts. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the TDK-Lambda Z36-6 without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the TDK-Lambda Z36-6 through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.