Multimeter · SCPI command reference

Siglent SDM3045X and SDM3055 SCPI commands

The Siglent SDM3045X and SDM3055 multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 116 commands for it, 31 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the SDM3045X on a real bench.

Commands
116
Queries
31
Subsystems
18
Models covered
2

Control the Siglent SDM3045X from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Siglent SDM3045X's VISA address

# Paste the address of your Siglent SDM3045X (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("CONFigure:VOLTage:DC")  # Conf DCV
print(inst.query("MEASure:VOLTage:DC?"))  # Measure DCV

inst.close()

Run a SDM3045X test without writing code

Describe the test in a sentence, for example “conf dcv on the SDM3045X, then measure dcv and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

116 commands in 18 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The SDM3045X and SDM3055 share this command set in TestFlow's library.

Sense (SENSe) (25)

CommandWhat it doesArguments
SENSe:VOLTage:DC:RANGe <Range>Set DCV Range. Sets the fixed DC voltage range in volts (0.2, 2, 20, 200 or 1000).Number: 0.6,6,60,600,1000
SENSe:VOLTage:AC:RANGe <Range>Set ACV Range. Sets the fixed AC voltage range in volts (0.2, 2, 20, 200 or 750).Number: 0.6,6,60,600,750
SENSe:CURRent:DC:RANGe <Range>Set DCI Range. Sets the fixed DC current range in amps.Number: 0.0006,0.006,0.06,0.6,6,10
SENSe:CURRent:AC:RANGe <Range>Set ACI Range. Sets the fixed AC current range in amps.Number: 0.06,0.6,6,10
SENSe:RESistance:RANGe <Range>Set 2W Res Range. Sets the fixed 2-wire resistance range in ohms.Number: 600,6000,60000,600000,6000000,60000000,100000000
SENSe:FRESistance:RANGe <Range>Set 4W Res Range. Sets the fixed 4-wire resistance range in ohms.Number: 600,6000,60000,600000,6000000,60000000,100000000
SENSe:CAPacitance:RANGe <Range>Set Cap Range. Sets the fixed capacitance range in farads.Number: 0.000000002,0.00000002,0.0000002,0.000002,0.00002,0.0002,0.01
SENSe:VOLTage:DC:RANGe:AUTO ONDCV Autorange On. Enables autorange for DC voltage measurement.None
SENSe:VOLTage:DC:RANGe:AUTO OFFDCV Autorange Off. Disables autorange for DC voltage measurement.None
SENSe:VOLTage:AC:RANGe:AUTO ONACV Autorange On. Enables autorange for AC voltage measurement.None
SENSe:VOLTage:AC:RANGe:AUTO OFFACV Autorange Off. Disables autorange for AC voltage measurement.None
SENSe:CURRent:DC:RANGe:AUTO ONDCI Autorange On. Enables autorange for DC current measurement.None
SENSe:CURRent:DC:RANGe:AUTO OFFDCI Autorange Off. Disables autorange for DC current measurement.None
SENSe:RESistance:RANGe:AUTO ONRes Autorange On. Enables autorange for 2-wire resistance measurement.None
SENSe:RESistance:RANGe:AUTO OFFRes Autorange Off. Disables autorange for 2-wire resistance measurement.None
SENSe:VOLTage:DC:NPLC <NPLC>Set DCV NPLC. Sets the integration time for DC voltage in number of power line cycles.Number: 0.3,1,10
SENSe:CURRent:DC:NPLC <NPLC>Set DCI NPLC. Sets the integration time for DC current in number of power line cycles.Number: 0.3,1,10
SENSe:RESistance:NPLC <NPLC>Set 2W Res NPLC. Sets the integration time for 2-wire resistance in number of power line cycles.Number: 0.3,1,10
SENSe:FRESistance:NPLC <NPLC>Set 4W Res NPLC. Sets the integration time for 4-wire resistance in number of power line cycles.Number: 0.3,1,10
SENSe:VOLTage:DC:IMPedance <Impedance>Set DCV Impedance. Sets the DC voltage input impedance to 10 MOhm or greater than 10 GOhm.Text: 10M,10G
SENSe:CONTinuity:THReshold:VALue <Ohm>Set Cont Threshold. Sets the continuity beep resistance threshold in ohms (1 to 2000).Number
SENSe:<Function>:NULL:STATe ONNull On. Enables the relative (null) math operation.Text: VOLTage:DC,VOLTage:AC,CURRent:DC,CURRent:AC,RESistance,FRESistance,FREQuency,CAPacitance,TEMPerature
SENSe:<Function>:NULL:STATe OFFNull Off. Disables the relative (null) math operation.Text: VOLTage:DC,VOLTage:AC,CURRent:DC,CURRent:AC,RESistance,FRESistance,FREQuency,CAPacitance,TEMPerature
SENSe:<Function>:NULL:VALue <Offset>Set Null Offset. Sets the reference (null) value subtracted from subsequent readings.Text: VOLTage:DC,VOLTage:AC,CURRent:DC,CURRent:AC,RESistance,FRESistance,FREQuency,CAPacitance,TEMPerature; Number: value
SENSe:<Function>:NULL:VALue?Read Null Offset. Returns the present null reference value.Text: VOLTage:DC,VOLTage:AC,CURRent:DC,CURRent:AC,RESistance,FRESistance,FREQuency,CAPacitance,TEMPerature

Configure (CONFigure) (20)

CommandWhat it doesArguments
CONFigure:VOLTage:DCConf DCV. Configures the instrument for DC voltage measurement using autorange.None
CONFigure:VOLTage:DC <Range>Conf DCV Range. Configures DC voltage measurement and sets the specified range or autorange.Text: 600mV,6V,60V,600V,1000V,AUTO
CONFigure:VOLTage:ACConf ACV. Configures the instrument for AC voltage measurement using autorange.None
CONFigure:VOLTage:AC <Range>Conf ACV Range. Configures AC voltage measurement and sets the specified range or autorange.Text: 600mV,6V,60V,600V,750V,AUTO
CONFigure:CURRent:DCConf DCI. Configures the instrument for DC current measurement using autorange.None
CONFigure:CURRent:DC <Range>Conf DCI Range. Configures DC current measurement and sets the specified range or autorange.Text: 600uA,6mA,60mA,600mA,6A,10A,AUTO
CONFigure:CURRent:ACConf ACI. Configures the instrument for AC current measurement using autorange.None
CONFigure:CURRent:AC <Range>Conf ACI Range. Configures AC current measurement and sets the specified range or autorange.Text: 60mA,600mA,6A,10A,AUTO
CONFigure:RESistanceConf 2W Resistance. Configures the instrument for 2-wire resistance measurement using autorange.None
CONFigure:RESistance <Range>Conf 2W Res Range. Configures 2-wire resistance measurement and sets the specified range in ohms or autorange.Text: 600,6000,60000,600000,6000000,60000000,100000000,AUTO
CONFigure:FRESistanceConf 4W Resistance. Configures the instrument for 4-wire resistance measurement using autorange.None
CONFigure:FRESistance <Range>Conf 4W Res Range. Configures 4-wire resistance measurement and sets the specified range in ohms or autorange.Text: 600,6000,60000,600000,6000000,60000000,100000000,AUTO
CONFigure:FREQuencyConf Frequency. Configures the instrument for frequency measurement.None
CONFigure:PERiodConf Period. Configures the instrument for period measurement.None
CONFigure:CAPacitanceConf Capacitance. Configures the instrument for capacitance measurement using autorange.None
CONFigure:CAPacitance <Range>Conf Cap Range. Configures capacitance measurement and sets the specified range or autorange.Text: 2nF,20nF,200nF,2uF,20uF,200uF,10000uF,AUTO
CONFigure:CONTinuityConf Continuity. Configures the instrument for continuity test.None
CONFigure:DIODeConf Diode. Configures the instrument for diode test.None
CONFigure:TEMPeratureConf Temperature. Configures the instrument for temperature measurement.None
CONFigure?Read Conf. Returns the present function and range configuration as a quoted string.None

Calculate (CALCulate) (16)

CommandWhat it doesArguments
CALCulate:SCALe:STATe ONScale On. Enables the scale (dB or dBm) math operation.None
CALCulate:SCALe:STATe OFFScale Off. Disables the scale (dB or dBm) math operation.None
CALCulate:SCALe:FUNCtion <Function>Set Scale Function. Selects whether the scale operation computes dB or dBm.Text: DB,DBM
CALCulate:SCALe:DB:REFerence <Reference>Set dB Reference. Sets the dB reference value in dBm.Number
CALCulate:SCALe:DBM:REFerence <Reference>Set dBm Reference. Sets the dBm reference resistance in ohms.Number: 50,75,93,110,124,125,135,150,250,300,500,600,800,900,1000,1200
CALCulate:AVERage:STATe ONStatistics On. Enables the statistics (min max average) math operation.None
CALCulate:AVERage:STATe OFFStatistics Off. Disables the statistics (min max average) math operation.None
CALCulate:AVERage:AVERage?Stat Average. Returns the average of all readings since statistics were enabled.None
CALCulate:AVERage:MINimum?Stat Min. Returns the minimum reading since statistics were enabled.None
CALCulate:AVERage:MAXimum?Stat Max. Returns the maximum reading since statistics were enabled.None
CALCulate:AVERage:COUNt?Stat Count. Returns the number of readings included in the statistics.None
CALCulate:AVERage:PTPeak?Stat PtoP. Returns the peak-to-peak value of the readings since statistics were enabled.None
CALCulate:LIMit:STATe ONLimit On. Enables the limit (pass/fail) math operation.None
CALCulate:LIMit:STATe OFFLimit Off. Disables the limit (pass/fail) math operation.None
CALCulate:LIMit:LOWer <Lower>Set Limit Lower. Sets the lower limit for the limit (pass/fail) operation.Number
CALCulate:LIMit:UPPer <Upper>Set Limit Upper. Sets the upper limit for the limit (pass/fail) operation.Number

Function (FUNCtion) (13)

CommandWhat it doesArguments
FUNCtion "VOLTage:DC"Func DCV. Selects DC voltage as the active measurement function.None
FUNCtion "VOLTage:AC"Func ACV. Selects AC voltage as the active measurement function.None
FUNCtion "CURRent:DC"Func DCI. Selects DC current as the active measurement function.None
FUNCtion "CURRent:AC"Func ACI. Selects AC current as the active measurement function.None
FUNCtion "RESistance"Func 2W Resistance. Selects 2-wire resistance as the active measurement function.None
FUNCtion "FRESistance"Func 4W Resistance. Selects 4-wire resistance as the active measurement function.None
FUNCtion "FREQuency"Func Frequency. Selects frequency as the active measurement function.None
FUNCtion "PERiod"Func Period. Selects period as the active measurement function.None
FUNCtion "CAPacitance"Func Capacitance. Selects capacitance as the active measurement function.None
FUNCtion "CONTinuity"Func Continuity. Selects continuity test as the active measurement function.None
FUNCtion "DIODe"Func Diode. Selects diode test as the active measurement function.None
FUNCtion "TEMPerature"Func Temperature. Selects temperature as the active measurement function.None
FUNCtion?Read Function. Returns the active measurement function as a quoted string.None

Measurement (MEASure) (12)

CommandWhat it doesArguments
MEASure:VOLTage:DC?Measure DCV. Configures DC voltage and returns one measurement in volts.None
MEASure:VOLTage:AC?Measure ACV. Configures AC voltage and returns one measurement in volts.None
MEASure:CURRent:DC?Measure DCI. Configures DC current and returns one measurement in amps.None
MEASure:CURRent:AC?Measure ACI. Configures AC current and returns one measurement in amps.None
MEASure:RESistance?Measure 2W Resistance. Configures 2-wire resistance and returns one measurement in ohms.None
MEASure:FRESistance?Measure 4W Resistance. Configures 4-wire resistance and returns one measurement in ohms.None
MEASure:FREQuency?Measure Frequency. Configures frequency and returns one measurement in Hz.None
MEASure:PERiod?Measure Period. Configures period and returns one measurement in seconds.None
MEASure:CAPacitance?Measure Capacitance. Configures capacitance and returns one measurement in farads.None
MEASure:CONTinuity?Measure Continuity. Configures continuity and returns the measured resistance in ohms.None
MEASure:DIODe?Measure Diode. Configures diode test and returns the measured forward voltage in volts.None
MEASure:TEMPerature?Measure Temperature. Configures temperature and returns one measurement in the configured units.None

Trigger (TRIGger) (7)

CommandWhat it doesArguments
TRIGger:SOURce <Source>Set Trigger Source. Sets the trigger source to immediate external or bus.Text: IMMediate,EXTernal,BUS
TRIGger:SOURce?Read Trigger Source. Returns the present trigger source setting.None
TRIGger:DELay <Seconds>Set Trigger Delay. Sets the delay in seconds between the trigger event and each measurement.Number
TRIGger:DELay:AUTO ONSet Trigger Delay Auto On. Enables automatic trigger delay determined by function and range.None
TRIGger:DELay:AUTO OFFSet Trigger Delay Auto Off. Disables automatic trigger delay so the manual delay value is used.None
TRIGger:COUNt <Count>Set Trigger Count. Sets the number of triggers accepted before returning to the idle state.Number
TRIGger:SLOPe <Slope>Set Trigger Slope. Sets the edge polarity used for the external trigger input.Text: POSitive,NEGative

System (SYSTem) (4)

CommandWhat it doesArguments
SYSTem:BEEPer:STATe ONBeeper On. Enables the beeper.None
SYSTem:BEEPer:STATe OFFBeeper Off. Disables the beeper.None
SYSTem:ERRor?Read Error. Reads and clears the oldest entry from the error queue (0,"No error" if empty).None
SYSTem:VERSion?Read SCPI Version. Returns the SCPI version that the instrument complies with.None

DATA subsystem (2)

CommandWhat it doesArguments
DATA:POINts?Read Data Points. Returns the number of readings currently stored in the reading memory.None
DATA:REMove? <Count>Remove Data. Reads and removes the specified number of oldest readings from the reading memory.Number

SAMP subsystem (2)

CommandWhat it doesArguments
SAMPle:COUNt <Count>Set Sample Count. Sets the number of samples taken per trigger.Number
SAMPle:COUNt?Read Sample Count. Returns the number of samples taken per trigger.None

Read (READ) (1)

CommandWhat it doesArguments
READ?Read Value. Starts a measurement using the present configuration and returns the reading.None

Fetch (FETCh) (1)

CommandWhat it doesArguments
FETCh?Fetch Value. Returns readings already stored in instrument memory from the last triggered acquisition.None

Initiate (INITiate) (1)

CommandWhat it doesArguments
INITiateInitiate. Changes state from idle to wait-for-trigger and stores readings in memory when triggered.None

UNIT subsystem (1)

CommandWhat it doesArguments
UNIT:TEMPerature <Unit>Set Temp Units. Sets the temperature measurement units to Celsius Fahrenheit or Kelvin.Text: C,F,K

Abort (1)

CommandWhat it doesArguments
ABORtAbort. Aborts a measurement in progress and returns the trigger system to idle.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (7)

CommandWhat it doesArguments
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 for pass or 1 for failure.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*TRGTrigger Bus. Generates a bus trigger when the instrument is in wait-for-trigger state.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Siglent's programming documentation for the SDM3045X and SDM3055 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Siglent's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the Siglent SDM3045X support SCPI remote control?

Yes. The Siglent SDM3045X accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 116 SCPI commands for it: 31 queries that read values or state, and 85 settings and actions.

How do I control the Siglent SDM3045X from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, CONFigure:VOLTage:DC and MEASure:VOLTage:DC?.

Which command reads a value from the Siglent SDM3045X?

MEASure:VOLTage:DC? (measure dcv). Configures DC voltage and returns one measurement in volts. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Siglent SDM3045X without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Siglent SDM3045X through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.