The Keysight U3606B multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 222 commands for it, 112 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the U3606B on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Keysight U3606B and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Keysight U3606B's VISA address
# Paste the address of your Keysight U3606B (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("OUTP:STAT ON") # Enable Output
print(inst.query("MEAS:FREQ?")) # Measure Frequency
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Keysight U3606B and writes the report from the measured results.
Describe the test in a sentence, for example “enable output on the U3606B, then measure frequency and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
222 commands in 27 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| VOLT:LIM <value>[,<save>] | Set Over-Voltage Limit. Sets the over-voltage limit for constant current output. | Number: 0 V to range maximum; Text: 0|1|OFF|ON, optional store flag |
| VOLT:LIM? | Read Over-Voltage Limit. Returns the over-voltage limit of constant current output. | None |
| VOLT:PROT <value>[,<save>] | Set Over-Voltage Protection. Sets the over-voltage protection that disables output when exceeded. | Number: 0 V to 33 V depending on range; Text: 0|1|OFF|ON, optional store flag |
| VOLT:PROT? | Read Over-Voltage Protection. Returns the over-voltage protection value. | None |
| VOLT <value> | Set Voltage Level. Sets the constant voltage output amplitude. | Number: 0 V to range maximum |
| VOLT? | Read Voltage Level. Returns the constant voltage output amplitude. | None |
| VOLT:RAMP <value> | Set Voltage Ramp End. Sets the amplitude end position of the voltage ramp. | Number: 0 V to range maximum |
| VOLT:RAMP? | Read Voltage Ramp End. Returns the voltage ramp amplitude end position. | None |
| VOLT:RAMP:STEP <value> | Set Voltage Ramp Steps. Sets the number of steps in the voltage ramp. | Number: 1 to 10000 |
| VOLT:RAMP:STEP? | Read Voltage Ramp Steps. Returns the number of steps in the voltage ramp. | None |
| VOLT:SCAN <value> | Set Voltage Scan End. Sets the amplitude end position of the voltage scan. | Number: 0 V to range maximum |
| VOLT:SCAN? | Read Voltage Scan End. Returns the voltage scan amplitude end position. | None |
| VOLT:SCAN:STEP <value> | Set Voltage Scan Steps. Sets the number of steps in the voltage scan. | Number: 1 to 100 |
| VOLT:SCAN:STEP? | Read Voltage Scan Steps. Returns the number of steps in the voltage scan. | None |
| VOLT:SCAN:DWEL <value> | Set Voltage Scan Dwell. Sets the dwelling time of the voltage scan. | Number: 1 to 99 (s) |
| VOLT:SCAN:DWEL? | Read Voltage Scan Dwell. Returns the dwelling time of the voltage scan. | None |
| VOLT:DC:RANG <range> | Set DC Voltage Range. Selects a fixed DC voltage range and disables autoranging. | Number: 0.02, 0.1, 1, 10, 100, 1000 V; MIN, MAX, DEF |
| VOLT:DC:RANG? | Query DC Voltage Range. Returns the DC voltage measurement range. | None |
| VOLT:DC:RANG:AUTO <mode> | Set DC Voltage Autorange. Enables or disables DC voltage autoranging. | Text: 0|1|OFF|ON |
| VOLT:DC:RANG:AUTO? | Query DC Voltage Autorange. Returns the DC voltage autoranging setting. | None |
| VOLT:DC:RES <resolution> | Set DC Voltage Resolution. Sets the DC voltage measurement resolution. | Number: 1e-7 to 1e-2 V; MIN, MAX, DEF |
| VOLT:DC:RES? | Query DC Voltage Resolution. Returns the DC voltage resolution. | None |
| VOLT:AC:RANG <range> | Set AC Voltage Range. Selects a fixed AC voltage range and disables autoranging. | Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF |
| VOLT:AC:RANG? | Query AC Voltage Range. Returns the AC voltage measurement range. | None |
| VOLT:AC:RANG:AUTO <mode> | Set AC Voltage Autorange. Enables or disables AC voltage autoranging. | Text: 0|1|OFF|ON |
| VOLT:AC:RANG:AUTO? | Query AC Voltage Autorange. Returns the AC voltage autoranging setting. | None |
| VOLT:AC:RES <resolution> | Set AC Voltage Resolution. Sets the AC voltage measurement resolution. | Number: 1e-6 to 1e-1 V; MIN, MAX, DEF |
| VOLT:AC:RES? | Query AC Voltage Resolution. Returns the AC voltage resolution. | None |
| VOLT:ACDC:RANG <range> | Set AC+DC Voltage Range. Selects a fixed AC+DC voltage range and disables autoranging. | Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF |
| VOLT:ACDC:RANG? | Query AC+DC Voltage Range. Returns the AC+DC voltage measurement range. | None |
| VOLT:ACDC:RANG:AUTO <mode> | Set AC+DC Voltage Autorange. Enables or disables AC+DC voltage autoranging. | Text: 0|1|OFF|ON |
| VOLT:ACDC:RANG:AUTO? | Query AC+DC Voltage Autorange. Returns the AC+DC voltage autoranging setting. | None |
| VOLT:ACDC:RES <resolution> | Set AC+DC Voltage Resolution. Sets the AC+DC voltage measurement resolution. | Number: 1e-6 to 1e-1 V; MIN, MAX, DEF |
| VOLT:ACDC:RES? | Query AC+DC Voltage Resolution. Returns the AC+DC voltage resolution. | None |
| Command | What it does | Arguments |
|---|---|---|
| CURR:LIM <value>[,<save>] | Set Over-Current Limit. Sets the over-current limit for constant voltage output. | Number: 0 A to range maximum; Text: 0|1|OFF|ON, optional store flag |
| CURR:LIM? | Read Over-Current Limit. Returns the over-current limit of constant voltage output. | None |
| CURR:PROT <value>[,<save>] | Set Over-Current Protection. Sets the over-current protection that disables output when exceeded. | Number: 0 A to 3.3 A depending on range; Text: 0|1|OFF|ON, optional store flag |
| CURR:PROT? | Read Over-Current Protection. Returns the over-current protection value. | None |
| CURR <value> | Set Current Level. Sets the constant current output amplitude. | Number: 0 A to range maximum |
| CURR? | Read Current Level. Returns the constant current output amplitude. | None |
| CURR:RAMP <value> | Set Current Ramp End. Sets the amplitude end position of the current ramp. | Number: 0 A to range maximum |
| CURR:RAMP? | Read Current Ramp End. Returns the current ramp amplitude end position. | None |
| CURR:RAMP:STEP <value> | Set Current Ramp Steps. Sets the number of steps in the current ramp. | Number: 1 to 10000 |
| CURR:RAMP:STEP? | Read Current Ramp Steps. Returns the number of steps in the current ramp. | None |
| CURR:SCAN <value> | Set Current Scan End. Sets the amplitude end position of the current scan. | Number: 0 A to range maximum |
| CURR:SCAN? | Read Current Scan End. Returns the current scan amplitude end position. | None |
| CURR:SCAN:STEP <value> | Set Current Scan Steps. Sets the number of steps in the current scan. | Number: 1 to 100 |
| CURR:SCAN:STEP? | Read Current Scan Steps. Returns the number of steps in the current scan. | None |
| CURR:SCAN:DWEL <value> | Set Current Scan Dwell. Sets the dwelling time of the current scan. | Number: 1 to 99 (s) |
| CURR:SCAN:DWEL? | Read Current Scan Dwell. Returns the dwelling time of the current scan. | None |
| CURR:DC:RANG <range> | Set DC Current Range. Selects a fixed DC current range and disables autoranging. | Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF |
| CURR:DC:RANG? | Query DC Current Range. Returns the DC current measurement range. | None |
| CURR:DC:RANG:AUTO <mode> | Set DC Current Autorange. Enables or disables DC current autoranging. | Text: 0|1|OFF|ON |
| CURR:DC:RANG:AUTO? | Query DC Current Autorange. Returns the DC current autoranging setting. | None |
| CURR:DC:RES <resolution> | Set DC Current Resolution. Sets the DC current measurement resolution. | Number: 1e-7 to 1e-3 A; MIN, MAX, DEF |
| CURR:DC:RES? | Query DC Current Resolution. Returns the DC current resolution. | None |
| CURR:AC:RANG <range> | Set AC Current Range. Selects a fixed AC current range and disables autoranging. | Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF |
| CURR:AC:RANG? | Query AC Current Range. Returns the AC current measurement range. | None |
| CURR:AC:RANG:AUTO <mode> | Set AC Current Autorange. Enables or disables AC current autoranging. | Text: 0|1|OFF|ON |
| CURR:AC:RANG:AUTO? | Query AC Current Autorange. Returns the AC current autoranging setting. | None |
| CURR:AC:RES <resolution> | Set AC Current Resolution. Sets the AC current measurement resolution. | Number: 1e-7 to 1e-3 A; MIN, MAX, DEF |
| CURR:AC:RES? | Query AC Current Resolution. Returns the AC current resolution. | None |
| CURR:ACDC:RANG <range> | Set AC+DC Current Range. Selects a fixed AC+DC current range and disables autoranging. | Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF |
| CURR:ACDC:RANG? | Query AC+DC Current Range. Returns the AC+DC current measurement range. | None |
| CURR:ACDC:RANG:AUTO <mode> | Set AC+DC Current Autorange. Enables or disables AC+DC current autoranging. | Text: 0|1|OFF|ON |
| CURR:ACDC:RANG:AUTO? | Query AC+DC Current Autorange. Returns the AC+DC current autoranging setting. | None |
| CURR:ACDC:RES <resolution> | Set AC+DC Current Resolution. Sets the AC+DC current measurement resolution. | Number: 1e-7 to 1e-3 A; MIN, MAX, DEF |
| CURR:ACDC:RES? | Query AC+DC Current Resolution. Returns the AC+DC current resolution. | None |
| Command | What it does | Arguments |
|---|---|---|
| MEAS:VOLT:DC? <range>,<resolution> | Measure DC Voltage. Configures and returns one DC voltage reading. | Number: 0.02, 0.1, 1, 10, 100, 1000 V; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-2 V; MIN, MAX, DEF |
| MEAS:VOLT:AC? <range>,<resolution> | Measure AC Voltage. Configures and returns one AC voltage reading. | Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 V; MIN, MAX, DEF |
| MEAS:VOLT:ACDC? <range>,<resolution> | Measure AC+DC Voltage. Configures and returns one AC+DC voltage reading. | Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 V; MIN, MAX, DEF |
| MEAS:CURR:DC? <range>,<resolution> | Measure DC Current. Configures and returns one DC current reading. | Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF |
| MEAS:CURR:AC? <range>,<resolution> | Measure AC Current. Configures and returns one AC current reading. | Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF |
| MEAS:CURR:ACDC? <range>,<resolution> | Measure AC+DC Current. Configures and returns one AC+DC current reading. | Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF |
| MEAS:RES? <range>,<resolution> | Measure Resistance. Configures and returns one 2-wire resistance reading. | Number: 100 to 100000000 ohm; MIN, MAX, DEF; omit for autorange; Number: 0.001 to 10000 ohm; MIN, MAX, DEF |
| MEAS:CONT? | Measure Continuity. Configures a continuity test and returns the measured resistance. | None |
| MEAS:LRES? <range>,<resolution> | Measure Low Resistance. Configures and returns one 4-wire low-resistance reading. | Number: 0.1 to 1000 ohm; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 ohm; MIN, MAX, DEF |
| MEAS:CAP? <range>,<resolution> | Measure Capacitance. Configures and returns one capacitance reading. | Number: 1e-9 to 1e-2 F; MIN, MAX, DEF; omit for autorange; Number: 1e-12 to 1e-6 F; MIN, MAX |
| MEAS:DIOD? | Measure Diode. Configures a diode test and returns the measured voltage. | None |
| MEAS:FREQ? | Measure Frequency. Configures a frequency measurement and returns one reading; the signal path must be set by a prior AC configuration. | None |
| MEAS:PWID? | Measure Pulse Width. Configures a pulse width measurement and returns one reading. | None |
| MEAS:DCYC? | Measure Duty Cycle. Configures a duty cycle measurement and returns one reading in percent. | None |
| MEAS:VOLT:DC? | Measure DC Voltage (MEAS:VOLT:DC?). Configures and measures DC voltage, returning the reading. | None |
| MEAS:VOLT:AC? | Measure AC Voltage (MEAS:VOLT:AC?). Configures and measures AC voltage, returning the reading. | None |
| MEAS:CURR:DC? | Measure DC Current (MEAS:CURR:DC?). Configures and measures DC current, returning the reading. | None |
| MEAS:CURR:AC? | Measure AC Current (MEAS:CURR:AC?). Configures and measures AC current, returning the reading. | None |
| MEAS:RES? | Measure Resistance (MEAS:RES?). Configures and measures 2-wire resistance, returning the reading. | None |
| Command | What it does | Arguments |
|---|---|---|
| CONF:AC <range>,<resolution> | Configure AC Voltage. Configures AC voltage measurements without starting a measurement. | Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 V; MIN, MAX, DEF |
| CONF:CURR:AC <range>,<resolution> | Configure AC Current. Configures AC current measurements without starting a measurement. | Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF |
| CONF:CURR:ACDC <range>,<resolution> | Configure AC+DC Current. Configures AC+DC current measurements without starting a measurement. | Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF |
| CONF:RES <range>,<resolution> | Configure Resistance. Configures 2-wire resistance measurements without starting a measurement. | Number: 100 to 100000000 ohm; MIN, MAX, DEF; omit for autorange; Number: 0.001 to 10000 ohm; MIN, MAX, DEF |
| CONF:CONT <range>,<resolution> | Configure Continuity. Configures continuity tests without starting a test. | Number: 100 to 100000000 ohm; MIN, MAX, DEF; Number: MAX, MIN |
| CONF:LRES <range>,<resolution> | Configure Low Resistance. Configures 4-wire low-resistance measurements without starting a measurement. | Number: 0.1 to 1000 ohm; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 ohm; MIN, MAX, DEF |
| CONF:CAP <range>,<resolution> | Configure Capacitance. Configures capacitance measurements without starting a measurement. | Number: 1e-9 to 1e-2 F; MIN, MAX, DEF; omit for autorange; Number: 1e-12 to 1e-6 F; MIN, MAX |
| CONF:DIOD | Configure Diode. Configures diode tests without starting a test. | None |
| CONF:FREQ | Configure Frequency. Configures frequency measurements on the path set by the preceding AC configuration. | None |
| CONF:PWID | Configure Pulse Width. Configures pulse width measurements on the path set by the preceding AC configuration. | None |
| CONF:DCYC | Configure Duty Cycle. Configures duty cycle measurements on the path set by the preceding AC configuration. | None |
| CONF? | Get Configuration. Returns the present measurement function, range and resolution. | None |
| CONF:VOLT:DC <range>,<resolution> | Configure DC Voltage. Configures DC voltage measurement without starting a measurement. | Number: range 0.02, 0.1, 1, 10, 100, 1000 V, or AUTO, MIN, MAX, DEF; Number: resolution in V, or MIN, MAX, DEF |
| CONF:VOLT:AC <range>,<resolution> | Configure AC Voltage (CONF:VOLT:AC). Configures AC voltage measurement without starting a measurement. | Number: range 0.1, 1, 10, 100, 750 V, or AUTO, MIN, MAX, DEF; Number: resolution in V, or MIN, MAX, DEF |
| CONF:CURR:DC <range>,<resolution> | Configure DC Current. Configures DC current measurement without starting a measurement. | Number: range 0.01, 0.1, 1, 3 A, or AUTO, MIN, MAX, DEF; Number: resolution in A, or MIN, MAX, DEF |
| Command | What it does | Arguments |
|---|---|---|
| SYST:ERR? | Read Error. Reads and clears one error from the error queue. | None |
| SYST:PRES | Instrument Preset. Resets the instrument to factory default state, same as *RST. | None |
| SYST:VERS? | Read SCPI Version. Returns the SCPI standard version the instrument complies with. | None |
| SYST:SMO:STAT <state> | Set Smooth Function. Enables or disables the smooth function. | Text: 0|1|OFF|ON |
| SYST:SMO:STAT? | Read Smooth Function. Returns the smooth function state. | None |
| SYST:SMO:FLUC <value> | Set Smooth Fluctuation. Sets the fluctuation count of the smooth function. | Number: 0.0 to 9.9 (%) |
| SYST:SMO:FLUC? | Read Smooth Fluctuation. Returns the smooth fluctuation count. | None |
| SYST:SMO:POIN <value> | Set Smooth Points. Sets the number of points for the smooth function. | Number: 2 to 1999 |
| SYST:SMO:POIN? | Read Smooth Points. Returns the number of smooth function points. | None |
| Command | What it does | Arguments |
|---|---|---|
| STAT:OPER:COND? | Read Operation Condition. Returns the Standard Operation condition register. | None |
| STAT:OPER:ENAB <enable_value> | Set Operation Enable. Sets the Standard Operation enable register mask. | Number: 0 to 65535 |
| STAT:OPER:ENAB? | Read Operation Enable. Returns the Standard Operation enable register mask. | None |
| STAT:OPER:EVEN? | Read Operation Event. Returns and clears the Standard Operation event register. | None |
| STAT:PRES | Preset Status. Clears the Questionable and Operation enable registers. | None |
| STAT:QUES:COND? | Read Questionable Condition. Returns the Questionable Data condition register. | None |
| STAT:QUES:ENAB <enable_value> | Set Questionable Enable. Sets the Questionable Data enable register mask. | Number: 0 to 65535 |
| STAT:QUES:ENAB? | Read Questionable Enable. Returns the Questionable Data enable register mask. | None |
| STAT:QUES:EVEN? | Read Questionable Event. Returns and clears the Questionable Data event register. | None |
| Command | What it does | Arguments |
|---|---|---|
| SQU:AMPL <value> | Set Square Wave Amplitude. Sets the square-wave output amplitude. | Number: 0 V to range maximum |
| SQU:AMPL? | Read Square Wave Amplitude. Returns the square-wave output amplitude. | None |
| SQU:FREQ <value> | Set Square Wave Frequency. Sets the square-wave output frequency. | Number: 0.5 to 4800 (Hz), predefined steps; S1S2 range 10 to 4800 |
| SQU:FREQ? | Read Square Wave Frequency. Returns the square-wave output frequency. | None |
| SQU:DCYC <value> | Set Square Wave Duty Cycle. Sets the square-wave duty cycle; not valid in S1S2 auto range. | Number: 0 to 100 (%) |
| SQU:DCYC? | Read Square Wave Duty Cycle. Returns the square-wave duty cycle. | None |
| SQU:PWID <second> | Set Square Wave Pulse Width. Sets the square-wave pulse width; not valid in S1S2 auto range. | Number: 0 s to MAX (depends on frequency) |
| SQU:PWID? | Read Square Wave Pulse Width. Returns the square-wave pulse width. | None |
| Command | What it does | Arguments |
|---|---|---|
| FREQ:VOLT:RANG <range> | Set Frequency Voltage Range. Selects the frequency measurement range via voltage path; disables autorange. | Number: 0.1 V|1 V|10 V|100 V|750 V|MIN|MAX|DEF |
| FREQ:VOLT:RANG? | Read Frequency Voltage Range. Returns the frequency voltage measurement range. | None |
| FREQ:VOLT:RANG:AUTO <mode> | Set Frequency Voltage Autorange. Disables or enables frequency autoranging via voltage path. | Text: 0|1|OFF|ON |
| FREQ:VOLT:RANG:AUTO? | Read Frequency Voltage Autorange. Returns the frequency voltage autoranging setting. | None |
| FREQ:CURR:RANG <range> | Set Frequency Current Range. Selects the frequency measurement range via current path; disables autorange. | Number: 0.01 A|0.1 A|1 A|3 A|MIN|MAX|DEF |
| FREQ:CURR:RANG? | Read Frequency Current Range. Returns the frequency current measurement range. | None |
| FREQ:CURR:RANG:AUTO <mode> | Set Frequency Current Autorange. Disables or enables frequency autoranging via current path. | Text: 0|1|OFF|ON |
| FREQ:CURR:RANG:AUTO? | Read Frequency Current Autorange. Returns the frequency current autoranging setting. | None |
| Command | What it does | Arguments |
|---|---|---|
| PWID:VOLT:RANG <range> | Set Pulse Width Voltage Range. Selects the pulse width measurement range via voltage path; disables autorange. | Number: 0.1 V|1 V|10 V|100 V|750 V|MIN|MAX|DEF |
| PWID:VOLT:RANG? | Read Pulse Width Voltage Range. Returns the pulse width voltage measurement range. | None |
| PWID:VOLT:RANG:AUTO <mode> | Set Pulse Width Voltage Autorange. Disables or enables pulse width autoranging via voltage path. | Text: 0|1|OFF|ON |
| PWID:VOLT:RANG:AUTO? | Read Pulse Width Voltage Autorange. Returns the pulse width voltage autoranging setting. | None |
| PWID:CURR:RANG <range> | Set Pulse Width Current Range. Selects the pulse width measurement range via current path; disables autorange. | Number: 0.01 A|0.1 A|1 A|3 A|MIN|MAX|DEF |
| PWID:CURR:RANG? | Read Pulse Width Current Range. Returns the pulse width current measurement range. | None |
| PWID:CURR:RANG:AUTO <mode> | Set Pulse Width Current Autorange. Disables or enables pulse width autoranging via current path. | Text: 0|1|OFF|ON |
| PWID:CURR:RANG:AUTO? | Read Pulse Width Current Autorange. Returns the pulse width current autoranging setting. | None |
| Command | What it does | Arguments |
|---|---|---|
| DCYC:VOLT:RANG <range> | Set Duty Cycle Voltage Range. Selects the duty cycle measurement range via voltage path; disables autorange. | Number: 0.1 V|1 V|10 V|100 V|750 V|MIN|MAX|DEF |
| DCYC:VOLT:RANG? | Read Duty Cycle Voltage Range. Returns the duty cycle voltage measurement range. | None |
| DCYC:VOLT:RANG:AUTO <mode> | Set Duty Cycle Voltage Autorange. Disables or enables duty cycle autoranging via voltage path. | Text: 0|1|OFF|ON |
| DCYC:VOLT:RANG:AUTO? | Read Duty Cycle Voltage Autorange. Returns the duty cycle voltage autoranging setting. | None |
| DCYC:CURR:RANG <range> | Set Duty Cycle Current Range. Selects the duty cycle measurement range via current path; disables autorange. | Number: 0.01 A|0.1 A|1 A|3 A|MIN|MAX|DEF |
| DCYC:CURR:RANG? | Read Duty Cycle Current Range. Returns the duty cycle current measurement range. | None |
| DCYC:CURR:RANG:AUTO <mode> | Set Duty Cycle Current Autorange. Disables or enables duty cycle autoranging via current path. | Text: 0|1|OFF|ON |
| DCYC:CURR:RANG:AUTO? | Read Duty Cycle Current Autorange. Returns the duty cycle current autoranging setting. | None |
| Command | What it does | Arguments |
|---|---|---|
| SOUR:SENS <mode> | Set Sense Source. Selects the sense signal source (external rear terminals or internal). | Text: EXTernal|INTernal |
| SOUR:SENS? | Read Sense Source. Returns the sense signal source. | None |
| SOUR:SENS:VOLT? | Read Sense Voltage. Returns the amplitude of the sensing voltage at the output. | None |
| SOUR:SENS:CURR? | Read Sense Current. Returns the amplitude of the sensing current at the output. | None |
| SOUR:VOLT:RANG <range> | Set Voltage Output Range. Sets the voltage output range; output must be off. | Text: 30 V|8 V|1 V|AUTO|MAX|MIN|DEF |
| SOUR:CURR:RANG <range> | Set Current Output Range. Sets the current output range; output must be off. | Text: 3 A|1 A|100 mA|AUTO|MAX|MIN|DEF |
| SOUR:SQU:RANG <range> | Set Square Wave Range. Sets the square-wave output range; output must be off. | Text: 30 V|8 V|AUTO|MAX|MIN|DEF |
| Command | What it does | Arguments |
|---|---|---|
| OUTP:STAT ON | Enable Output. Turns the source output on. | None |
| OUTP:STAT OFF | Disable Output. Turns the source output off (standby). | None |
| OUTP:STAT? | Read Output State. Returns the output state. | None |
| OUTP <state> | Set Output State. Enables or disables the instrument output. | Text: 0|1|OFF|ON |
| OUTP? | Query Output State. Returns 1 when the output is active and 0 when on standby. | None |
| OUTP ON | Output On. Turns the source output on. | None |
| OUTP OFF | Output Off. Turns the source output off. | None |
| Command | What it does | Arguments |
|---|---|---|
| LRES:RANG <range> | Set Low-Resistance Range. Selects a fixed 4-wire low-resistance range and disables autoranging. | Number: 0.1, 1, 10, 100, 1000 ohm; MIN, MAX, DEF |
| LRES:RANG? | Query Low-Resistance Range. Returns the 4-wire low-resistance range. | None |
| LRES:RANG:AUTO <mode> | Set Low-Resistance Autorange. Enables or disables 4-wire low-resistance autoranging. | Text: 0|1|OFF|ON |
| LRES:RANG:AUTO? | Query Low-Resistance Autorange. Returns the 4-wire low-resistance autoranging setting. | None |
| LRES:RES <resolution> | Set Low-Resistance Resolution. Sets the 4-wire low-resistance measurement resolution. | Number: 1e-6 to 1e-1 ohm; MIN, MAX, DEF |
| LRES:RES? | Query Low-Resistance Resolution. Returns the 4-wire low-resistance resolution. | None |
| LRES:COMP | Compensate Low-Resistance. Runs the environment compensation for 4-wire low-resistance measurements. | None |
| Command | What it does | Arguments |
|---|---|---|
| RES:RANG <range> | Set Resistance Range. Selects a fixed 2-wire resistance range and disables autoranging. | Number: 100, 1000, 10000, 100000, 1000000, 10000000, 100000000 ohm; MIN, MAX, DEF |
| RES:RANG? | Query Resistance Range. Returns the 2-wire resistance range. | None |
| RES:RANG:AUTO <mode> | Set Resistance Autorange. Enables or disables 2-wire resistance autoranging. | Text: 0|1|OFF|ON |
| RES:RANG:AUTO? | Query Resistance Autorange. Returns the 2-wire resistance autoranging setting. | None |
| RES:RES <resolution> | Set Resistance Resolution. Sets the 2-wire resistance measurement resolution. | Number: 0.001 to 10000 ohm; MIN, MAX, DEF |
| RES:RES? | Query Resistance Resolution. Returns the 2-wire resistance resolution. | None |
| Command | What it does | Arguments |
|---|---|---|
| INIT | Initiate Measurement. Places the instrument in the wait-for-trigger state. | None |
| INIT:CONT <mode> | Set Initiate Continuous. Enables or disables free-run continuous measurement mode. | Text: 0|1|OFF|ON |
| INIT:CONT? | Query Initiate Continuous. Returns the initiate continuous mode state. | None |
| INIT:CONT <state> | Set Continuous Initiation. Enables or disables free-run measurement mode. | Text: ON or OFF |
| Command | What it does | Arguments |
|---|---|---|
| CONT:RANG <range> | Set Continuity Range. Selects the fixed continuity test range. | Number: 100, 1000, 10000, 100000, 1000000, 10000000, 100000000 ohm; MIN, MAX, DEF |
| CONT:RANG? | Query Continuity Range. Returns the continuity test range. | None |
| CONT:RANG:AUTO <mode> | Set Continuity Autorange. Enables or disables continuity autoranging. | Text: 0|1|OFF|ON |
| CONT:RANG:AUTO? | Query Continuity Autorange. Returns the continuity autoranging setting. | None |
| Command | What it does | Arguments |
|---|---|---|
| CAP:RANG <range> | Set Capacitance Range. Selects a fixed capacitance range and disables autoranging. | Number: 1e-9, 1e-8, 1e-7, 1e-6, 1e-5, 1e-4, 1e-3, 1e-2 F; MIN, MAX, DEF |
| CAP:RANG? | Query Capacitance Range. Returns the capacitance measurement range. | None |
| CAP:RANG:AUTO <mode> | Set Capacitance Autorange. Enables or disables capacitance autoranging. | Text: 0|1|OFF|ON |
| CAP:RANG:AUTO? | Query Capacitance Autorange. Returns the capacitance autoranging setting. | None |
| Command | What it does | Arguments |
|---|---|---|
| LOG:STAT <mode> | Set Data Logging. Starts or stops the data logging operation. | Text: 0|1|OFF|ON |
| LOG:STAT? | Query Data Logging. Returns the data logging status. | None |
| LOG:DATA? | Read Logged Data. Returns the next stored logging value; END is returned when all data has been read. | None |
| Command | What it does | Arguments |
|---|---|---|
| SENS:VOLT:DC:RANG:AUTO <state> | Set DC Voltage Autorange (SENS:VOLT:DC:RANG:AUTO). Enables or disables autoranging for DC voltage. | Text: ON or OFF |
| SENS:VOLT:DC:RANG:UPP <range> | Set DC Voltage Range (SENS:VOLT:DC:RANG:UPP). Selects the fixed DC voltage measurement range. | Number: 0.02, 0.1, 1, 10, 100, 1000 V, or MIN, MAX, DEF |
| SENS:VOLT:DC:RES <resolution> | Set DC Voltage Resolution (SENS:VOLT:DC:RES). Sets the DC voltage measurement resolution. | Number: resolution in V, or MIN, MAX, DEF |
| Command | What it does | Arguments |
|---|---|---|
| PROT:STAT <state> | Set Output Protection State. Enables or disables the output circuit trip. | Text: 0|1|OFF|ON |
| PROT:STAT? | Read Output Protection State. Returns the output protection state. | None |
| Command | What it does | Arguments |
|---|---|---|
| SSTart:STEP <value> | Set Soft Start Step. Sets the soft start step for constant voltage or current output. | Number: 1 to 10000 |
| SSTart:STEP? | Read Soft Start Step. Returns the soft start step setting. | None |
| Command | What it does | Arguments |
|---|---|---|
| TRIG:SOUR <source> | Set Trigger Source. Selects the trigger source for measurements. | Text: IMMediate|BUS |
| TRIG:SOUR? | Read Trigger Source. Returns the selected trigger source. | None |
| Command | What it does | Arguments |
|---|---|---|
| FUNC "<function>" | Select Function. Selects the measurement function. | Text: CAPacitance, CONTinuity, CURRent:AC, CURRent:ACDC, CURRent[:DC], DIODe, FREQuency, PWIDth, DCYCle, RESistance, LRESistance, VOLTage:AC, VOLTage:ACDC, VOLTage[:DC] |
| FUNC? | Query Function. Returns the present measurement function short name. | None |
| Command | What it does | Arguments |
|---|---|---|
| ABOR | Abort Measurement. Aborts a measurement in progress. | None |
| Command | What it does | Arguments |
|---|---|---|
| FETC? | Fetch Reading. Transfers the latest reading from instrument memory to the output buffer. | None |
| Command | What it does | Arguments |
|---|---|---|
| READ? | Read Reading. Triggers a measurement and returns the reading (IMMediate trigger source only). | None |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Returns manufacturer, model, serial number and firmware revision. | None |
| *RST | Reset. Resets the instrument to its factory default state. | None |
| *CLS | Clear Status. Clears the event registers and the error queue. | None |
| *TST? | Self Test. Runs the self-test and returns 0 if all tests pass, 1 if any fail. | None |
| *OPC | Operation Complete. Sets the operation complete bit when pending operations finish. | None |
| *OPC? | Operation Complete Query. Returns 1 when the current operation completes. | None |
| *WAI | Wait To Continue. Waits until all pending operations complete before continuing. | None |
| *ESE <enable_value> | Set Standard Event Enable. Sets the standard event enable register mask. | Number: 0 to 255 |
| *ESE? | Read Standard Event Enable. Returns the standard event enable register mask. | None |
| *ESR? | Read Standard Event Register. Returns and clears the standard event register. | None |
| *SRE <value> | Set Service Request Enable. Sets the status byte service request enable mask. | Number: 0 to 255 |
| *SRE? | Read Service Request Enable. Returns the service request enable mask. | None |
| *STB? | Read Status Byte. Returns the status byte condition register. | None |
| *TRG | Bus Trigger. Triggers the instrument when TRIG:SOUR is BUS and it is waiting for a trigger. | None |
This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the U3606B and checked when the agent runs it. The library marks this instrument as supporting *OPC? for operation-complete synchronisation. It reads errors back with SYSTem:ERRor?. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated October 29, 2026.
Yes. The Keysight U3606B accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 222 SCPI commands for it: 112 queries that read values or state, and 110 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTP:STAT ON and MEAS:FREQ?.
MEAS:FREQ? (measure frequency). Configures a frequency measurement and returns one reading; the signal path must be set by a prior AC configuration. The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight U3606B through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.