Multimeter · SCPI command reference

Keysight U3606B SCPI commands

The Keysight U3606B multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 222 commands for it, 112 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the U3606B on a real bench.

Commands
222
Queries
112
Subsystems
27
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight U3606B and writes the report from the measured results.

Get started

Control the Keysight U3606B from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keysight U3606B's VISA address

# Paste the address of your Keysight U3606B (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("OUTP:STAT ON")  # Enable Output
print(inst.query("MEAS:FREQ?"))  # Measure Frequency

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight U3606B and writes the report from the measured results.

Get started

Run a U3606B test without writing code

Describe the test in a sentence, for example “enable output on the U3606B, then measure frequency and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

222 commands in 27 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Voltage (VOLTage) (34)

CommandWhat it doesArguments
VOLT:LIM <value>[,<save>]Set Over-Voltage Limit. Sets the over-voltage limit for constant current output.Number: 0 V to range maximum; Text: 0|1|OFF|ON, optional store flag
VOLT:LIM?Read Over-Voltage Limit. Returns the over-voltage limit of constant current output.None
VOLT:PROT <value>[,<save>]Set Over-Voltage Protection. Sets the over-voltage protection that disables output when exceeded.Number: 0 V to 33 V depending on range; Text: 0|1|OFF|ON, optional store flag
VOLT:PROT?Read Over-Voltage Protection. Returns the over-voltage protection value.None
VOLT <value>Set Voltage Level. Sets the constant voltage output amplitude.Number: 0 V to range maximum
VOLT?Read Voltage Level. Returns the constant voltage output amplitude.None
VOLT:RAMP <value>Set Voltage Ramp End. Sets the amplitude end position of the voltage ramp.Number: 0 V to range maximum
VOLT:RAMP?Read Voltage Ramp End. Returns the voltage ramp amplitude end position.None
VOLT:RAMP:STEP <value>Set Voltage Ramp Steps. Sets the number of steps in the voltage ramp.Number: 1 to 10000
VOLT:RAMP:STEP?Read Voltage Ramp Steps. Returns the number of steps in the voltage ramp.None
VOLT:SCAN <value>Set Voltage Scan End. Sets the amplitude end position of the voltage scan.Number: 0 V to range maximum
VOLT:SCAN?Read Voltage Scan End. Returns the voltage scan amplitude end position.None
VOLT:SCAN:STEP <value>Set Voltage Scan Steps. Sets the number of steps in the voltage scan.Number: 1 to 100
VOLT:SCAN:STEP?Read Voltage Scan Steps. Returns the number of steps in the voltage scan.None
VOLT:SCAN:DWEL <value>Set Voltage Scan Dwell. Sets the dwelling time of the voltage scan.Number: 1 to 99 (s)
VOLT:SCAN:DWEL?Read Voltage Scan Dwell. Returns the dwelling time of the voltage scan.None
VOLT:DC:RANG <range>Set DC Voltage Range. Selects a fixed DC voltage range and disables autoranging.Number: 0.02, 0.1, 1, 10, 100, 1000 V; MIN, MAX, DEF
VOLT:DC:RANG?Query DC Voltage Range. Returns the DC voltage measurement range.None
VOLT:DC:RANG:AUTO <mode>Set DC Voltage Autorange. Enables or disables DC voltage autoranging.Text: 0|1|OFF|ON
VOLT:DC:RANG:AUTO?Query DC Voltage Autorange. Returns the DC voltage autoranging setting.None
VOLT:DC:RES <resolution>Set DC Voltage Resolution. Sets the DC voltage measurement resolution.Number: 1e-7 to 1e-2 V; MIN, MAX, DEF
VOLT:DC:RES?Query DC Voltage Resolution. Returns the DC voltage resolution.None
VOLT:AC:RANG <range>Set AC Voltage Range. Selects a fixed AC voltage range and disables autoranging.Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF
VOLT:AC:RANG?Query AC Voltage Range. Returns the AC voltage measurement range.None
VOLT:AC:RANG:AUTO <mode>Set AC Voltage Autorange. Enables or disables AC voltage autoranging.Text: 0|1|OFF|ON
VOLT:AC:RANG:AUTO?Query AC Voltage Autorange. Returns the AC voltage autoranging setting.None
VOLT:AC:RES <resolution>Set AC Voltage Resolution. Sets the AC voltage measurement resolution.Number: 1e-6 to 1e-1 V; MIN, MAX, DEF
VOLT:AC:RES?Query AC Voltage Resolution. Returns the AC voltage resolution.None
VOLT:ACDC:RANG <range>Set AC+DC Voltage Range. Selects a fixed AC+DC voltage range and disables autoranging.Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF
VOLT:ACDC:RANG?Query AC+DC Voltage Range. Returns the AC+DC voltage measurement range.None
VOLT:ACDC:RANG:AUTO <mode>Set AC+DC Voltage Autorange. Enables or disables AC+DC voltage autoranging.Text: 0|1|OFF|ON
VOLT:ACDC:RANG:AUTO?Query AC+DC Voltage Autorange. Returns the AC+DC voltage autoranging setting.None
VOLT:ACDC:RES <resolution>Set AC+DC Voltage Resolution. Sets the AC+DC voltage measurement resolution.Number: 1e-6 to 1e-1 V; MIN, MAX, DEF
VOLT:ACDC:RES?Query AC+DC Voltage Resolution. Returns the AC+DC voltage resolution.None

Current (CURRent) (34)

CommandWhat it doesArguments
CURR:LIM <value>[,<save>]Set Over-Current Limit. Sets the over-current limit for constant voltage output.Number: 0 A to range maximum; Text: 0|1|OFF|ON, optional store flag
CURR:LIM?Read Over-Current Limit. Returns the over-current limit of constant voltage output.None
CURR:PROT <value>[,<save>]Set Over-Current Protection. Sets the over-current protection that disables output when exceeded.Number: 0 A to 3.3 A depending on range; Text: 0|1|OFF|ON, optional store flag
CURR:PROT?Read Over-Current Protection. Returns the over-current protection value.None
CURR <value>Set Current Level. Sets the constant current output amplitude.Number: 0 A to range maximum
CURR?Read Current Level. Returns the constant current output amplitude.None
CURR:RAMP <value>Set Current Ramp End. Sets the amplitude end position of the current ramp.Number: 0 A to range maximum
CURR:RAMP?Read Current Ramp End. Returns the current ramp amplitude end position.None
CURR:RAMP:STEP <value>Set Current Ramp Steps. Sets the number of steps in the current ramp.Number: 1 to 10000
CURR:RAMP:STEP?Read Current Ramp Steps. Returns the number of steps in the current ramp.None
CURR:SCAN <value>Set Current Scan End. Sets the amplitude end position of the current scan.Number: 0 A to range maximum
CURR:SCAN?Read Current Scan End. Returns the current scan amplitude end position.None
CURR:SCAN:STEP <value>Set Current Scan Steps. Sets the number of steps in the current scan.Number: 1 to 100
CURR:SCAN:STEP?Read Current Scan Steps. Returns the number of steps in the current scan.None
CURR:SCAN:DWEL <value>Set Current Scan Dwell. Sets the dwelling time of the current scan.Number: 1 to 99 (s)
CURR:SCAN:DWEL?Read Current Scan Dwell. Returns the dwelling time of the current scan.None
CURR:DC:RANG <range>Set DC Current Range. Selects a fixed DC current range and disables autoranging.Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF
CURR:DC:RANG?Query DC Current Range. Returns the DC current measurement range.None
CURR:DC:RANG:AUTO <mode>Set DC Current Autorange. Enables or disables DC current autoranging.Text: 0|1|OFF|ON
CURR:DC:RANG:AUTO?Query DC Current Autorange. Returns the DC current autoranging setting.None
CURR:DC:RES <resolution>Set DC Current Resolution. Sets the DC current measurement resolution.Number: 1e-7 to 1e-3 A; MIN, MAX, DEF
CURR:DC:RES?Query DC Current Resolution. Returns the DC current resolution.None
CURR:AC:RANG <range>Set AC Current Range. Selects a fixed AC current range and disables autoranging.Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF
CURR:AC:RANG?Query AC Current Range. Returns the AC current measurement range.None
CURR:AC:RANG:AUTO <mode>Set AC Current Autorange. Enables or disables AC current autoranging.Text: 0|1|OFF|ON
CURR:AC:RANG:AUTO?Query AC Current Autorange. Returns the AC current autoranging setting.None
CURR:AC:RES <resolution>Set AC Current Resolution. Sets the AC current measurement resolution.Number: 1e-7 to 1e-3 A; MIN, MAX, DEF
CURR:AC:RES?Query AC Current Resolution. Returns the AC current resolution.None
CURR:ACDC:RANG <range>Set AC+DC Current Range. Selects a fixed AC+DC current range and disables autoranging.Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF
CURR:ACDC:RANG?Query AC+DC Current Range. Returns the AC+DC current measurement range.None
CURR:ACDC:RANG:AUTO <mode>Set AC+DC Current Autorange. Enables or disables AC+DC current autoranging.Text: 0|1|OFF|ON
CURR:ACDC:RANG:AUTO?Query AC+DC Current Autorange. Returns the AC+DC current autoranging setting.None
CURR:ACDC:RES <resolution>Set AC+DC Current Resolution. Sets the AC+DC current measurement resolution.Number: 1e-7 to 1e-3 A; MIN, MAX, DEF
CURR:ACDC:RES?Query AC+DC Current Resolution. Returns the AC+DC current resolution.None

Measurement (MEASure) (19)

CommandWhat it doesArguments
MEAS:VOLT:DC? <range>,<resolution>Measure DC Voltage. Configures and returns one DC voltage reading.Number: 0.02, 0.1, 1, 10, 100, 1000 V; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-2 V; MIN, MAX, DEF
MEAS:VOLT:AC? <range>,<resolution>Measure AC Voltage. Configures and returns one AC voltage reading.Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 V; MIN, MAX, DEF
MEAS:VOLT:ACDC? <range>,<resolution>Measure AC+DC Voltage. Configures and returns one AC+DC voltage reading.Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 V; MIN, MAX, DEF
MEAS:CURR:DC? <range>,<resolution>Measure DC Current. Configures and returns one DC current reading.Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF
MEAS:CURR:AC? <range>,<resolution>Measure AC Current. Configures and returns one AC current reading.Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF
MEAS:CURR:ACDC? <range>,<resolution>Measure AC+DC Current. Configures and returns one AC+DC current reading.Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF
MEAS:RES? <range>,<resolution>Measure Resistance. Configures and returns one 2-wire resistance reading.Number: 100 to 100000000 ohm; MIN, MAX, DEF; omit for autorange; Number: 0.001 to 10000 ohm; MIN, MAX, DEF
MEAS:CONT?Measure Continuity. Configures a continuity test and returns the measured resistance.None
MEAS:LRES? <range>,<resolution>Measure Low Resistance. Configures and returns one 4-wire low-resistance reading.Number: 0.1 to 1000 ohm; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 ohm; MIN, MAX, DEF
MEAS:CAP? <range>,<resolution>Measure Capacitance. Configures and returns one capacitance reading.Number: 1e-9 to 1e-2 F; MIN, MAX, DEF; omit for autorange; Number: 1e-12 to 1e-6 F; MIN, MAX
MEAS:DIOD?Measure Diode. Configures a diode test and returns the measured voltage.None
MEAS:FREQ?Measure Frequency. Configures a frequency measurement and returns one reading; the signal path must be set by a prior AC configuration.None
MEAS:PWID?Measure Pulse Width. Configures a pulse width measurement and returns one reading.None
MEAS:DCYC?Measure Duty Cycle. Configures a duty cycle measurement and returns one reading in percent.None
MEAS:VOLT:DC?Measure DC Voltage (MEAS:VOLT:DC?). Configures and measures DC voltage, returning the reading.None
MEAS:VOLT:AC?Measure AC Voltage (MEAS:VOLT:AC?). Configures and measures AC voltage, returning the reading.None
MEAS:CURR:DC?Measure DC Current (MEAS:CURR:DC?). Configures and measures DC current, returning the reading.None
MEAS:CURR:AC?Measure AC Current (MEAS:CURR:AC?). Configures and measures AC current, returning the reading.None
MEAS:RES?Measure Resistance (MEAS:RES?). Configures and measures 2-wire resistance, returning the reading.None

Configure (CONFigure) (15)

CommandWhat it doesArguments
CONF:AC <range>,<resolution>Configure AC Voltage. Configures AC voltage measurements without starting a measurement.Number: 0.1, 1, 10, 100, 750 V; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 V; MIN, MAX, DEF
CONF:CURR:AC <range>,<resolution>Configure AC Current. Configures AC current measurements without starting a measurement.Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF
CONF:CURR:ACDC <range>,<resolution>Configure AC+DC Current. Configures AC+DC current measurements without starting a measurement.Number: 0.01, 0.1, 1, 3 A; MIN, MAX, DEF; omit for autorange; Number: 1e-7 to 1e-3 A; MIN, MAX, DEF
CONF:RES <range>,<resolution>Configure Resistance. Configures 2-wire resistance measurements without starting a measurement.Number: 100 to 100000000 ohm; MIN, MAX, DEF; omit for autorange; Number: 0.001 to 10000 ohm; MIN, MAX, DEF
CONF:CONT <range>,<resolution>Configure Continuity. Configures continuity tests without starting a test.Number: 100 to 100000000 ohm; MIN, MAX, DEF; Number: MAX, MIN
CONF:LRES <range>,<resolution>Configure Low Resistance. Configures 4-wire low-resistance measurements without starting a measurement.Number: 0.1 to 1000 ohm; MIN, MAX, DEF; omit for autorange; Number: 1e-6 to 1e-1 ohm; MIN, MAX, DEF
CONF:CAP <range>,<resolution>Configure Capacitance. Configures capacitance measurements without starting a measurement.Number: 1e-9 to 1e-2 F; MIN, MAX, DEF; omit for autorange; Number: 1e-12 to 1e-6 F; MIN, MAX
CONF:DIODConfigure Diode. Configures diode tests without starting a test.None
CONF:FREQConfigure Frequency. Configures frequency measurements on the path set by the preceding AC configuration.None
CONF:PWIDConfigure Pulse Width. Configures pulse width measurements on the path set by the preceding AC configuration.None
CONF:DCYCConfigure Duty Cycle. Configures duty cycle measurements on the path set by the preceding AC configuration.None
CONF?Get Configuration. Returns the present measurement function, range and resolution.None
CONF:VOLT:DC <range>,<resolution>Configure DC Voltage. Configures DC voltage measurement without starting a measurement.Number: range 0.02, 0.1, 1, 10, 100, 1000 V, or AUTO, MIN, MAX, DEF; Number: resolution in V, or MIN, MAX, DEF
CONF:VOLT:AC <range>,<resolution>Configure AC Voltage (CONF:VOLT:AC). Configures AC voltage measurement without starting a measurement.Number: range 0.1, 1, 10, 100, 750 V, or AUTO, MIN, MAX, DEF; Number: resolution in V, or MIN, MAX, DEF
CONF:CURR:DC <range>,<resolution>Configure DC Current. Configures DC current measurement without starting a measurement.Number: range 0.01, 0.1, 1, 3 A, or AUTO, MIN, MAX, DEF; Number: resolution in A, or MIN, MAX, DEF

System (SYSTem) (9)

CommandWhat it doesArguments
SYST:ERR?Read Error. Reads and clears one error from the error queue.None
SYST:PRESInstrument Preset. Resets the instrument to factory default state, same as *RST.None
SYST:VERS?Read SCPI Version. Returns the SCPI standard version the instrument complies with.None
SYST:SMO:STAT <state>Set Smooth Function. Enables or disables the smooth function.Text: 0|1|OFF|ON
SYST:SMO:STAT?Read Smooth Function. Returns the smooth function state.None
SYST:SMO:FLUC <value>Set Smooth Fluctuation. Sets the fluctuation count of the smooth function.Number: 0.0 to 9.9 (%)
SYST:SMO:FLUC?Read Smooth Fluctuation. Returns the smooth fluctuation count.None
SYST:SMO:POIN <value>Set Smooth Points. Sets the number of points for the smooth function.Number: 2 to 1999
SYST:SMO:POIN?Read Smooth Points. Returns the number of smooth function points.None

Status (STATus) (9)

CommandWhat it doesArguments
STAT:OPER:COND?Read Operation Condition. Returns the Standard Operation condition register.None
STAT:OPER:ENAB <enable_value>Set Operation Enable. Sets the Standard Operation enable register mask.Number: 0 to 65535
STAT:OPER:ENAB?Read Operation Enable. Returns the Standard Operation enable register mask.None
STAT:OPER:EVEN?Read Operation Event. Returns and clears the Standard Operation event register.None
STAT:PRESPreset Status. Clears the Questionable and Operation enable registers.None
STAT:QUES:COND?Read Questionable Condition. Returns the Questionable Data condition register.None
STAT:QUES:ENAB <enable_value>Set Questionable Enable. Sets the Questionable Data enable register mask.Number: 0 to 65535
STAT:QUES:ENAB?Read Questionable Enable. Returns the Questionable Data enable register mask.None
STAT:QUES:EVEN?Read Questionable Event. Returns and clears the Questionable Data event register.None

SQU subsystem (8)

CommandWhat it doesArguments
SQU:AMPL <value>Set Square Wave Amplitude. Sets the square-wave output amplitude.Number: 0 V to range maximum
SQU:AMPL?Read Square Wave Amplitude. Returns the square-wave output amplitude.None
SQU:FREQ <value>Set Square Wave Frequency. Sets the square-wave output frequency.Number: 0.5 to 4800 (Hz), predefined steps; S1S2 range 10 to 4800
SQU:FREQ?Read Square Wave Frequency. Returns the square-wave output frequency.None
SQU:DCYC <value>Set Square Wave Duty Cycle. Sets the square-wave duty cycle; not valid in S1S2 auto range.Number: 0 to 100 (%)
SQU:DCYC?Read Square Wave Duty Cycle. Returns the square-wave duty cycle.None
SQU:PWID <second>Set Square Wave Pulse Width. Sets the square-wave pulse width; not valid in S1S2 auto range.Number: 0 s to MAX (depends on frequency)
SQU:PWID?Read Square Wave Pulse Width. Returns the square-wave pulse width.None

Frequency (FREQuency) (8)

CommandWhat it doesArguments
FREQ:VOLT:RANG <range>Set Frequency Voltage Range. Selects the frequency measurement range via voltage path; disables autorange.Number: 0.1 V|1 V|10 V|100 V|750 V|MIN|MAX|DEF
FREQ:VOLT:RANG?Read Frequency Voltage Range. Returns the frequency voltage measurement range.None
FREQ:VOLT:RANG:AUTO <mode>Set Frequency Voltage Autorange. Disables or enables frequency autoranging via voltage path.Text: 0|1|OFF|ON
FREQ:VOLT:RANG:AUTO?Read Frequency Voltage Autorange. Returns the frequency voltage autoranging setting.None
FREQ:CURR:RANG <range>Set Frequency Current Range. Selects the frequency measurement range via current path; disables autorange.Number: 0.01 A|0.1 A|1 A|3 A|MIN|MAX|DEF
FREQ:CURR:RANG?Read Frequency Current Range. Returns the frequency current measurement range.None
FREQ:CURR:RANG:AUTO <mode>Set Frequency Current Autorange. Disables or enables frequency autoranging via current path.Text: 0|1|OFF|ON
FREQ:CURR:RANG:AUTO?Read Frequency Current Autorange. Returns the frequency current autoranging setting.None

PWID subsystem (8)

CommandWhat it doesArguments
PWID:VOLT:RANG <range>Set Pulse Width Voltage Range. Selects the pulse width measurement range via voltage path; disables autorange.Number: 0.1 V|1 V|10 V|100 V|750 V|MIN|MAX|DEF
PWID:VOLT:RANG?Read Pulse Width Voltage Range. Returns the pulse width voltage measurement range.None
PWID:VOLT:RANG:AUTO <mode>Set Pulse Width Voltage Autorange. Disables or enables pulse width autoranging via voltage path.Text: 0|1|OFF|ON
PWID:VOLT:RANG:AUTO?Read Pulse Width Voltage Autorange. Returns the pulse width voltage autoranging setting.None
PWID:CURR:RANG <range>Set Pulse Width Current Range. Selects the pulse width measurement range via current path; disables autorange.Number: 0.01 A|0.1 A|1 A|3 A|MIN|MAX|DEF
PWID:CURR:RANG?Read Pulse Width Current Range. Returns the pulse width current measurement range.None
PWID:CURR:RANG:AUTO <mode>Set Pulse Width Current Autorange. Disables or enables pulse width autoranging via current path.Text: 0|1|OFF|ON
PWID:CURR:RANG:AUTO?Read Pulse Width Current Autorange. Returns the pulse width current autoranging setting.None

DCYC subsystem (8)

CommandWhat it doesArguments
DCYC:VOLT:RANG <range>Set Duty Cycle Voltage Range. Selects the duty cycle measurement range via voltage path; disables autorange.Number: 0.1 V|1 V|10 V|100 V|750 V|MIN|MAX|DEF
DCYC:VOLT:RANG?Read Duty Cycle Voltage Range. Returns the duty cycle voltage measurement range.None
DCYC:VOLT:RANG:AUTO <mode>Set Duty Cycle Voltage Autorange. Disables or enables duty cycle autoranging via voltage path.Text: 0|1|OFF|ON
DCYC:VOLT:RANG:AUTO?Read Duty Cycle Voltage Autorange. Returns the duty cycle voltage autoranging setting.None
DCYC:CURR:RANG <range>Set Duty Cycle Current Range. Selects the duty cycle measurement range via current path; disables autorange.Number: 0.01 A|0.1 A|1 A|3 A|MIN|MAX|DEF
DCYC:CURR:RANG?Read Duty Cycle Current Range. Returns the duty cycle current measurement range.None
DCYC:CURR:RANG:AUTO <mode>Set Duty Cycle Current Autorange. Disables or enables duty cycle autoranging via current path.Text: 0|1|OFF|ON
DCYC:CURR:RANG:AUTO?Read Duty Cycle Current Autorange. Returns the duty cycle current autoranging setting.None

Source (SOURce) (7)

CommandWhat it doesArguments
SOUR:SENS <mode>Set Sense Source. Selects the sense signal source (external rear terminals or internal).Text: EXTernal|INTernal
SOUR:SENS?Read Sense Source. Returns the sense signal source.None
SOUR:SENS:VOLT?Read Sense Voltage. Returns the amplitude of the sensing voltage at the output.None
SOUR:SENS:CURR?Read Sense Current. Returns the amplitude of the sensing current at the output.None
SOUR:VOLT:RANG <range>Set Voltage Output Range. Sets the voltage output range; output must be off.Text: 30 V|8 V|1 V|AUTO|MAX|MIN|DEF
SOUR:CURR:RANG <range>Set Current Output Range. Sets the current output range; output must be off.Text: 3 A|1 A|100 mA|AUTO|MAX|MIN|DEF
SOUR:SQU:RANG <range>Set Square Wave Range. Sets the square-wave output range; output must be off.Text: 30 V|8 V|AUTO|MAX|MIN|DEF

Output (OUTPut) (7)

CommandWhat it doesArguments
OUTP:STAT ONEnable Output. Turns the source output on.None
OUTP:STAT OFFDisable Output. Turns the source output off (standby).None
OUTP:STAT?Read Output State. Returns the output state.None
OUTP <state>Set Output State. Enables or disables the instrument output.Text: 0|1|OFF|ON
OUTP?Query Output State. Returns 1 when the output is active and 0 when on standby.None
OUTP ONOutput On. Turns the source output on.None
OUTP OFFOutput Off. Turns the source output off.None

LRES subsystem (7)

CommandWhat it doesArguments
LRES:RANG <range>Set Low-Resistance Range. Selects a fixed 4-wire low-resistance range and disables autoranging.Number: 0.1, 1, 10, 100, 1000 ohm; MIN, MAX, DEF
LRES:RANG?Query Low-Resistance Range. Returns the 4-wire low-resistance range.None
LRES:RANG:AUTO <mode>Set Low-Resistance Autorange. Enables or disables 4-wire low-resistance autoranging.Text: 0|1|OFF|ON
LRES:RANG:AUTO?Query Low-Resistance Autorange. Returns the 4-wire low-resistance autoranging setting.None
LRES:RES <resolution>Set Low-Resistance Resolution. Sets the 4-wire low-resistance measurement resolution.Number: 1e-6 to 1e-1 ohm; MIN, MAX, DEF
LRES:RES?Query Low-Resistance Resolution. Returns the 4-wire low-resistance resolution.None
LRES:COMPCompensate Low-Resistance. Runs the environment compensation for 4-wire low-resistance measurements.None

RES subsystem (6)

CommandWhat it doesArguments
RES:RANG <range>Set Resistance Range. Selects a fixed 2-wire resistance range and disables autoranging.Number: 100, 1000, 10000, 100000, 1000000, 10000000, 100000000 ohm; MIN, MAX, DEF
RES:RANG?Query Resistance Range. Returns the 2-wire resistance range.None
RES:RANG:AUTO <mode>Set Resistance Autorange. Enables or disables 2-wire resistance autoranging.Text: 0|1|OFF|ON
RES:RANG:AUTO?Query Resistance Autorange. Returns the 2-wire resistance autoranging setting.None
RES:RES <resolution>Set Resistance Resolution. Sets the 2-wire resistance measurement resolution.Number: 0.001 to 10000 ohm; MIN, MAX, DEF
RES:RES?Query Resistance Resolution. Returns the 2-wire resistance resolution.None

Initiate (INITiate) (4)

CommandWhat it doesArguments
INITInitiate Measurement. Places the instrument in the wait-for-trigger state.None
INIT:CONT <mode>Set Initiate Continuous. Enables or disables free-run continuous measurement mode.Text: 0|1|OFF|ON
INIT:CONT?Query Initiate Continuous. Returns the initiate continuous mode state.None
INIT:CONT <state>Set Continuous Initiation. Enables or disables free-run measurement mode.Text: ON or OFF

CONT subsystem (4)

CommandWhat it doesArguments
CONT:RANG <range>Set Continuity Range. Selects the fixed continuity test range.Number: 100, 1000, 10000, 100000, 1000000, 10000000, 100000000 ohm; MIN, MAX, DEF
CONT:RANG?Query Continuity Range. Returns the continuity test range.None
CONT:RANG:AUTO <mode>Set Continuity Autorange. Enables or disables continuity autoranging.Text: 0|1|OFF|ON
CONT:RANG:AUTO?Query Continuity Autorange. Returns the continuity autoranging setting.None

CAP subsystem (4)

CommandWhat it doesArguments
CAP:RANG <range>Set Capacitance Range. Selects a fixed capacitance range and disables autoranging.Number: 1e-9, 1e-8, 1e-7, 1e-6, 1e-5, 1e-4, 1e-3, 1e-2 F; MIN, MAX, DEF
CAP:RANG?Query Capacitance Range. Returns the capacitance measurement range.None
CAP:RANG:AUTO <mode>Set Capacitance Autorange. Enables or disables capacitance autoranging.Text: 0|1|OFF|ON
CAP:RANG:AUTO?Query Capacitance Autorange. Returns the capacitance autoranging setting.None

LOG subsystem (3)

CommandWhat it doesArguments
LOG:STAT <mode>Set Data Logging. Starts or stops the data logging operation.Text: 0|1|OFF|ON
LOG:STAT?Query Data Logging. Returns the data logging status.None
LOG:DATA?Read Logged Data. Returns the next stored logging value; END is returned when all data has been read.None

Sense (SENSe) (3)

CommandWhat it doesArguments
SENS:VOLT:DC:RANG:AUTO <state>Set DC Voltage Autorange (SENS:VOLT:DC:RANG:AUTO). Enables or disables autoranging for DC voltage.Text: ON or OFF
SENS:VOLT:DC:RANG:UPP <range>Set DC Voltage Range (SENS:VOLT:DC:RANG:UPP). Selects the fixed DC voltage measurement range.Number: 0.02, 0.1, 1, 10, 100, 1000 V, or MIN, MAX, DEF
SENS:VOLT:DC:RES <resolution>Set DC Voltage Resolution (SENS:VOLT:DC:RES). Sets the DC voltage measurement resolution.Number: resolution in V, or MIN, MAX, DEF

PROT subsystem (2)

CommandWhat it doesArguments
PROT:STAT <state>Set Output Protection State. Enables or disables the output circuit trip.Text: 0|1|OFF|ON
PROT:STAT?Read Output Protection State. Returns the output protection state.None

SST subsystem (2)

CommandWhat it doesArguments
SSTart:STEP <value>Set Soft Start Step. Sets the soft start step for constant voltage or current output.Number: 1 to 10000
SSTart:STEP?Read Soft Start Step. Returns the soft start step setting.None

Trigger (TRIGger) (2)

CommandWhat it doesArguments
TRIG:SOUR <source>Set Trigger Source. Selects the trigger source for measurements.Text: IMMediate|BUS
TRIG:SOUR?Read Trigger Source. Returns the selected trigger source.None

Function (FUNCtion) (2)

CommandWhat it doesArguments
FUNC "<function>"Select Function. Selects the measurement function.Text: CAPacitance, CONTinuity, CURRent:AC, CURRent:ACDC, CURRent[:DC], DIODe, FREQuency, PWIDth, DCYCle, RESistance, LRESistance, VOLTage:AC, VOLTage:ACDC, VOLTage[:DC]
FUNC?Query Function. Returns the present measurement function short name.None

Abort (1)

CommandWhat it doesArguments
ABORAbort Measurement. Aborts a measurement in progress.None

Fetch (FETCh) (1)

CommandWhat it doesArguments
FETC?Fetch Reading. Transfers the latest reading from instrument memory to the output buffer.None

Read (READ) (1)

CommandWhat it doesArguments
READ?Read Reading. Triggers a measurement and returns the reading (IMMediate trigger source only).None

IEEE 488.2 common commands (14)

CommandWhat it doesArguments
*IDN?Identify. Returns manufacturer, model, serial number and firmware revision.None
*RSTReset. Resets the instrument to its factory default state.None
*CLSClear Status. Clears the event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 if all tests pass, 1 if any fail.None
*OPCOperation Complete. Sets the operation complete bit when pending operations finish.None
*OPC?Operation Complete Query. Returns 1 when the current operation completes.None
*WAIWait To Continue. Waits until all pending operations complete before continuing.None
*ESE <enable_value>Set Standard Event Enable. Sets the standard event enable register mask.Number: 0 to 255
*ESE?Read Standard Event Enable. Returns the standard event enable register mask.None
*ESR?Read Standard Event Register. Returns and clears the standard event register.None
*SRE <value>Set Service Request Enable. Sets the status byte service request enable mask.Number: 0 to 255
*SRE?Read Service Request Enable. Returns the service request enable mask.None
*STB?Read Status Byte. Returns the status byte condition register.None
*TRGBus Trigger. Triggers the instrument when TRIG:SOUR is BUS and it is waiting for a trigger.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the U3606B and checked when the agent runs it. The library marks this instrument as supporting *OPC? for operation-complete synchronisation. It reads errors back with SYSTem:ERRor?. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated October 29, 2026.

Frequently asked questions

Does the Keysight U3606B support SCPI remote control?

Yes. The Keysight U3606B accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 222 SCPI commands for it: 112 queries that read values or state, and 110 settings and actions.

How do I control the Keysight U3606B from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTP:STAT ON and MEAS:FREQ?.

Which command reads a value from the Keysight U3606B?

MEAS:FREQ? (measure frequency). Configures a frequency measurement and returns one reading; the signal path must be set by a prior AC configuration. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Keysight U3606B without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight U3606B through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.