The inTEST ATS-545 and ATS-645 thermal forcing system is controlled remotely with SCPI commands. TestFlow's instrument library holds 72 commands for it, 34 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the ATS-545 on a real bench.
Commands
72
Queries
34
Subsystems
37
Models covered
2
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your inTEST ATS-545 or ATS-645 and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
Describe the test in a sentence, for example “select setpoint on the ATS-545, then read selected setpoint and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
72 commands in 37 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The ATS-545 and ATS-645 share this command set in TestFlow's library.
Other subsystem (4)
Command
What it does
Arguments
%RM
Remote Mode. Enters remote mode (front panel shows a Return to Local button).
None
%LL
Local Lockout. Enters local lockout (front panel disabled).
None
%GL
Go To Local. Returns the system to local (front panel) control.
None
<Output Variable>=<Equation>
Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)
Number; Text
FLOW subsystem (3)
Command
What it does
Arguments
FLOW 1
Turn On Flow. Turns on the main nozzle air flow.
None
FLOW 0
Turn Off Flow. Turns off the main nozzle air flow.
None
FLOW?
Read Flow State. Reads the main air flow state (1 on or 0 off).
None
HEAD subsystem (3)
Command
What it does
Arguments
HEAD 1
Head Down. Lowers the test head (ignored if the head is locked).
None
HEAD 0
Head Up. Raises the test head (ignored if the head is locked).
None
HEAD?
Read Head State. Reads the head position (1 down or 0 up).
None
COOL subsystem (3)
Command
What it does
Arguments
COOL 1
Turn On Compressor. Turns on the compressor (about 60 s before cooling is ready).
None
COOL 0
Turn Off Compressor. Turns off the compressor.
None
COOL?
Read Compressor State. Reads the compressor state (1 on or 0 off).
None
DUTM subsystem (3)
Command
What it does
Arguments
DUTM 1
Control On DUT. Switches to DUT control mode (temperature controlled on the DUT sensor).
None
DUTM 0
Control On Air. Switches to air control mode (temperature controlled on the air sensor).
None
DUTM?
Read Control Mode. Reads the control mode (1 DUT or 0 air).
None
LRNM subsystem (3)
Command
What it does
Arguments
LRNM 1
Turn On Learn Mode. Turns on DUT automatic tuning (learning mode).
None
LRNM 0
Turn Off Learn Mode. Turns off DUT automatic tuning (learning mode).
None
LRNM?
Read Learn Mode. Reads the DUT learning mode state (1 on or 0 off).
None
CYCL subsystem (3)
Command
What it does
Arguments
CYCL 1
Start Cycling. Starts temperature cycling through the enabled setpoints.
None
CYCL 0
Stop Cycling. Stops cycling (also required after end of cycling or stop-on-fail).
None
CYCL?
Read Cycle Number. Reads the current cycle number.
None
SETN subsystem (2)
Command
What it does
Arguments
SETN <Setpoint_Num>
Select Setpoint. Selects the active setpoint: 0 hot, 1 ambient, 2 cold on the Operator screen (0-17 on the Cycle screen).
Number: 0,17
SETN?
Read Selected Setpoint. Reads the number of the currently selected setpoint.
None
SETP subsystem (2)
Command
What it does
Arguments
SETP <Temp>
Set Setpoint. Sets the temperature of the selected setpoint in °C (must be within LLIM/ULIM).
Number: -99.9,225.0
SETP?
Read Setpoint. Reads the temperature of the selected setpoint in °C.
None
RAMP subsystem (2)
Command
What it does
Arguments
RAMP <Rate>
Set Ramp Rate. Sets the ramp rate of the selected setpoint in °C/min (0-99.9 in 0.1 steps or 100-9999).
Number: 0,9999
RAMP?
Read Ramp Rate. Reads the ramp rate of the selected setpoint in °C/min.
None
SOAK subsystem (2)
Command
What it does
Arguments
SOAK <Seconds>
Set Soak Time. Sets the soak time of the selected setpoint in seconds (0-9999).
Number: 0,9999
SOAK?
Read Soak Time. Reads the soak time of the selected setpoint in seconds.
None
WNDW subsystem (2)
Command
What it does
Arguments
WNDW <Window>
Set Window. Sets the at-temperature window (±°C) of the selected setpoint (0.1-9.9).
Number: 0.1,9.9
WNDW?
Read Window. Reads the at-temperature window of the selected setpoint.
None
ULIM subsystem (2)
Command
What it does
Arguments
ULIM <Temp>
Set Upper Limit. Sets the upper air temperature limit in °C.
Number: 25,225
ULIM?
Read Upper Limit. Reads the upper air temperature limit in °C.
None
LLIM subsystem (2)
Command
What it does
Arguments
LLIM <Temp>
Set Lower Limit. Sets the lower air temperature limit in °C.
Number: -80,25
LLIM?
Read Lower Limit. Reads the lower air temperature limit in °C.
None
ADMD subsystem (2)
Command
What it does
Arguments
ADMD <Temp>
Set Max Air DUT Difference. Sets the maximum allowed air-to-DUT temperature difference in °C (10-300).
Number: 10,300
ADMD?
Read Max Air DUT Difference. Reads the maximum allowed air-to-DUT temperature difference in °C.
None
DSNS subsystem (2)
Command
What it does
Arguments
DSNS <Type>
Set DUT Sensor Type. Sets the DUT sensor type (0 none, 1 T-type thermocouple, 2 K-type thermocouple).
Number: 0,1,2
DSNS?
Read DUT Sensor Type. Reads the DUT sensor type.
None
DUTC subsystem (2)
Command
What it does
Arguments
DUTC <Constant>
Set DUT Thermal Constant. Sets the DUT thermal constant (20-500; higher values for larger mass and less overshoot).
Number: 20,500
DUTC?
Read DUT Thermal Constant. Reads the DUT thermal constant.
None
TTIM subsystem (2)
Command
What it does
Arguments
TTIM <Seconds>
Set Test Time. Sets the maximum test time per setpoint in seconds (0-9999).
Number: 0,9999
TTIM?
Read Test Time. Reads the maximum test time per setpoint in seconds.
None
CYCC subsystem (2)
Command
What it does
Arguments
CYCC <Count>
Set Cycle Count. Sets the number of temperature cycles to run (1-9999).
Number: 1,9999
CYCC?
Read Cycle Count. Reads the configured cycle count.
None
SFIL subsystem (2)
Command
What it does
Arguments
SFIL <File>
Load Setup File. Loads a stored setup file (1-12); wait 2 s after.
Number: 1,12
SFIL?
Read Setup File. Reads the number of the loaded setup file.
None
SETD subsystem (1)
Command
What it does
Arguments
SETD?
Read Dynamic Setpoint. Reads the live (ramping) setpoint in °C.
None
TEMP subsystem (1)
Command
What it does
Arguments
TEMP?
Read Temperature. Reads the main control temperature in °C (air in air mode or DUT in DUT mode).
None
TMPA subsystem (1)
Command
What it does
Arguments
TMPA?
Read Air Temperature. Reads the air temperature in °C regardless of control mode.
None
TMPD subsystem (1)
Command
What it does
Arguments
TMPD?
Read DUT Temperature. Reads the DUT temperature in °C regardless of control mode.
None
FLWR subsystem (1)
Command
What it does
Arguments
FLWR?
Read Flow Rate SCFM. Reads the measured main air flow rate in scfm.
None
FLRL subsystem (1)
Command
What it does
Arguments
FLRL?
Read Flow Rate LPS. Reads the measured main air flow rate in liters per second.
None
RSTO subsystem (1)
Command
What it does
Arguments
RSTO
Reset Operator. Forces the Operator screen and resets errors with setpoint 1 (ambient) active; wait 4 s after.
None
CLER subsystem (1)
Command
What it does
Arguments
CLER
Clear Errors. Clears the device error register (EROR); wait 4 s after.
None
TECR subsystem (1)
Command
What it does
Arguments
TECR?
Read Temp Condition. Reads the temperature condition register (b0 at temp, b1 not at temp, b2 end of test, b3 end of cycle, b4 end of all cycles, b5 stopped, b7 datalog).
None
TESR subsystem (1)
Command
What it does
Arguments
TESR?
Read Temp Event. Reads the latched temperature event register (same bits as TECR; cleared when read).
None
EROR subsystem (1)
Command
What it does
Arguments
EROR?
Read Error Register. Reads the 16-bit device error register (overheat, low flow, low air pressure, sensor open etc.).
None
AUXC subsystem (1)
Command
What it does
Arguments
AUXC?
Read Aux Condition. Reads the auxiliary condition register (head, compressor, DUT mode, flow, ready, manual/program, ramp mode).
None
WHAT subsystem (1)
Command
What it does
Arguments
WHAT?
Read System State. Reads what the system is currently doing (operator/manual or cycle/program).
None
CFIL subsystem (1)
Command
What it does
Arguments
CFIL <File>
Copy Setup File. Copies the active setup to setup file n (1-12).
Number: 1,12
WAIT subsystem (1)
Command
What it does
Arguments
wait(<ms>)
Wait. Waits for number of milliseconds
Number
SAVE subsystem (1)
Command
What it does
Arguments
save2var <Variable_name>
Save2Variable. Saves last measurement to a variable
Text
IEEE 488.2 common commands (8)
Command
What it does
Arguments
*IDN?
Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).
None
*RST
Reset. Resets to the Cycle screen with errors and limits reset and setpoint 5 active; wait 4 s after.
None
*CLS
Clear Status. Clears the ESR and TESR status registers.
Read Event Status. Reads and clears the standard event status register (command, execution, device and query errors).
None
*ESE <Mask>
Set Event Status Enable. Sets the standard event status enable mask (0-255).
Number: 0,255
*STB?
Read Status Byte. Reads the status byte (ready, MSS, ESB, MAV, TESR summary, EROR summary).
None
*SRE <Mask>
Set Service Request Enable. Sets the service request enable mask (0-255).
Number: 0,255
Where these commands come from
This reference is generated from TestFlow's instrument library, which is drafted from inTEST's programming documentation for the ATS-545 and ATS-645 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by inTEST's programming manual: check it before you rely on an edge case. Last updated September 23, 2026.
Frequently asked questions
Does the inTEST ATS-545 support SCPI remote control?
Yes. The inTEST ATS-545 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 72 SCPI commands for it: 34 queries that read values or state, and 38 settings and actions.
How do I control the inTEST ATS-545 from Python?
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, SETN 0 and SETN?.
Which command reads a value from the inTEST ATS-545?
SETN? (read selected setpoint). Reads the number of the currently selected setpoint. The full list of queries is in the command reference on this page, grouped by subsystem.