Function Generator · SCPI command comparison

Keithley 3390 vs Keysight 33511B: SCPI commands and remote control compared

In TestFlow's instrument library, the Keithley 3390 takes 88 SCPI commands, 10 of them queries, and the Keysight 33511B takes 117 SCPI commands, 17 of them queries. That is 29 more on the 33511B. Counting distinct command headers (the command path without its arguments), 18 are accepted by both, 56 exist only on the 3390, and 83 only on the 33511B. 24% of the 3390's command headers also exist on the 33511B.

Both libraries have commands under OUTPut, SYSTem, and TRIGger. Only the 3390 has VOLTage, FREQuency, APPLy, BURSt, FUNCtion, and AM and 6 more. Only the 33511B has SOURce, DISPlay, ROSCillator, and MMEMory.

Both libraries synchronise with *OPC?. Both libraries read errors with SYSTem:ERRor?.

TestFlow's library gives the 33519B, 33521B, 33611A, and 33621A the same command set as the 33511B, so the 33511B column applies to them too.

3390 commands
88
33511B commands
117
Shared headers
18
Unique headers
139

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Keithley 3390 or Keysight 33511B and writes the report from the measured results.

Get started

3390 vs 33511B at a glance

Counts come from TestFlow's instrument library, the command sets its AI agent uses to drive each instrument. A distinct command header is the command path without its arguments, so :OUTPut ON and :OUTPut OFF count once.

Keithley 3390Keysight 33511B
Commands in TestFlow's library88117
Queries (read a value or state)1017
Settings and actions78100
Subsystems1911
Distinct command headers74101
Headers the other lacks5683
Command languageSCPISCPI
Operation complete*OPC?*OPC?
Error queueSYSTem:ERRor?SYSTem:ERRor?
Models on this command set339033511B, 33519B, 33521B, 33611A, 33621A

Commands both accept

18 command headers are in both libraries. Here are 12, taken across subsystems. Where the 33511B library writes the command differently, its form is shown under the 3390's.

CommandWhat it does
OUTPut:LOAD <Ohm>33511B: OUTPut1:LOAD <Ohm>Set Output Load. Sets the output load in ohms (1 to 10k); use INF for high-Z.
SYSTem:BEEPer:STATe ONEnable Beeper. Enables the beeper.
TRIGger:SOURce <Src>Set Trigger Source. Sets the trigger source.
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).
OUTPut:POLarity NORMal33511B: OUTPut1:POLarity NORMalSet Output Polarity Normal. Sets output polarity to normal.
SYSTem:BEEPerBeep Now. Sounds the beeper once.
TRIGger:SLOPe <Slope>Set Trigger Slope. Sets the external trigger slope.
*RSTReset. Resets the instrument to its default state.
SYSTem:ERRor?Read Error. Reads and clears the next error-queue entry.
TRIGgerTrigger Now. Issues an immediate trigger.
*CLSClear Status. Clears all event registers and the error queue.
SYSTem:VERSion?Read SCPI Version. Returns the SCPI version.

Only on the Keithley 3390

56 command headers in the 3390 library have no match in the 33511B library. Below are 24 of them from the 6 largest subsystems; the full list is in the Keithley 3390 command reference.

VOLTage (7 only on the 3390)

CommandWhat it does
VOLTage <Amp>Set Amplitude. Sets the output amplitude (Vpp).
VOLTage?Read Amplitude. Reads the output amplitude.
VOLTage:OFFSet <Volt>Set Voltage Offset. Sets the DC offset voltage.
VOLTage:OFFSet?Read Voltage Offset. Reads the DC offset voltage.

APPLy (6 only on the 3390)

CommandWhat it does
APPLy:SINusoid <Freq>,<Amp>,<Offset>Apply Sine. Sets sine with the given frequency, amplitude and offset.
APPLy:SQUare <Freq>,<Amp>,<Offset>Apply Square. Sets square with the given frequency, amplitude and offset.
APPLy:RAMP <Freq>,<Amp>,<Offset>Apply Ramp. Sets ramp with the given frequency, amplitude and offset.
APPLy:PULSe <Freq>,<Amp>,<Offset>Apply Pulse. Sets pulse with the given frequency, amplitude and offset.

FREQuency (6 only on the 3390)

CommandWhat it does
FREQuency <Freq>Set Frequency. Sets the output frequency in Hz.
FREQuency?Read Frequency. Reads the output frequency.
FREQuency:STARt <Freq>Set Sweep Start Freq. Sets the sweep start frequency.
FREQuency:STOP <Freq>Set Sweep Stop Freq. Sets the sweep stop frequency.

BURSt (5 only on the 3390)

CommandWhat it does
BURSt:STATe ONEnable Burst. Enables burst mode.
BURSt:MODE <Mode>Set Burst Mode. Sets the burst mode.
BURSt:NCYCles <N>Set Burst Cycles. Sets the number of cycles per burst.
BURSt:INTernal:PERiod <Sec>Set Burst Period. Sets the internal burst period.

FUNCtion (5 only on the 3390)

CommandWhat it does
FUNCtion <Shape>Set Function. Sets the output waveform shape.
FUNCtion?Read Function. Reads the output waveform shape.
FUNCtion:RAMP:SYMMetry <Pct>Set Ramp Symmetry. Sets the ramp symmetry percentage.
FUNCtion:SQUare:DCYCle <Pct>Set Square Duty Cycle. Sets the square-wave duty cycle percentage.

AM (4 only on the 3390)

CommandWhat it does
AM:STATe ONEnable AM. Enables AM modulation.
AM:DEPTh <Pct>Set AM Deviation. Sets the AM depth percentage.
AM:SOURce <Src>Set AM Source. Sets the AM modulation source.
AM:INTernal:FREQuency <Freq>Set AM Int Freq. Sets the internal AM modulating frequency.

Only on the Keysight 33511B

83 command headers in the 33511B library have no match in the 3390 library. Below are 13 of them from the 6 largest subsystems; the full list is in the Keysight 33511B command reference.

SOURce (72 only on the 33511B)

CommandWhat it does
SOURce1:FUNCtion <Shape>Set Function CH1. Sets the output waveform shape for channel 1.
SOURce1:FUNCtion?Read Function CH1. Reads the output waveform shape for channel 1.
SOURce1:FREQuency <Freq>Set Frequency CH1. Sets the output frequency in Hz for channel 1.
SOURce1:FREQuency?Read Frequency CH1. Reads the output frequency for channel 1.

OUTPut (6 only on the 33511B)

CommandWhat it does
OUTPut1:STATe ONTurn On Output CH1. Enables channel 1 output.
OUTPut1:STATe?Read Output State CH1. Reads whether channel 1 output is on.
OUTPut1:LOAD?Read Output Load CH1. Reads the output load setting for channel 1.
OUTPut1:POLarity?Read Output Polarity CH1. Reads channel 1 output polarity.

MMEMory (2 only on the 33511B)

CommandWhat it does
MMEMory:STORe:STATe <File>Save State. Stores the instrument state to a file.
MMEMory:LOAD:STATe <File>Load State. Loads an instrument state file.

DISPlay (1 only on the 33511B)

CommandWhat it does
DISPlay ONEnable Display. Turns the front-panel display on.

IEEE 488.2 common (*) (1 only on the 33511B)

CommandWhat it does
*OPT?Read Options. Queries the installed options.

ROSCillator (1 only on the 33511B)

CommandWhat it does
ROSCillator:SOURce INTernalSet Clock Internal. Uses the internal 10 MHz reference clock.

Control both from Python

With PyVISA and any VISA backend, this opens both instruments, identifies them, sends one setting, and reads one value from each. For the setting, both libraries hold OUTPut:POLarity (set output polarity normal); the 33511B library writes it as OUTPut1:POLarity NORMal. For the reading, both libraries hold SYSTem:ERRor? (read error), and the same line runs on each. For the basics, see the PyVISA tutorial.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find both VISA addresses

a = rm.open_resource("PASTE_3390_ADDRESS")  # Keithley 3390
b = rm.open_resource("PASTE_33511B_ADDRESS")  # Keysight 33511B
for inst in (a, b):
    inst.timeout = 5000  # ms
    print(inst.query("*IDN?"))

# The same setting on both: Set Output Polarity Normal
a.write("OUTPut:POLarity NORMal")
b.write("OUTPut1:POLarity NORMal")

# The same reading on both: Read Error
print(a.query("SYSTem:ERRor?"))
print(b.query("SYSTem:ERRor?"))

a.close()
b.close()

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Keithley 3390 or Keysight 33511B and writes the report from the measured results.

Get started

Frequently asked questions

What is the difference between the Keithley 3390 and the Keysight 33511B for remote control?

In TestFlow's library the 3390 has 88 SCPI commands and the 33511B has 117 SCPI commands. 18 command headers are common to both, 56 are only on the 3390 (mostly VOLTage and APPLy), and 83 are only on the 33511B (mostly SOURce and OUTPut).

Can I use the same SCPI code for the Keithley 3390 and the Keysight 33511B?

Partly. 18 command headers are accepted by both, including OUTPut:LOAD, SYSTem:BEEPer:STATe, and TRIGger:SOURce. 139 are not, for example VOLTage on the 3390 and SOURce1:FUNCtion on the 33511B, so a script that uses those needs a separate branch for each instrument.

Which has more remote commands, the 3390 or the 33511B?

The 33511B: 88 commands for the 3390 against 117 for the 33511B in TestFlow's library. Of those, 10 on the 3390 and 17 on the 33511B are queries that read a value or state, and the library splits them into 19 and 11 subsystems.

How do the 3390 and 33511B signal operation complete and report errors?

Both libraries synchronise with *OPC?. Both libraries read errors with SYSTem:ERRor?. Each command named here is listed in the tables on this page and on each instrument's command reference.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.