In TestFlow's instrument library, the inTEST ATS-710E takes 72 SCPI commands, 34 of them queries, and the SP Scientific THJ80 takes 62 SCPI commands, 31 of them queries. That is 10 more on the ATS-710E. Counting distinct command headers (the command path without its arguments), 8 are accepted by both, 55 exist only on the ATS-710E, and 51 only on the THJ80. 13% of the ATS-710E's command headers also exist on the THJ80.
Only the ATS-710E has %RM, %LL, %GL, FLOW, HEAD, and COOL and 30 more. Only the THJ80 has SP, PV, TA, TD, AF, and SK and 38 more.
Only the THJ80 library has an operation-complete command (*OPC?); the ATS-710E library has none. Neither library has an error-queue read such as SYSTem:ERRor?.
TestFlow's library gives the ATS-730E, ATS-750E, ATS-770E, ATS-810E, ATS-830E, ATS-850E, and ATS-870E the same command set as the ATS-710E, so the ATS-710E column applies to them too. TestFlow's library gives the ThermoJet ES the same command set as the THJ80, so the THJ80 column applies to it too.
Describe the test; TestFlow's agent uses these command sets to drive your inTEST ATS-710E or SP Scientific THJ80 and writes the report from the measured results.
Counts come from TestFlow's instrument library, the command sets its AI agent uses to drive each instrument. A distinct command header is the command path without its arguments, so :OUTPut ON and :OUTPut OFF count once.
| inTEST ATS-710E | SP Scientific THJ80 | |
|---|---|---|
| Commands in TestFlow's library | 72 | 62 |
| Queries (read a value or state) | 34 | 31 |
| Settings and actions | 38 | 31 |
| Subsystems | 37 | 25 |
| Distinct command headers | 63 | 59 |
| Headers the other lacks | 55 | 51 |
| Command language | SCPI | SCPI |
| Operation complete | Not in library | *OPC? |
| Error queue | Not in library | Not in library |
| Models on this command set | ATS-710E, ATS-730E, ATS-750E, ATS-770E, ATS-810E, ATS-830E, ATS-850E, ATS-870E | THJ80, ThermoJet ES |
8 command headers are in both libraries.
| Command | What it does |
|---|---|
| *IDN? | Identify. Returns the instrument identification string (manufacturer, model, serial, firmware). |
| *RST | Reset. Resets to the Cycle screen with errors and limits reset and setpoint 5 active; wait 4 s after. |
| *CLS | Clear Status. Clears the ESR and TESR status registers. |
| *TST? | Self Test. Self-test query (dummy; always returns 0). |
| *ESR? | Read Event Status. Reads and clears the standard event status register (command, execution, device and query errors). |
| *ESE <Mask> | Set Event Status Enable. Sets the standard event status enable mask (0-255). |
| *STB? | Read Status Byte. Reads the status byte (ready, MSS, ESB, MAV, TESR summary, EROR summary). |
| *SRE <Mask> | Set Service Request Enable. Sets the service request enable mask (0-255). |
55 command headers in the ATS-710E library have no match in the THJ80 library. Below are 12 of them from the 6 largest subsystems; the full list is in the inTEST ATS-710E command reference.
| Command | What it does |
|---|---|
| ADMD <Temp> | Set Max Air DUT Difference. Sets the maximum allowed air-to-DUT temperature difference in °C (10-300). |
| ADMD? | Read Max Air DUT Difference. Reads the maximum allowed air-to-DUT temperature difference in °C. |
| Command | What it does |
|---|---|
| COOL 1 | Turn On Compressor. Turns on the compressor (about 60 s before cooling is ready). |
| COOL? | Read Compressor State. Reads the compressor state (1 on or 0 off). |
| Command | What it does |
|---|---|
| CYCC <Count> | Set Cycle Count. Sets the number of temperature cycles to run (1-9999). |
| CYCC? | Read Cycle Count. Reads the configured cycle count. |
| Command | What it does |
|---|---|
| CYCL 1 | Start Cycling. Starts temperature cycling through the enabled setpoints. |
| CYCL? | Read Cycle Number. Reads the current cycle number. |
| Command | What it does |
|---|---|
| DSNS <Type> | Set DUT Sensor Type. Sets the DUT sensor type (0 none, 1 T-type thermocouple, 2 K-type thermocouple). |
| DSNS? | Read DUT Sensor Type. Reads the DUT sensor type. |
| Command | What it does |
|---|---|
| DUTC <Constant> | Set DUT Thermal Constant. Sets the DUT thermal constant (20-500; higher values for larger mass and less overshoot). |
| DUTC? | Read DUT Thermal Constant. Reads the DUT thermal constant. |
51 command headers in the THJ80 library have no match in the ATS-710E library. Below are 12 of them from the 6 largest subsystems; the full list is in the SP Scientific THJ80 command reference.
| Command | What it does |
|---|---|
| ?AF | Read Airflow Setpoint. Reads the airflow setpoint. |
| AF<Flow> | Set Airflow. Changes the airflow setpoint in SCFM (5-20). |
| Command | What it does |
|---|---|
| ?CY | Read Program Cycles. Reads the number of program cycles. |
| CY<Cycles> | Set Program Cycles. Changes the number of program cycles. |
| Command | What it does |
|---|---|
| ?HL | Read Program Hold. Reads whether the program is holding (1) or running (0). |
| HL | Halt Program. Halts (holds) the running program. |
| Command | What it does |
|---|---|
| ?SK | Read Soak Setpoint. Reads the soak timer setpoint in seconds. |
| SK<Seconds> | Set Soak Time. Changes the soak time in seconds (1-9999). |
| Command | What it does |
|---|---|
| ?SP | Read Setpoint. Reads the current temperature setpoint in °C. |
| SP<Setpoint_Num><Temp> | Set Setpoint. Changes manual setpoint n (1-3) to the given temperature in °C (-80.0 to 225.0); manual mode only. |
| Command | What it does |
|---|---|
| ?WL | Read Window Limit. Reads the window limit. |
| WL<Window> | Set Window Limit. Changes the at-temperature window limit (0.5-10.0). |
With PyVISA and any VISA backend, this opens both instruments, identifies them, sends one setting, and reads one value from each. For the setting, the two libraries have no common command with a worked example on both, so each instrument uses its own: SETN 0 on the ATS-710E and SP1-80.0 on the THJ80. For the reading, the two libraries have no common command with a worked example on both, so each instrument uses its own: SETN? on the ATS-710E and ?SP on the THJ80. For the basics, see the PyVISA tutorial.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find both VISA addresses
a = rm.open_resource("PASTE_ATS_710E_ADDRESS") # inTEST ATS-710E
b = rm.open_resource("PASTE_THJ80_ADDRESS") # SP Scientific THJ80
for inst in (a, b):
inst.timeout = 5000 # ms
print(inst.query("*IDN?"))
# No shared setting with an example on both, so each uses its own command
a.write("SETN 0") # Select Setpoint
b.write("SP1-80.0") # Set Setpoint
# No shared reading with an example on both, so each uses its own command
print(a.query("SETN?")) # Read Selected Setpoint
print(b.query("?SP")) # Read Setpoint
a.close()
b.close()Describe the test; TestFlow's agent uses these command sets to drive your inTEST ATS-710E or SP Scientific THJ80 and writes the report from the measured results.
In TestFlow's library the ATS-710E has 72 SCPI commands and the THJ80 has 62 SCPI commands. 8 command headers are common to both, 55 are only on the ATS-710E (mostly ADMD and COOL), and 51 are only on the THJ80 (mostly AF and CY).
Partly. 8 command headers are accepted by both. 106 are not, for example ADMD on the ATS-710E and ?AF on the THJ80, so a script that uses those needs a separate branch for each instrument.
The ATS-710E: 72 commands for the ATS-710E against 62 for the THJ80 in TestFlow's library. Of those, 34 on the ATS-710E and 31 on the THJ80 are queries that read a value or state, and the library splits them into 37 and 25 subsystems.
Only the THJ80 library has an operation-complete command (*OPC?); the ATS-710E library has none. Neither library has an error-queue read such as SYSTem:ERRor?. Each command named here is listed in the tables on this page and on each instrument's command reference.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.