Thermal Forcing System · SCPI command reference

SP Scientific THJ80 and ThermoJet ES SCPI commands

The SP Scientific THJ80 and ThermoJet ES thermal forcing system is controlled remotely with SCPI commands. TestFlow's instrument library holds 62 commands for it, 31 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the THJ80 on a real bench.

Commands
62
Queries
31
Subsystems
25
Models covered
2

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your SP Scientific THJ80 or ThermoJet ES and writes the report from the measured results.

Get started

Control the SP Scientific THJ80 from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the SP Scientific THJ80's VISA address

# Paste the address of your SP Scientific THJ80 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("SP1-80.0")  # Set Setpoint
print(inst.query("?SP"))  # Read Setpoint

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your SP Scientific THJ80 or ThermoJet ES and writes the report from the measured results.

Get started

Run a THJ80 test without writing code

Describe the test in a sentence, for example “set setpoint on the THJ80, then read setpoint and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

62 commands in 25 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The THJ80 and ThermoJet ES share this command set in TestFlow's library.

Other subsystem (27)

CommandWhat it doesArguments
?SPRead Setpoint. Reads the current temperature setpoint in °C.None
?PVRead Temperature. Reads the system temperature in °C (DUT in DUT mode or nozzle in nozzle mode).None
?TARead Air Temperature. Reads the nozzle (air) temperature in °C.None
?TDRead DUT Temperature. Reads the DUT temperature in °C (returns 999 when not in DUT mode).None
?AFRead Airflow Setpoint. Reads the airflow setpoint.None
?SKRead Soak Setpoint. Reads the soak timer setpoint in seconds.None
?STRead Soak Timer. Reads the running soak timer in seconds.None
?WLRead Window Limit. Reads the window limit.None
?TCRead Control Mode. Reads the control mode (0 air or 1 DUT).None
?DARead Purge State. Reads the purge state (1 on or 0 off).None
?HPRead Head Position. Reads the head position (0 up or 1 down).None
?HKRead Head Lock. Reads the head lock state (0 unlocked or 1 locked).None
?MSRead Manual Status. Reads the manual status (0 not manual or 1-3 active manual setpoint).None
?CYRead Program Cycles. Reads the number of program cycles.None
?HLRead Program Hold. Reads whether the program is holding (1) or running (0).None
?PIRead Program State. Reads the program status (1 on or 0 off).None
?SIRead Program Step. Reads the program step index (0 not in program; odd transitioning and even soaking).None
?PTRead Program Temps. Reads program temperatures 1-8 as signed hexadecimal values.None
?PMRead Program Ramps. Reads program ramp rates 1-8.None
?PSRead Program Soaks. Reads program soak times 1-8 (divide by 10).None
?PRRead Mode. Reads the operating mode (IDLE, MAN or PRO).None
?MERead System Status. Reads system status (1 idle, 2 transitioning, 3 soaking, 4 at temp, 5 defrosting, 6 holding, 7 auto-tuning, 8 initializing).None
?ALRead Alarm. Reads the active alarm (None, AP low air pressure, TC open nozzle TC, DUT open DUT TC, AF low air flow, DEFR defrost).None
?RS0Read Manual Status String. Reads the manual status string (soak time, air/DUT temps, flow, status, setpoints).None
?RS1Read Program Status String. Reads the program status string (program on, cycles, hold, step).None
?RS2Read Utility Status String. Reads the utility status string (control mode, head position, head lock, purge, window, soak, ramp).None
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

TC subsystem (2)

CommandWhat it doesArguments
TC0Control On Air. Selects control on the nozzle (air) sensor.None
TC1Control On DUT. Selects control on the DUT sensor.None

DA subsystem (2)

CommandWhat it doesArguments
DA1Turn On Purge. Turns the dry-air purge on.None
DA0Turn Off Purge. Turns the dry-air purge off.None

HK subsystem (2)

CommandWhat it doesArguments
HK1Lock Head. Locks the head.None
HK0Unlock Head. Unlocks the head.None

SP subsystem (1)

CommandWhat it doesArguments
SP<Setpoint_Num><Temp>Set Setpoint. Changes manual setpoint n (1-3) to the given temperature in °C (-80.0 to 225.0); manual mode only.Number: 1,2,3; Number: -80.0,225.0

AF subsystem (1)

CommandWhat it doesArguments
AF<Flow>Set Airflow. Changes the airflow setpoint in SCFM (5-20).Number: 5,20

SK subsystem (1)

CommandWhat it doesArguments
SK<Seconds>Set Soak Time. Changes the soak time in seconds (1-9999).Number: 1,9999

WL subsystem (1)

CommandWhat it doesArguments
WL<Window>Set Window Limit. Changes the at-temperature window limit (0.5-10.0).Number: 0.5,10.0

RR subsystem (1)

CommandWhat it doesArguments
RR<Rate>Set Ramp Rate. Sets the manual-mode ramp rate (0-999).Number: 0,999

AU subsystem (1)

CommandWhat it doesArguments
AUHead Up. Raises the arm (head).None

AD subsystem (1)

CommandWhat it doesArguments
ADHead Down. Lowers the arm (head).None

RM subsystem (1)

CommandWhat it doesArguments
RM<Setpoint_Num>Run Manual. Starts manual mode on setpoint n (1-3).Number: 1,2,3

CY subsystem (1)

CommandWhat it doesArguments
CY<Cycles>Set Program Cycles. Changes the number of program cycles.Number

RP subsystem (1)

CommandWhat it doesArguments
RPRun Program. Begins running the stored program.None

HL subsystem (1)

CommandWhat it doesArguments
HLHalt Program. Halts (holds) the running program.None

CN subsystem (1)

CommandWhat it doesArguments
CNContinue Program. Continues a halted program.None

ID subsystem (1)

CommandWhat it doesArguments
IDGo To Idle. Stops the current operation and returns to idle.None

DF subsystem (1)

CommandWhat it doesArguments
DFStart Defrost. Begins a defrost cycle.None

OFF subsystem (1)

CommandWhat it doesArguments
OFFShutdown. Requests shutdown mode.None

SL subsystem (1)

CommandWhat it doesArguments
SLRemote Display On. Shows the remote-control screen on the touchscreen.None

SU subsystem (1)

CommandWhat it doesArguments
SURemote Display Off. Returns the touchscreen to the manual screen.None

GP subsystem (1)

CommandWhat it doesArguments
GPComm Test. Toggles the LED on the Command Set screen to verify communication.None

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (9)

CommandWhat it doesArguments
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns the result.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*ESR?Read Event Status. Reads and clears the standard event status register.None
*ESE <Mask>Set Event Status Enable. Sets the standard event status enable mask (0-255).Number: 0,255
*STB?Read Status Byte. Reads the status byte.None
*SRE <Mask>Set Service Request Enable. Sets the service request enable mask (0-255).Number: 0,255

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from SP Scientific's programming documentation for the THJ80 and ThermoJet ES and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by SP Scientific's programming manual: check it before you rely on an edge case. Last updated September 23, 2026.

Frequently asked questions

Does the SP Scientific THJ80 support SCPI remote control?

Yes. The SP Scientific THJ80 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 62 SCPI commands for it: 31 queries that read values or state, and 31 settings and actions.

How do I control the SP Scientific THJ80 from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, SP1-80.0 and ?SP.

Which command reads a value from the SP Scientific THJ80?

?SP (read setpoint). Reads the current temperature setpoint in °C. The full list of queries is in the command reference on this page, grouped by subsystem.