The Anritsu MS2026C, MS2027C, MS2028C, MS2036C, MS2037C, and MS2038C vector network analyzer is controlled remotely with SCPI commands. TestFlow's instrument library holds 29 commands for it, 6 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the MS2026C on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Anritsu MS2026C, MS2027C, MS2028C, MS2036C, MS2037C, or MS2038C and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Anritsu MS2026C's VISA address
# Paste the address of your Anritsu MS2026C (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write(":SOURce:POWer LOW") # Set Power Level
print(inst.query(":SENSe:FREQuency:STARt?")) # Read Start Freq
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Anritsu MS2026C, MS2027C, MS2028C, MS2036C, MS2037C, or MS2038C and writes the report from the measured results.
Describe the test in a sentence, for example “set power level on the MS2026C, then read start freq and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
29 commands in 9 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The MS2026C, MS2027C, MS2028C, MS2036C, MS2037C, and MS2038C share this command set in TestFlow's library.
| Command | What it does | Arguments |
|---|---|---|
| :SENSe:FREQuency:STARt <Frequency> | Set Start Freq. Sets the sweep start frequency in Hz. | Number |
| :SENSe:FREQuency:STOP <Frequency> | Set Stop Freq. Sets the sweep stop frequency in Hz. | Number |
| :SENSe:FREQuency:CENTer <Frequency> | Set Center Freq. Sets the sweep center frequency in Hz. | Number |
| :SENSe:FREQuency:SPAN <Frequency> | Set Span Freq. Sets the sweep frequency span in Hz. | Number |
| :SENSe:FREQuency:STARt? | Read Start Freq. Returns the start frequency in Hz. | None |
| :SENSe:FREQuency:STOP? | Read Stop Freq. Returns the stop frequency in Hz. | None |
| :SENSe:SWEep:POINts <Points> | Set Points. Sets the number of data points per sweep. | Number |
| :SENSe:SWEep:IFBW <Bandwidth> | Set IF Bandwidth. Sets the IF bandwidth in Hz. | Number: 10,20,50,100,200,500,1000,2000,5000,10000,20000,50000,100000 |
| :SENSe:AVERage:COUNt <Count> | Set Average Count. Sets the number of sweeps to average. | Number |
| :SENSe:TRACe<Trace>:SELect | Set Active Trace. Selects the active trace number. | Number: 1-4 |
| :SENSe:TRACe<Trace>:SPARams <Parameter> | Define Measurement. Assigns an S-parameter to the active trace. | Number: 1-4; Text: S11,S21,S12,S22 |
| Command | What it does | Arguments |
|---|---|---|
| :CALCulate<Trace>:FORMat <Format> | Set Format. Sets the display format of the active trace. | Number: 1-4; Text: LMAGnitude,SWR,PHASe,REAL,IMAGinary,GDELay,SMITh,ISMith,LM/2,LINPolar,LOGPolar,RIMPedance,IIMPedance |
| :CALCulate<Trace>:DATA? FDATa | Read Formatted Data. Returns the formatted trace data array. | Number: 1-4 |
| :CALCulate:MARKer<Marker>:TYPE REFerence | Marker On. Turns on the specified marker. | Number: 1-12 |
| :CALCulate:MARKer<Marker>:TYPE OFF | Marker Off. Turns off the specified marker. | Number: 1-12 |
| :CALCulate:MARKer<Marker>:X <Frequency> | Set Marker X. Sets the marker stimulus position in Hz. | Number: 1,2,3,4,5,6; Number |
| :CALCulate:MARKer<Marker>:Y? | Read Marker Y. Returns the marker response value. | Number: 1,2,3,4,5,6 |
| Command | What it does | Arguments |
|---|---|---|
| :INITiate:CONTinuous ON | Continuous On. Enables continuous sweeping. | None |
| :INITiate:CONTinuous OFF | Continuous Off. Stops continuous sweeping (hold). | None |
| :INITiate:IMMediate | Trigger Single. Triggers a single sweep. | None |
| :INITiate:HOLD ON | Hold. Holds (stops) the current sweep. | None |
| Command | What it does | Arguments |
|---|---|---|
| :SOURce:POWer <Power> | Set Power Level. Sets the source output power level in dBm. | Text: LOW,HIGH |
| Command | What it does | Arguments |
|---|---|---|
| :SYSTem:PRESet | Preset. Presets the instrument to default state. | None |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Returns the instrument identification string (manufacturer, model, serial, firmware). | None |
| *RST | Reset. Resets the instrument to its default state. | None |
| *OPC? | Operation Complete. Returns 1 when all pending operations are complete. | None |
This reference is generated from TestFlow's instrument library, which is drafted from Anritsu's programming documentation for the MS2026C, MS2027C, MS2028C, MS2036C, MS2037C, and MS2038C and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Anritsu's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.
Yes. The Anritsu MS2026C accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 29 SCPI commands for it: 6 queries that read values or state, and 23 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, :SOURce:POWer LOW and :SENSe:FREQuency:STARt?.
:SENSe:FREQuency:STARt? (read start freq). Returns the start frequency in Hz. The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Anritsu MS2026C through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.