The Viavi 2945B radio test set is controlled remotely with SCPI commands. TestFlow's instrument library holds 92 commands for it, 36 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 2945B on a real bench.
Commands
92
Queries
36
Subsystems
20
Models covered
1
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Viavi 2945B and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Viavi 2945B's VISA address
# Paste the address of your Viavi 2945B (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write(":RECEiver:Autotune ON") # Turn On Autotune
print(inst.query(":MEASUre:TXFreq?")) # Measure TX Frequency
inst.close()
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Viavi 2945B and writes the report from the measured results.
Describe the test in a sentence, for example “turn on autotune on the 2945B, then measure tx frequency and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
Turn On Measure Cycle. Starts the instrument measure cycle.
None
:MEASCycl OFF
Turn Off Measure Cycle. Stops the instrument measure cycle.
None
AUDF subsystem (2)
Command
What it does
Arguments
:AUDFilt:Psoph ON
Turn On Psophometric Filter. Turns the receiver psophometric audio filter on.
None
:AUDFilt:Psoph OFF
Turn Off Psophometric Filter. Turns the receiver psophometric audio filter off.
None
G subsystem (1)
Command
What it does
Arguments
:Genswitch <Port>
Set Generator Output Port. Selects the RF output routing.
Text: GEN_N,GEN_BNC
MODT subsystem (1)
Command
What it does
Arguments
:MODType <Type>
Set Modulation Type. Sets the modulation type of the signal generator.
Text: AM,FM
RECS subsystem (1)
Command
What it does
Arguments
:RECSwitch <Input>
Set Receiver Input. Selects the RF input routing from the transmitter under test.
Text: REC_N,REC_ANT
DEM subsystem (1)
Command
What it does
Arguments
:DEModtype <Type>
Set Demodulation Type. Sets the demodulation type used on the received signal.
Text: AM,FM,SSB
AFI subsystem (1)
Command
What it does
Arguments
:AFInput <Coupling>
Set Audio Input Coupling. Sets the audio input socket coupling.
Text: AC,DC
RECA subsystem (1)
Command
What it does
Arguments
:RECAll <Store>
Recall Store. Recalls instrument settings from the given store number.
Number
DEV subsystem (1)
Command
What it does
Arguments
:DEVerror?
Read Device Error. Returns the last device error generated by the instrument.
None
Other subsystem (1)
Command
What it does
Arguments
<Output Variable>=<Equation>
Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)
Number; Text
WAIT subsystem (1)
Command
What it does
Arguments
wait(<ms>)
Wait. Waits for number of milliseconds
Number
SAVE subsystem (1)
Command
What it does
Arguments
save2var <Variable_name>
Save2Variable. Saves last measurement to a variable
Text
IEEE 488.2 common commands (8)
Command
What it does
Arguments
*IDN?
Identify. Returns the identification string (manufacturer, type number, serial number, firmware version).
None
*RST
Reset. Sets the instrument functions to the factory default power-up state.
None
*CLS
Clear Status. Clears all status event registers.
None
*TST?
Self Test. Runs the self test and returns 0 if all tests passed.
None
*OPC?
Operation Complete. Returns 1 when the preceding operation has completed.
None
*WAI
Wait To Continue. Inhibits execution of an overlapped command until the preceding operation is complete.
None
*OPT?
Read Options. Returns the state of enabled options in 32 data fields.
None
*STB?
Read Status Byte. Returns the value of the status byte.
None
Where these commands come from
This reference is generated from TestFlow's instrument library, which is drafted from Viavi's programming documentation for the 2945B and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Viavi's programming manual: check it before you rely on an edge case. Last updated September 27, 2026.
Frequently asked questions
Does the Viavi 2945B support SCPI remote control?
Yes. The Viavi 2945B accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 92 SCPI commands for it: 36 queries that read values or state, and 56 settings and actions.
How do I control the Viavi 2945B from Python?
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, :RECEiver:Autotune ON and :MEASUre:TXFreq?.
Which command reads a value from the Viavi 2945B?
:MEASUre:TXFreq? (measure tx frequency). Returns the RF counter reading in MHz (1 Hz resolution). The full list of queries is in the command reference on this page, grouped by subsystem.
Can I automate the Viavi 2945B without writing code?