The UNI-T UT8805E multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 123 commands for it, 37 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the UT8805E on a real bench.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the UNI-T UT8805E's VISA address
# Paste the address of your UNI-T UT8805E (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("FUNC \"VOLT:DC\"") # Func DCV
print(inst.query("MEAS:VOLT:DC?")) # Measure DCV
inst.close()Describe the test in a sentence, for example “func dcv on the UT8805E, then measure dcv and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
123 commands in 26 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| CALC:LIM:LOW <Lower> | Set Limit Lower. Lower limit of the limit test (manual 2.2.2). | Number: -1E15 to 1E15 |
| CALC:LIM:UPP <Upper> | Set Limit Upper. Upper limit of the limit test (manual 2.2.2). | Number: -1E15 to 1E15 |
| CALC:LIM:STAT ON | Limit On. Turn the limit test on (manual 2.2.3). | None |
| CALC:LIM:STAT OFF | Limit Off. Turn the limit test off (manual 2.2.3). | None |
| CALC:LIM:CLE | Limit Clear. Clear the limit-failure indication (manual 2.2.1). | None |
| CALC:AVER:STAT ON | Statistic On. Turn statistics on; they restart (manual 2.5.1). | None |
| CALC:AVER:STAT OFF | Statistic Off. Turn statistics off (manual 2.5.1). | None |
| CALC:AVER:CLE | Statistic Clear. Clear the statistics (manual 2.5.3). | None |
| CALC:AVER:AVER? | Stat Average. Return the mean of the readings since the statistics were cleared (manual 2.5.2). | None |
| CALC:AVER:MIN? | Stat Min. Return the minimum reading (manual 2.5.2). | None |
| CALC:AVER:MAX? | Stat Max. Return the maximum reading (manual 2.5.2). | None |
| CALC:AVER:COUN? | Stat Count. Return the number of readings in the statistics (manual 2.5.2). | None |
| CALC:AVER:PTP? | Stat PeakToPeak. Return the peak-to-peak spread of the readings (manual 2.5.2). | None |
| CALC:AVER:SDEV? | Stat StdDev. Return the standard deviation of the readings (manual 2.5.2). | None |
| CALC:AVER:ALL? | Stat All. Return mean, standard deviation, maximum and minimum, comma separated (manual 2.5.2). | None |
| CALC:SCAL:FUNC <Function> | Set Scale Function. dB or dBm scaling of voltage readings (ACV and DCV only) (manual 2.4.3). | Text: DB,DBM |
| CALC:SCAL:STAT ON | Scale On. Turn dB/dBm scaling on (manual 2.4.5). | None |
| CALC:SCAL:STAT OFF | Scale Off. Turn dB/dBm scaling off (manual 2.4.5). | None |
| CALC:SCAL:DB:REF <Reference> | Set dB Reference. dB reference in dBm; turns the automatic reference off (manual 2.4.1). | Number: -200 to 200 |
| CALC:SCAL:DBM:REF <Ohm> | Set dBm Reference. Reference resistance for dBm (manual 2.4.2). | Number: 50,75,93,110,124,125,135,150,250,300,500,600,800,900,1000,1200,8000 |
| CALC:SCAL:REF:AUTO <State> | Set Scale Auto Reference. Use the first reading as the dB reference (manual 2.4.4). | Text: ON,OFF |
| CALC:CLE | Clear Calculations. Clear limits, histogram, statistics and readings in memory (manual 2.1). | None |
| Command | What it does | Arguments |
|---|---|---|
| CONF:VOLT:DC | Configure DCV. Set up DC voltage with default settings and auto range; measure with Read (manual 3.8). | None |
| CONF:VOLT:AC | Configure ACV. Set up AC voltage with default settings and auto range; measure with Read (manual 3.8). | None |
| CONF:CURR:DC | Configure DCI. Set up DC current with default settings and auto range; measure with Read (manual 3.3). | None |
| CONF:CURR:AC | Configure ACI. Set up AC current with default settings and auto range; measure with Read (manual 3.3). | None |
| CONF:RES | Configure 2W Resistance. Set up 2-wire resistance with default settings and auto range; measure with Read (manual 3.6). | None |
| CONF:FRES | Configure 4W Resistance. Set up 4-wire resistance with default settings and auto range; measure with Read (manual 3.6). | None |
| CONF:FREQ | Configure Frequency. Set up frequency with default settings; measure with Read (manual 3.5). | None |
| CONF:PER | Configure Period. Set up period with default settings; measure with Read (manual 3.5). | None |
| CONF:CONT | Configure Continuity. Set up continuity (fixed 1 kOhm range); measure with Read (manual 3.2). | None |
| CONF:DIOD | Configure Diode. Set up diode test (fixed 2 V range); measure with Read (manual 3.4). | None |
| CONF:CAP | Configure Capacitance. Set up capacitance with default settings and auto range; measure with Read (manual 3.9). | None |
| CONF:TEMP TC,<Type> | Configure Thermocouple. Set up a thermocouple temperature measurement of this type, other settings at default; set the cold junction next if needed, then Read (manual 3.7, 6.5.5). | Text: E,J,K,N,R,T |
| CONF:TEMP <Probe>,<Type> | Configure Temperature. Set up temperature for this probe (TC thermocouple, RTD/FRTD 2/4-wire RTD, THER/FTH 2/4-wire thermistor) and type (E/J/K/N/R/T for TC, PT100 for RTD); then Read (manual 3.7). | Text: TC,RTD,FRTD,THER,FTH; Text: E,J,K,N,R,T,PT100,DEF |
| CONF? | Read Configuration. Return the present function and range, e.g. "VOLT +2.00000000E-01" (manual 3.1). | None |
| Command | What it does | Arguments |
|---|---|---|
| FUNC "VOLT:DC" | Func DCV. Select DC voltage, keeping its settings (manual 6.1). | None |
| FUNC "VOLT:AC" | Func ACV. Select AC voltage, keeping its settings (manual 6.1). | None |
| FUNC "CURR:DC" | Func DCI. Select DC current, keeping its settings (manual 6.1). | None |
| FUNC "CURR:AC" | Func ACI. Select AC current, keeping its settings (manual 6.1). | None |
| FUNC "RES" | Func 2W Resistance. Select 2-wire resistance, keeping its settings (manual 6.1). | None |
| FUNC "FRES" | Func 4W Resistance. Select 4-wire resistance, keeping its settings (manual 6.1). | None |
| FUNC "FREQ" | Func Frequency. Select frequency, keeping its settings (manual 6.1). | None |
| FUNC "PER" | Func Period. Select period, keeping its settings (manual 6.1). | None |
| FUNC "CONT" | Func Continuity. Select continuity, keeping its settings (manual 6.1). | None |
| FUNC "DIOD" | Func Diode. Select diode test, keeping its settings (manual 6.1). | None |
| FUNC "CAP" | Func Capacitance. Select capacitance, keeping its settings (manual 6.1). | None |
| FUNC "TEMP" | Func Temperature. Select temperature with the probe settings already made; set the probe with Set_Temperature_Probe (manual 6.1, 6.5.4). | None |
| FUNC? | Read Function. Return the selected measurement function (manual 6.1). | None |
| Command | What it does | Arguments |
|---|---|---|
| MEAS:VOLT:DC? | Measure DCV. Measure DC voltage with default settings and auto range; every other setting is reset (manual 5.7). | None |
| MEAS:VOLT:AC? | Measure ACV. Measure AC voltage with default settings and auto range; every other setting is reset (manual 5.7). | None |
| MEAS:CURR:DC? | Measure DCI. Measure DC current with default settings and auto range; every other setting is reset (manual 5.2). | None |
| MEAS:CURR:AC? | Measure ACI. Measure AC current with default settings and auto range; every other setting is reset (manual 5.2). | None |
| MEAS:RES? | Measure 2W Resistance. Measure 2-wire resistance with default settings and auto range (manual 5.5). | None |
| MEAS:FRES? | Measure 4W Resistance. Measure 4-wire resistance with default settings and auto range (manual 5.5). | None |
| MEAS:FREQ? | Measure Frequency. Measure frequency with default settings (manual 5.4). | None |
| MEAS:PER? | Measure Period. Measure period with default settings (manual 5.4). | None |
| MEAS:CONT? | Measure Continuity. Measure continuity resistance (manual 5.1). | None |
| MEAS:DIOD? | Measure Diode. Measure the diode forward voltage (manual 5.3). | None |
| MEAS:CAP? | Measure Capacitance. Measure capacitance with default settings and auto range (manual 5.8). | None |
| MEAS:TEMP? | Measure Temperature. Measure temperature with the DEFAULT probe settings (MEASure resets them): for a thermocouple use Configure_Thermocouple, then Read (manual 5.6). | None |
| Command | What it does | Arguments |
|---|---|---|
| TEMP:NPLC <Rate> | Set Temperature Rate. Reading rate for temperature (manual 6.5). | Text: Slow,Medium,Fast |
| TEMP:TRAN:TYPE <Probe> | Set Temperature Probe. Probe for temperature: TC thermocouple, RTD/FRTD 2/4-wire RTD, THER/FTH 2/4-wire thermistor (manual 6.5.4). | Text: TC,RTD,FRTD,THER,FTH |
| TEMP:TRAN:TYPE? | Read Temperature Probe. Return the temperature probe type (manual 6.5.4). | None |
| TEMP:TRAN:TC:TYPE <Type> | Set Thermocouple Type. Thermocouple type when the probe is TC (manual 6.5.8). | Text: E,J,K,N,R,T |
| TEMP:TRAN:TC:TYPE? | Read Thermocouple Type. Return the thermocouple type (manual 6.5.8). | None |
| TEMP:TRAN:TC:RJUN:TYPE <Junction> | Set TC Reference Junction. Cold-junction compensation: INT uses the meter's own terminal temperature (default), FIX uses Set_TC_Fixed_Junction_Temp; the manual recommends FIX at a known temperature for best accuracy (manual 6.5.7). | Text: INT,FIX |
| TEMP:TRAN:TC:RJUN:TYPE? | Read TC Reference Junction. Return the cold-junction source, INT or FIX (manual 6.5.7). | None |
| TEMP:TRAN:TC:RJUN <Temperature> | Set TC Fixed Junction Temp. Cold-junction temperature in deg C used when the junction is FIX; default 0 (manual 6.5.5). | Number: -20 to 80 |
| TEMP:TRAN:TC:RJUN? | Read TC Fixed Junction Temp. Return the fixed cold-junction temperature (manual 6.5.5). | None |
| TEMP:TRAN:TC:RJUN:OFFS:ADJ <Offset> | Set TC Junction Offset. Correction in deg C added to the internal cold-junction temperature (manual 6.5.6). | Number: -20 to 20 |
| TEMP:TRAN:TC:RJUN:OFFS:ADJ? | Read TC Junction Offset. Return the internal cold-junction correction (manual 6.5.6). | None |
| Command | What it does | Arguments |
|---|---|---|
| TRIG:SOUR <Source> | Set Trigger Source. IMM measures at once, EXT waits for the rear Ext Trig input, BUS waits for Single_Trigger (manual 8.5). | Text: IMM,EXT,BUS |
| TRIG:SOUR? | Read Trigger Source. Return the trigger source (manual 8.5). | None |
| TRIG:COUN <Count> | Set Trigger Count. Triggers accepted per measurement; readings = sample count x trigger count (manual 8.1). | Number: 1 to 2147483647 |
| TRIG:DEL <Seconds> | Set Trigger Delay. Delay between the trigger and the first reading; turns the automatic delay off (manual 8.2). | Number: seconds, up to 3600 |
| TRIG:DEL:AUTO <State> | Set Trigger Delay Auto. Let the meter choose the trigger delay from the function and range (manual 8.3). | Text: ON,OFF |
| TRIG:SLOP <Slope> | Set Trigger Slope. Edge of the rear Ext Trig input that triggers (manual 8.4). | Text: POS,NEG |
| Command | What it does | Arguments |
|---|---|---|
| VOLT:DC:RANG <Range> | Set DCV Range. Fixed DC voltage range in volts; turns auto range off (manual 6.6.4). | Number: 0.2,2,20,200,1000 |
| VOLT:AC:RANG <Range> | Set ACV Range. Fixed AC voltage range in volts; turns auto range off (manual 6.6.4). | Number: 0.2,2,20,200,750 |
| VOLT:DC:NPLC <Rate> | Set DCV Rate. Reading rate for DC voltage (manual 6.6.6). | Text: Slow,Medium,Fast |
| VOLT:AC:NPLC <Rate> | Set ACV Rate. Reading rate for AC voltage (manual 6.6.6). | Text: Slow,Medium,Fast |
| VOLT:IMP:AUTO <State> | Set DCV Impedance. OFF: 10 MOhm input on every DC range. ON: over 10 GOhm (Hi-Z) on the 200 mV, 2 V and 20 V ranges (manual 6.6.7). | Text: ON,OFF |
| Command | What it does | Arguments |
|---|---|---|
| CURR:DC:RANG <Range> | Set DCI Range. Fixed DC current range in amperes; turns auto range off (manual 6.2.4). | Number: 0.0002,0.002,0.02,0.2,2,10 |
| CURR:AC:RANG <Range> | Set ACI Range. Fixed AC current range in amperes, 20 mA at least; turns auto range off (manual 6.2.4). | Number: 0.02,0.2,2,10 |
| CURR:DC:NPLC <Rate> | Set DCI Rate. Reading rate for DC current (manual 6.2.6). | Text: Slow,Medium,Fast |
| CURR:AC:NPLC <Rate> | Set ACI Rate. Reading rate for AC current (manual 6.2.6). | Text: Slow,Medium,Fast |
| CURR:<Mode>:TERM <Terminal> | Set Current Terminals. Current input to measure on: Small is the 200 mA terminal, Big the 10 A terminal (manual 6.2.7). | Text: DC,AC; Text: Small,Big |
| Command | What it does | Arguments |
|---|---|---|
| <Function>:RANG:AUTO <State> | Set Auto Range. Turn auto range on or off for one function (manual 6.2.5, 6.3.5, 6.4.6, 6.6.5, 6.7.5). | Text: VOLT:DC,VOLT:AC,CURR:DC,CURR:AC,RES,FRES,CAP,FREQ:VOLT; Text: ON,OFF |
| <Function>:NULL:STAT <State> | Set Null State. Subtract the null value from this function's readings; turning it on also turns automatic null on (manual 6.2.1 to 6.7.1). | Text: VOLT:DC,VOLT:AC,CURR:DC,CURR:AC,RES,FRES,FREQ,PER,CAP,TEMP; Text: ON,OFF |
| <Function>:NULL:VAL <Value> | Set Null Value. Null value for this function, in its unit; turns automatic null off (manual 6.2.2 to 6.7.2). | Text: VOLT:DC,VOLT:AC,CURR:DC,CURR:AC,RES,FRES,FREQ,PER,CAP,TEMP; Number: in the function's unit |
| <Function>:NULL:VAL:AUTO <State> | Set Null Auto. Use the first reading as this function's null value (manual 6.2.3 to 6.7.3). | Text: VOLT:DC,VOLT:AC,CURR:DC,CURR:AC,RES,FRES,FREQ,PER,CAP,TEMP; Text: ON,OFF |
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| SYST:ERR? | Read Error. Read and remove the oldest entry of the error queue, e.g. -113,"Undefined header" (manual 7.1.1). | None |
| SYST:BEEP:STAT ON | Beeper On. Turn the beeper on (manual 7.1.2). | None |
| SYST:BEEP:STAT OFF | Beeper Off. Turn the beeper off; key clicks are not affected (manual 7.1.2). | None |
| Command | What it does | Arguments |
|---|---|---|
| DATA:LAST? | Read Last. Return the latest reading with its unit, e.g. -4.79221344E-04 VDC (manual 4.1). | None |
| DATA:POIN? | Reading Count. Return how many readings are in memory (manual 4.2). | None |
| DATA:REM? <Count> | Remove Readings. Return and erase the oldest readings from memory (manual 4.3). | Number: 1 to 10000 |
| Command | What it does | Arguments |
|---|---|---|
| SAMP:COUN <Count> | Set Sample Count. Readings per trigger (manual 1.11). | Number: 1 to 100000 |
| SAMP:COUN? | Read Sample Count. Return the readings per trigger (manual 1.11). | None |
| Command | What it does | Arguments |
|---|---|---|
| RES:RANG <Range> | Set 2W Res Range. Fixed 2-wire resistance range in ohms; turns auto range off (manual 6.4.5). | Number: 200,2000,20000,200000,2000000,10000000,100000000 |
| RES:NPLC <Rate> | Set 2W Res Rate. Reading rate for 2-wire resistance (manual 6.4.1). | Text: Slow,Medium,Fast |
| Command | What it does | Arguments |
|---|---|---|
| FRES:RANG <Range> | Set 4W Res Range. Fixed 4-wire resistance range in ohms; turns auto range off (manual 6.4.5). | Number: 200,2000,20000,200000,2000000,10000000,100000000 |
| FRES:NPLC <Rate> | Set 4W Res Rate. Reading rate for 4-wire resistance (manual 6.4.1). | Text: Slow,Medium,Fast |
| Command | What it does | Arguments |
|---|---|---|
| UNIT:TEMP <Unit> | Set Temperature Unit. Unit of temperature readings (manual 1.12). | Text: C,F,K |
| UNIT:TEMP? | Read Temperature Unit. Return the temperature unit (manual 1.12). | None |
| Command | What it does | Arguments |
|---|---|---|
| READ? | Read. Start the configured measurement and return its readings, comma separated when there are several (manual 1.10). | None |
| Command | What it does | Arguments |
|---|---|---|
| INIT | Initiate. Arm the trigger system; readings go to memory, collect them with Fetch (manual 1.8). | None |
| Command | What it does | Arguments |
|---|---|---|
| FETC? | Fetch. Wait for the measurement started by Initiate and return its readings; memory keeps them (manual 1.7). | None |
| Command | What it does | Arguments |
|---|---|---|
| ABOR | Abort. Stop a measurement in progress and return to idle (manual 1.6). | None |
| Command | What it does | Arguments |
|---|---|---|
| OUTP:TRIG:SLOP <Slope> | Set VM Complete Slope. Pulse polarity of the rear VM Comp output after each reading (manual 1.9). | Text: POS,NEG |
| Command | What it does | Arguments |
|---|---|---|
| CAP:RANG <Range> | Set Capacitance Range. Fixed capacitance range in farads; turns auto range off (manual 6.7.4). | Number: 2E-9,2E-8,2E-7,2E-6,2E-5,2E-4,2E-3 |
| Command | What it does | Arguments |
|---|---|---|
| FREQ:VOLT:RANG <Range> | Set Frequency Voltage Range. Input voltage range in volts for frequency and period (manual 6.3.4). | Number: 0.2,2,20,200,750 |
| Command | What it does | Arguments |
|---|---|---|
| CONT:THR:VAL <Ohm> | Set Continuity Threshold. Continuity beeps at or below this resistance (manual 6.8.1). | Number: 0 to 2000 |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Return the identification string (IEEE 488.2 common command, manual 1.4). | None |
| *RST | Reset. Restore the factory settings (manual 1.4). | None |
| *CLS | Clear Status. Clear the status registers and the error queue (manual 1.4). | None |
| *OPC? | Operation Complete. Return 1 once every pending command has finished (manual 1.4). | None |
| *UNREMOTE | Return To Local. Leave remote mode and unlock the front-panel keys (manual 1.13). | None |
| *TRG | Single Trigger. Send a bus trigger: with Set_Trigger_Source BUS and Initiate it starts one measurement (manual 1.8). | None |
This reference is generated from TestFlow's instrument library, which is drafted from UNI-T's programming documentation for the UT8805E and checked when the agent runs it. The library marks this instrument as supporting *OPC? for operation-complete synchronisation. It reads errors back with SYSTem:ERRor?. Ranges, timing, and option-dependent commands are defined by UNI-T's programming manual: check it before you rely on an edge case. Last updated October 25, 2026.
Yes. The UNI-T UT8805E accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 123 SCPI commands for it: 37 queries that read values or state, and 86 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, FUNC "VOLT:DC" and MEAS:VOLT:DC?.
MEAS:VOLT:DC? (measure dcv). Measure DC voltage with default settings and auto range; every other setting is reset (manual 5.7). The full list of queries is in the command reference on this page, grouped by subsystem.