The Rohde & Schwarz ZNA26, ZNA43, ZNA50, ZNA67, ZNB20, ZNB26, ZNB3000, ZNB4, ZNB40, ZNB43, ZNB8, ZNBT20, ZNBT26, ZNBT8, and ZND vector network analyzer is controlled remotely with SCPI commands. TestFlow's instrument library holds 59 commands for it, 13 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the ZNA26 on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Rohde & Schwarz ZNA26, ZNA43, ZNA50, ZNA67, ZNB20, ZNB26, ZNB3000, ZNB4, ZNB40, ZNB43, ZNB8, ZNBT20, ZNBT26, ZNBT8, or ZND and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Rohde & Schwarz ZNA26's VISA address
# Paste the address of your Rohde & Schwarz ZNA26 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("OUTPut:STATe ON") # RF Power On
print(inst.query("SENSe:FREQuency:STARt?")) # Read Start Freq
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Rohde & Schwarz ZNA26, ZNA43, ZNA50, ZNA67, ZNB20, ZNB26, ZNB3000, ZNB4, ZNB40, ZNB43, ZNB8, ZNBT20, ZNBT26, ZNBT8, or ZND and writes the report from the measured results.
Describe the test in a sentence, for example “rf power on on the ZNA26, then read start freq and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
59 commands in 14 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The ZNA26, ZNA43, ZNA50, ZNA67, ZNB20, ZNB26, ZNB3000, ZNB4, ZNB40, ZNB43, ZNB8, ZNBT20, ZNBT26, ZNBT8, and ZND share this command set in TestFlow's library.
| Command | What it does | Arguments |
|---|---|---|
| SENSe:FREQuency:STARt <Frequency> | Set Start Freq. Sets the sweep start frequency in Hz. | Number |
| SENSe:FREQuency:STOP <Frequency> | Set Stop Freq. Sets the sweep stop frequency in Hz. | Number |
| SENSe:FREQuency:CENTer <Frequency> | Set Center Freq. Sets the sweep center frequency in Hz. | Number |
| SENSe:FREQuency:SPAN <Frequency> | Set Span Freq. Sets the sweep frequency span in Hz. | Number |
| SENSe:FREQuency:CW <Frequency> | Set CW Freq. Sets the CW (fixed) frequency in Hz. | Number |
| SENSe:FREQuency:STARt? | Read Start Freq. Returns the current start frequency in Hz. | None |
| SENSe:FREQuency:STOP? | Read Stop Freq. Returns the current stop frequency in Hz. | None |
| SENSe:SWEep:POINts <Points> | Set Points. Sets the number of sweep points. | Number |
| SENSe:SWEep:POINts? | Read Points. Returns the number of sweep points. | None |
| SENSe:SWEep:TYPE <Type> | Set Sweep Type. Sets the sweep type. | Text: LINear,LOGarithmic,SEGMent,POWer,CW,POINt |
| SENSe:SWEep:TIME <Seconds> | Set Sweep Time. Sets the sweep time in seconds. | Number |
| SENSe:BANDwidth <Bandwidth> | Set IF Bandwidth. Sets the measurement (IF) bandwidth in Hz. | Number |
| SENSe:BANDwidth? | Read IF Bandwidth. Returns the measurement bandwidth in Hz. | None |
| SENSe:AVERage:STATe ON | Average On. Enables trace averaging for the channel. | None |
| SENSe:AVERage:STATe OFF | Average Off. Disables trace averaging for the channel. | None |
| SENSe:AVERage:COUNt <Count> | Set Average Count. Sets the averaging factor (number of sweeps). | Number |
| SENSe:AVERage:CLEar | Average Clear. Restarts the averaging measurement. | None |
| SENSe:CORRection:STATe ON | Correction On. Turns system error correction on. | None |
| SENSe:CORRection:STATe OFF | Correction Off. Turns system error correction off. | None |
| Command | What it does | Arguments |
|---|---|---|
| CALCulate:PARameter:SDEFine '<Name>','<Parameter>' | Define Trace. Creates a trace with the given name and S-parameter (R&S SDEFine syntax). | Text; Text: S11,S12,S13,S14,S21,S22,S23,S24,S31,S32,S33,S34,S41,S42,S43,S44 |
| CALCulate:PARameter:SELect '<Name>' | Select Trace. Selects the trace as active for subsequent CALCulate commands. | Text |
| CALCulate:PARameter:DELete '<Name>' | Delete Trace. Deletes the named trace. | Text |
| CALCulate:PARameter:DELete:ALL | Delete All Traces. Deletes all traces in all channels. | None |
| CALCulate:FORMat <Format> | Set Format. Sets the display format of the active trace. | Text: MLOGarithmic,MLINear,PHASe,UPHase,SMITh,ISMith,POLar,SWR,GDELay,REAL,IMAGinary,COMPlex |
| CALCulate:FORMat? | Read Format. Returns the display format of the active trace. | None |
| CALCulate:DATA? FDATa | Read Formatted Data. Returns the formatted trace data of the active trace. | None |
| CALCulate:DATA? SDATa | Read Complex Data. Returns the complex (real/imag) trace data of the active trace. | None |
| CALCulate:MARKer<Marker>:STATe ON | Marker On. Turns on the specified marker. | Number: 1,2,3,4,5,6,7,8,9,10 |
| CALCulate:MARKer<Marker>:STATe OFF | Marker Off. Turns off the specified marker. | Number: 1,2,3,4,5,6,7,8,9,10 |
| CALCulate:MARKer<Marker>:X <Frequency> | Set Marker X. Sets the stimulus position of the marker in Hz. | Number: 1,2,3,4,5,6,7,8,9,10; Number |
| CALCulate:MARKer<Marker>:Y? | Read Marker Y. Returns the response value at the marker. | Number: 1,2,3,4,5,6,7,8,9,10 |
| CALCulate:MARKer<Marker>:FUNCtion:EXECute MAX | Marker Search Max. Moves the marker to the trace maximum. | Number: 1,2,3,4,5,6,7,8,9,10 |
| CALCulate:MARKer<Marker>:FUNCtion:EXECute MIN | Marker Search Min. Moves the marker to the trace minimum. | Number: 1,2,3,4,5,6,7,8,9,10 |
| Command | What it does | Arguments |
|---|---|---|
| INITiate:CONTinuous:ALL ON | Continuous On. Enables continuous sweep on all channels. | None |
| INITiate:CONTinuous:ALL OFF | Continuous Off. Sets all channels to single-sweep (hold) mode. | None |
| INITiate:IMMediate | Trigger Single. Starts a single sweep when in single-sweep mode. | None |
| Command | What it does | Arguments |
|---|---|---|
| DISPlay:TRACe:Y:SCALe:AUTO ONCE | Autoscale. Autoscales the active trace once to fit the diagram. | None |
| DISPlay:TRACe:Y:SCALe:PDIVision <Per_Div> | Set Y Scale. Sets the vertical scale per division of the active trace. | Number |
| DISPlay:TRACe:Y:SCALe:RLEVel <Level> | Set Ref Level. Sets the reference level of the active trace. | Number |
| Command | What it does | Arguments |
|---|---|---|
| SOURce:POWer <Power> | Set Power Level. Sets the internal source power level in dBm. | Number |
| SOURce:POWer? | Read Power Level. Returns the source power level in dBm. | None |
| Command | What it does | Arguments |
|---|---|---|
| OUTPut:STATe ON | RF Power On. Turns the RF source power on. | None |
| OUTPut:STATe OFF | RF Power Off. Turns the RF source power off. | None |
| Command | What it does | Arguments |
|---|---|---|
| MMEMory:STORe:STATe 1,'<Filename>' | Save State. Saves the instrument setup to the named recall set file. | Text |
| MMEMory:LOAD:STATe 1,'<Filename>' | Recall State. Loads an instrument setup from the named file. | Text |
| Command | What it does | Arguments |
|---|---|---|
| SYSTem:PRESet | Preset. Presets the analyzer to the default setup. | None |
| SYSTem:ERRor:NEXT? | Read Error. Reads the next entry from the error queue. | None |
| Command | What it does | Arguments |
|---|---|---|
| TRIGger:SOURce <Source> | Set Trigger Source. Sets the trigger source for the sweep. | Text: IMMediate,EXTernal,MANual,MULTiple |
| Command | What it does | Arguments |
|---|---|---|
| HCOPy:DESTination 'MMEM';:MMEMory:NAME '<Filename>';:HCOPy:IMMediate | Save Screenshot. Saves a screenshot of the display to the named file. | Text |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Returns the instrument identification string (manufacturer, model, serial, firmware). | None |
| *RST | Reset. Resets the instrument to its default state. | None |
| *CLS | Clear Status. Clears all event registers and the error queue. | None |
| *TST? | Self Test. Runs the self-test and returns 0 for pass or 1 for failure. | None |
| *OPC? | Operation Complete. Returns 1 when all pending operations are complete. | None |
| *WAI | Wait To Continue. Waits until all pending operations are complete before continuing. | None |
| *TRG | Trigger Bus. Generates a bus trigger when the instrument is in wait-for-trigger state. | None |
This reference is generated from TestFlow's instrument library, which is drafted from Rohde & Schwarz's programming documentation for the ZNA26, ZNA43, ZNA50, ZNA67, ZNB20, ZNB26, ZNB3000, ZNB4, ZNB40, ZNB43, ZNB8, ZNBT20, ZNBT26, ZNBT8, and ZND and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Rohde & Schwarz's programming manual: check it before you rely on an edge case. Last updated November 12, 2026.
Yes. The Rohde & Schwarz ZNA26 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 59 SCPI commands for it: 13 queries that read values or state, and 46 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTPut:STATe ON and SENSe:FREQuency:STARt?.
SENSe:FREQuency:STARt? (read start freq). Returns the current start frequency in Hz. The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Rohde & Schwarz ZNA26 through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.