| CONFigure:GPRF:MEASurement1:RFSettings:FREQuency <Freq> | Set Analyzer Frequency. Sets the center frequency of the RF analyzer in Hz. | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:FREQuency? | Read Analyzer Frequency. Reads the center frequency of the RF analyzer in Hz. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:ENPower <Power> | Set Analyzer Expected Power. Sets the expected nominal power of the measured RF signal in dBm. | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:ENPower? | Read Analyzer Expected Power. Reads the expected nominal power of the measured RF signal in dBm. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:EATTenuation <Atten> | Set Analyzer External Attenuation. Sets the external attenuation (or gain, if negative) at the RF input connector in dB (-50 dB to 90 dB). | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:EATTenuation? | Read Analyzer External Attenuation. Reads the external attenuation configured for the RF input path in dB. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:UMARgin <Margin> | Set Analyzer User Margin. Sets the margin added to the expected nominal power to determine the reference power in dB. | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:UMARgin? | Read Analyzer User Margin. Reads the user margin used to determine the reference power in dB. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:MLOFfset <Offset> | Set Analyzer Mixer Level Offset. Varies the input level of the mixer in the analyzer path in dB (-10 dB to 16 dB). | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:MLOFfset? | Read Analyzer Mixer Level Offset. Reads the mixer level offset of the analyzer path in dB. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:FOFFset <Offset> | Set Analyzer Frequency Offset. Sets a positive or negative frequency offset added to the analyzer center frequency in Hz (-100 kHz to 100 kHz). | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:FOFFset? | Read Analyzer Frequency Offset. Reads the analyzer frequency offset in Hz. | None |
| CONFigure:GPRF:MEASurement1:POWer:MODE <Mode> | Set Power Measurement Mode. Selects the power measurement mode used when list mode is off. | Text: POWer,STATistic |
| CONFigure:GPRF:MEASurement1:POWer:MODE? | Read Power Measurement Mode. Reads the selected power measurement mode. | None |
| CONFigure:GPRF:MEASurement1:POWer:REPetition <Repetition> | Set Power Repetition. Selects whether the power measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:POWer:REPetition? | Read Power Repetition. Reads the repetition mode of the power measurement. | None |
| CONFigure:GPRF:MEASurement1:POWer:SCOunt <Count> | Set Power Statistic Count. Specifies the number of measurement intervals per measurement cycle (1 to 100000). | Number |
| CONFigure:GPRF:MEASurement1:POWer:SCOunt? | Read Power Statistic Count. Reads the statistic count of the power measurement. | None |
| CONFigure:GPRF:MEASurement1:POWer:MLENgth <Seconds> | Set Power Measurement Length. Sets the evaluation interval used to calculate one set of current power results (RMS, minimum, maximum) in seconds. | Number |
| CONFigure:GPRF:MEASurement1:POWer:MLENgth? | Read Power Measurement Length. Reads the power measurement evaluation interval in seconds. | None |
| CONFigure:GPRF:MEASurement1:POWer:SLENgth <Seconds> | Set Power Step Length. Sets the time between the beginnings of two consecutive measurement lengths in seconds. | Number |
| CONFigure:GPRF:MEASurement1:POWer:SLENgth? | Read Power Step Length. Reads the step length of the power measurement in seconds. | None |
| CONFigure:GPRF:MEASurement1:POWer:TOUT <Seconds> | Set Power Timeout. Defines a timeout for the power measurement in seconds; 0 s means an infinite timeout. | Number |
| CONFigure:GPRF:MEASurement1:POWer:TOUT? | Read Power Timeout. Reads the power measurement timeout in seconds. | None |
| CONFigure:GPRF:MEASurement1:POWer:PDEFset <Name> | Set Power Predefined Set. Loads a predefined parameter set for the power measurement. | Text |
| CONFigure:GPRF:MEASurement1:POWer:PDEFset? | Read Power Predefined Set. Reads the name of the loaded predefined power parameter set. | None |
| CONFigure:GPRF:MEASurement1:DISPlay <Tab> | Set Measurement Display Tab. Selects the measurement tab shown on the instrument display during remote control. | Text: POWer,SPECtrum,FFTSanalyzer,IQRecorder,IQVSlot,EPSensor |
| CONFigure:GPRF:MEASurement1:DISPlay? | Read Measurement Display Tab. Reads the measurement tab currently shown on the instrument display. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:FSPan <Span> | Set FFT Frequency Span. Configures the frequency span of the FFT spectrum analyzer in Hz. | Number: 1250000,2500000,5000000,10000000,20000000,40000000,80000000,160000000 |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:FSPan? | Read FFT Frequency Span. Reads the frequency span of the FFT spectrum analyzer in Hz. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:FFTLength <Length> | Set FFT Length. Selects the number of samples recorded per measurement interval of the FFT spectrum analyzer. | Number: 1024,2048,4096,8192,16384 |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:FFTLength? | Read FFT Length. Reads the configured FFT length. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:DETector <Detector> | Set FFT Detector. Defines how the spectrum diagram is calculated from the frequency domain samples. | Text: PEAK,RMS |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:DETector? | Read FFT Detector. Reads the detector used by the FFT spectrum analyzer. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:AMODe <Mode> | Set FFT Averaging Mode. Selects whether the average traces are calculated from linear or logarithmic power values. | Text: LINear,LOGarithmic |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:AMODe? | Read FFT Averaging Mode. Reads the averaging mode of the FFT spectrum analyzer. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:REPetition <Repetition> | Set FFT Repetition. Selects whether the FFT measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:REPetition? | Read FFT Repetition. Reads the repetition mode of the FFT spectrum analyzer. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:SCOunt <Count> | Set FFT Statistic Count. Specifies the number of measurement intervals per single shot for the FFT measurement (1 to 1000). | Number |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:SCOunt? | Read FFT Statistic Count. Reads the statistic count of the FFT measurement. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:TOUT <Seconds> | Set FFT Timeout. Defines a timeout for the FFT measurement in seconds. | Number |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:TOUT? | Read FFT Timeout. Reads the FFT measurement timeout in seconds. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:MOEXception ON | Turn On FFT Measure On Exception. Accepts faulty FFT results instead of rejecting them. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:MOEXception OFF | Turn Off FFT Measure On Exception. Rejects faulty FFT results. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:MOEXception? | Read FFT Measure On Exception. Reads whether faulty FFT results are accepted or rejected. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:AMODe <Mode> | Set Spectrum Averaging Mode. Defines how the average traces are calculated from the current spectrum results. | Text: LINear,LOGarithmic |
| CONFigure:GPRF:MEASurement1:SPECtrum:AMODe? | Read Spectrum Averaging Mode. Reads the averaging mode of the spectrum measurement. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:REPetition <Repetition> | Set Spectrum Repetition. Selects whether the spectrum measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:SPECtrum:REPetition? | Read Spectrum Repetition. Reads the repetition mode of the spectrum measurement. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:SCOunt <Count> | Set Spectrum Statistic Count. Specifies the number of measurement intervals per single shot for the spectrum measurement (1 to 1000). | Number |
| CONFigure:GPRF:MEASurement1:SPECtrum:SCOunt? | Read Spectrum Statistic Count. Reads the statistic count of the spectrum measurement. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:TOUT <Seconds> | Set Spectrum Timeout. Defines a timeout for the spectrum measurement in seconds; 0 s means an infinite timeout. | Number |
| CONFigure:GPRF:MEASurement1:SPECtrum:TOUT? | Read Spectrum Timeout. Reads the spectrum measurement timeout in seconds. | None |
| CONFigure:GPRF:MEASurement1:IQVSlot:SCOunt <Count> | Set IQVSlot Step Count. Defines the number of measurement steps per subsweep of the I/Q vs slot measurement (1 to 3000). | Number |
| CONFigure:GPRF:MEASurement1:IQVSlot:SCOunt? | Read IQVSlot Step Count. Reads the number of measurement steps of the I/Q vs slot measurement. | None |
| CONFigure:GPRF:MEASurement1:IQVSlot:MLENgth <Seconds> | Set IQVSlot Measurement Length. Sets the length of the evaluation intervals of the I/Q vs slot measurement in seconds. | Number |
| CONFigure:GPRF:MEASurement1:IQVSlot:MLENgth? | Read IQVSlot Measurement Length. Reads the evaluation interval length of the I/Q vs slot measurement in seconds. | None |
| CONFigure:GPRF:MEASurement1:IQVSlot:SLENgth <Seconds> | Set IQVSlot Step Length. Sets the time between the beginnings of two consecutive measurement steps in seconds. | Number |
| CONFigure:GPRF:MEASurement1:IQVSlot:SLENgth? | Read IQVSlot Step Length. Reads the step length of the I/Q vs slot measurement in seconds. | None |
| CONFigure:GPRF:MEASurement1:IQVSlot:FTYPe <Filter> | Set IQVSlot Filter Type. Selects the IF filter type used by the I/Q vs slot measurement. | Text: GAUSs,NYQuist,NY1Mhz |
| CONFigure:GPRF:MEASurement1:IQVSlot:FTYPe? | Read IQVSlot Filter Type. Reads the IF filter type of the I/Q vs slot measurement. | None |
| CONFigure:GPRF:MEASurement1:IQVSlot:FELimit <Limit> | Set IQVSlot Frequency Estimation Limit. Defines the frequency estimation limit as signal level relative to the nominal power in dB (-100 dB to 0 dB). | Number |
| CONFigure:GPRF:MEASurement1:IQVSlot:FELimit? | Read IQVSlot Frequency Estimation Limit. Reads the frequency estimation limit of the I/Q vs slot measurement in dB. | None |
| CONFigure:GPRF:MEASurement1:IQVSlot:REPetition <Repetition> | Set IQVSlot Repetition. Selects whether the I/Q vs slot measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:IQVSlot:REPetition? | Read IQVSlot Repetition. Reads the repetition mode of the I/Q vs slot measurement. | None |
| CONFigure:GPRF:MEASurement1:IQVSlot:TOUT <Seconds> | Set IQVSlot Timeout. Defines a timeout for the I/Q vs slot measurement in seconds. | Number |
| CONFigure:GPRF:MEASurement1:IQVSlot:TOUT? | Read IQVSlot Timeout. Reads the I/Q vs slot measurement timeout in seconds. | None |
| CONFigure:GPRF:MEASurement1:EPSensor:FREQuency <Freq> | Set Power Sensor Frequency. Specifies the input frequency at the external power sensor in Hz for frequency correction. | Number |
| CONFigure:GPRF:MEASurement1:EPSensor:FREQuency? | Read Power Sensor Frequency. Reads the correction frequency of the external power sensor in Hz. | None |
| CONFigure:GPRF:MEASurement1:EPSensor:RESolution <Resolution> | Set Power Sensor Resolution. Defines the number of displayed digits of the external power sensor results. | Text: PD0,PD1,PD2,PD3 |
| CONFigure:GPRF:MEASurement1:EPSensor:RESolution? | Read Power Sensor Resolution. Reads the result resolution of the external power sensor measurement. | None |
| CONFigure:GPRF:MEASurement1:EPSensor:REPetition <Repetition> | Set Power Sensor Repetition. Selects whether the external power sensor measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:EPSensor:REPetition? | Read Power Sensor Repetition. Reads the repetition mode of the external power sensor measurement. | None |
| CONFigure:GPRF:MEASurement1:EPSensor:SCOunt <Count> | Set Power Sensor Statistic Count. Specifies the statistic count of the external power sensor measurement (1 to 1000). | Number |
| CONFigure:GPRF:MEASurement1:EPSensor:SCOunt? | Read Power Sensor Statistic Count. Reads the statistic count of the external power sensor measurement. | None |
| CONFigure:GPRF:MEASurement1:EPSensor:TOUT <Seconds> | Set Power Sensor Timeout. Defines a timeout for the external power sensor measurement in seconds. | Number |
| CONFigure:GPRF:MEASurement1:EPSensor:TOUT? | Read Power Sensor Timeout. Reads the external power sensor measurement timeout in seconds. | None |
| CONFigure:LTE:SIGNaling1:PCC:BAND <Band> | Set LTE Operating Band. Selects the operating band of the LTE primary component carrier; the allowed range depends on the duplex mode. | Text: OB1,OB2,OB3,OB4,OB5,OB7,OB8,OB12,OB13,OB17,OB20,OB25,OB28,OB38,OB39,OB40,OB41,OB42 |
| CONFigure:LTE:SIGNaling1:PCC:BAND? | Read LTE Operating Band. Reads the selected LTE operating band of the primary component carrier. | None |
| CONFigure:LTE:SIGNaling1:PCC:DMODe <Mode> | Set LTE Duplex Mode. Selects the duplex mode of the LTE primary component carrier. | Text: FDD,TDD |
| CONFigure:LTE:SIGNaling1:PCC:DMODe? | Read LTE Duplex Mode. Reads the duplex mode of the LTE primary component carrier. | None |
| CONFigure:LTE:SIGNaling1:CELL:BANDwidth:PCC:DL <Bandwidth> | Set LTE Cell Bandwidth. Defines the DL cell bandwidth (1.4, 3, 5, 10, 15 or 20 MHz); the PCC DL bandwidth is also used for the UL. | Text: B014,B030,B050,B100,B150,B200 |
| CONFigure:LTE:SIGNaling1:CELL:BANDwidth:PCC:DL? | Read LTE Cell Bandwidth. Reads the configured LTE downlink cell bandwidth. | None |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:CHANnel:DL <Channel> | Set LTE Downlink Channel. Sets the downlink EARFCN of the primary component carrier; the uplink channel is calculated automatically. | Number |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:CHANnel:DL? | Read LTE Downlink Channel. Reads the downlink EARFCN of the primary component carrier. | None |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:CHANnel:UL <Channel> | Set LTE Uplink Channel. Sets the uplink EARFCN of the primary component carrier; the downlink channel is calculated automatically. | Number |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:CHANnel:UL? | Read LTE Uplink Channel. Reads the uplink EARFCN of the primary component carrier. | None |
| CONFigure:LTE:SIGNaling1:DL:PCC:RSEPre:LEVel <Level> | Set LTE RS EPRE Level. Defines the energy per resource element of the cell-specific reference signal in dBm/15 kHz. | Number |
| CONFigure:LTE:SIGNaling1:DL:PCC:RSEPre:LEVel? | Read LTE RS EPRE Level. Reads the RS EPRE downlink power level in dBm/15 kHz. | None |
| CONFigure:LTE:SIGNaling1:DL:PCC:AWGN <Level> | Set LTE AWGN Level. Specifies the level of the additive white Gaussian noise added to the downlink signal in dBm/15 kHz. | Number |
| CONFigure:LTE:SIGNaling1:DL:PCC:AWGN? | Read LTE AWGN Level. Reads the AWGN noise level of the downlink signal in dBm/15 kHz. | None |
| CONFigure:LTE:SIGNaling1:DL:PCC:OCNG ON | Turn On LTE OCNG. Enables the OFDMA channel noise generator on the downlink. | None |
| CONFigure:LTE:SIGNaling1:DL:PCC:OCNG OFF | Turn Off LTE OCNG. Disables the OFDMA channel noise generator on the downlink. | None |
| CONFigure:LTE:SIGNaling1:DL:PCC:OCNG? | Read LTE OCNG State. Reads whether the OFDMA channel noise generator is enabled. | None |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:ENPower <Power> | Set LTE Expected Nominal Power. Sets the expected nominal power of the uplink signal in dBm for manual nominal power mode. | Number |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:ENPower? | Read LTE Expected Nominal Power. Reads the expected nominal power of the uplink signal in dBm. | None |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:ENPMode <Mode> | Set LTE Nominal Power Mode. Selects whether the expected nominal power is set manually or derived from the uplink power control settings. | Text: MANual,ULPC |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:ENPMode? | Read LTE Nominal Power Mode. Reads the expected nominal power mode of the LTE signaling application. | None |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:UMARgin <Margin> | Set LTE User Margin. Sets the margin added to the expected nominal power to determine the reference level in dB. | Number |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:UMARgin? | Read LTE User Margin. Reads the user margin of the LTE signaling application in dB. | None |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:MLOFfset <Offset> | Set LTE Mixer Level Offset. Varies the input level of the mixer in the analyzer path in dB (-10 dB to 10 dB). | Number |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:MLOFfset? | Read LTE Mixer Level Offset. Reads the mixer level offset of the LTE signaling analyzer path in dB. | None |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:UDSeparation <Freq> | Set LTE UL DL Separation. Configures the uplink/downlink frequency separation in Hz; this setting is fixed for most operating bands. | Number |
| CONFigure:LTE:SIGNaling1:RFSettings:PCC:UDSeparation? | Read LTE UL DL Separation. Reads the configured uplink/downlink frequency separation in Hz. | None |
| CONFigure:LTE:SIGNaling1:CELL:CPRefix <Prefix> | Set LTE Cyclic Prefix. Defines whether a normal or an extended cyclic prefix is used. | Text: NORMal,EXTended |
| CONFigure:LTE:SIGNaling1:CELL:CPRefix? | Read LTE Cyclic Prefix. Reads the configured cyclic prefix type. | None |
| CONFigure:LTE:SIGNaling1:CELL:MCC <MCC> | Set LTE Mobile Country Code. Specifies the three-digit mobile country code of the simulated cell (0 to 999). | Number |
| CONFigure:LTE:SIGNaling1:CELL:MCC? | Read LTE Mobile Country Code. Reads the mobile country code of the simulated cell. | None |
| CONFigure:LTE:SIGNaling1:CELL:TAC <TAC> | Set LTE Tracking Area Code. Specifies the tracking area code of the simulated cell (0 to 65535). | Number |
| CONFigure:LTE:SIGNaling1:CELL:TAC? | Read LTE Tracking Area Code. Reads the tracking area code of the simulated cell. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:TRANsmission <Mode> | Set LTE Transmission Mode. Selects the LTE downlink transmission mode of the primary component carrier. | Text: TM1,TM2,TM3,TM4,TM6,TM7,TM8,TM9 |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:TRANsmission? | Read LTE Transmission Mode. Reads the selected LTE downlink transmission mode. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:DCIFormat <Format> | Set LTE DCI Format. Selects the downlink control information format; the allowed values depend on the transmission mode. | Text: D1,D1A,D1B,D2,D2A,D2B,D2C,D61 |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:DCIFormat? | Read LTE DCI Format. Reads the selected downlink control information format. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:NENBantennas <Antennas> | Set LTE Number Of eNB Antennas. Selects the number of downlink transmit antennas for transmission modes 1 to 6. | Text: ONE,TWO,FOUR |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:NENBantennas? | Read LTE Number Of eNB Antennas. Reads the number of configured downlink transmit antennas. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:HDUPlex ON | Turn On LTE Half Duplex. Selects half-duplex operation for the primary component carrier. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:HDUPlex OFF | Turn Off LTE Half Duplex. Selects full-duplex operation for the primary component carrier. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:HDUPlex? | Read LTE Half Duplex State. Reads whether half-duplex operation is enabled. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:TTIBundling ON | Turn On LTE TTI Bundling. Enables TTI bundling for the uplink. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:TTIBundling OFF | Turn Off LTE TTI Bundling. Disables TTI bundling for the uplink. | None |
| CONFigure:LTE:SIGNaling1:CONNection:PCC:TTIBundling? | Read LTE TTI Bundling State. Reads whether TTI bundling is enabled for the uplink. | None |
| CONFigure:LTE:MEASurement1:RFSettings:EATTenuation <Atten> | Set LTE Meas External Attenuation. Sets the external attenuation (or gain, if negative) at the LTE measurement input connector in dB (-50 dB to 90 dB). | Number |
| CONFigure:LTE:MEASurement1:RFSettings:EATTenuation? | Read LTE Meas External Attenuation. Reads the external attenuation of the LTE measurement input path in dB. | None |
| CONFigure:LTE:MEASurement1:RFSettings:ENPower <Power> | Set LTE Meas Expected Power. Sets the expected nominal power of the measured LTE uplink signal in dBm. | Number |
| CONFigure:LTE:MEASurement1:RFSettings:ENPower? | Read LTE Meas Expected Power. Reads the expected nominal power of the measured LTE uplink signal in dBm. | None |
| CONFigure:LTE:MEASurement1:RFSettings:UMARgin <Margin> | Set LTE Meas User Margin. Sets the margin added to the expected nominal power to determine the reference power in dB. | Number |
| CONFigure:LTE:MEASurement1:RFSettings:UMARgin? | Read LTE Meas User Margin. Reads the user margin of the LTE measurement in dB. | None |
| CONFigure:LTE:MEASurement1:RFSettings:FOFFset <Offset> | Set LTE Meas Frequency Offset. Specifies a frequency offset added to the carrier center frequency in Hz (-100 kHz to 100 kHz). | Number |
| CONFigure:LTE:MEASurement1:RFSettings:FOFFset? | Read LTE Meas Frequency Offset. Reads the frequency offset of the LTE measurement in Hz. | None |
| CONFigure:LTE:MEASurement1:RFSettings:MLOFfset <Offset> | Set LTE Meas Mixer Level Offset. Varies the input level of the mixer in the LTE analyzer path in dB (-10 dB to 10 dB). | Number |
| CONFigure:LTE:MEASurement1:RFSettings:MLOFfset? | Read LTE Meas Mixer Level Offset. Reads the mixer level offset of the LTE analyzer path in dB. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MMODe <Mode> | Set LTE Meas Mode. Selects the LTE multi-evaluation measurement mode: normal, TPC, or multi-evaluation list mode. | Text: NORMal,TMODe,MELMode |
| CONFigure:LTE:MEASurement1:MEValuation:MMODe? | Read LTE Meas Mode. Reads the LTE multi-evaluation measurement mode. | None |
| CONFigure:LTE:MEASurement1:MEValuation:REPetition <Repetition> | Set LTE Meas Repetition. Selects whether the LTE multi-evaluation measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:LTE:MEASurement1:MEValuation:REPetition? | Read LTE Meas Repetition. Reads the repetition mode of the LTE multi-evaluation measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:TOUT <Seconds> | Set LTE Meas Timeout. Defines a timeout for the LTE multi-evaluation measurement in seconds. | Number |
| CONFigure:LTE:MEASurement1:MEValuation:TOUT? | Read LTE Meas Timeout. Reads the LTE multi-evaluation measurement timeout in seconds. | None |
| CONFigure:LTE:MEASurement1:MEValuation:SCONdition <Condition> | Set LTE Meas Stop Condition. Qualifies whether the LTE measurement is stopped after a limit failure or continues regardless. | Text: NONE,SLFail |
| CONFigure:LTE:MEASurement1:MEValuation:SCONdition? | Read LTE Meas Stop Condition. Reads the stop condition of the LTE multi-evaluation measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:CPRefix <Prefix> | Set LTE Meas Cyclic Prefix. Selects the cyclic prefix type of the measured LTE uplink signal. | Text: NORMal,EXTended |
| CONFigure:LTE:MEASurement1:MEValuation:CPRefix? | Read LTE Meas Cyclic Prefix. Reads the cyclic prefix type used by the LTE measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:CTYPe <Type> | Set LTE Meas Channel Type. Configures the uplink channel type detection of the LTE multi-evaluation measurement. | Text: AUTO,PUSCh,PUCCh |
| CONFigure:LTE:MEASurement1:MEValuation:CTYPe? | Read LTE Meas Channel Type. Reads the configured uplink channel type of the LTE measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:PFORmat <Format> | Set LTE Meas PUCCH Format. Specifies the PUCCH format of the measured uplink signal according to 3GPP TS 36.211. | Text: F1,F1A,F1B,F2,F2A,F2B,F3 |
| CONFigure:LTE:MEASurement1:MEValuation:PFORmat? | Read LTE Meas PUCCH Format. Reads the configured PUCCH format of the LTE measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MSLot <Slot> | Set LTE Meas Slot. Selects which slots of a subframe are measured by the LTE multi-evaluation measurement. | Text: MS0,MS1,ALL |
| CONFigure:LTE:MEASurement1:MEValuation:MSLot? | Read LTE Meas Slot. Reads the slots measured by the LTE multi-evaluation measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:SCOunt:MODulation <Count> | Set LTE Meas Modulation Statistic Count. Specifies the statistic count of the modulation measurement in slots (1 to 1000). | Number |
| CONFigure:LTE:MEASurement1:MEValuation:SCOunt:MODulation? | Read LTE Meas Modulation Statistic Count. Reads the statistic count of the LTE modulation measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:SCOunt:POWer <Count> | Set LTE Meas Power Statistic Count. Specifies the statistic count of the power measurement in subframes (1 to 1000). | Number |
| CONFigure:LTE:MEASurement1:MEValuation:SCOunt:POWer? | Read LTE Meas Power Statistic Count. Reads the statistic count of the LTE power measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MOEXception ON | Turn On LTE Meas On Exception. Accepts faulty LTE measurement results instead of rejecting them. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MOEXception OFF | Turn Off LTE Meas On Exception. Rejects faulty LTE measurement results. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MOEXception? | Read LTE Meas On Exception. Reads whether faulty LTE measurement results are accepted or rejected. | None |