The Rohde & Schwarz CMW100 radio test set is controlled remotely with SCPI commands. TestFlow's instrument library holds 305 commands for it, 160 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the CMW100 on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Rohde & Schwarz CMW100 and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Rohde & Schwarz CMW100's VISA address
# Paste the address of your Rohde & Schwarz CMW100 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("SOURce:GPRF:GENerator1:STATe ON") # Turn On Generator
print(inst.query("CONFigure:GPRF:MEASurement1:RFSettings:FREQuency?")) # Read Analyzer Frequency
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Rohde & Schwarz CMW100 and writes the report from the measured results.
Describe the test in a sentence, for example “turn on generator on the CMW100, then read analyzer frequency and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
305 commands in 15 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| CONFigure:GPRF:GENerator1:CMWS:USAGe:TX <Port>, ON | Enable Generator Port. Activates an individual RF output port of the connector bench for the generated signal. | Text: R11,R12,R13,R14,R15,R16,R17,R18 |
| CONFigure:GPRF:GENerator1:CMWS:USAGe:TX <Port>, OFF | Disable Generator Port. Deactivates an individual RF output port of the connector bench. | Text: R11,R12,R13,R14,R15,R16,R17,R18 |
| CONFigure:GPRF:GENerator1:CMWS:USAGe:TX? <Port> | Read Generator Port Usage. Queries whether an individual RF output port is active. | Text: R11,R12,R13,R14,R15,R16,R17,R18 |
| CONFigure:GPRF:MEASurement1:RFSettings:FREQuency <Freq> | Set Analyzer Frequency. Sets the center frequency of the RF analyzer in Hz. | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:FREQuency? | Read Analyzer Frequency. Reads the center frequency of the RF analyzer in Hz. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:ENPower <Power> | Set Analyzer Expected Power. Sets the expected nominal power of the measured RF signal in dBm. | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:ENPower? | Read Analyzer Expected Power. Reads the expected nominal power of the measured RF signal in dBm. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:EATTenuation <Atten> | Set Analyzer External Attenuation. Sets the external attenuation (or gain, if negative) at the RF input connector in dB (-50 dB to 90 dB). | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:EATTenuation? | Read Analyzer External Attenuation. Reads the external attenuation configured for the RF input path in dB. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:UMARgin <Margin> | Set Analyzer User Margin. Sets the margin added to the expected nominal power to determine the reference power in dB. | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:UMARgin? | Read Analyzer User Margin. Reads the user margin used to determine the reference power in dB. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:MLOFfset <Offset> | Set Analyzer Mixer Level Offset. Varies the input level of the mixer in the analyzer path in dB (-10 dB to 16 dB). | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:MLOFfset? | Read Analyzer Mixer Level Offset. Reads the mixer level offset of the analyzer path in dB. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:FOFFset <Offset> | Set Analyzer Frequency Offset. Sets a positive or negative frequency offset added to the analyzer center frequency in Hz (-100 kHz to 100 kHz). | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:FOFFset? | Read Analyzer Frequency Offset. Reads the analyzer frequency offset in Hz. | None |
| CONFigure:GPRF:MEASurement1:POWer:MODE <Mode> | Set Power Measurement Mode. Selects the power measurement mode used when list mode is off. | Text: POWer,STATistic |
| CONFigure:GPRF:MEASurement1:POWer:MODE? | Read Power Measurement Mode. Reads the selected power measurement mode. | None |
| CONFigure:GPRF:MEASurement1:POWer:REPetition <Repetition> | Set Power Repetition. Selects whether the power measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:POWer:REPetition? | Read Power Repetition. Reads the repetition mode of the power measurement. | None |
| CONFigure:GPRF:MEASurement1:POWer:SCOunt <Count> | Set Power Statistic Count. Specifies the number of measurement intervals per measurement cycle (1 to 100000). | Number |
| CONFigure:GPRF:MEASurement1:POWer:SCOunt? | Read Power Statistic Count. Reads the statistic count of the power measurement. | None |
| CONFigure:GPRF:MEASurement1:POWer:MLENgth <Seconds> | Set Power Measurement Length. Sets the evaluation interval used to calculate one set of current power results (RMS, minimum, maximum) in seconds. | Number |
| CONFigure:GPRF:MEASurement1:POWer:MLENgth? | Read Power Measurement Length. Reads the power measurement evaluation interval in seconds. | None |
| CONFigure:GPRF:MEASurement1:POWer:SLENgth <Seconds> | Set Power Step Length. Sets the time between the beginnings of two consecutive measurement lengths in seconds. | Number |
| CONFigure:GPRF:MEASurement1:POWer:SLENgth? | Read Power Step Length. Reads the step length of the power measurement in seconds. | None |
| CONFigure:GPRF:MEASurement1:POWer:TOUT <Seconds> | Set Power Timeout. Defines a timeout for the power measurement in seconds; 0 s means an infinite timeout. | Number |
| CONFigure:GPRF:MEASurement1:POWer:TOUT? | Read Power Timeout. Reads the power measurement timeout in seconds. | None |
| CONFigure:GPRF:MEASurement1:POWer:PDEFset <Name> | Set Power Predefined Set. Loads a predefined parameter set for the power measurement. | Text |
| CONFigure:GPRF:MEASurement1:POWer:PDEFset? | Read Power Predefined Set. Reads the name of the loaded predefined power parameter set. | None |
| CONFigure:GPRF:MEASurement1:DISPlay <Tab> | Set Measurement Display Tab. Selects the measurement tab shown on the instrument display during remote control. | Text: POWer,SPECtrum,FFTSanalyzer,IQRecorder,IQVSlot,EPSensor |
| CONFigure:GPRF:MEASurement1:DISPlay? | Read Measurement Display Tab. Reads the measurement tab currently shown on the instrument display. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:FSPan <Span> | Set FFT Frequency Span. Configures the frequency span of the FFT spectrum analyzer in Hz. | Number: 1250000,2500000,5000000,10000000,20000000,40000000,80000000,160000000 |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:FSPan? | Read FFT Frequency Span. Reads the frequency span of the FFT spectrum analyzer in Hz. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:FFTLength <Length> | Set FFT Length. Selects the number of samples recorded per measurement interval of the FFT spectrum analyzer. | Number: 1024,2048,4096,8192,16384 |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:FFTLength? | Read FFT Length. Reads the configured FFT length. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:DETector <Detector> | Set FFT Detector. Defines how the spectrum diagram is calculated from the frequency domain samples. | Text: PEAK,RMS |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:DETector? | Read FFT Detector. Reads the detector used by the FFT spectrum analyzer. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:AMODe <Mode> | Set FFT Averaging Mode. Selects whether the average traces are calculated from linear or logarithmic power values. | Text: LINear,LOGarithmic |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:AMODe? | Read FFT Averaging Mode. Reads the averaging mode of the FFT spectrum analyzer. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:REPetition <Repetition> | Set FFT Repetition. Selects whether the FFT measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:REPetition? | Read FFT Repetition. Reads the repetition mode of the FFT spectrum analyzer. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:SCOunt <Count> | Set FFT Statistic Count. Specifies the number of measurement intervals per single shot for the FFT measurement (1 to 1000). | Number |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:SCOunt? | Read FFT Statistic Count. Reads the statistic count of the FFT measurement. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:TOUT <Seconds> | Set FFT Timeout. Defines a timeout for the FFT measurement in seconds. | Number |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:TOUT? | Read FFT Timeout. Reads the FFT measurement timeout in seconds. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:MOEXception ON | Turn On FFT Measure On Exception. Accepts faulty FFT results instead of rejecting them. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:MOEXception OFF | Turn Off FFT Measure On Exception. Rejects faulty FFT results. | None |
| CONFigure:GPRF:MEASurement1:FFTSanalyzer:MOEXception? | Read FFT Measure On Exception. Reads whether faulty FFT results are accepted or rejected. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:AMODe <Mode> | Set Spectrum Averaging Mode. Defines how the average traces are calculated from the current spectrum results. | Text: LINear,LOGarithmic |
| CONFigure:GPRF:MEASurement1:SPECtrum:AMODe? | Read Spectrum Averaging Mode. Reads the averaging mode of the spectrum measurement. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:REPetition <Repetition> | Set Spectrum Repetition. Selects whether the spectrum measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:SPECtrum:REPetition? | Read Spectrum Repetition. Reads the repetition mode of the spectrum measurement. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:SCOunt <Count> | Set Spectrum Statistic Count. Specifies the number of measurement intervals per single shot for the spectrum measurement (1 to 1000). | Number |
| CONFigure:GPRF:MEASurement1:SPECtrum:SCOunt? | Read Spectrum Statistic Count. Reads the statistic count of the spectrum measurement. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:TOUT <Seconds> | Set Spectrum Timeout. Defines a timeout for the spectrum measurement in seconds; 0 s means an infinite timeout. | Number |
| CONFigure:GPRF:MEASurement1:SPECtrum:TOUT? | Read Spectrum Timeout. Reads the spectrum measurement timeout in seconds. | None |
| CONFigure:LTE:MEASurement1:RFSettings:EATTenuation <Atten> | Set LTE Meas External Attenuation. Sets the external attenuation (or gain, if negative) at the LTE measurement input connector in dB (-50 dB to 90 dB). | Number |
| CONFigure:LTE:MEASurement1:RFSettings:EATTenuation? | Read LTE Meas External Attenuation. Reads the external attenuation of the LTE measurement input path in dB. | None |
| CONFigure:LTE:MEASurement1:RFSettings:ENPower <Power> | Set LTE Meas Expected Power. Sets the expected nominal power of the measured LTE uplink signal in dBm. | Number |
| CONFigure:LTE:MEASurement1:RFSettings:ENPower? | Read LTE Meas Expected Power. Reads the expected nominal power of the measured LTE uplink signal in dBm. | None |
| CONFigure:LTE:MEASurement1:RFSettings:UMARgin <Margin> | Set LTE Meas User Margin. Sets the margin added to the expected nominal power to determine the reference power in dB. | Number |
| CONFigure:LTE:MEASurement1:RFSettings:UMARgin? | Read LTE Meas User Margin. Reads the user margin of the LTE measurement in dB. | None |
| CONFigure:LTE:MEASurement1:RFSettings:FOFFset <Offset> | Set LTE Meas Frequency Offset. Specifies a frequency offset added to the carrier center frequency in Hz (-100 kHz to 100 kHz). | Number |
| CONFigure:LTE:MEASurement1:RFSettings:FOFFset? | Read LTE Meas Frequency Offset. Reads the frequency offset of the LTE measurement in Hz. | None |
| CONFigure:LTE:MEASurement1:RFSettings:MLOFfset <Offset> | Set LTE Meas Mixer Level Offset. Varies the input level of the mixer in the LTE analyzer path in dB (-10 dB to 10 dB). | Number |
| CONFigure:LTE:MEASurement1:RFSettings:MLOFfset? | Read LTE Meas Mixer Level Offset. Reads the mixer level offset of the LTE analyzer path in dB. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MMODe <Mode> | Set LTE Meas Mode. Selects the LTE multi-evaluation measurement mode: normal, TPC, or multi-evaluation list mode. | Text: NORMal,TMODe,MELMode |
| CONFigure:LTE:MEASurement1:MEValuation:MMODe? | Read LTE Meas Mode. Reads the LTE multi-evaluation measurement mode. | None |
| CONFigure:LTE:MEASurement1:MEValuation:REPetition <Repetition> | Set LTE Meas Repetition. Selects whether the LTE multi-evaluation measurement is stopped after one cycle or repeated continuously. | Text: SINGleshot,CONTinuous |
| CONFigure:LTE:MEASurement1:MEValuation:REPetition? | Read LTE Meas Repetition. Reads the repetition mode of the LTE multi-evaluation measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:TOUT <Seconds> | Set LTE Meas Timeout. Defines a timeout for the LTE multi-evaluation measurement in seconds. | Number |
| CONFigure:LTE:MEASurement1:MEValuation:TOUT? | Read LTE Meas Timeout. Reads the LTE multi-evaluation measurement timeout in seconds. | None |
| CONFigure:LTE:MEASurement1:MEValuation:SCONdition <Condition> | Set LTE Meas Stop Condition. Qualifies whether the LTE measurement is stopped after a limit failure or continues regardless. | Text: NONE,SLFail |
| CONFigure:LTE:MEASurement1:MEValuation:SCONdition? | Read LTE Meas Stop Condition. Reads the stop condition of the LTE multi-evaluation measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:CPRefix <Prefix> | Set LTE Meas Cyclic Prefix. Selects the cyclic prefix type of the measured LTE uplink signal. | Text: NORMal,EXTended |
| CONFigure:LTE:MEASurement1:MEValuation:CPRefix? | Read LTE Meas Cyclic Prefix. Reads the cyclic prefix type used by the LTE measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:CTYPe <Type> | Set LTE Meas Channel Type. Configures the uplink channel type detection of the LTE multi-evaluation measurement. | Text: AUTO,PUSCh,PUCCh |
| CONFigure:LTE:MEASurement1:MEValuation:CTYPe? | Read LTE Meas Channel Type. Reads the configured uplink channel type of the LTE measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:PFORmat <Format> | Set LTE Meas PUCCH Format. Specifies the PUCCH format of the measured uplink signal according to 3GPP TS 36.211. | Text: F1,F1A,F1B,F2,F2A,F2B,F3 |
| CONFigure:LTE:MEASurement1:MEValuation:PFORmat? | Read LTE Meas PUCCH Format. Reads the configured PUCCH format of the LTE measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MSLot <Slot> | Set LTE Meas Slot. Selects which slots of a subframe are measured by the LTE multi-evaluation measurement. | Text: MS0,MS1,ALL |
| CONFigure:LTE:MEASurement1:MEValuation:MSLot? | Read LTE Meas Slot. Reads the slots measured by the LTE multi-evaluation measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:SCOunt:MODulation <Count> | Set LTE Meas Modulation Statistic Count. Specifies the statistic count of the modulation measurement in slots (1 to 1000). | Number |
| CONFigure:LTE:MEASurement1:MEValuation:SCOunt:MODulation? | Read LTE Meas Modulation Statistic Count. Reads the statistic count of the LTE modulation measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:SCOunt:POWer <Count> | Set LTE Meas Power Statistic Count. Specifies the statistic count of the power measurement in subframes (1 to 1000). | Number |
| CONFigure:LTE:MEASurement1:MEValuation:SCOunt:POWer? | Read LTE Meas Power Statistic Count. Reads the statistic count of the LTE power measurement. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MOEXception ON | Turn On LTE Meas On Exception. Accepts faulty LTE measurement results instead of rejecting them. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MOEXception OFF | Turn Off LTE Meas On Exception. Rejects faulty LTE measurement results. | None |
| CONFigure:LTE:MEASurement1:MEValuation:MOEXception? | Read LTE Meas On Exception. Reads whether faulty LTE measurement results are accepted or rejected. | None |
| CONFigure:NRSub:MEASurement1:BAND <Band> | Set NR Meas Band. Selects the NR operating band (allowed values depend on the duplex mode). | Text: OB1,OB2,OB3,OB5,OB7,OB8,OB12,OB20,OB28,OB38,OB40,OB41,OB66,OB71,OB77,OB78,OB79 |
| CONFigure:NRSub:MEASurement1:BAND? | Read NR Meas Band. Reads the selected NR operating band (allowed values depend on the duplex mode). | None |
| CONFigure:NRSub:MEASurement1:MEValuation:DMODe <Mode> | Set NR Meas Duplex Mode. Selects the duplex mode of the NR signal. | Text: FDD,TDD |
| CONFigure:NRSub:MEASurement1:MEValuation:DMODe? | Read NR Meas Duplex Mode. Reads the selected duplex mode of the NR signal. | None |
| CONFigure:NRSub:MEASurement1:RFSettings:FREQuency <Frequency> | Set NR Meas Frequency. Sets the analyzer center frequency in Hz. | Number |
| CONFigure:NRSub:MEASurement1:RFSettings:FREQuency? | Read NR Meas Frequency. Reads the analyzer center frequency in Hz. | None |
| CONFigure:NRSub:MEASurement1:RFSettings:ENPower <Power> | Set NR Meas Expected Power. Sets the expected nominal power of the measured NR signal in dBm. | Number |
| CONFigure:NRSub:MEASurement1:RFSettings:ENPower? | Read NR Meas Expected Power. Reads the expected nominal power of the measured NR signal in dBm. | None |
| CONFigure:NRSub:MEASurement1:RFSettings:EATTenuation <Atten> | Set NR Meas External Attenuation. Sets the external attenuation at the RF input in dB. | Number: -50,90 |
| CONFigure:NRSub:MEASurement1:RFSettings:EATTenuation? | Read NR Meas External Attenuation. Reads the external attenuation at the RF input in dB. | None |
| CONFigure:NRSub:MEASurement1:RFSettings:UMARgin <Margin> | Set NR Meas User Margin. Sets the user margin added to the expected nominal power in dB. | Number |
| CONFigure:NRSub:MEASurement1:RFSettings:UMARgin? | Read NR Meas User Margin. Reads the user margin added to the expected nominal power in dB. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:CBANdwidth <Bandwidth> | Set NR Meas Channel Bandwidth. Selects the NR channel bandwidth. | Text: B005,B010,B015,B020,B025,B030,B040,B050,B060,B070,B080,B090,B100 |
| CONFigure:NRSub:MEASurement1:MEValuation:CBANdwidth? | Read NR Meas Channel Bandwidth. Reads the selected NR channel bandwidth. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:SCSPacing <Spacing> | Set NR Meas Subcarrier Spacing. Selects the subcarrier spacing. | Text: S15K,S30K,S60K |
| CONFigure:NRSub:MEASurement1:MEValuation:SCSPacing? | Read NR Meas Subcarrier Spacing. Reads the selected subcarrier spacing. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:PLCid <Cell_ID> | Set NR Meas Physical Cell ID. Sets the physical layer cell ID of the measured signal. | Number: 0,1007 |
| CONFigure:NRSub:MEASurement1:MEValuation:PLCid? | Read NR Meas Physical Cell ID. Reads the physical layer cell ID of the measured signal. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:DFTPrecoding ON | Turn On NR Meas DFT Precoding. Enables transform precoding (DFT-s-OFDM) of the measured PUSCH. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:DFTPrecoding OFF | Turn Off NR Meas DFT Precoding. Disables transform precoding (CP-OFDM). | None |
| CONFigure:NRSub:MEASurement1:MEValuation:NSUBframes <Number> | Set NR Meas Subframes. Sets the number of subframes to be evaluated (1 to 250). | Number: 1,250 |
| CONFigure:NRSub:MEASurement1:MEValuation:NSUBframes? | Read NR Meas Subframes. Reads the number of subframes to be evaluated (1 to 250). | None |
| CONFigure:NRSub:MEASurement1:MEValuation:REPetition <Repetition> | Set NR Meas Repetition. Selects single-shot or continuous NR multi-evaluation measurement. | Text: SINGleshot,CONTinuous |
| CONFigure:NRSub:MEASurement1:MEValuation:REPetition? | Read NR Meas Repetition. Reads the selection: single-shot or continuous NR multi-evaluation measurement. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:SCOunt:MODulation <Count> | Set NR Meas Modulation Count. Sets the statistic count of modulation measurements in slots. | Number: 1,1000 |
| CONFigure:NRSub:MEASurement1:MEValuation:SCOunt:MODulation? | Read NR Meas Modulation Count. Reads the statistic count of modulation measurements in slots. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:SCOunt:POWer <Count> | Set NR Meas Power Count. Sets the statistic count of power measurements. | Number |
| CONFigure:NRSub:MEASurement1:MEValuation:SCOunt:POWer? | Read NR Meas Power Count. Reads the statistic count of power measurements. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:TOUT <Seconds> | Set NR Meas Timeout. Defines a timeout for the measurement in seconds. | Number |
| CONFigure:NRSub:MEASurement1:MEValuation:TOUT? | Read NR Meas Timeout. Reads the timeout for the measurement in seconds. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:SCONdition <Condition> | Set NR Meas Stop Condition. Defines whether the measurement stops on a limit failure. | Text: NONE,SLFail |
| CONFigure:NRSub:MEASurement1:MEValuation:SCONdition? | Read NR Meas Stop Condition. Reads whether the measurement stops on a limit failure. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:MOEXception ON | Turn On NR Meas On Exception. Results identified as faulty are never rejected. | None |
| CONFigure:NRSub:MEASurement1:MEValuation:MOEXception OFF | Turn Off NR Meas On Exception. Faulty results are rejected. | None |
| CONFigure:WLAN:MEASurement1:ISIGnal:STANdard <Standard> | Set WLAN Meas Standard. Selects the IEEE 802.11 standard of the measured signal (b/g DSSS, a/g OFDM, n, ac, ax, p). | Text: DSSS,LOFDm,HTOFdm,VHTofdm,HEOFdm,POFDm |
| CONFigure:WLAN:MEASurement1:ISIGnal:STANdard? | Read WLAN Meas Standard. Reads the selected IEEE 802.11 standard of the measured signal (b/g DSSS, a/g OFDM, n, ac, ax, p). | None |
| CONFigure:WLAN:MEASurement1:ISIGnal:BWIDth <Bandwidth> | Set WLAN Meas Bandwidth. Selects the channel bandwidth of the measured signal. | Text: BW05mhz,BW10mhz,BW20mhz,BW40mhz,BW80mhz,BW16mhz,BW88mhz |
| CONFigure:WLAN:MEASurement1:ISIGnal:BWIDth? | Read WLAN Meas Bandwidth. Reads the selected channel bandwidth of the measured signal. | None |
| CONFigure:WLAN:MEASurement1:RFSettings:FREQuency <Frequency> | Set WLAN Meas Frequency. Sets the analyzer center frequency in Hz (center of the received WLAN channel). | Number |
| CONFigure:WLAN:MEASurement1:RFSettings:FREQuency? | Read WLAN Meas Frequency. Reads the analyzer center frequency in Hz (center of the received WLAN channel). | None |
| CONFigure:WLAN:MEASurement1:RFSettings:FREQuency:BAND <Band> | Set WLAN Meas Frequency Band. Selects the WLAN frequency band. | Text: B24Ghz,B4GHz,B5GHz,B6GHz |
| CONFigure:WLAN:MEASurement1:RFSettings:FREQuency:BAND? | Read WLAN Meas Frequency Band. Reads the selected WLAN frequency band. | None |
| CONFigure:WLAN:MEASurement1:RFSettings:ENPower <Power> | Set WLAN Meas Expected Power. Sets the expected nominal power of the measured RF signal in dBm. | Number |
| CONFigure:WLAN:MEASurement1:RFSettings:ENPower? | Read WLAN Meas Expected Power. Reads the expected nominal power of the measured RF signal in dBm. | None |
| CONFigure:WLAN:MEASurement1:RFSettings:EATTenuation <Atten> | Set WLAN Meas External Attenuation. Sets the external attenuation at the RF input in dB (-50 to 90). | Number: -50,90 |
| CONFigure:WLAN:MEASurement1:RFSettings:EATTenuation? | Read WLAN Meas External Attenuation. Reads the external attenuation at the RF input in dB (-50 to 90). | None |
| CONFigure:WLAN:MEASurement1:RFSettings:UMARgin <Margin> | Set WLAN Meas User Margin. Sets the margin added to the expected nominal power to determine the reference power in dB. | Number |
| CONFigure:WLAN:MEASurement1:RFSettings:UMARgin? | Read WLAN Meas User Margin. Reads the margin added to the expected nominal power to determine the reference power in dB. | None |
| CONFigure:WLAN:MEASurement1:MEValuation:REPetition <Repetition> | Set WLAN Meas Repetition. Selects single-shot or continuous WLAN multi-evaluation measurement. | Text: SINGleshot,CONTinuous |
| CONFigure:WLAN:MEASurement1:MEValuation:REPetition? | Read WLAN Meas Repetition. Reads the selection: single-shot or continuous WLAN multi-evaluation measurement. | None |
| CONFigure:WLAN:MEASurement1:MEValuation:SCOunt:MODulation <Count> | Set WLAN Meas Modulation Count. Sets the statistic count for modulation measurements (1 to 2000). | Number: 1,2000 |
| CONFigure:WLAN:MEASurement1:MEValuation:SCOunt:MODulation? | Read WLAN Meas Modulation Count. Reads the statistic count for modulation measurements (1 to 2000). | None |
| CONFigure:WLAN:MEASurement1:MEValuation:SCOunt:TSMask <Count> | Set WLAN Meas Spectrum Mask Count. Sets the statistic count for the transmit spectrum mask measurement (1 to 1000). | Number: 1,1000 |
| CONFigure:WLAN:MEASurement1:MEValuation:SCOunt:TSMask? | Read WLAN Meas Spectrum Mask Count. Reads the statistic count for the transmit spectrum mask measurement (1 to 1000). | None |
| CONFigure:WLAN:MEASurement1:MEValuation:TOUT <Seconds> | Set WLAN Meas Timeout. Defines a timeout for the WLAN measurement in seconds. | Number |
| CONFigure:WLAN:MEASurement1:MEValuation:TOUT? | Read WLAN Meas Timeout. Reads the timeout for the WLAN measurement in seconds. | None |
| CONFigure:WLAN:MEASurement1:MEValuation:SCONdition <Condition> | Set WLAN Meas Stop Condition. Defines whether the measurement stops on a limit failure. | Text: NONE,SLFail |
| CONFigure:WLAN:MEASurement1:MEValuation:SCONdition? | Read WLAN Meas Stop Condition. Reads whether the measurement stops on a limit failure. | None |
| CONFigure:WLAN:MEASurement1:MEValuation:MOEXception ON | Turn On WLAN Meas On Exception. Results identified as faulty are never rejected. | None |
| CONFigure:WLAN:MEASurement1:MEValuation:MOEXception OFF | Turn Off WLAN Meas On Exception. Faulty results are rejected. | None |
| CONFigure:BASE:FDCorrection:CTABle:CREate <Table_Name>, <Freq_Loss_Pairs> | Create Correction Table. Creates a frequency-dependent attenuation table; pairs are <Frequency Hz>, <Correction dB> (e.g. 1E9, 0.5, 2E9, 0.7). | Text; Text |
| CONFigure:BASE:FDCorrection:CTABle:DELete <Table_Name> | Delete Correction Table. Deletes the named correction table. | Text |
| CONFigure:BASE:FDCorrection:CTABle:CATalog? | Read Correction Table Catalog. Returns the names of all existing correction tables. | None |
| CONFigure:CMWS:FDCorrection:ACTivate:RX <Port>, <Table_Name> | Activate Port Correction RX. Activates a correction table for the input direction of an RF port. | Text: R11,R12,R13,R14,R15,R16,R17,R18; Text |
| CONFigure:CMWS:FDCorrection:ACTivate:TX <Port>, <Table_Name> | Activate Port Correction TX. Activates a correction table for the output direction of an RF port. | Text: R11,R12,R13,R14,R15,R16,R17,R18; Text |
| CONFigure:CMWS:FDCorrection:DEACtivate:RX <Port> | Deactivate Port Correction RX. Deactivates the correction table of an RF port in input direction. | Text: R11,R12,R13,R14,R15,R16,R17,R18 |
| CONFigure:CMWS:FDCorrection:DEACtivate:TX <Port> | Deactivate Port Correction TX. Deactivates the correction table of an RF port in output direction. | Text: R11,R12,R13,R14,R15,R16,R17,R18 |
| CONFigure:CMWS:FDCorrection:DEACtivate:ALL | Deactivate All Port Corrections. Deactivates all port correction tables in both directions. | None |
| Command | What it does | Arguments |
|---|---|---|
| SOURce:GPRF:GENerator1:STATe ON | Turn On Generator. Turns the GPRF RF generator on. | None |
| SOURce:GPRF:GENerator1:STATe OFF | Turn Off Generator. Turns the GPRF RF generator off. | None |
| SOURce:GPRF:GENerator1:STATe? | Read Generator State. Reads the generator state (OFF / PENDing / ON / RDY). | None |
| SOURce:GPRF:GENerator1:STATe:ALL? | Read Generator State All. Reads detailed generator state information including the main state and the synchronization state. | None |
| SOURce:GPRF:GENerator1:BBMode <Mode> | Set Generator Baseband Mode. Selects the baseband mode of the generator signal: unmodulated CW, dual tone, or ARB waveform playback. | Text: CW,DTONe,ARB |
| SOURce:GPRF:GENerator1:BBMode? | Read Generator Baseband Mode. Reads the currently selected generator baseband mode. | None |
| SOURce:GPRF:GENerator1:RFSettings:FREQuency <Freq> | Set Generator Frequency. Sets the frequency of the unmodulated RF carrier in Hz (70 MHz to 6 GHz depending on installed options). | Number |
| SOURce:GPRF:GENerator1:RFSettings:FREQuency? | Read Generator Frequency. Reads the RF generator carrier frequency in Hz. | None |
| SOURce:GPRF:GENerator1:RFSettings:LEVel <Level> | Set Generator Level. Sets the base RMS level of the RF generator in dBm. | Number |
| SOURce:GPRF:GENerator1:RFSettings:LEVel? | Read Generator Level. Reads the base RMS output level of the RF generator in dBm. | None |
| SOURce:GPRF:GENerator1:RFSettings:EATTenuation <Atten> | Set Generator External Attenuation. Sets the external attenuation (or gain, if negative) at the RF output connector in dB (-50 dB to 90 dB). | Number |
| SOURce:GPRF:GENerator1:RFSettings:EATTenuation? | Read Generator External Attenuation. Reads the external attenuation configured for the RF output path in dB. | None |
| SOURce:GPRF:GENerator1:RFSettings:DGAin <Gain> | Set Generator Digital Gain. Sets the digital gain of the RF generator in dB. | Number |
| SOURce:GPRF:GENerator1:RFSettings:DGAin? | Read Generator Digital Gain. Reads the digital gain of the RF generator in dB. | None |
| SOURce:GPRF:GENerator1:RFSettings:PEPower? | Read Generator Peak Envelope Power. Reads the peak envelope power of the generated signal in dBm. | None |
| SOURce:GPRF:GENerator1:ARB:FILE <Filename> | Select ARB Waveform File. Selects the ARB waveform file (.wv) to be played back in ARB baseband mode. | Text |
| SOURce:GPRF:GENerator1:ARB:FILE? | Read ARB Waveform File. Reads the name of the currently loaded ARB waveform file. | None |
| SOURce:GPRF:GENerator1:ARB:FILE:DATE? | Read ARB File Date. Reads the creation date stored in the loaded ARB waveform file. | None |
| SOURce:GPRF:GENerator1:ARB:FILE:VERSion? | Read ARB File Version. Reads the version of the loaded ARB waveform file. | None |
| SOURce:GPRF:GENerator1:ARB:FILE:OPTion? | Read ARB File Options. Reads the list of instrument options required to play back the loaded ARB waveform file. | None |
| SOURce:GPRF:GENerator1:ARB:REPetition <Repetition> | Set ARB Repetition. Defines whether the ARB waveform file is processed continuously or for a defined number of cycles. | Text: CONTinuous,SINGle |
| SOURce:GPRF:GENerator1:ARB:REPetition? | Read ARB Repetition. Reads the ARB repetition mode. | None |
| SOURce:GPRF:GENerator1:ARB:CYCLes <Cycles> | Set ARB Cycles. Defines how often the ARB file is processed in single-shot repetition mode (1 to 10000). | Number |
| SOURce:GPRF:GENerator1:ARB:CYCLes? | Read ARB Cycles. Reads the configured number of ARB playback cycles. | None |
| SOURce:GPRF:GENerator1:ARB:FOFFset <Freq> | Set ARB Frequency Offset. Sets the frequency offset imposed at the baseband during ARB generation in Hz. | Number |
| SOURce:GPRF:GENerator1:ARB:FOFFset? | Read ARB Frequency Offset. Reads the ARB baseband frequency offset in Hz. | None |
| SOURce:GPRF:GENerator1:ARB:ASAMples <Samples> | Set ARB Additional Samples. Extends the processing time of the waveform file by the specified number of samples. | Number |
| SOURce:GPRF:GENerator1:ARB:ASAMples? | Read ARB Additional Samples. Reads the number of additional samples appended to the waveform file. | None |
| SOURce:GPRF:GENerator1:ARB:CRATe? | Read ARB Clock Rate. Reads the clock rate of the loaded ARB waveform file in Hz. | None |
| SOURce:GPRF:GENerator1:ARB:POFFset? | Read ARB Peak Offset. Reads the peak offset of the loaded ARB waveform file in dB. | None |
| SOURce:GPRF:GENerator1:ARB:LOFFset? | Read ARB Level Offset. Reads the level offset (peak to average ratio) of the loaded ARB waveform file in dB. | None |
| SOURce:GPRF:GENerator1:ARB:CRCProtect? | Read ARB CRC Protection. Reads whether the loaded ARB waveform file contains a CRC checksum. | None |
| SOURce:GPRF:GENerator1:ARB:STATus? | Read ARB Segment Status. Reads the number of the ARB segment currently being processed. | None |
| SOURce:GPRF:GENerator1:ARB:SCOunt? | Read ARB Sample Count. Reads the progress of the ARB file processing as elapsed time and sample count. | None |
| SOURce:GPRF:GENerator1:LIST ON | Turn On List Mode. Enables the frequency/level list mode of the RF generator. | None |
| SOURce:GPRF:GENerator1:LIST OFF | Turn Off List Mode. Disables the frequency/level list mode of the RF generator. | None |
| SOURce:GPRF:GENerator1:LIST? | Read List Mode State. Reads whether the generator list mode is enabled. | None |
| SOURce:GPRF:GENerator1:LIST:STARt <Index> | Set List Start Index. Defines the index of the first frequency/level step processed in list mode (0 to 1999). | Number |
| SOURce:GPRF:GENerator1:LIST:STARt? | Read List Start Index. Reads the index of the first frequency/level step in list mode. | None |
| SOURce:GPRF:GENerator1:LIST:STOP <Index> | Set List Stop Index. Defines the index of the last frequency/level step processed in list mode (0 to 1999). | Number |
| SOURce:GPRF:GENerator1:LIST:STOP? | Read List Stop Index. Reads the index of the last frequency/level step in list mode. | None |
| SOURce:GPRF:GENerator1:LIST:GOTO <Index> | Set List Goto Index. Defines the start index used for the subsequent list cycles in continuous repetition mode (1 to 2000). | Number |
| SOURce:GPRF:GENerator1:LIST:REPetition <Repetition> | Set List Repetition. Defines how often the RF generator runs through the frequency/level list. | Text: CONTinuous,SINGle |
| SOURce:GPRF:GENerator1:LIST:REPetition? | Read List Repetition. Reads the list mode repetition setting. | None |
| SOURce:GPRF:GENerator1:LIST:AINDex? | Read List Active Index. Reads the list index currently being generated (0 to 19). | None |
| SOURce:GPRF:GENerator1:LIST:COUNt? | Read List Step Count. Reads the number of frequency/level steps configured in list mode (1 to 2000). | None |
| Command | What it does | Arguments |
|---|---|---|
| FETCh:GPRF:MEASurement1:POWer:STATe? | Read Power Measurement State. Reads the main state of the power measurement (OFF / RUN / RDY). | None |
| FETCh:GPRF:MEASurement1:POWer:STATe:ALL? | Read Power Measurement State All. Reads the detailed power measurement state including the main state and the synchronization state. | None |
| FETCh:GPRF:MEASurement1:POWer:CURRent? | Fetch Power Current. Fetches the current power results of all measured segments in dBm without restarting the measurement. | None |
| FETCh:GPRF:MEASurement1:POWer:CURRent:RMS? | Fetch Power Current RMS. Fetches the current RMS power results in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:AVERage? | Fetch Power Average. Fetches the average power results of all measured segments in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:AVERage:RMS? | Fetch Power Average RMS. Fetches the average RMS power results in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:MAXimum:CURRent? | Fetch Power Maximum. Fetches the current-maximum power results (one value per measured segment) in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:MINimum:CURRent? | Fetch Power Minimum. Fetches the current-minimum power results (one value per measured segment) in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:PEAK:MAXimum? | Fetch Power Peak Maximum. Fetches the peak maximum power results in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:PEAK:MINimum? | Fetch Power Peak Minimum. Fetches the peak minimum power results in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:SDEViation? | Fetch Power Standard Deviation. Fetches the standard deviation of the power results in dB. | None |
| FETCh:GPRF:MEASurement1:FFTSanalyzer:STATe? | Read FFT Measurement State. Reads the main state of the FFT spectrum analyzer measurement. | None |
| FETCh:GPRF:MEASurement1:FFTSanalyzer:POWer:CURRent? | Fetch FFT Power Current. Fetches the current FFT spectrum trace as a list of 801 power values in dBm. | None |
| FETCh:GPRF:MEASurement1:FFTSanalyzer:POWer:AVERage? | Fetch FFT Power Average. Fetches the average FFT spectrum trace as a list of 801 power values in dBm. | None |
| FETCh:GPRF:MEASurement1:FFTSanalyzer:POWer:MAXimum? | Fetch FFT Power Maximum. Fetches the maximum-hold FFT spectrum trace in dBm. | None |
| FETCh:GPRF:MEASurement1:FFTSanalyzer:POWer:MINimum? | Fetch FFT Power Minimum. Fetches the minimum-hold FFT spectrum trace in dBm. | None |
| FETCh:GPRF:MEASurement1:SPECtrum:STATe? | Read Spectrum Measurement State. Reads the main state of the spectrum measurement. | None |
| FETCh:GPRF:MEASurement1:SPECtrum:RMS:CURRent? | Fetch Spectrum RMS Current. Fetches the current RMS spectrum trace in dBm. | None |
| FETCh:GPRF:MEASurement1:SPECtrum:RMS:AVERage? | Fetch Spectrum RMS Average. Fetches the average RMS spectrum trace in dBm. | None |
| FETCh:GPRF:MEASurement1:SPECtrum:RMS:MAXimum? | Fetch Spectrum RMS Maximum. Fetches the maximum-hold RMS spectrum trace in dBm. | None |
| FETCh:GPRF:MEASurement1:SPECtrum:RMS:MINimum? | Fetch Spectrum RMS Minimum. Fetches the minimum-hold RMS spectrum trace in dBm. | None |
| FETCh:LTE:MEASurement1:MEValuation:STATe? | Read LTE MultiEval State. Reads the main state of the LTE multi-evaluation measurement. | None |
| FETCh:LTE:MEASurement1:MEValuation:MODulation:AVERage? | Fetch LTE Modulation Average. Fetches the average modulation results: EVM, magnitude error, phase error, I/Q offset, frequency error, timing error and TX power. | None |
| FETCh:NRSub:MEASurement1:MEValuation:STATe? | Read NR MultiEval State. Returns the NR FR1 multi-evaluation measurement state: OFF, RUN or RDY. | None |
| FETCh:NRSub:MEASurement1:MEValuation:MODulation:AVERage? | Fetch NR Modulation Average. Returns the average single-value modulation results (EVM, magnitude/phase error, frequency error...). | None |
| FETCh:NRSub:MEASurement1:MEValuation:TXPower:AVERage? | Fetch NR TX Power Average. Returns the total TX power results. | None |
| FETCh:NRSub:MEASurement1:MEValuation:ACLR:AVERage? | Fetch NR ACLR Average. Returns the relative ACLR values. | None |
| FETCh:NRSub:MEASurement1:MEValuation:SEMask:AVERage? | Fetch NR Spectrum Emission Average. Returns the spectrum emission mask single-value results. | None |
| FETCh:NRSub:MEASurement1:MEValuation:ESFLatness:AVERage? | Fetch NR Spectrum Flatness Average. Returns the equalizer spectrum flatness single-value results. | None |
| FETCh:NRSub:MEASurement1:MEValuation:IEMission:MARGin:AVERage? | Fetch NR Inband Emission Margin. Returns the in-band emission limit line margins (negative = limit exceeded). | None |
| FETCh:WLAN:MEASurement1:MEValuation:STATe? | Read WLAN MultiEval State. Returns the WLAN multi-evaluation measurement state: OFF, RUN or RDY. | None |
| FETCh:WLAN:MEASurement1:MEValuation:MODulation:AVERage? | Fetch WLAN Modulation Average. Returns the average single-value modulation results (EVM, frequency/clock error, burst power...) for OFDM SISO signals. | None |
| FETCh:WLAN:MEASurement1:MEValuation:MODulation:CURRent? | Fetch WLAN Modulation Current. Returns the current single-value modulation results for OFDM SISO signals. | None |
| FETCh:WLAN:MEASurement1:MEValuation:MODulation:DSSS:AVERage? | Fetch WLAN DSSS Modulation Average. Returns the average single-value modulation results for DSSS (802.11b) signals. | None |
| FETCh:WLAN:MEASurement1:MEValuation:TSMask:AVERage? | Fetch WLAN Spectrum Mask Average. Returns the limit line margins of the transmit spectrum mask (positive = limit exceeded). | None |
| FETCh:WLAN:MEASurement1:MEValuation:SFLatness:AVERage? | Fetch WLAN Spectral Flatness Average. Returns the spectrum flatness margin values (positive = limit violated). | None |
| FETCh:WLAN:MEASurement1:MEValuation:PVTime:REDGe:AVERage? | Fetch WLAN PvT Rising Edge Average. Returns the rising edge ramp durations of the power vs time measurement. | None |
| Command | What it does | Arguments |
|---|---|---|
| SYSTem:BASE:REFerence:FREQuency:SOURce <Source> | Set Reference Frequency Source. Selects the internal or an external reference frequency source for the instrument. | Text: INTernal,EXTernal |
| SYSTem:BASE:REFerence:FREQuency:SOURce? | Read Reference Frequency Source. Reads the selected reference frequency source. | None |
| SYSTem:BASE:REFerence:FREQuency <Freq> | Set External Reference Frequency. Sets the expected external reference frequency in Hz when the external reference source is selected. | Number |
| SYSTem:BASE:REFerence:FREQuency? | Read External Reference Frequency. Reads the expected external reference frequency in Hz. | None |
| SYSTem:ERRor:ALL? | Read Error Queue. Queries and deletes all entries in the error queue. | None |
| SYSTem:ERRor:COUNt? | Read Error Count. Queries the number of entries currently in the error queue. | None |
| SYSTem:RESet:BASE | Reset Base Settings. Resets the base settings to the default values suitable for remote operation. | None |
| SYSTem:RESet:ALL | Reset All Settings. Resets the entire instrument including the base settings and all firmware applications. | None |
| SYSTem:PRESet:BASE | Preset Base Settings. Presets the base settings to the default values suitable for manual operation. | None |
| SYSTem:PRESet:ALL | Preset All Settings. Presets the entire instrument including the base settings and all firmware applications. | None |
| SYSTem:VERSion? | Read SCPI Version. Queries the SCPI version number the instrument complies with. | None |
| SYSTem:BASE:OPTion:LIST? | Read Option List. Returns the list of installed software/hardware options and firmware applications. | None |
| SYSTem:GENerator:ALL:OFF | All Generators Off. Switches off all running generators. | None |
| SYSTem:MEASurement:ALL:OFF | All Measurements Off. Switches off all running measurements. | None |
| Command | What it does | Arguments |
|---|---|---|
| ROUTe:GPRF:GENerator1:SCENario:SALone <Connector>, <Converter> | Route Generator Standalone. Selects the TX path of the generated signal: connector bench R118 (RF 1 to RF 8) and TX module; enable individual ports with Enable_Generator_Port. | Text: R118; Text: TX1 |
| ROUTe:GPRF:GENerator1:SCENario? | Read Generator Scenario. Reads the active routing scenario of the GPRF generator. | None |
| ROUTe:GPRF:MEASurement1:SCENario:SALone <Connector>, <Converter> | Route Measurement Standalone. Activates the standalone scenario and selects the RF input port (RF 1 to RF 8 = R11 to R18) and RX module. | Text: R11,R12,R13,R14,R15,R16,R17,R18; Text: RX1 |
| ROUTe:GPRF:MEASurement1:SCENario? | Read Measurement Scenario. Reads the active routing scenario of the GPRF measurement. | None |
| ROUTe:LTE:MEASurement1:SCENario:SALone <Connector>,<Converter> | Route LTE Measurement Standalone. Activates the standalone scenario and selects the RF input path for the LTE uplink measurement. | Text: R11,R12,R13,R14,R15,R16,R17,R18; Text: RX1 |
| ROUTe:LTE:MEASurement1:SCENario? | Read LTE Measurement Scenario. Reads the active routing scenario of the LTE measurement application. | None |
| ROUTe:NRSub:MEASurement1:SCENario:SALone <Connector>, <Converter> | Route NR Measurement Standalone. Activates the standalone scenario and selects the RF input path for the NR FR1 uplink measurement. | Text: R11,R12,R13,R14,R15,R16,R17,R18; Text: RX1 |
| ROUTe:WLAN:MEASurement1:SCENario:SALone <Connector>, <Converter> | Route WLAN Measurement Standalone. Activates the standalone scenario and selects the RF input connector and RX module for WLAN measurements. | Text: R11,R12,R13,R14,R15,R16,R17,R18; Text: RX1 |
| ROUTe:WLAN:MEASurement1:SCENario? | Read WLAN Measurement Scenario. Returns the active WLAN measurement scenario. | None |
| Command | What it does | Arguments |
|---|---|---|
| INITiate:GPRF:MEASurement1:POWer | Start Power Measurement. Starts or restarts the GPRF power measurement and puts it into the RUN state. | None |
| INITiate:GPRF:MEASurement1:FFTSanalyzer | Start FFT Measurement. Starts or restarts the FFT spectrum analyzer measurement. | None |
| INITiate:GPRF:MEASurement1:SPECtrum | Start Spectrum Measurement. Starts or restarts the spectrum measurement. | None |
| INITiate:LTE:MEASurement1:MEValuation | Start LTE MultiEval Measurement. Starts or restarts the LTE uplink multi-evaluation measurement. | None |
| INITiate:NRSub:MEASurement1:MEValuation | Start NR MultiEval. Starts or restarts the NR FR1 multi-evaluation measurement (state RUN). | None |
| INITiate:WLAN:MEASurement1:MEValuation | Start WLAN MultiEval. Starts or restarts the WLAN multi-evaluation measurement (state RUN). | None |
| Command | What it does | Arguments |
|---|---|---|
| STOP:GPRF:MEASurement1:POWer | Stop Power Measurement. Halts the power measurement and puts it into the RDY state while keeping the results. | None |
| STOP:GPRF:MEASurement1:FFTSanalyzer | Stop FFT Measurement. Halts the FFT measurement and keeps the results. | None |
| STOP:GPRF:MEASurement1:SPECtrum | Stop Spectrum Measurement. Halts the spectrum measurement and keeps the results. | None |
| STOP:LTE:MEASurement1:MEValuation | Stop LTE MultiEval Measurement. Halts the LTE multi-evaluation measurement and keeps the results. | None |
| STOP:NRSub:MEASurement1:MEValuation | Stop NR MultiEval. Halts the NR FR1 multi-evaluation measurement; results are kept (state RDY). | None |
| STOP:WLAN:MEASurement1:MEValuation | Stop WLAN MultiEval. Halts the WLAN multi-evaluation measurement; results are kept (state RDY). | None |
| Command | What it does | Arguments |
|---|---|---|
| ABORt:GPRF:MEASurement1:POWer | Abort Power Measurement. Halts the power measurement and puts it into the OFF state; all results are set to NAV. | None |
| ABORt:GPRF:MEASurement1:FFTSanalyzer | Abort FFT Measurement. Halts the FFT measurement and clears all results. | None |
| ABORt:GPRF:MEASurement1:SPECtrum | Abort Spectrum Measurement. Halts the spectrum measurement and clears all results. | None |
| ABORt:LTE:MEASurement1:MEValuation | Abort LTE MultiEval Measurement. Halts the LTE multi-evaluation measurement and clears all results. | None |
| ABORt:NRSub:MEASurement1:MEValuation | Abort NR MultiEval. Aborts the NR FR1 multi-evaluation measurement and releases resources (state OFF). | None |
| ABORt:WLAN:MEASurement1:MEValuation | Abort WLAN MultiEval. Aborts the WLAN multi-evaluation measurement and releases resources (state OFF). | None |
| Command | What it does | Arguments |
|---|---|---|
| READ:GPRF:MEASurement1:POWer:CURRent? | Read Power Current. Starts the measurement and returns the current power results of all measured segments in dBm. | None |
| READ:GPRF:MEASurement1:POWer:CURRent:RMS? | Read Power Current RMS. Starts the measurement and returns the current RMS power results in dBm. | None |
| READ:GPRF:MEASurement1:POWer:AVERage? | Read Power Average. Starts the measurement and returns the average power results of all measured segments in dBm. | None |
| READ:GPRF:MEASurement1:POWer:SDEViation? | Read Power Standard Deviation. Starts the measurement and returns the standard deviation of the power results in dB. | None |
| READ:GPRF:MEASurement1:FFTSanalyzer:POWer:AVERage? | Read FFT Power Average. Starts the measurement and returns the average FFT spectrum trace in dBm. | None |
| READ:LTE:MEASurement1:MEValuation:MODulation:AVERage? | Read LTE Modulation Average. Starts the measurement and returns the average LTE uplink modulation results. | None |
| Command | What it does | Arguments |
|---|---|---|
| CALCulate:GPRF:MEASurement1:POWer:AVERage? | Calculate Power Average Limits. Returns the limit check results for the average power instead of the measured values. | None |
| CALCulate:LTE:MEASurement1:MEValuation:MODulation:AVERage? | Calculate LTE Modulation Average Limits. Returns the limit check results for the average LTE modulation measurement instead of the measured values. | None |
| Command | What it does | Arguments |
|---|---|---|
| SENSe:BASE:REFerence:FREQuency:LOCKed? | Read Reference Locked. Queries whether the instrument is locked to the selected reference frequency. | None |
| SENSe:BASE:TEMPerature:OPERating:INTernal? | Read Internal Temperature. Queries the average internal temperature measured at the RF modules in degrees C. | None |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Returns the instrument identification string (manufacturer, model, serial, firmware). | None |
| *RST | Reset. Resets the current sub-instrument to its default state. | None |
| *CLS | Clear Status. Clears the status byte, event registers and the output buffer. | None |
| *ESR? | Read Event Status Register. Returns the contents of the standard event status register (0 to 255) and clears it. | None |
| *STB? | Read Status Byte. Returns the contents of the status byte in decimal form. | None |
| *OPC | Set Operation Complete Bit. Sets bit 0 of the event status register when all preceding commands are complete. | None |
| *OPC? | Operation Complete. Returns 1 when all pending operations are complete. | None |
| *WAI | Wait To Continue. Waits until all pending operations are complete before continuing. | None |
| *SAV <Memory> | Save State. Stores the current instrument settings in intermediate memory 0 to 99. | Number: 0,99 |
| *RCL <Memory> | Recall State. Recalls instrument settings from intermediate memory 0 to 99. | Number: 0,99 |
| *TRG | Trigger Bus. Generates a software (bus) trigger event. | None |
This reference is generated from TestFlow's instrument library, which is drafted from Rohde & Schwarz's programming documentation for the CMW100 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Rohde & Schwarz's programming manual: check it before you rely on an edge case. Last updated September 27, 2026.
Yes. The Rohde & Schwarz CMW100 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 305 SCPI commands for it: 160 queries that read values or state, and 145 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, SOURce:GPRF:GENerator1:STATe ON and CONFigure:GPRF:MEASurement1:RFSettings:FREQuency?.
CONFigure:GPRF:MEASurement1:RFSettings:FREQuency? (read analyzer frequency). Reads the center frequency of the RF analyzer in Hz. The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Rohde & Schwarz CMW100 through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.