The Rohde & Schwarz CMU200 radio test set is controlled remotely with SCPI commands. TestFlow's instrument library holds 115 commands for it, 54 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the CMU200 on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Rohde & Schwarz CMU200 and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Rohde & Schwarz CMU200's VISA address
# Paste the address of your Rohde & Schwarz CMU200 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("SOURce:RFGenerator:PULSe:STATe ON") # Turn On RF Generator Pulse
print(inst.query("READ:RFANalyzer:POWer?")) # Measure Analyzer Power
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Rohde & Schwarz CMU200 and writes the report from the measured results.
Describe the test in a sentence, for example “turn on rf generator pulse on the CMU200, then measure analyzer power and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
115 commands in 16 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| SENSe:CORRection:LOSS:INPut<Connector_Nr> <Atten> | Set Input External Attenuation. Sets the external attenuation of RF input connector RF<nr> (1, 2 or 4 = RF4 IN) in dB (-50 to +90). | Number: 1,2,4; Number: -50,90 |
| SENSe:CORRection:LOSS:INPut<Connector_Nr>? | Read Input External Attenuation. Reads the external attenuation of RF input connector RF<nr> in dB. | Number: 1,2,4 |
| SENSe:LEVel:MODE <Mode> | Set Max Input Level Mode. Selects manual setting or automatic setting (from average signal power) of the maximum input level. | Text: MANual,AUTomatic |
| SENSe:LEVel:MODE? | Read Max Input Level Mode. Reads the selection: manual setting or automatic setting (from average signal power) of the maximum input level. | None |
| SENSe:LEVel:MAXimum <Level> | Set Max Input Level. Sets the maximum expected input level in dBm (RF1 -40 to +53, RF2 -54 to +39, RF4 IN -77 to 0). | Number |
| SENSe:LEVel:MAXimum? | Read Max Input Level. Reads the maximum expected input level in dBm (RF1 -40 to +53, RF2 -54 to +39, RF4 IN -77 to 0). | None |
| SENSe:LEVel:ATTenuation <Mode> | Set Input Attenuation Mode. Selects normal, low noise or low distortion setting of the input path. | Text: NORMal,LNOise,LDIStortion |
| SENSe:LEVel:ATTenuation? | Read Input Attenuation Mode. Reads the selection: normal, low noise or low distortion setting of the input path. | None |
| SENSe:RFANalyzer:FREQuency <Frequency> | Set Analyzer Frequency. Sets the analyzer frequency in Hz (50 kHz to 2.7 GHz). | Number: 50E3,2.7E9 |
| SENSe:RFANalyzer:FREQuency? | Read Analyzer Frequency. Reads the analyzer frequency in Hz (50 kHz to 2.7 GHz). | None |
| SENSe:RFANalyzer:BANDwidth <Bandwidth> | Set Analyzer Bandwidth. Sets the analyzer bandwidth: 10 Hz to 1 MHz (1-2-3-5 steps) or WIDE. | Text |
| SENSe:RFANalyzer:BANDwidth? | Read Analyzer Bandwidth. Reads the analyzer bandwidth: 10 Hz to 1 MHz (1-2-3-5 steps) or WIDE. | None |
| SENSe:NPOWer:BWIDth <Bandwidth> | Set NPower Bandwidth. Sets the narrow-band power measurement bandwidth: 10 Hz to 1 MHz, TDMA or CDMA filter. | Text |
| SENSe:NPOWer:BWIDth? | Read NPower Bandwidth. Reads the narrow-band power measurement bandwidth: 10 Hz to 1 MHz, TDMA or CDMA filter. | None |
| SENSe:POWer:FREQuency:CENTer <Frequency> | Set Power Center Frequency. Sets the center frequency of the power vs time measurement in Hz. | Number |
| SENSe:POWer:FREQuency:CENTer? | Read Power Center Frequency. Reads the center frequency of the power vs time measurement in Hz. | None |
| SENSe:POWer:FREQuency:BANDwidth <Bandwidth> | Set Power Bandwidth. Sets the resolution bandwidth of the power vs time measurement in Hz. | Number: 10,1E6 |
| SENSe:POWer:FREQuency:BANDwidth? | Read Power Bandwidth. Reads the resolution bandwidth of the power vs time measurement in Hz. | None |
| SENSe:POWer:TIME:SPAN <Span> | Set Power Time Span. Sets the span (total evaluation time) of the power measurement in s. | Number |
| SENSe:POWer:TIME:SPAN? | Read Power Time Span. Reads the span (total evaluation time) of the power measurement in s. | None |
| SENSe:POWer:TIME:DELay <Delay> | Set Power Time Delay. Sets the trigger delay of the power measurement in s. | Number |
| SENSe:POWer:TIME:DELay? | Read Power Time Delay. Reads the trigger delay of the power measurement in s. | None |
| SENSe:SPECtrum:FREQuency:CENTer <Frequency> | Set Spectrum Center Frequency. Sets the spectrum center frequency in Hz. | Number: 10E6,2.7E9 |
| SENSe:SPECtrum:FREQuency:CENTer? | Read Spectrum Center Frequency. Reads the spectrum center frequency in Hz. | None |
| SENSe:SPECtrum:FREQuency:SPAN <Span> | Set Spectrum Span. Sets the spectrum frequency span in Hz. | Number |
| SENSe:SPECtrum:FREQuency:SPAN? | Read Spectrum Span. Reads the spectrum frequency span in Hz. | None |
| SENSe:SPECtrum:FREQuency:STARt <Frequency> | Set Spectrum Start Frequency. Sets the spectrum start frequency in Hz. | Number |
| SENSe:SPECtrum:FREQuency:STARt? | Read Spectrum Start Frequency. Reads the spectrum start frequency in Hz. | None |
| SENSe:SPECtrum:FREQuency:STOP <Frequency> | Set Spectrum Stop Frequency. Sets the spectrum stop frequency in Hz. | Number |
| SENSe:SPECtrum:FREQuency:STOP? | Read Spectrum Stop Frequency. Reads the spectrum stop frequency in Hz. | None |
| SENSe:SPECtrum:FREQuency:BANDwidth <Bandwidth> | Set Spectrum RBW. Sets the spectrum resolution bandwidth: 10 Hz to 1 MHz or AUTO. | Text |
| SENSe:SPECtrum:FREQuency:BANDwidth? | Read Spectrum RBW. Reads the spectrum resolution bandwidth: 10 Hz to 1 MHz or AUTO. | None |
| SENSe:SPECtrum:DETector <Detector> | Set Spectrum Detector. Selects the peak or RMS detector for the spectrum trace. | Text: PEAK,RMS |
| SENSe:SPECtrum:DETector? | Read Spectrum Detector. Reads the selected peak or RMS detector for the spectrum trace. | None |
| SENSe:SYNChronize:FREQuency:REFerence:LOCKed? | Read Reference Locked. Queries whether the instrument is locked to the reference frequency. | None |
| Command | What it does | Arguments |
|---|---|---|
| SOURce:CORRection:LOSS:OUTPut<Connector_Nr> <Atten> | Set Output External Attenuation. Sets the external attenuation of RF output connector RF<nr> (1, 2 or 3 = RF3 OUT) in dB (-50 to +90). | Number: 1,2,3; Number: -50,90 |
| SOURce:CORRection:LOSS:OUTPut<Connector_Nr>? | Read Output External Attenuation. Reads the external attenuation of RF output connector RF<nr> in dB. | Number: 1,2,3 |
| SOURce:RFGenerator:FREQuency <Frequency> | Set RF Generator Frequency. Sets the RF generator output frequency in Hz (100 kHz to 2.7 GHz). | Number: 100E3,2.7E9 |
| SOURce:RFGenerator:FREQuency? | Read RF Generator Frequency. Reads the RF generator output frequency in Hz (100 kHz to 2.7 GHz). | None |
| SOURce:RFGenerator:LEVel <Level> | Set RF Generator Level. Sets the RF generator level in dBm (RF1 -137 to -27, RF2 -137 to -10, RF3 OUT -90 to +13). | Number: -137,13 |
| SOURce:RFGenerator:LEVel? | Read RF Generator Level. Reads the RF generator level in dBm (RF1 -137 to -27, RF2 -137 to -10, RF3 OUT -90 to +13). | None |
| SOURce:RFGenerator:MODulation <Modulation> | Set RF Generator Modulation. Selects the RF generator modulation: none (CW), SSB frequency shift, AM or FM. | Text: OFF,SSB,AM,FM |
| SOURce:RFGenerator:MODulation? | Read RF Generator Modulation. Reads the selected RF generator modulation: none (CW), SSB frequency shift, AM or FM. | None |
| SOURce:RFGenerator:MODulation:AM:INDex <Index> | Set AM Modulation Index. Sets the AM modulation index in % (0 to 100). | Number: 0,100 |
| SOURce:RFGenerator:MODulation:AM:INDex? | Read AM Modulation Index. Reads the AM modulation index in % (0 to 100). | None |
| SOURce:RFGenerator:MODulation:FM:DEViation <Deviation> | Set FM Deviation. Sets the FM deviation in Hz (0 to 500 kHz). | Number: 0,500000 |
| SOURce:RFGenerator:MODulation:FM:DEViation? | Read FM Deviation. Reads the FM deviation in Hz (0 to 500 kHz). | None |
| SOURce:RFGenerator:MODulation:FM:FREQuency <Frequency> | Set FM Modulation Frequency. Sets the FM modulation frequency in Hz (1 Hz to 50 kHz). | Number: 1,50000 |
| SOURce:RFGenerator:MODulation:FM:FREQuency? | Read FM Modulation Frequency. Reads the FM modulation frequency in Hz (1 Hz to 50 kHz). | None |
| SOURce:RFGenerator:MODulation:SSB:FREQuency <Offset> | Set SSB Frequency Offset. Sets the SSB frequency offset in Hz (-300 kHz to +300 kHz). | Number: -300000,300000 |
| SOURce:RFGenerator:MODulation:SSB:FREQuency? | Read SSB Frequency Offset. Reads the SSB frequency offset in Hz (-300 kHz to +300 kHz). | None |
| SOURce:RFGenerator:PULSe:STATe ON | Turn On RF Generator Pulse. Switches to a pulsed signal with 577 us burst length. | None |
| SOURce:RFGenerator:PULSe:STATe OFF | Turn Off RF Generator Pulse. Switches to a CW (unpulsed) signal. | None |
| SOURce:RFGenerator:PULSe:STATe? | Read RF Generator Pulse. Reads the RF generator pulse state. | None |
| Command | What it does | Arguments |
|---|---|---|
| FETCh:RFGenerator:STATus? | Read RF Generator Status. Returns the RF generator status: OFF, RUN or ERR. | None |
| FETCh:RFANalyzer:STATus? | Read Analyzer Status. Returns the measurement status (OFF, RUN, STOP, ERR, STEP, RDY) and counters. | None |
| FETCh:RFANalyzer:POWer? | Fetch Analyzer Power. Reads the last RMS power result in dBm (unsynchronized). | None |
| FETCh:NPOWer:STATus? | Read NPower Status. Returns the narrow-band power measurement status. | None |
| FETCh:NPOWer? | Fetch NPower. Reads the narrow-band power results (unsynchronized). | None |
| FETCh:POWer:STATus? | Read Power Status. Returns the power measurement status. | None |
| FETCh:ARRay:POWer:CURRent? | Fetch Power Trace Current. Reads the current power vs time trace in dBm. | None |
| FETCh:ARRay:POWer:MAXimum? | Fetch Power Trace Maximum. Reads the maximum power vs time trace in dBm. | None |
| FETCh:SPECtrum:STATus? | Read Spectrum Status. Returns the spectrum measurement status. | None |
| FETCh:ARRay:SPECtrum:AVERage? | Fetch Spectrum Trace Average. Reads the average spectrum trace in dBm. | None |
| FETCh:SPECtrum:MARKer:PEAK? | Fetch Spectrum Peak. Returns the frequency and level of the spectrum peak. | None |
| Command | What it does | Arguments |
|---|---|---|
| SYSTem:REMote:ADDRess:SECondary <Address>, "<Function_Group>" | Map Secondary Address. Base system (secondary address 0): assigns a GPIB secondary address 1 to 29 to a function group, e.g. 1,"RF_NSig". | Number: 1,29; Text: RF_NSig,BASE |
| SYSTem:RESet:CURRent | Reset Function Group. Sets all parameters of the current function group and test mode to default values. | None |
| SYSTem:RESet:ALL | Reset All. Resets the base system and all function groups. | None |
| SYSTem:ERRor? | Read Error. Returns the oldest entry of the error queue. | None |
| SYSTem:VERSion? | Read SCPI Version. Returns the SCPI version of the instrument. | None |
| SYSTem:OPTions? | Read Options. Returns the installed hardware/software options and firmware versions. | None |
| Command | What it does | Arguments |
|---|---|---|
| CONFigure:RFANalyzer:CONTrol:REPetition <Repetition>, <Stop_Condition>, <Step_Mode> | Set Analyzer Repetition. Sets repetition (CONTinuous, SINGleshot or 1 to 10000 cycles), stop condition and step mode of the analyzer power measurement. | Text: CONTinuous,SINGleshot; Text: SONerror,NONE; Text: STEP,NONE |
| CONFigure:NPOWer:CONTrol:REPetition <Repetition>, <Stop_Condition>, <Step_Mode> | Set NPower Repetition. Sets repetition, stop condition and step mode of the narrow-band power measurement. | Text: CONTinuous,SINGleshot; Text: SONerror,NONE; Text: STEP,NONE |
| CONFigure:POWer:CONTrol:REPetition <Repetition>, <Stop_Condition>, <Step_Mode> | Set Power Repetition. Sets repetition, stop condition and step mode of the power vs time measurement. | Text: CONTinuous,SINGleshot; Text: SONerror,NONE; Text: STEP,NONE |
| CONFigure:SPECtrum:CONTrol:REPetition <Repetition>, <Stop_Condition>, <Step_Mode> | Set Spectrum Repetition. Sets repetition, stop condition and step mode of the spectrum measurement. | Text: CONTinuous,SINGleshot; Text: SONerror,NONE; Text: STEP,NONE |
| CONFigure:SYNChronize:FREQuency:REFerence:MODE <Mode> | Set Reference Frequency Mode. Selects the internal or external reference frequency. | Text: INTernal,EXTernal |
| CONFigure:SYNChronize:FREQuency:REFerence:MODE? | Read Reference Frequency Mode. Reads the selected internal or external reference frequency. | None |
| Command | What it does | Arguments |
|---|---|---|
| INITiate:RFGenerator | Start RF Generator. Starts the RF generator (Tx signal) and reserves resources. | None |
| INITiate:RFANalyzer | Start Analyzer Power. Starts the analyzer power measurement. | None |
| INITiate:NPOWer | Start NPower. Starts the narrow-band power measurement. | None |
| INITiate:POWer | Start Power. Starts the power vs time measurement. | None |
| INITiate:SPECtrum | Start Spectrum. Starts the spectrum measurement. | None |
| Command | What it does | Arguments |
|---|---|---|
| ABORt:RFGenerator | Stop RF Generator. Switches the RF generator off and releases resources. | None |
| ABORt:RFANalyzer | Abort Analyzer Power. Aborts the analyzer power measurement and releases resources. | None |
| ABORt:NPOWer | Abort NPower. Aborts the narrow-band power measurement. | None |
| ABORt:POWer | Abort Power. Aborts the power vs time measurement. | None |
| ABORt:SPECtrum | Abort Spectrum. Aborts the spectrum measurement. | None |
| Command | What it does | Arguments |
|---|---|---|
| STOP:RFANalyzer | Stop Analyzer Power. Stops the analyzer power measurement. | None |
| STOP:NPOWer | Stop NPower. Stops the narrow-band power measurement. | None |
| STOP:POWer | Stop Power. Stops the power vs time measurement. | None |
| STOP:SPECtrum | Stop Spectrum. Stops the spectrum measurement. | None |
| Command | What it does | Arguments |
|---|---|---|
| READ:RFANalyzer:POWer? | Measure Analyzer Power. Starts a single-shot measurement and returns the RMS power of the RF input signal in dBm. | None |
| READ:NPOWer? | Measure NPower. Starts a single shot and returns avg/min/max power results in dBm and the frequency. | None |
| READ:ARRay:POWer:AVERage? | Measure Power Trace Average. Starts a single shot and returns the average power vs time trace in dBm. | None |
| READ:ARRay:SPECtrum:CURRent? | Measure Spectrum Trace. Starts a single shot and returns the current spectrum trace (560 values in dBm). | None |
| Command | What it does | Arguments |
|---|---|---|
| INPut:STATe <Connector> | Set RF Input. Selects the RF input connector: RF1, RF2 or RF4 IN (RF function group). | Text: RF1,RF2,RF4 |
| INPut:STATe? | Read RF Input. Reads the selected RF input connector: RF1, RF2 or RF4 IN (RF function group). | None |
| Command | What it does | Arguments |
|---|---|---|
| OUTPut:STATe <Connector> | Set RF Output. Selects the RF output connector: RF1, RF2 or RF3 OUT (RF function group). | Text: RF1,RF2,RF3 |
| OUTPut:STATe? | Read RF Output. Reads the selected RF output connector: RF1, RF2 or RF3 OUT (RF function group). | None |
| Command | What it does | Arguments |
|---|---|---|
| DEFault:RFGenerator ON | Default RF Generator. Sets all RF generator parameters to default values. | None |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Returns the instrument identification string. | None |
| *RST | Reset. Resets the instrument (addressed function group) to its default state. | None |
| *CLS | Clear Status. Clears the status byte, event registers and output buffer. | None |
| *ESR? | Read Event Status Register. Returns and clears the standard event status register (0 to 255). | None |
| *ESE <Mask> | Set Event Status Enable. Sets the event status enable register (0 to 255). | Number: 0,255 |
| *STB? | Read Status Byte. Returns the status byte in decimal form. | None |
| *SRE <Mask> | Set Service Request Enable. Sets the service request enable register (0 to 255). | Number: 0,255 |
| *OPT? | Read Installed Options. Returns the installed options. | None |
| *TST? | Self Test. Runs the instrument self-test and returns the result. | None |
| *OPC | Set Operation Complete Bit. Sets bit 0 of the event status register when all preceding commands are executed. | None |
| *OPC? | Operation Complete. Returns 1 when all pending operations are complete. | None |
| *WAI | Wait To Continue. Waits until all pending operations are complete before continuing. | None |
This reference is generated from TestFlow's instrument library, which is drafted from Rohde & Schwarz's programming documentation for the CMU200 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Rohde & Schwarz's programming manual: check it before you rely on an edge case. Last updated November 12, 2026.
Yes. The Rohde & Schwarz CMU200 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 115 SCPI commands for it: 54 queries that read values or state, and 61 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, SOURce:RFGenerator:PULSe:STATe ON and READ:RFANalyzer:POWer?.
READ:RFANalyzer:POWer? (measure analyzer power). Starts a single-shot measurement and returns the RMS power of the RF input signal in dBm. The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Rohde & Schwarz CMU200 through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.