The Rohde & Schwarz CMA180 radio test set is controlled remotely with SCPI commands. TestFlow's instrument library holds 145 commands for it, 72 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the CMA180 on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Rohde & Schwarz CMA180 and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Rohde & Schwarz CMA180's VISA address
# Paste the address of your Rohde & Schwarz CMA180 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("SOURce:AFRF:GENerator1:STATe ON") # Turn On Generator
print(inst.query("CONFigure:AFRF:MEASurement1:RFSettings:FREQuency?")) # Read Analyzer Frequency
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Rohde & Schwarz CMA180 and writes the report from the measured results.
Describe the test in a sentence, for example “turn on generator on the CMA180, then read analyzer frequency and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
145 commands in 12 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| CONFigure:BASE:SCENario <Scenario> | Set Test Scenario. Selects the test scenario (analog TX/RX/duplex/spectrum/expert/audio/avionics or digital); select it before configuring an application. | Text: TXTest,RXTest,DXTest,SPECtrum,EXPert,AUDio,AVIonics,DTXTest,DRXTest,DSPectrum,DEXPert |
| CONFigure:BASE:SCENario? | Read Test Scenario. Reads the selected test scenario (analog TX/RX/duplex/spectrum/expert/audio/avionics or digital); select it before configuring an application. | None |
| CONFigure:AFRF:MEASurement1:RFSettings:FREQuency <Frequency> | Set Analyzer Frequency. Sets the center frequency of the RF analyzer in Hz (100 kHz to 3 GHz). | Number: 100E3,3E9 |
| CONFigure:AFRF:MEASurement1:RFSettings:FREQuency? | Read Analyzer Frequency. Reads the center frequency of the RF analyzer in Hz (100 kHz to 3 GHz). | None |
| CONFigure:AFRF:MEASurement1:RFSettings:CONNector <Connector> | Set Analyzer Connector. Selects the RF input connector: RF COM or RF IN. | Text: RFCom,RFIN |
| CONFigure:AFRF:MEASurement1:RFSettings:CONNector? | Read Analyzer Connector. Reads the selected RF input connector: RF COM or RF IN. | None |
| CONFigure:AFRF:MEASurement1:RFSettings:ENPower <Power> | Set Analyzer Expected Power. Sets the expected nominal power of the measured RF signal in dBm. | Number |
| CONFigure:AFRF:MEASurement1:RFSettings:ENPower? | Read Analyzer Expected Power. Reads the expected nominal power of the measured RF signal in dBm. | None |
| CONFigure:AFRF:MEASurement1:RFSettings:EATTenuation <Atten> | Set Analyzer External Attenuation. Sets the external attenuation in the RF input path in dB (negative values = gain). | Number: -50,90 |
| CONFigure:AFRF:MEASurement1:RFSettings:EATTenuation? | Read Analyzer External Attenuation. Reads the external attenuation in the RF input path in dB (negative values = gain). | None |
| CONFigure:AFRF:MEASurement1:DEModulation <Type> | Set Demodulation Type. Selects the type of demodulation. | Text: FM,AM,PM,USB,LSB,FMSTereo |
| CONFigure:AFRF:MEASurement1:DEModulation? | Read Demodulation Type. Reads the selected type of demodulation. | None |
| CONFigure:AFRF:MEASurement1:DEModulation:FILTer:LPASs <Filter> | Set Demod Lowpass Filter. Configures the lowpass filter in the demodulation path (cutoff 255 Hz to 15 kHz). | Text: OFF,F255,F3K,F3K4,F4K,F15K |
| CONFigure:AFRF:MEASurement1:DEModulation:FILTer:LPASs? | Read Demod Lowpass Filter. Reads back the setting: configures the lowpass filter in the demodulation path (cutoff 255 Hz to 15 kHz). | None |
| CONFigure:AFRF:MEASurement1:DEModulation:FILTer:HPASs <Filter> | Set Demod Highpass Filter. Configures the highpass filter in the demodulation path (cutoff 6, 50 or 300 Hz). | Text: OFF,F6,F50,F300 |
| CONFigure:AFRF:MEASurement1:DEModulation:FILTer:HPASs? | Read Demod Highpass Filter. Reads back the setting: configures the highpass filter in the demodulation path (cutoff 6, 50 or 300 Hz). | None |
| CONFigure:AFRF:MEASurement1:DEModulation:FILTer:WEIGhting <Filter> | Set Demod Weighting Filter. Selects the weighting filter in the demodulation path. | Text: OFF,AWEighting,CCITt,CMESsage |
| CONFigure:AFRF:MEASurement1:DEModulation:FILTer:WEIGhting? | Read Demod Weighting Filter. Reads the selected weighting filter in the demodulation path. | None |
| CONFigure:AFRF:MEASurement1:DEModulation:FILTer:DEEMphasis <Filter> | Set Demod Deemphasis. Configures the de-emphasis filter (time constant 50, 75 or 750 us). | Text: OFF,T50,T75,T750 |
| CONFigure:AFRF:MEASurement1:DEModulation:FILTer:DEEMphasis? | Read Demod Deemphasis. Reads back the setting: configures the de-emphasis filter (time constant 50, 75 or 750 us). | None |
| CONFigure:AFRF:MEASurement1:AIN1:ENABle ON | Turn On AF1 Input. Enables the AF1 IN connector. | None |
| CONFigure:AFRF:MEASurement1:AIN1:ENABle OFF | Turn Off AF1 Input. Disables the AF1 IN connector. | None |
| CONFigure:AFRF:MEASurement1:AIN1:MLEVel <Level> | Set AF1 Input Max Level. Sets the maximum expected level of the AF1 IN connector in V (used when auto ranging is disabled). | Number |
| CONFigure:AFRF:MEASurement1:AIN1:MLEVel? | Read AF1 Input Max Level. Reads the maximum expected level of the AF1 IN connector in V (used when auto ranging is disabled). | None |
| CONFigure:AFRF:MEASurement1:AIN1:ICOupling <Coupling> | Set AF1 Input Coupling. Selects AC or DC coupling of the AF1 IN connector. | Text: AC,DC |
| CONFigure:AFRF:MEASurement1:AIN1:ICOupling? | Read AF1 Input Coupling. Reads the selection: AC or DC coupling of the AF1 IN connector. | None |
| CONFigure:AFRF:MEASurement1:MEValuation:REPetition <Repetition> | Set Meas Repetition. Selects single-shot or continuous multi-evaluation measurement. | Text: SINGleshot,CONTinuous |
| CONFigure:AFRF:MEASurement1:MEValuation:REPetition? | Read Meas Repetition. Reads the selection: single-shot or continuous multi-evaluation measurement. | None |
| CONFigure:AFRF:MEASurement1:MEValuation:TOUT <Seconds> | Set Meas Timeout. Defines a timeout for the measurement in seconds. | Number |
| CONFigure:AFRF:MEASurement1:MEValuation:TOUT? | Read Meas Timeout. Reads the timeout for the measurement in seconds. | None |
| CONFigure:AFRF:MEASurement1:MEValuation:RF:SCOunt <Count> | Set RF Statistic Count. Sets the number of measurement intervals per cycle for RF results. | Number: 1,1000 |
| CONFigure:AFRF:MEASurement1:MEValuation:RF:SCOunt? | Read RF Statistic Count. Reads the number of measurement intervals per cycle for RF results. | None |
| CONFigure:AFRF:MEASurement1:MEValuation:AF:SCOunt <Count> | Set AF Statistic Count. Sets the number of measurement intervals per cycle for AF results. | Number: 1,1000 |
| CONFigure:AFRF:MEASurement1:MEValuation:AF:SCOunt? | Read AF Statistic Count. Reads the number of measurement intervals per cycle for AF results. | None |
| CONFigure:AFRF:MEASurement1:FREQuency:COUNter:MODE <Mode> | Set Frequency Counter Mode. Selects whether the frequency search procedure is used in analog or digital scenarios. | Text: ANALog,DIGital |
| CONFigure:AFRF:MEASurement1:FREQuency:COUNter:MODE? | Read Frequency Counter Mode. Reads the selection: whether the frequency search procedure is used in analog or digital scenarios. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:FREQuency <Frequency> | Set Power Meter Frequency. Sets the center frequency of the GPRF power measurement in Hz. | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:FREQuency? | Read Power Meter Frequency. Reads the center frequency of the GPRF power measurement in Hz. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:CONNector <Connector> | Set Power Meter Connector. Selects the RF input connector of the GPRF measurement. | Text: RFCom,RFIN |
| CONFigure:GPRF:MEASurement1:RFSettings:CONNector? | Read Power Meter Connector. Reads the selected RF input connector of the GPRF measurement. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:ENPower <Power> | Set Power Meter Expected Power. Sets the expected nominal input power in dBm. | Number |
| CONFigure:GPRF:MEASurement1:RFSettings:ENPower? | Read Power Meter Expected Power. Reads the expected nominal input power in dBm. | None |
| CONFigure:GPRF:MEASurement1:RFSettings:EATTenuation <Atten> | Set Power Meter External Attenuation. Sets the external input attenuation in dB. | Number: -50,90 |
| CONFigure:GPRF:MEASurement1:RFSettings:EATTenuation? | Read Power Meter External Attenuation. Reads the external input attenuation in dB. | None |
| CONFigure:GPRF:MEASurement1:POWer:FILTer:TYPE <Type> | Set Power Filter Type. Selects the IF filter type of the power measurement. | Text: BANDpass,GAUSs |
| CONFigure:GPRF:MEASurement1:POWer:FILTer:TYPE? | Read Power Filter Type. Reads the selected IF filter type of the power measurement. | None |
| CONFigure:GPRF:MEASurement1:POWer:MLENgth <Seconds> | Set Power Measurement Length. Sets the power measurement length in seconds. | Number |
| CONFigure:GPRF:MEASurement1:POWer:MLENgth? | Read Power Measurement Length. Reads the power measurement length in seconds. | None |
| CONFigure:GPRF:MEASurement1:POWer:REPetition <Repetition> | Set Power Repetition. Selects single-shot or continuous power measurement. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:POWer:REPetition? | Read Power Repetition. Reads the selection: single-shot or continuous power measurement. | None |
| CONFigure:GPRF:MEASurement1:POWer:SCOunt <Count> | Set Power Statistic Count. Sets the statistic count of the power measurement. | Number |
| CONFigure:GPRF:MEASurement1:POWer:SCOunt? | Read Power Statistic Count. Reads the statistic count of the power measurement. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:FREQuency:CENTer <Frequency> | Set Spectrum Center Frequency. Sets the spectrum analyzer center frequency in Hz. | Number |
| CONFigure:GPRF:MEASurement1:SPECtrum:FREQuency:CENTer? | Read Spectrum Center Frequency. Reads the spectrum analyzer center frequency in Hz. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:FREQuency:SPAN <Span> | Set Spectrum Span. Sets the spectrum analyzer span in Hz. | Number |
| CONFigure:GPRF:MEASurement1:SPECtrum:FREQuency:SPAN? | Read Spectrum Span. Reads the spectrum analyzer span in Hz. | None |
| CONFigure:GPRF:MEASurement1:SPECtrum:REPetition <Repetition> | Set Spectrum Repetition. Selects single-shot or continuous spectrum measurement. | Text: SINGleshot,CONTinuous |
| CONFigure:GPRF:MEASurement1:SPECtrum:REPetition? | Read Spectrum Repetition. Reads the selection: single-shot or continuous spectrum measurement. | None |
| Command | What it does | Arguments |
|---|---|---|
| SOURce:AFRF:GENerator1:STATe ON | Turn On Generator. Starts the AFRF generator. | None |
| SOURce:AFRF:GENerator1:STATe OFF | Turn Off Generator. Stops the AFRF generator. | None |
| SOURce:AFRF:GENerator1:STATe? | Read Generator. Reads the AFRF generator state (OFF, ON, PENDing). | None |
| SOURce:AFRF:GENerator1:MSCHeme <Scheme> | Set RF Signal Mode. Selects the RF modulation scheme for analog scenarios (FM, AM, PM, SSB upper/lower, CW, FM stereo, ARB). | Text: FM,AM,PM,USB,LSB,CW,FMSTereo,ARB |
| SOURce:AFRF:GENerator1:MSCHeme? | Read RF Signal Mode. Reads the selected RF modulation scheme for analog scenarios (FM, AM, PM, SSB upper/lower, CW, FM stereo, ARB). | None |
| SOURce:AFRF:GENerator1:DSOurce <Source> | Set Digital Source. Selects the data source for digital scenarios. | Text: DMR,ARB,NXDN,POCSag,P25,UDEFined,ZIGBee,DPMR |
| SOURce:AFRF:GENerator1:DSOurce? | Read Digital Source. Reads the selected data source for digital scenarios. | None |
| SOURce:AFRF:GENerator1:RFSettings:FREQuency <Frequency> | Set Generator Frequency. Sets the center frequency of the RF carrier in Hz (100 kHz to 3 GHz). | Number: 100E3,3E9 |
| SOURce:AFRF:GENerator1:RFSettings:FREQuency? | Read Generator Frequency. Reads the center frequency of the RF carrier in Hz (100 kHz to 3 GHz). | None |
| SOURce:AFRF:GENerator1:RFSettings:LEVel <Level> | Set Generator Level. Sets the RMS level of the unmodulated RF signal in dBm. | Number |
| SOURce:AFRF:GENerator1:RFSettings:LEVel? | Read Generator Level. Reads the RMS level of the unmodulated RF signal in dBm. | None |
| SOURce:AFRF:GENerator1:RFSettings:CONNector <Connector> | Set Generator Connector. Selects the RF output connector: RF COM or RF OUT. | Text: RFCom,RFOut |
| SOURce:AFRF:GENerator1:RFSettings:CONNector? | Read Generator Connector. Reads the selected RF output connector: RF COM or RF OUT. | None |
| SOURce:AFRF:GENerator1:RFSettings:EATTenuation <Atten> | Set Generator External Attenuation. Sets the external attenuation in the RF output path in dB (negative values = gain). | Number: -50,90 |
| SOURce:AFRF:GENerator1:RFSettings:EATTenuation? | Read Generator External Attenuation. Reads the external attenuation in the RF output path in dB (negative values = gain). | None |
| SOURce:AFRF:GENerator1:RFSettings:RF:ENABle ON | Turn On Generator RF Output. Enables the RF signal without changing the generator state. | None |
| SOURce:AFRF:GENerator1:RFSettings:RF:ENABle OFF | Turn Off Generator RF Output. Disables the RF signal without changing the generator state. | None |
| SOURce:AFRF:GENerator1:RFSettings:RF:ENABle? | Read Generator RF Output. Reads whether the RF signal is enabled. | None |
| SOURce:AFRF:GENerator1:RFSettings:PEPower? | Read Generator PEP. Returns the peak envelope power of the generated signal in dBm. | None |
| SOURce:AFRF:GENerator1:MODulator <Source> | Set Modulator Source. Selects the audio source modulated onto the RF carrier (internal audio generators, AF IN, SPDIF IN). | Text: GEN3,GEN4,GENB,AFI1,AFI2,AFIB,SPIL,SPIR,SPIN |
| SOURce:AFRF:GENerator1:MODulator? | Read Modulator Source. Reads the selected audio source modulated onto the RF carrier (internal audio generators, AF IN, SPDIF IN). | None |
| SOURce:AFRF:GENerator1:MODulator:FDEViation <Deviation> | Set FM Deviation. Sets the maximum frequency deviation for FM in Hz (0 to 100 kHz). | Number: 0,100000 |
| SOURce:AFRF:GENerator1:MODulator:FDEViation? | Read FM Deviation. Reads the maximum frequency deviation for FM in Hz (0 to 100 kHz). | None |
| SOURce:AFRF:GENerator1:MODulator:MDEPth <Depth> | Set AM Modulation Depth. Sets the AM modulation depth in % (0 to 100). | Number: 0,100 |
| SOURce:AFRF:GENerator1:MODulator:MDEPth? | Read AM Modulation Depth. Reads the AM modulation depth in % (0 to 100). | None |
| SOURce:AFRF:GENerator1:MODulator:PDEViation <Deviation> | Set PM Deviation. Sets the maximum phase deviation for PM in rad (0 to 10). | Number: 0,10 |
| SOURce:AFRF:GENerator1:MODulator:PDEViation? | Read PM Deviation. Reads the maximum phase deviation for PM in rad (0 to 10). | None |
| SOURce:AFRF:GENerator1:IGENerator1:TMODe <Mode> | Set Audio Generator Tone Mode. Selects the tone mode of internal audio generator 1. | Text: STONe,DTONe,MTONe,NOISe,DTMF,SELCall,FDIaling,SCAL,SQUare |
| SOURce:AFRF:GENerator1:IGENerator1:TMODe? | Read Audio Generator Tone Mode. Reads the selected tone mode of internal audio generator 1. | None |
| SOURce:AFRF:GENerator1:IGENerator1:FREQuency <Frequency> | Set Audio Generator Frequency. Sets the single-tone/square frequency of internal audio generator 1 in Hz (0 to 21 kHz). | Number: 0,21000 |
| SOURce:AFRF:GENerator1:IGENerator1:FREQuency? | Read Audio Generator Frequency. Reads the single-tone/square frequency of internal audio generator 1 in Hz (0 to 21 kHz). | None |
| SOURce:AFRF:GENerator1:IGENerator1:ENABle? | Read Audio Generator Enabled. Queries whether internal audio generator 1 is assigned to an output path. | None |
| SOURce:AFRF:GENerator1:AOUT1 <Source> | Set AF1 Output Source. Selects the audio signal source for the AF1 OUT connector. | Text: GEN1,AFI1,SPIL |
| SOURce:AFRF:GENerator1:AOUT1? | Read AF1 Output Source. Reads the selected audio signal source for the AF1 OUT connector. | None |
| SOURce:AFRF:GENerator1:AOUT1:ENABle ON | Turn On AF1 Output. Enables the AF1 OUT connector. | None |
| SOURce:AFRF:GENerator1:AOUT1:ENABle OFF | Turn Off AF1 Output. Disables the AF1 OUT connector. | None |
| SOURce:AFRF:GENerator1:AOUT1:LEVel <Level> | Set AF1 Output Level. Sets the output level of the AF1 OUT connector. | Number |
| SOURce:AFRF:GENerator1:AOUT1:LEVel? | Read AF1 Output Level. Reads the output level of the AF1 OUT connector. | None |
| Command | What it does | Arguments |
|---|---|---|
| FETCh:AFRF:MEASurement1:MEValuation:STATe? | Read Measurement State. Returns the AFRF multi-evaluation measurement state: OFF, RUN or RDY. | None |
| FETCh:AFRF:MEASurement1:MEValuation:RFCarrier:CURRent? | Fetch RF Carrier Current. Returns the current RF carrier results (power, PEP, frequency error...). | None |
| FETCh:AFRF:MEASurement1:MEValuation:RFCarrier:AVERage? | Fetch RF Carrier Average. Returns the average RF carrier results. | None |
| FETCh:AFRF:MEASurement1:MEValuation:RFCarrier:FREQuency:AVERage? | Fetch RF Carrier Frequency. Returns the average RF carrier frequency in Hz. | None |
| FETCh:AFRF:MEASurement1:MEValuation:DEModulation:FDEViation:AVERage? | Fetch FM Deviation Average. Returns the FM frequency deviation statistics of the demodulated signal. | None |
| FETCh:AFRF:MEASurement1:MEValuation:DEModulation:MDEPth:AVERage? | Fetch AM Depth Average. Returns the AM modulation depth statistics of the demodulated signal. | None |
| FETCh:AFRF:MEASurement1:MEValuation:DEModulation:PDEViation:AVERage? | Fetch PM Deviation Average. Returns the PM phase deviation statistics of the demodulated signal. | None |
| FETCh:AFRF:MEASurement1:MEValuation:AIN1:AFSignal:AVERage? | Fetch AF1 Signal Average. Returns AF frequency and level results measured at AF1 IN. | None |
| FETCh:AFRF:MEASurement1:FREQuency:COUNter? | Fetch Frequency Counter. Returns the results of the RF frequency search procedure. | None |
| FETCh:GPRF:MEASurement1:POWer:STATe? | Read Power Measurement State. Returns the GPRF power measurement state: OFF, RUN or RDY. | None |
| FETCh:GPRF:MEASurement1:POWer:AVERage? | Fetch Power Average. Returns the average RMS power in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:CURRent? | Fetch Power Current. Returns the current RMS power in dBm. | None |
| FETCh:GPRF:MEASurement1:POWer:PEAK:MAXimum? | Fetch Power Peak Maximum. Returns the maximum peak power in dBm. | None |
| FETCh:GPRF:MEASurement1:SPECtrum:STATe? | Read Spectrum Measurement State. Returns the spectrum analyzer measurement state: OFF, RUN or RDY. | None |
| FETCh:GPRF:MEASurement1:SPECtrum:RMS:AVERage? | Fetch Spectrum RMS Average. Returns the average RMS spectrum trace. | None |
| FETCh:GPRF:MEASurement1:SPECtrum:REFMarker:NPEak? | Fetch Spectrum Peak Marker. Moves the reference marker to the next peak and returns its position and level. | None |
| Command | What it does | Arguments |
|---|---|---|
| SYSTem:BASE:REFerence:FREQuency:SOURce <Source> | Set Reference Frequency Source. Selects the internal or external reference frequency source. | Text: INTernal,EXTernal |
| SYSTem:BASE:REFerence:FREQuency:SOURce? | Read Reference Frequency Source. Reads the selected internal or external reference frequency source. | None |
| SYSTem:COMMunicate:NET:IPADdress? | Read IP Address. Returns the IP address of the LAN interface. | None |
| SYSTem:RESet:BASE | Reset Base Settings. Resets the base settings. | None |
| SYSTem:RESet:ALL | Reset All Settings. Resets the base settings and all applications. | None |
| SYSTem:PRESet:ALL | Preset All Settings. Presets the base settings and all applications. | None |
| SYSTem:ERRor:ALL? | Read Error Queue. Returns all entries of the error queue and clears it. | None |
| SYSTem:ERRor:COUNt? | Read Error Count. Returns the number of entries in the error queue. | None |
| SYSTem:BASE:VERSion? | Read Software Version. Returns the version of the base software. | None |
| Command | What it does | Arguments |
|---|---|---|
| INITiate:AFRF:MEASurement1:MEValuation | Start Measurement. Starts or restarts the AFRF multi-evaluation measurement (state RUN). | None |
| INITiate:AFRF:MEASurement1:FREQuency:COUNter | Start Frequency Counter. Starts the RF frequency search procedure. | None |
| INITiate:GPRF:MEASurement1:POWer | Start Power Measurement. Starts or restarts the GPRF power measurement (state RUN). | None |
| INITiate:GPRF:MEASurement1:SPECtrum | Start Spectrum Measurement. Starts or restarts the spectrum analyzer measurement (state RUN). | None |
| Command | What it does | Arguments |
|---|---|---|
| ABORt:AFRF:MEASurement1:MEValuation | Abort Measurement. Aborts the AFRF multi-evaluation measurement and releases resources (state OFF). | None |
| ABORt:AFRF:MEASurement1:FREQuency:COUNter | Abort Frequency Counter. Aborts the RF frequency search procedure. | None |
| ABORt:GPRF:MEASurement1:POWer | Abort Power Measurement. Aborts the GPRF power measurement and releases resources (state OFF). | None |
| ABORt:GPRF:MEASurement1:SPECtrum | Abort Spectrum Measurement. Aborts the spectrum analyzer measurement and releases resources (state OFF). | None |
| Command | What it does | Arguments |
|---|---|---|
| STOP:AFRF:MEASurement1:MEValuation | Stop Measurement. Halts the AFRF multi-evaluation measurement; results are kept (state RDY). | None |
| STOP:GPRF:MEASurement1:POWer | Stop Power Measurement. Halts the GPRF power measurement; results are kept (state RDY). | None |
| STOP:GPRF:MEASurement1:SPECtrum | Stop Spectrum Measurement. Halts the spectrum analyzer measurement; results are kept (state RDY). | None |
| Command | What it does | Arguments |
|---|---|---|
| SENSe:BASE:REFerence:FREQuency:LOCKed? | Read Reference Locked. Queries whether the reference frequency is locked. | None |
| SENSe:BASE:POWer:OMODe? | Read Power Supply Mode. Returns whether the instrument is powered by BATTery or MAINs. | None |
| SENSe:BASE:BATTery:CAPacity? | Read Battery Capacity. Returns the total remaining battery capacity in %. | None |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Returns the instrument identification string (manufacturer, model, serial, firmware). | None |
| *RST | Reset. Resets the instrument to its default state. | None |
| *CLS | Clear Status. Clears all event registers and the error queue. | None |
| *OPC? | Operation Complete. Returns 1 when all pending operations are complete. | None |
| *WAI | Wait To Continue. Waits until all pending operations are complete before continuing. | None |
| *SAV <Memory> | Save State. Stores the current settings in an intermediate memory. | Number |
| *RCL <Memory> | Recall State. Recalls settings from an intermediate memory. | Number |
This reference is generated from TestFlow's instrument library, which is drafted from Rohde & Schwarz's programming documentation for the CMA180 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Rohde & Schwarz's programming manual: check it before you rely on an edge case. Last updated September 27, 2026.
Yes. The Rohde & Schwarz CMA180 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 145 SCPI commands for it: 72 queries that read values or state, and 73 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, SOURce:AFRF:GENerator1:STATe ON and CONFigure:AFRF:MEASurement1:RFSettings:FREQuency?.
CONFigure:AFRF:MEASurement1:RFSettings:FREQuency? (read analyzer frequency). Reads the center frequency of the RF analyzer in Hz (100 kHz to 3 GHz). The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Rohde & Schwarz CMA180 through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.