Radio Test Set · SCPI command reference

Keysight E5515C and E5515E SCPI commands

The Keysight E5515C and E5515E radio test set is controlled remotely with SCPI commands. TestFlow's instrument library holds 172 commands for it, 96 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the E5515C on a real bench.

Commands
172
Queries
96
Subsystems
15
Models covered
2

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight E5515C or E5515E and writes the report from the measured results.

Get started

Control the Keysight E5515C from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keysight E5515C's VISA address

# Paste the address of your Keysight E5515C (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("RFANalyzer:CONTrol:POWer:AUTO ON")  # Turn On Analyzer Auto Power
print(inst.query("RFANalyzer:CONTrol:MEASurement:FREQuency:AUTO?"))  # Read Analyzer Auto Frequency State

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight E5515C or E5515E and writes the report from the measured results.

Get started

Run a E5515C test without writing code

Describe the test in a sentence, for example “turn on analyzer auto power on the E5515C, then read analyzer auto frequency state and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

172 commands in 15 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The E5515C and E5515E share this command set in TestFlow's library.

CALL subsystem (63)

CommandWhat it doesArguments
CALL:OPERating:MODE <Mode>Set Operating Mode. Sets the test set operating mode: Active Cell (CALL), GSM BCH or BCH+TCH test mode, GPRS BCH or BCH+PDTCH test mode, continuous wave (CW) or OFF.Text: CALL,GBTest,GBTTest,PBTest,PBPTest,CW,OFF
CALL:OPERating:MODE?Read Operating Mode. Reads the test set operating mode: Active Cell (CALL), GSM BCH or BCH+TCH test mode, GPRS BCH or BCH+PDTCH test mode, continuous wave (CW) or OFF.None
CALL:CELL:BAND <Band>Set Cell Band. Sets the band in which the broadcast channel (BCH) is transmitted.Text: PGSM,EGSM,DCS,PCS,GSM450,GSM480,GSM750,GSM850,RGSM
CALL:CELL:BAND?Read Cell Band. Reads the band in which the broadcast channel (BCH) is transmitted.None
CALL:CELL:BCHannel:ARFCn <Channel>Set BCH ARFCN. Sets the broadcast channel ARFCN for the selected broadcast band (e.g. PGSM 1-124, DCS 512-885, PCS 512-810, GSM850 128-251).Number
CALL:CELL:BCHannel:ARFCn?Read BCH ARFCN. Reads the broadcast channel ARFCN for the selected broadcast band (e.g. PGSM 1-124, DCS 512-885, PCS 512-810, GSM850 128-251).None
CALL:CELL:BCHannel:SCELl <Cell>Set Serving Cell. Sets what information is broadcast on the BCH (GSM, GPRS or EGPRS serving cell).Text: GSM,GPRS,EGPRS
CALL:CELL:BCHannel:SCELl?Read Serving Cell. Reads what information is broadcast on the BCH (GSM, GPRS or EGPRS serving cell).None
CALL:CELL:POWer <Level>Set Cell Power. Sets the cell (BCH) RF power in dBm for the selected format and turns the cell power state on.Number
CALL:CELL:POWer?Read Cell Power. Reads the cell (BCH) RF power in dBm for the selected format.None
CALL:CELL:POWer:STATe ONTurn On Cell Power. Turns the RF power of the cell on.None
CALL:CELL:POWer:STATe OFFTurn Off Cell Power. Turns the RF power of the cell off.None
CALL:CELL:POWer:STATe?Read Cell Power State. Reads the on/off state set by Turn_On_Cell_Power / Turn_Off_Cell_Power.None
CALL:CELL:RFGenerator:FREQuency <Freq>Set CW Generator Frequency. Sets the RF generator frequency in Hz used in CW operating mode (292 MHz to 2700 MHz).Number
CALL:CELL:RFGenerator:FREQuency?Read CW Generator Frequency. Reads the RF generator frequency in Hz used in CW operating mode (292 MHz to 2700 MHz).None
CALL:CELL:MCCode <MCC>Set Mobile Country Code. Sets the Mobile Country Code broadcast by the cell (0 to 999); cell operating mode must be OFF.Number
CALL:CELL:MCCode?Read Mobile Country Code. Reads the Mobile Country Code broadcast by the cell (0 to 999).None
CALL:CELL:MNCode <MNC>Set Mobile Network Code. Sets the Mobile Network Code broadcast by the cell (0 to 99); cell operating mode must be OFF.Number
CALL:CELL:MNCode?Read Mobile Network Code. Reads the Mobile Network Code broadcast by the cell (0 to 99).None
CALL:CELL:LACode <LAC>Set Location Area Code. Sets the cell Location Area Code (0 to 65535); cell operating mode must be OFF.Number
CALL:CELL:LACode?Read Location Area Code. Reads the cell Location Area Code (0 to 65535).None
CALL:CELL:BCCode <BCC>Set Base Station Colour Code. Sets the Base Station Colour Code (0 to 7); cell operating mode must be OFF.Number: 0,1,2,3,4,5,6,7
CALL:CELL:BCCode?Read Base Station Colour Code. Reads the Base Station Colour Code (0 to 7).None
CALL:CELL:NCCode <NCC>Set Network Colour Code. Sets the Network Colour Code (0 to 7); cell operating mode must be OFF.Number: 0,1,2,3,4,5,6,7
CALL:CELL:NCCode?Read Network Colour Code. Reads the Network Colour Code (0 to 7).None
CALL:BURSt:TYPE <Burst>Set Expected Burst. Sets the expected burst type used for measurement synchronization in test operating modes.Text: RACH,TSC0,TSC1,TSC2,TSC3,TSC4,TSC5,TSC6,TSC7
CALL:BURSt:TYPE?Read Expected Burst. Reads the expected burst type used for measurement synchronization in test operating modes.None
CALL:TCHannel:BAND <Band>Set TCH Band. Sets the GSM band used by the BS emulator for the traffic channel (TCH).Text: PGSM,EGSM,DCS,PCS,GSM450,GSM480,GSM750,GSM850,RGSM
CALL:TCHannel:BAND?Read TCH Band. Reads the GSM band used by the BS emulator for the traffic channel (TCH).None
CALL:TCHannel:ARFCn <Channel>Set TCH ARFCN. Sets the downlink/uplink TCH channel number for the selected TCH band.Number
CALL:TCHannel:ARFCn?Read TCH ARFCN. Reads the downlink/uplink TCH channel number for the selected TCH band.None
CALL:TCHannel:TSLot <Timeslot>Set TCH Timeslot. Sets the timeslot used for the downlink and uplink traffic channel (1 to 7).Number: 1,2,3,4,5,6,7
CALL:TCHannel:TSLot?Read TCH Timeslot. Reads the timeslot used for the downlink and uplink traffic channel (1 to 7).None
CALL:TCHannel:CMODe <Mode>Set TCH Channel Mode. Sets the speech channel mode: full rate, enhanced full rate or half rate speech.Text: FRSPeech,EFRSpeech,HRSPeech
CALL:TCHannel:CMODe?Read TCH Channel Mode. Reads the speech channel mode: full rate, enhanced full rate or half rate speech.None
CALL:TCHannel:LOOPback <Loop>Set TCH Loopback. Sets the MS traffic channel loopback type (A residual, B non-residual, C burst-by-burst) required for bit error measurements.Text: OFF,A,B,C
CALL:TCHannel:LOOPback?Read TCH Loopback. Reads the MS traffic channel loopback type (A residual, B non-residual, C burst-by-burst) required for bit error measurements.None
CALL:TCHannel:DOWNlink:SPEech <Speech>Set Downlink Speech. Sets the kind of speech data transmitted on the downlink TCH.Text: ECHO,NONE,PRBS15,SIN300,SIN1000,SIN3000
CALL:TCHannel:DOWNlink:SPEech?Read Downlink Speech. Reads the kind of speech data transmitted on the downlink TCH.None
CALL:TCHannel:POWer?Read TCH Burst Power. Reads the downlink TCH used burst power amplitude in dB (0 to 30 dB).None
CALL:MS:TXLevel <Level>Set MS TX Level. Sets the mobile station uplink power control level for the selected band (0 to 31).Number
CALL:MS:TXLevel?Read MS TX Level. Reads the mobile station uplink power control level for the selected band (0 to 31).None
CALL:MS:TADVance <TA>Set MS Timing Advance. Sets the mobile station uplink TCH timing advance for the selected band (0 to 63).Number
CALL:MS:TADVance?Read MS Timing Advance. Reads the mobile station uplink TCH timing advance for the selected band (0 to 63).None
CALL:MS:DTX:STATe ONTurn On MS DTX. Enables discontinuous transmission (DTX) for the mobile station.None
CALL:MS:DTX:STATe OFFTurn Off MS DTX. Disables discontinuous transmission (DTX) for the mobile station.None
CALL:MS:DTX:STATe?Read MS DTX State. Reads the on/off state set by Turn_On_MS_DTX / Turn_Off_MS_DTX.None
CALL:ORIGinateOriginate Call. Performs a base-station originated call to the mobile station.None
CALL:ENDEnd Call. Terminates the active voice or CSD call.None
CALL:STATus:STATe?Read Call Status. Reads the call status (IDLE / SREQ / ALER / CONN / DISC).None
CALL:CONNected:STATe?Read Call Connected. Returns 1 when the call is connected or 0 when idle; hangs during transitions until the timeout.None
CALL:CONNected:ARMArm Call Connected Detector. Arms the call-state-change detector used with Read_Call_Connected for call synchronization.None
CALL:CONNected:TIMeout <Seconds>Set Call Connected Timeout. Sets the maximum time in seconds the test set waits for a hanging connected-state query to complete.Number
CALL:CONNected:TIMeout?Read Call Connected Timeout. Reads the maximum time in seconds the test set waits for a hanging connected-state query to complete.None
CALL:ATTached:STATe?Read Data Attached. Returns 1 if the GPRS data connection is in the attached state, otherwise 0.None
CALL:PAGing:IMSI '<IMSI>'Set Paging IMSI. Sets the IMSI used for paging the mobile station.Text
CALL:PAGing:IMSI?Read Paging IMSI. Reads the IMSI used for paging the mobile station.None
CALL:MS:REPorted:IMSI?Read MS IMSI. Reads the IMSI of the SIM reported by the mobile station.None
CALL:MS:REPorted:IMEI?Read MS IMEI. Reads the IMEI reported by the mobile equipment.None
CALL:MS:REPorted:RXLevel?Read MS Reported RX Level. Reads the TCH received level reported by the mobile (0 to 63, relative to -110 dBm).None
CALL:MS:REPorted:RXQuality?Read MS Reported RX Quality. Reads the received signal quality reported by the mobile (0 to 7).None
CALL:MS:REPorted:TXLevel?Read MS Reported TX Level. Reads the last transmit level reported by the mobile station (0 to 31).None
CALL:MS:REPorted:TADVance?Read MS Reported Timing Advance. Reads the last TCH timing advance used by the mobile station (0 to 63).None

System (SYSTem) (31)

CommandWhat it doesArguments
SYSTem:APPLication:SELect:NAME '<Application>'Select Test Application. Selects the test application to run after reboot (name as shown in the Test Application Setup menu); the test set reboots in about 1 minute.Text
SYSTem:APPLication:SELect:NAME?Read Selected Test Application. Reads the test application that is selected and will run after the next reboot.None
SYSTem:APPLication:CURRent:NAME?Read Current Application Name. Reads the name of the test application currently running.None
SYSTem:APPLication:CURRent:REVision?Read Current Application Revision. Reads the revision of the test application currently running.None
SYSTem:APPLication:CATalog:NAME?Read Application Catalog. Reads the list of test applications installed in the test set.None
SYSTem:APPLication:FORMat:NAME '<Format>'Set Application Format. Switches between licensed formats of a fast switching or integrated test application (for example GSM or GPRS).Text: GSM,GPRS
SYSTem:APPLication:FORMat:NAME?Read Application Format. Reads the name of the currently active format.None
SYSTem:CURRent:TA:NAME?Read Current TA Name. Reads the name of the currently running test application.None
SYSTem:CURRent:TA:MODel?Read Current TA Model. Reads the model number of the currently running test application.None
SYSTem:CORRection:FREQuency <Freq_List>Set Amplitude Offset Frequencies. Sets up to 20 comma-separated frequencies for the RF IN/OUT amplitude offset table (Hz by default).Text
SYSTem:CORRection:FREQuency?Read Amplitude Offset Frequencies. Reads up to 20 comma-separated frequencies for the RF IN/OUT amplitude offset table (Hz by default).None
SYSTem:CORRection:GAIN <Offset_List>Set Amplitude Offset Gains. Sets up to 20 comma-separated amplitude offset values in dB paired with the offset frequencies.Text
SYSTem:CORRection:GAIN?Read Amplitude Offset Gains. Reads up to 20 comma-separated amplitude offset values in dB paired with the offset frequencies.None
SYSTem:CORRection:STATe ONTurn On Amplitude Offset. Turns the RF IN/OUT amplitude offset function on.None
SYSTem:CORRection:STATe OFFTurn Off Amplitude Offset. Turns the RF IN/OUT amplitude offset function off.None
SYSTem:CORRection:STATe?Read Amplitude Offset State. Reads the on/off state set by Turn_On_Amplitude_Offset / Turn_Off_Amplitude_Offset.None
SYSTem:MEASurement:RESetReset Measurements. Aborts all measurements and clears results; continuous measurements restart.None
SYSTem:PRESet2Full Preset. Performs a full preset of the test set (equivalent to SHIFT Preset).None
SYSTem:PRESet3Partial Preset. Performs a partial preset: disconnects any call and aborts measurements without changing measurement settings.None
SYSTem:ERRor?Read Error. Reads and removes the oldest entry in the error queue.None
SYSTem:ROSCillator:TIMebase?Read Timebase Source. Reads whether the internal or an external 10 MHz reference drives the time base.None
SYSTem:ROSCillator:LOCKed?Read Timebase Locked. Reads whether the reference oscillator is locked (1) or unlocked (0).None
SYSTem:COMMunicate:GPIB:SELF:ADDRess <Address>Set GPIB Address. Sets the GPIB address of the test set (0 to 30).Number
SYSTem:COMMunicate:GPIB:SELF:ADDRess?Read GPIB Address. Reads the GPIB address of the test set (0 to 30).None
SYSTem:COMMunicate:LAN:SELF:ADDRess <IP>Set LAN IP Address. Sets the test set LAN IP address (string, e.g. '192.168.0.10').Text
SYSTem:COMMunicate:LAN:SELF:ADDRess?Read LAN IP Address. Reads the test set LAN IP address (string, e.g. '192.168.0.10').None
SYSTem:BEEPer:STATe ONTurn On Beeper. Turns the beeper on.None
SYSTem:BEEPer:STATe OFFTurn Off Beeper. Turns the beeper off.None
SYSTem:BEEPer:STATe?Read Beeper State. Reads the on/off state set by Turn_On_Beeper / Turn_Off_Beeper.None
SYSTem:DATE?Read System Date. Reads the system date.None
SYSTem:TIME?Read System Time. Reads the system time.None

SET subsystem (28)

CommandWhat it doesArguments
SETup:ALL:CONTinuous ONTurn On All Continuous Trigger. Sets the trigger arm of all measurements to continuous.None
SETup:ALL:CONTinuous OFFTurn Off All Continuous Trigger. Sets the trigger arm of all measurements to single.None
SETup:TXPower:CONTinuous ONTurn On TX Power Continuous. Sets the TX power measurement trigger arm to continuous.None
SETup:TXPower:CONTinuous OFFTurn Off TX Power Continuous. Sets the TX power measurement trigger arm to single.None
SETup:TXPower:COUNt <Count>Set TX Power Count. Sets the TX power multi-measurement count (1 to 999) and turns the count state on.Number
SETup:TXPower:COUNt?Read TX Power Count. Reads the TX power multi-measurement count (1 to 999).None
SETup:TXPower:TRIGger:SOURce <Source>Set TX Power Trigger Source. Selects the TX power measurement trigger source.Text: AUTO,PROTocol,RISE,IMMediate
SETup:TXPower:TRIGger:SOURce?Read TX Power Trigger Source. Reads the TX power measurement trigger source.None
SETup:TXPower:TIMeout <Seconds>Set TX Power Timeout. Sets the TX power measurement timeout in seconds (0.1 to 999 s).Number
SETup:TXPower:TIMeout?Read TX Power Timeout. Reads the TX power measurement timeout in seconds (0.1 to 999 s).None
SETup:PFERror:COUNt <Count>Set PFER Count. Sets the phase and frequency error multi-measurement count (1 to 999).Number
SETup:PFERror:COUNt?Read PFER Count. Reads the phase and frequency error multi-measurement count (1 to 999).None
SETup:PFERror:SYNC <Sync>Set PFER Burst Sync. Sets the burst synchronization method for phase and frequency error.Text: MIDamble,AMPLitude,NONE
SETup:PFERror:SYNC?Read PFER Burst Sync. Reads the burst synchronization method for phase and frequency error.None
SETup:PVTime:COUNt <Count>Set PVT Count. Sets the power versus time multi-measurement count (1 to 999).Number
SETup:PVTime:COUNt?Read PVT Count. Reads the power versus time multi-measurement count (1 to 999).None
SETup:PVTime:TIME <Offset_List>Set PVT Time Offsets. Sets up to 12 comma-separated PvT time offsets (-50 us to 593 us) at which power is reported.Text
SETup:PVTime:TIME?Read PVT Time Offsets. Reads up to 12 comma-separated PvT time offsets (-50 us to 593 us) at which power is reported.None
SETup:ORFSpectrum:MODulation:COUNt <Count>Set ORFS Modulation Count. Sets the ORFS due to modulation multi-measurement count (1 to 999).Number
SETup:ORFSpectrum:MODulation:COUNt?Read ORFS Modulation Count. Reads the ORFS due to modulation multi-measurement count (1 to 999).None
SETup:ORFSpectrum:MODulation:FREQuency <Offset_List>Set ORFS Modulation Offsets. Sets the list of ORFS due to modulation frequency offsets (Hz by default).Text
SETup:ORFSpectrum:MODulation:FREQuency?Read ORFS Modulation Offsets. Reads the list of ORFS due to modulation frequency offsets (Hz by default).None
SETup:ORFSpectrum:SWITching:FREQuency <Offset_List>Set ORFS Switching Offsets. Sets up to 8 comma-separated ORFS due to switching frequency offsets (-1.8 MHz to -10 Hz and +10 Hz to +1.8 MHz).Text
SETup:ORFSpectrum:SWITching:FREQuency?Read ORFS Switching Offsets. Reads up to 8 comma-separated ORFS due to switching frequency offsets (-1.8 MHz to -10 Hz and +10 Hz to +1.8 MHz).None
SETup:BERRor:TYPE <Type>Set BER Type. Sets the bit error measurement type (Type IA/IB/II, residual or non-residual).Text: TYPEIA,TYPEIB,TYPEII,RESTYPEIA,RESTYPEIB,RESTYPEII
SETup:BERRor:TYPE?Read BER Type. Reads the bit error measurement type (Type IA/IB/II, residual or non-residual).None
SETup:BERRor:COUNt <Bits>Set BER Count. Sets the number of bits tested by the BER measurement (1 to 999000).Number
SETup:BERRor:COUNt?Read BER Count. Reads the number of bits tested by the BER measurement (1 to 999000).None

Fetch (FETCh) (15)

CommandWhat it doesArguments
FETCh:TXPower?Fetch TX Power. Fetches the integrity indicator and average transmit power in dBm.None
FETCh:TXPower:POWer:AVERage?Fetch TX Power Average. Fetches the average TX power result in dBm.None
FETCh:TXPower:POWer:MAXimum?Fetch TX Power Maximum. Fetches the maximum TX power result in dBm.None
FETCh:TXPower:POWer:MINimum?Fetch TX Power Minimum. Fetches the minimum TX power result in dBm.None
FETCh:TXPower:INTegrity?Fetch TX Power Integrity. Fetches the TX power measurement integrity indicator (0 = normal).None
FETCh:PFERror?Fetch PFER. Fetches integrity, maximum RMS phase error, maximum peak phase error and worst frequency error.None
FETCh:PFERror:RMS:MAXimum?Fetch PFER RMS Phase Max. Fetches the maximum RMS phase error in degrees.None
FETCh:PFERror:PEAK:MAXimum?Fetch PFER Peak Phase Max. Fetches the maximum peak phase error in degrees.None
FETCh:PFERror:FERRor:WORSt?Fetch PFER Frequency Error Worst. Fetches the worst-case frequency error in Hz.None
FETCh:PVTime:MASK?Fetch PVT Mask. Fetches the power versus time mask pass/fail indicator (0 pass, 1 fail).None
FETCh:PVTime:TXPower?Fetch PVT TX Power. Fetches the average TX power result of the PvT measurement in dBm.None
FETCh:ORFSpectrum:MODulation?Fetch ORFS Modulation. Fetches the ORFS due to modulation results in dB.None
FETCh:ORFSpectrum:SWITChing?Fetch ORFS Switching. Fetches the ORFS due to switching results in dBm.None
FETCh:BERRor?Fetch BER. Fetches integrity, bits tested, bit error ratio and bit error count.None
FETCh:BERRor:RATio?Fetch BER Ratio. Fetches the bit error ratio in percent.None

RFAN subsystem (10)

CommandWhat it doesArguments
RFANalyzer:CONTrol:POWer:AUTO ONTurn On Analyzer Auto Power. Sets the receiver expected power control to automatic.None
RFANalyzer:CONTrol:POWer:AUTO OFFTurn Off Analyzer Auto Power. Sets the receiver expected power control to manual.None
RFANalyzer:CONTrol:POWer:AUTO?Read Analyzer Auto Power State. Reads the on/off state set by Turn_On_Analyzer_Auto_Power / Turn_Off_Analyzer_Auto_Power.None
RFANalyzer:CW:EXPected:POWer <Level>Set CW Expected Power. Sets the power level in dBm the mobile is expected to transmit in CW mode (-60 to +53 dBm).Number
RFANalyzer:CW:EXPected:POWer?Read CW Expected Power. Reads the power level in dBm the mobile is expected to transmit in CW mode (-60 to +53 dBm).None
RFANalyzer:CONTrol:MEASurement:FREQuency:AUTO ONTurn On Analyzer Auto Frequency. Sets the measuring receiver frequency control to automatic.None
RFANalyzer:CONTrol:MEASurement:FREQuency:AUTO OFFTurn Off Analyzer Auto Frequency. Sets the measuring receiver frequency control to manual.None
RFANalyzer:CONTrol:MEASurement:FREQuency:AUTO?Read Analyzer Auto Frequency State. Reads the on/off state set by Turn_On_Analyzer_Auto_Frequency / Turn_Off_Analyzer_Auto_Frequency.None
RFANalyzer:MANual:MEASurement <Freq>Set Analyzer Manual Frequency. Sets the measuring receiver frequency in Hz in manual mode (292.5 MHz to 2700 MHz).Number
RFANalyzer:MANual:MEASurement?Read Analyzer Manual Frequency. Reads the measuring receiver frequency in Hz in manual mode (292.5 MHz to 2700 MHz).None

Read (READ) (5)

CommandWhat it doesArguments
READ:TXPower?Read TX Power. Initiates and fetches one TX power measurement (integrity indicator and average TX power in dBm).None
READ:PFERror?Read PFER. Initiates and fetches one phase/frequency error measurement (integrity, RMS phase, peak phase, worst frequency error).None
READ:PVTime?Read PVT. Initiates and fetches one power versus time measurement.None
READ:ORFSpectrum?Read ORFS. Initiates and fetches one output RF spectrum measurement.None
READ:BERRor?Read BER. Initiates and fetches one bit error measurement.None

Initiate (INITiate) (4)

CommandWhat it doesArguments
INITiate:<Measurement>Initiate Measurement. Starts (activates) the selected measurement.Text: TXPower,PFERror,PVTime,ORFSpectrum,BERRor,FBERror,DPOWer,ETXPower,IQTuning,AAUDio,DAUDio,SMONitor
INITiate:<Measurement>:OFFDeactivate Measurement. Deactivates the selected measurement.Text: TXPower,PFERror,PVTime,ORFSpectrum,BERRor,FBERror,DPOWer,ETXPower,IQTuning,AAUDio,DAUDio,SMONitor
INITiate:DONE?Read Measurement Done. Returns the mnemonic of a measurement that has completed or timed out (NONE if none).None
INITiate:ON?Read Active Measurements. Returns the list of measurements that are currently ON.None

RFG subsystem (2)

CommandWhat it doesArguments
RFGenerator:OUTPut <Port>Set RF Output Port. Selects the RF generator output port: RF IN/OUT (IO) or RF OUT ONLY (OUT, requires the second RF output hardware).Text: IO,OUT
RFGenerator:OUTPut?Read RF Output Port. Reads the RF generator output port: RF IN/OUT (IO) or RF OUT ONLY (OUT, requires the second RF output hardware).None

Display (DISPlay) (2)

CommandWhat it doesArguments
DISPlay:MODE <Mode>Set Display Mode. Sets the display mode: FAST (display off for speed) or TRACK.Text: FAST,TRACK
DISPlay:MODE?Read Display Mode. Reads the display mode: FAST (display off for speed) or TRACK.None

Abort (1)

CommandWhat it doesArguments
ABORt:ALLAbort All Measurements. Stops all active measurements.None

Status (STATus) (1)

CommandWhat it doesArguments
STATus:PRESetStatus Preset. Presets the status subsystem enable and transition registers.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (7)

CommandWhat it doesArguments
*IDN?Identify. Returns the identification string (manufacturer, model, serial number, 0).None
*RSTReset. Resets the test set to its default state.None
*CLSClear Status. Clears the status registers and closes the error message screen.None
*OPT?Read Options. Returns the installed options.None
*ESR?Read Event Status. Reads and clears the standard event status register.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the E5515C and E5515E and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated September 27, 2026.

Frequently asked questions

Does the Keysight E5515C support SCPI remote control?

Yes. The Keysight E5515C accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 172 SCPI commands for it: 96 queries that read values or state, and 76 settings and actions.

How do I control the Keysight E5515C from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, RFANalyzer:CONTrol:POWer:AUTO ON and RFANalyzer:CONTrol:MEASurement:FREQuency:AUTO?.

Which command reads a value from the Keysight E5515C?

RFANalyzer:CONTrol:MEASurement:FREQuency:AUTO? (read analyzer auto frequency state). Reads the on/off state set by Turn_On_Analyzer_Auto_Frequency / Turn_Off_Analyzer_Auto_Frequency. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Keysight E5515C without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight E5515C through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today