Phase Noise Analyzer · SCPI command reference

Keysight E5055A SCPI commands

The Keysight E5055A phase noise analyzer is controlled remotely with SCPI commands. TestFlow's instrument library holds 204 commands for it, 35 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the E5055A on a real bench.

Commands
204
Queries
35
Subsystems
16
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight E5055A and writes the report from the measured results.

Get started

Control the Keysight E5055A from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keysight E5055A's VISA address

# Paste the address of your Keysight E5055A (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:STATe ON")  # Carrier Search On
print(inst.query("SENSe:CLASs:NAME?"))  # Read Class Name

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight E5055A and writes the report from the measured results.

Get started

Run a E5055A test without writing code

Describe the test in a sentence, for example “carrier search on on the E5055A, then read class name and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

204 commands in 16 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Sense (SENSe) (84)

CommandWhat it doesArguments
SENSe:CLASs:NAME?Read Class Name. Returns the measurement class name of channel 1.None
SENSe:PN:NTYPe <Type>Set Noise Type. Selects phase noise / residual (additive) noise / baseband noise measurement.Text: PNOise,RESidual,BASeband
SENSe:PN:PORT <Port>Set Input Port. Selects RF input 1 or 2 for the phase noise measurement.Number: 1,2
SENSe:PN:SWEep:CARRier:FREQuency <Frequency>Set Carrier Frequency. Sets the carrier frequency in Hz (*RST 1 GHz).Number
SENSe:PN:SWEep:CARRier:FREQuency?Read Carrier Frequency Setting. Returns the carrier frequency setting in Hz.None
SENSe:FREQuency:STARt <Frequency>Set Start Offset. Sets the start offset frequency of the phase noise measurement in Hz.Number
SENSe:FREQuency:STOP <Frequency>Set Stop Offset. Sets the stop offset frequency of the phase noise measurement in Hz.Number
SENSe:FREQuency:STARt?Read Start Offset. Returns the start offset frequency in Hz.None
SENSe:FREQuency:STOP?Read Stop Offset. Returns the stop offset frequency in Hz.None
SENSe:PN:SWEep:TYPE <Type>Set Sweep Type. Selects logarithmic or segment phase noise sweep.Text: LOGarithmic,SEGMent
SENSe:PN:BWIDth:RESolution:RATio <Percent>Set RBW Ratio. Sets the RBW as a percentage of every half decade offset frequency (*RST 10 %).Number
SENSe:PN:CORRelation:COUNt <Count>Set XCorr Count. Sets the cross-correlation count (1 to 100000); more correlations lower the noise floor.Number: 1,100000
SENSe:PN:CORRelation:MODE <Mode>Set XCorr Mode. Selects normal or fast cross-correlation mode.Text: NORMal,FAST
SENSe:PN:SWEep:OFFSet:MODE <Mode>Set Offset Mode. Selects the offset frequency range mode.Text: NORMal,EXTend,MAXimum
SENSe:PN:SWEep:OFFSet:EXTend ONOffset Extend On. Turns the extended offset frequency range on.None
SENSe:PN:SWEep:OFFSet:EXTend OFFOffset Extend Off. Turns the extended offset frequency range off.None
SENSe:PN:SWEep:DETector:FUNCtion <Function>Set Detector Function. Selects sample averaging or average detector mode (maximum offset mode).Text: SAVerage,AVERage
SENSe:PN:SEGMent<Segment>:BWIDth:RESolution <Bandwidth>Set Segment RBW. Sets the resolution bandwidth in Hz of a phase noise segment.Number; Number
SENSe:PN:SEGMent<Segment>:CORRelation:COUNt <Count>Set Segment XCorr. Sets the cross-correlation count (1 to 100000) of a phase noise segment.Number; Number: 1,100000
SENSe:PN:SEGMent:COUNt?Read Segment Count. Returns the number of phase noise segments currently defined.None
SENSe:PN:SEGMent<Segment>:FREQuency:STARt?Read Segment Start. Returns the start offset frequency in Hz of a segment.Number
SENSe:PN:SEGMent<Segment>:FREQuency:STOP?Read Segment Stop. Returns the stop offset frequency in Hz of a segment.Number
SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:STATe ONCarrier Search On. Turns the broadband carrier search on.None
SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:STATe OFFCarrier Search Off. Turns the broadband carrier search off.None
SENSe:PN:ADJust:CONFigure:FREQuency:LIMit:CENTer <Frequency>Set Search Center. Sets the center frequency in Hz of the carrier search range.Number
SENSe:PN:ADJust:CONFigure:FREQuency:LIMit:SPAN <Frequency>Set Search Span. Sets the span in Hz of the carrier search range.Number
SENSe:PN:ADJust:CONFigure:FREQuency:LIMit:STARt <Frequency>Set Search Start. Sets the start frequency in Hz of the carrier search range.Number
SENSe:PN:ADJust:CONFigure:FREQuency:LIMit:STOP <Frequency>Set Search Stop. Sets the stop frequency in Hz of the carrier search range.Number
SENSe:PN:ADJust:CONFigure:LEVel:THReshold <Level>Set Search Threshold. Sets the carrier search threshold in dBm (*RST -20 dBm).Number
SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:RESult?Search Carrier Once. Executes the carrier search and returns found flag, carrier frequency and power.None
SENSe:PN:POWer:INPut<Input>:LEVel:AUTO ONAuto Level On. Turns the auto leveling of RF input 1 or 2 at pre-measurement on.Number: 1,2
SENSe:PN:POWer:INPut<Input>:LEVel:AUTO OFFAuto Level Off. Turns the auto leveling of RF input 1 or 2 at pre-measurement off.Number: 1,2
SENSe:PN:POWer:INPut<Input>:LEVel:MAXimum <Level>Set Max Input Level. Sets the maximum input level in dBm (-45 to 27) used to calculate the attenuation.Number: 1,2; Number: -45,27
SENSe:PN:POWer:INPut<Input>:ATTenuation:AUTO ONAuto Attenuation On. Turns the automatic attenuator setting from the maximum input level on.Number: 1,2
SENSe:PN:POWer:INPut<Input>:ATTenuation:AUTO OFFAuto Attenuation Off. Turns the automatic attenuator setting from the maximum input level off.Number: 1,2
SENSe:PN:POWer:INPut<Input>:ATTenuation <Attenuation>Set Attenuation. Sets the manual attenuation in dB of RF input 1 or 2 (0 to 30 dB in 1 dB steps).Number: 1,2; Number: 0,30
SENSe:PN:POWer:INPut<Input>:CORRection:STATe ONPower Correction On. Turns the power correction of RF input 1 or 2 on.Number: 1,2
SENSe:PN:POWer:INPut<Input>:CORRection:STATe OFFPower Correction Off. Turns the power correction of RF input 1 or 2 off.Number: 1,2
SENSe:PN:POWer:INPut<Input>:CORRection:LOAD <FileName>Load Power Correction. Loads power correction data of RF input 1 or 2 from a csv file (string in quotes).Number: 1,2; Text
SENSe:PN:PULSe:STATe ONPulse Mode On. Turns pulsed phase noise measurement on.None
SENSe:PN:PULSe:STATe OFFPulse Mode Off. Turns pulsed phase noise measurement off.None
SENSe:PN:PULSe:PERiod <Period>Set Pulse Period. Sets the pulse modulation period in s.Number
SENSe:PN:PULSe:PORT<Port>:GATE:STATe ONPulse Gate On. Turns pulse modulation gating of the RF port on.Number: 1,2
SENSe:PN:PULSe:PORT<Port>:GATE:STATe OFFPulse Gate Off. Turns pulse modulation gating of the RF port off.Number: 1,2
SENSe:PN:PULSe:PORT<Port>:GATE:LEVel <Level>Set Pulse Gate Level. Sets the pulse gate level in dBm.Number: 1,2; Number
SENSe:PN:ROSCillator:SOURce:FIXed:STATe ONFixed Internal Reference On. Turns the always-use-internal-reference for phase noise measurements on.None
SENSe:PN:ROSCillator:SOURce:FIXed:STATe OFFFixed Internal Reference Off. Turns the always-use-internal-reference for phase noise measurements off.None
SENSe:PN:MEXTernal:STATe ONExternal Mixer On. Turns the external mixer measurement on.None
SENSe:PN:MEXTernal:STATe OFFExternal Mixer Off. Turns the external mixer measurement off.None
SENSe:PN:MEXTernal:HARMonic <Harmonic>Set Mixer Harmonic. Sets the LO harmonic number used for external mixing.Number
SENSe:PN:MEXTernal:LO<LO>:STATe ONMixer LO Output On. Turns the LO 1 or 2 output power on.Number: 1,2
SENSe:PN:MEXTernal:LO<LO>:STATe OFFMixer LO Output Off. Turns the LO 1 or 2 output power off.Number: 1,2
SENSe:PN:MEXTernal:LO<LO>:LEVel <Level>Set Mixer LO Level. Sets the LO 1 or 2 output power level.Number: 1,2; Number
SENSe:PN:MIXer:STATe ONFrequency Translation On. Turns frequency translation (DUT count 2 residual measurement) on.None
SENSe:PN:MIXer:STATe OFFFrequency Translation Off. Turns frequency translation (DUT count 2 residual measurement) off.None
SENSe:PN:MIXer:LO:FREQuency <Frequency>Set Translation LO Frequency. Sets the external LO frequency in Hz for frequency conversion.Number
SENSe:PN:MIXer:INPut:FREQuency:NUMerator <M>Set Translation Multiplier. Sets the multiplier M of the frequency conversion equation.Number
SENSe:PN:MIXer:INPut:FREQuency:DENominator <N>Set Translation Divisor. Sets the divisor N of the frequency conversion equation.Number
SENSe:PN:PDELay:AUTO:STATe ONPath Delay Auto On. Turns automatic path delay compensation on.None
SENSe:PN:PDELay:AUTO:STATe OFFPath Delay Auto Off. Turns automatic path delay compensation off.None
SENSe:PN:PDELay:VALue <Delay>Set Path Delay. Sets the manual path delay in s.Number
SENSe:PN:RESidual:DUT:COUNt <Count>Set Residual DUT Count. Sets the DUT count (1 or 2) for residual measurement.Number: 1,2
SENSe:PN:RESidual:SUBTract:STATe ONSubtract 3dB On. Turns the subtract 3 dB (identical DUT) correction on.None
SENSe:PN:RESidual:SUBTract:STATe OFFSubtract 3dB Off. Turns the subtract 3 dB (identical DUT) correction off.None
SENSe:PN:SOURce:FREQuency <Frequency>Set Residual Source Frequency. Sets the source 1/2 output frequency in Hz for residual measurement.Number
SENSe:PN:SOURce:POWer:LEVel <Level>Set Residual Source Power. Sets the source 1/2 output power in dBm for residual measurement.Number
SENSe:PN:SOURce:RESolution <Resolution>Set Residual Source Resolution. Selects the source power resolution (HIGH 0.1 dB / NORMal 0.5 dB).Text: HIGH,NORMal
SENSe:PN:BASeband:INPut:COUPling <Coupling>Set Baseband Coupling. Selects the DC block capacitor type for baseband measurement.Text: ACLF,ACHF
SENSe:PN:BASeband:GAIN <Gain>Set Baseband Gain. Sets the baseband path gain.Number
SENSe:PN:BASeband:DISChargeDischarge DC Block. Discharges the baseband DC block capacitor.None
SENSe:PN:XGINdicator:STATe ONXCorr Gain Indicator On. Turns the cross-correlation gain indicator on.None
SENSe:PN:XGINdicator:STATe OFFXCorr Gain Indicator Off. Turns the cross-correlation gain indicator off.None
SENSe:PN:XCOunter:STATe ONXCorr Counter On. Turns the cross-correlation counter on.None
SENSe:PN:XCOunter:STATe OFFXCorr Counter Off. Turns the cross-correlation counter off.None
SENSe:PN:SPURious:SYSTem:OMISsion:STATe ONOmit System Spurs On. Turns omission of carrier derived system spurs on.None
SENSe:PN:SPURious:SYSTem:OMISsion:STATe OFFOmit System Spurs Off. Turns omission of carrier derived system spurs off.None
SENSe:SWEep:MODE <Mode>Set Sweep Mode. Sets the number of triggers channel 1 accepts.Text: HOLD,CONTinuous,GROups,SINGle
SENSe:SWEep:POINts?Read Sweep Points. Returns the number of data points of the measurement.None
SENSe:AVERage:STATe ONAveraging On. Turns trace averaging on.None
SENSe:AVERage:STATe OFFAveraging Off. Turns trace averaging off.None
SENSe:AVERage:COUNt <Count>Set Average Count. Sets the number of measurements to average (1 to 65536).Number: 1,65536
SENSe:AVERage:CLEarClear Average. Clears and restarts averaging.None
SENSe:ROSCillator:SOURce <Source>Set Reference Source. Selects the internal or external reference oscillator.Text: INTernal,EXTernal
SENSe:ROSCillator:SOURce:CONDition?Read Reference Condition. Returns whether the analyzer is locked to the external reference.None

Calculate (CALCulate) (59)

CommandWhat it doesArguments
CALCulate:MEASure<Meas>:DEFine "<Parameter>"Define Measurement. Creates a phase noise class measurement with the given measurement number (string in quotes e.g. 'PN:Phase Noise').Number; Text: PN:Phase Noise,AM:Phase Noise,PN+AM:Phase Noise,L(f):Phase Noise,Sphi(f):Phase Noise,Sv(f):Phase Noise,Sy(f):Phase Noise,AVAR:Phase Noise,ADEV:Phase Noise
CALCulate:MEASure<Meas>:DELeteDelete Measurement. Deletes the specified measurement.Number
CALCulate:MEASure:DELete:ALLDelete All Measurements. Deletes all measurements on the analyzer.None
CALCulate:MEASure<Meas>:PN:CARRier:FREQuency?Measure Carrier Frequency. Returns the measured carrier frequency in Hz.Number
CALCulate:MEASure<Meas>:PN:CARRier:LEVel?Measure Carrier Level. Returns the measured carrier power level in dBm.Number
CALCulate:MEASure<Meas>:DATA:X?Read Trace X. Returns the x-axis (offset frequency) values of the measurement.Number
CALCulate:MEASure<Meas>:DATA:FDATA?Read Trace Data. Returns the formatted trace data of the measurement (e.g. dBc/Hz).Number
CALCulate:MEASure<Meas>:DATA:FMEM?Read Memory Data. Returns the formatted memory trace data of the measurement.Number
CALCulate:MEASure<Meas>:PN:DATA:PDATa?Read PN Data. Returns the dBc data array of the measurement.Number
CALCulate:MEASure<Meas>:PN:DATA:SPData?Read Spur Flags. Returns the spurious flag (0 or 1) for each data point.Number
CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:TYPE <Type>Set Integration Type. Sets integration range 1-4 type (off / full span / custom).Number; Number: 1,2,3,4; Text: OFF,FULL,CUSTom
CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:STARt <Frequency>Set Integration Start. Sets the start offset in Hz of a custom integration range.Number; Number: 1,2,3,4; Number
CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:STOP <Frequency>Set Integration Stop. Sets the stop offset in Hz of a custom integration range.Number; Number: 1,2,3,4; Number
CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:WEIGhting <Filter>Set Integration Weighting. Selects a weighting filter file for the integration range (string in quotes).Number; Number: 1,2,3,4; Text
CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:DATA? <Result>Measure Integrated Result. Returns integrated noise / residual FM / AM / PM / RMS jitter / RMS radian / RMS degree for the range.Number; Number: 1,2,3,4; Text: IPN,RFM,RAM,RPM,RMSJ,RMSR,RMSD
CALCulate:MEASure<Meas>:PN:AVARiance:DEViation? <AvgTime>,<Cutoff>Measure Allan Deviation. Returns the Allan deviation for the average time in s and cut-off frequency in Hz.Number; Number; Number
CALCulate:MEASure<Meas>:PN:AVARiance:VARiance? <AvgTime>,<Cutoff>Measure Allan Variance. Returns the Allan variance for the average time in s and cut-off frequency in Hz.Number; Number; Number
CALCulate:MEASure<Meas>:PN:AVARiance:JITTer? <AvgTime>,<Cutoff>Measure Allan Jitter. Returns the jitter for the average time in s and cut-off frequency in Hz.Number; Number; Number
CALCulate:MEASure<Meas>:PN:SNOise:STATe ONSpot Noise On. Turns spot noise calculation on.Number
CALCulate:MEASure<Meas>:PN:SNOise:STATe OFFSpot Noise Off. Turns spot noise calculation off.Number
CALCulate:MEASure<Meas>:PN:SNOise:DECades:STATe ONSpot Noise Decades On. Turns spot noise calculation at every decade offset on.Number
CALCulate:MEASure<Meas>:PN:SNOise:DECades:STATe OFFSpot Noise Decades Off. Turns spot noise calculation at every decade offset off.Number
CALCulate:MEASure<Meas>:PN:SNOise:DECades:X?Read Spot Noise Decades X. Returns the decade offset frequencies in Hz of the spot noise.Number
CALCulate:MEASure<Meas>:PN:SNOise:DECades:Y?Read Spot Noise Decades Y. Returns the spot noise values at the decade offset frequencies.Number
CALCulate:MEASure<Meas>:PN:SNOise:USER<Position>:STATe ONSpot Noise User On. Turns the user defined spot noise offset 1-6 on.Number; Number: 1,2,3,4,5,6
CALCulate:MEASure<Meas>:PN:SNOise:USER<Position>:STATe OFFSpot Noise User Off. Turns the user defined spot noise offset 1-6 off.Number; Number: 1,2,3,4,5,6
CALCulate:MEASure<Meas>:PN:SNOise:USER<Position>:X <Frequency>Set Spot Noise Offset. Sets the user defined spot noise offset frequency in Hz.Number; Number: 1,2,3,4,5,6; Number
CALCulate:MEASure<Meas>:PN:SNOise:USER<Position>:Y?Measure Spot Noise. Returns the spot noise value at the user defined offset frequency.Number; Number: 1,2,3,4,5,6
CALCulate:MEASure<Meas>:PN:SPURious:ANALysis:STATe ONSpur Analysis On. Turns spurious analysis on.Number
CALCulate:MEASure<Meas>:PN:SPURious:ANALysis:STATe OFFSpur Analysis Off. Turns spurious analysis off.Number
CALCulate:MEASure<Meas>:PN:SPURious:DATA?Measure Spurs. Returns the list of detected spurs (frequency, power and jitter per spur).Number
CALCulate:MEASure<Meas>:PN:SPURious:OMISsion:STATe ONSpur Omission On. Turns omission of displayed spurs on.Number
CALCulate:MEASure<Meas>:PN:SPURious:OMISsion:STATe OFFSpur Omission Off. Turns omission of displayed spurs off.Number
CALCulate:MEASure<Meas>:PN:SPURious:SENSibility <Value>Set Spur Sensibility. Sets the spurious sensibility (*RST 3 = 3 x standard deviation).Number; Number
CALCulate:MEASure<Meas>:PN:SPURious:SORT <Order>Set Spur Sort. Selects the spurious table sort order.Number; Text: POWer,OFFSet
CALCulate:MEASure<Meas>:PN:SPURious:THReshold:LEVel:MINimum <Level>Set Spur Min Level. Sets the minimum spurious threshold level in dBc (-500 to 500).Number; Number: -500,500
CALCulate:MEASure<Meas>:PN:SPURious:THReshold:TABle:DELeteClear Spur Threshold Table. Deletes all entries of the spurious threshold table.Number
CALCulate:MEASure<Meas>:PN:SPURious:OSSPur:STATe ONUser Spur Omission On. Turns omission of user specified spurs on.Number
CALCulate:MEASure<Meas>:PN:SPURious:OSSPur:STATe OFFUser Spur Omission Off. Turns omission of user specified spurs off.Number
CALCulate:MEASure<Meas>:PN:SPURious:OSSPur:DELeteClear User Spur Table. Deletes all entries of the user spur table.Number
CALCulate:MEASure<Meas>:PN:QTRace ONQuick Trace On. Turns the quick trace (intermediate trace update) function on.Number
CALCulate:MEASure<Meas>:PN:QTRace OFFQuick Trace Off. Turns the quick trace (intermediate trace update) function off.Number
CALCulate:MEASure<Meas>:FORMat <Format>Set Display Format. Sets the display format of the measurement.Number; Text: MLOGarithmic,MLINear,PHASe,UPHase,REAL,IMAGinary,FREQuency,DFRequency
CALCulate:MEASure<Meas>:SMOothing:STATe ONSmoothing On. Turns trace smoothing on.Number
CALCulate:MEASure<Meas>:SMOothing:STATe OFFSmoothing Off. Turns trace smoothing off.Number
CALCulate:MEASure<Meas>:SMOothing:APERture <Percent>Set Smoothing Aperture. Sets the smoothing aperture in percent of the data points (1 to 25).Number; Number: 1,25
CALCulate:MEASure<Meas>:MATH:MEMorizeData To Memory. Stores the measurement trace into memory.Number
CALCulate:MEASure<Meas>:MARKer<Marker>:STATe ONMarker On. Turns the marker (1 to 15) on.Number; Number: 1,15
CALCulate:MEASure<Meas>:MARKer<Marker>:STATe OFFMarker Off. Turns the marker (1 to 15) off.Number; Number: 1,15
CALCulate:MEASure<Meas>:MARKer:AOFFAll Markers Off. Turns all markers of the measurement off.Number
CALCulate:MEASure<Meas>:MARKer<Marker>:X <Position>Set Marker X. Moves the marker to the given x position (offset frequency in Hz).Number; Number: 1,15; Number
CALCulate:MEASure<Meas>:MARKer<Marker>:X?Read Marker X. Returns the marker x position (offset frequency in Hz).Number; Number: 1,15
CALCulate:MEASure<Meas>:MARKer<Marker>:Y?Read Marker Y. Returns the marker y value (value and 0).Number; Number: 1,15
CALCulate:MEASure<Meas>:MARKer<Marker>:FUNCtion:EXECute <Function>Marker Search. Executes a marker search function.Number; Number: 1,15; Text: MAXimum,MINimum,RPEak,LPEak,NPEak,TARGet,LTARget,RTARget,PEAK,SPURious
CALCulate:MEASure<Meas>:MARKer<Marker>:DELTa ONDelta Marker On. Turns delta (relative to reference marker) mode of the marker on.Number; Number: 1,15
CALCulate:MEASure<Meas>:MARKer<Marker>:DELTa OFFDelta Marker Off. Turns delta (relative to reference marker) mode of the marker off.Number; Number: 1,15
CALCulate:MEASure<Meas>:LIMit:STATe ONLimit Test On. Turns the limit line test on.Number
CALCulate:MEASure<Meas>:LIMit:STATe OFFLimit Test Off. Turns the limit line test off.Number
CALCulate:MEASure<Meas>:LIMit:FAIL?Read Limit Fail. Returns the limit test result: 0 = pass and 1 = fail.Number

Display (DISPlay) (16)

CommandWhat it doesArguments
DISPlay:WINDow<Window>:STATe ONCreate Window. Creates the display window with the given number.Number
DISPlay:WINDow<Window>:STATe OFFDelete Window. Deletes the display window with the given number.Number
DISPlay:MEASure<Meas>:FEED <Window>Feed Measurement To Window. Displays the measurement in the given window.Number; Number
DISPlay:WINDow<Window>:TABLe:SPURious:ENABle ONSpur Table On. Turns the spurious table display on.Number
DISPlay:WINDow<Window>:TABLe:SPURious:ENABle OFFSpur Table Off. Turns the spurious table display off.Number
DISPlay:WINDow<Window>:TABLe:INOise:ENABle ONIntegrated Noise Table On. Turns the integrated noise table display on.Number
DISPlay:WINDow<Window>:TABLe:INOise:ENABle OFFIntegrated Noise Table Off. Turns the integrated noise table display off.Number
DISPlay:WINDow<Window>:TABLe:SNOise:ENABle ONSpot Noise Table On. Turns the spot noise table display on.Number
DISPlay:WINDow<Window>:TABLe:SNOise:ENABle OFFSpot Noise Table Off. Turns the spot noise table display off.Number
DISPlay:WINDow<Window>:TRACe<Trace>:Y:AUTOAutoscale Trace. Autoscales the specified trace in the window once.Number; Number
DISPlay:WINDow<Window>:Y:AUTOAutoscale All. Autoscales all traces in the window once.Number
DISPlay:MEASure<Meas>:Y:PDIVision <Value>Set Scale Per Division. Sets the y-axis scale per division of the measurement trace.Number; Number
DISPlay:MEASure<Meas>:Y:RLEVel <Level>Set Reference Level. Sets the y-axis reference level of the measurement trace.Number; Number
DISPlay:MEASure<Meas>:Y:RPOSition <Position>Set Reference Position. Sets the reference position (graticule from the bottom) of the measurement trace.Number; Number
DISPlay:UPDate:STATe ONDisplay Update On. Turns display updates on.None
DISPlay:UPDate:STATe OFFDisplay Update Off. Turns display updates off.None

Source (SOURce) (12)

CommandWhat it doesArguments
SOURce:DC:CATalog?Read DC Sources. Returns the names of the configured and active DC sources.None
SOURce:DC:STATe "<Name>",ONDC Source On. Turns the specified DC source on (name string in quotes).Text: VControl,VSupply1,VSupply2
SOURce:DC:STATe <Name>,OFFDC Source Off. Turns the specified DC source off (name string in quotes).Text: VControl,VSupply1,VSupply2
SOURce:DC:OUTPut <Name>,<State>Set DC Output State. Sets the DC channel output state of the specified DC source.Text: VControl,VSupply1,VSupply2; Text: ON,OFF,DEFault
SOURce:DC:STARt <Name>,<Value>Set DC Start. Sets the start DC value of the source (same start and stop for a fixed voltage).Text: VControl,VSupply1,VSupply2; Number
SOURce:DC:STOP <Name>,<Value>Set DC Stop. Sets the stop DC value of the source.Text: VControl,VSupply1,VSupply2; Number
SOURce:DC:CENTer <Name>,<Value>Set DC Center. Sets the center DC value of the source (use span 0 for a fixed voltage).Text: VControl,VSupply1,VSupply2; Number
SOURce:DC:SPAN <Name>,<Value>Set DC Span. Sets the span DC value of the source.Text: VControl,VSupply1,VSupply2; Number
SOURce:DC:LIMit:MAXimum <Name>,<Value>Set DC Limit Max. Sets the maximum DC limit of the source.Text: VControl,VSupply1,VSupply2; Number
SOURce:DC:LIMit:MINimum <Name>,<Value>Set DC Limit Min. Sets the minimum DC limit of the source.Text: VControl,VSupply1,VSupply2; Number
SOURce:DC:LNOise 'VControl',ONLow Noise VControl On. Turns the low noise V Control on.None
SOURce:DC:LNOise 'VControl',OFFLow Noise VControl Off. Turns the low noise V Control off.None

System (SYSTem) (6)

CommandWhat it doesArguments
SYSTem:FPResetPreset Blank. Performs a standard preset and then deletes all traces, measurements and windows (blank screen).None
SYSTem:DC:ENABle ONDC Outputs On. Turns all configured DC sources for all channels on.None
SYSTem:DC:ENABle OFFDC Outputs Off. Turns all configured DC sources for all channels off.None
SYSTem:ERRor?Read Error. Returns the next error in the error queue.None
SYSTem:ERRor:COUNt?Read Error Count. Returns the number of errors in the error queue.None
SYSTem:PRESetPreset. Presets the analyzer to factory defaults (same as *RST except data format).None

Mass memory (MMEMory) (6)

CommandWhat it doesArguments
MMEMory:STORe:STATe <FileName>Save State. Saves the instrument state to a .sta file (string in quotes).Text
MMEMory:STORe:CSARchive <FileName>Save State And Cal. Saves instrument state and calibration data to a .csa file (string in quotes).Text
MMEMory:LOAD:STATe <FileName>Load State. Loads an instrument state file (string in quotes).Text
MMEMory:STORe:DATA "<FileName>","CSV Formatted Data","<Scope>","Displayed",<Selector>Save Trace CSV. Saves formatted trace data to a csv file (file and scope strings in quotes; selector = measurement number or -1).Text; Text: Trace,Channel,Auto; Number
MMEMory:LOAD:PN:SPURious:THReshold <FileName>Load Spur Threshold Table. Loads the spurious threshold table from a file (string in quotes).Text
MMEMory:LOAD:PN:SPURious:OSSPur <FileName>Load User Spur Table. Loads the user spur omission table from a file (string in quotes).Text

Initiate (INITiate) (3)

CommandWhat it doesArguments
INITiate:CONTinuous ONContinuous Measurement. Sets internal (continuous) triggering.None
INITiate:CONTinuous OFFManual Trigger Mode. Sets manual triggering (use INITiate:IMMediate to trigger).None
INITiate:IMMediateStart Measurement. Sends one trigger to start a measurement (trigger source manual); synchronize with *OPC?.None

FORM subsystem (3)

CommandWhat it doesArguments
FORMat:DATA ASCii,0Set Data Format ASCII. Selects ASCII data transfer format.None
FORMat:DATA REAL,64Set Data Format Real64. Selects 64-bit binary data transfer format.None
FORMat:BORDer <Order>Set Byte Order. Selects the byte order of binary data transfers.Text: NORMal,SWAPped

Output (OUTPut) (2)

CommandWhat it doesArguments
OUTPut:STATe ONRF Output On. Turns the RF source power on.None
OUTPut:STATe OFFRF Output Off. Turns the RF source power off.None

Trigger (TRIGger) (1)

CommandWhat it doesArguments
TRIGger:SOURce <Source>Set Trigger Source. Selects the sweep trigger source.Text: IMMediate,EXTernal,MANual

Abort (1)

CommandWhat it doesArguments
ABORtAbort Measurement. Stops all sweeps and returns the trigger system to idle.None

HCOP subsystem (1)

CommandWhat it doesArguments
HCOPy:FILE <FileName>Save Screenshot. Saves the screen image to a .png/.bmp/.jpg file (string in quotes).Text

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (7)

CommandWhat it doesArguments
*OPT?Read Options. Returns the analyzer option configuration.None
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Returns the hardware status: 0 = no failures, otherwise the sum of failure weights.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the E5055A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the Keysight E5055A support SCPI remote control?

Yes. The Keysight E5055A accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 204 SCPI commands for it: 35 queries that read values or state, and 169 settings and actions.

How do I control the Keysight E5055A from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:STATe ON and SENSe:CLASs:NAME?.

Which command reads a value from the Keysight E5055A?

SENSe:CLASs:NAME? (read class name). Returns the measurement class name of channel 1. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Keysight E5055A without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight E5055A through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.