The Keysight E5055A phase noise analyzer is controlled remotely with SCPI commands. TestFlow's instrument library holds 204 commands for it, 35 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the E5055A on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your Keysight E5055A and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Keysight E5055A's VISA address
# Paste the address of your Keysight E5055A (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:STATe ON") # Carrier Search On
print(inst.query("SENSe:CLASs:NAME?")) # Read Class Name
inst.close()Describe the test; TestFlow's agent uses this command set to drive your Keysight E5055A and writes the report from the measured results.
Describe the test in a sentence, for example “carrier search on on the E5055A, then read class name and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
204 commands in 16 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
| Command | What it does | Arguments |
|---|---|---|
| SENSe:CLASs:NAME? | Read Class Name. Returns the measurement class name of channel 1. | None |
| SENSe:PN:NTYPe <Type> | Set Noise Type. Selects phase noise / residual (additive) noise / baseband noise measurement. | Text: PNOise,RESidual,BASeband |
| SENSe:PN:PORT <Port> | Set Input Port. Selects RF input 1 or 2 for the phase noise measurement. | Number: 1,2 |
| SENSe:PN:SWEep:CARRier:FREQuency <Frequency> | Set Carrier Frequency. Sets the carrier frequency in Hz (*RST 1 GHz). | Number |
| SENSe:PN:SWEep:CARRier:FREQuency? | Read Carrier Frequency Setting. Returns the carrier frequency setting in Hz. | None |
| SENSe:FREQuency:STARt <Frequency> | Set Start Offset. Sets the start offset frequency of the phase noise measurement in Hz. | Number |
| SENSe:FREQuency:STOP <Frequency> | Set Stop Offset. Sets the stop offset frequency of the phase noise measurement in Hz. | Number |
| SENSe:FREQuency:STARt? | Read Start Offset. Returns the start offset frequency in Hz. | None |
| SENSe:FREQuency:STOP? | Read Stop Offset. Returns the stop offset frequency in Hz. | None |
| SENSe:PN:SWEep:TYPE <Type> | Set Sweep Type. Selects logarithmic or segment phase noise sweep. | Text: LOGarithmic,SEGMent |
| SENSe:PN:BWIDth:RESolution:RATio <Percent> | Set RBW Ratio. Sets the RBW as a percentage of every half decade offset frequency (*RST 10 %). | Number |
| SENSe:PN:CORRelation:COUNt <Count> | Set XCorr Count. Sets the cross-correlation count (1 to 100000); more correlations lower the noise floor. | Number: 1,100000 |
| SENSe:PN:CORRelation:MODE <Mode> | Set XCorr Mode. Selects normal or fast cross-correlation mode. | Text: NORMal,FAST |
| SENSe:PN:SWEep:OFFSet:MODE <Mode> | Set Offset Mode. Selects the offset frequency range mode. | Text: NORMal,EXTend,MAXimum |
| SENSe:PN:SWEep:OFFSet:EXTend ON | Offset Extend On. Turns the extended offset frequency range on. | None |
| SENSe:PN:SWEep:OFFSet:EXTend OFF | Offset Extend Off. Turns the extended offset frequency range off. | None |
| SENSe:PN:SWEep:DETector:FUNCtion <Function> | Set Detector Function. Selects sample averaging or average detector mode (maximum offset mode). | Text: SAVerage,AVERage |
| SENSe:PN:SEGMent<Segment>:BWIDth:RESolution <Bandwidth> | Set Segment RBW. Sets the resolution bandwidth in Hz of a phase noise segment. | Number; Number |
| SENSe:PN:SEGMent<Segment>:CORRelation:COUNt <Count> | Set Segment XCorr. Sets the cross-correlation count (1 to 100000) of a phase noise segment. | Number; Number: 1,100000 |
| SENSe:PN:SEGMent:COUNt? | Read Segment Count. Returns the number of phase noise segments currently defined. | None |
| SENSe:PN:SEGMent<Segment>:FREQuency:STARt? | Read Segment Start. Returns the start offset frequency in Hz of a segment. | Number |
| SENSe:PN:SEGMent<Segment>:FREQuency:STOP? | Read Segment Stop. Returns the stop offset frequency in Hz of a segment. | Number |
| SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:STATe ON | Carrier Search On. Turns the broadband carrier search on. | None |
| SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:STATe OFF | Carrier Search Off. Turns the broadband carrier search off. | None |
| SENSe:PN:ADJust:CONFigure:FREQuency:LIMit:CENTer <Frequency> | Set Search Center. Sets the center frequency in Hz of the carrier search range. | Number |
| SENSe:PN:ADJust:CONFigure:FREQuency:LIMit:SPAN <Frequency> | Set Search Span. Sets the span in Hz of the carrier search range. | Number |
| SENSe:PN:ADJust:CONFigure:FREQuency:LIMit:STARt <Frequency> | Set Search Start. Sets the start frequency in Hz of the carrier search range. | Number |
| SENSe:PN:ADJust:CONFigure:FREQuency:LIMit:STOP <Frequency> | Set Search Stop. Sets the stop frequency in Hz of the carrier search range. | Number |
| SENSe:PN:ADJust:CONFigure:LEVel:THReshold <Level> | Set Search Threshold. Sets the carrier search threshold in dBm (*RST -20 dBm). | Number |
| SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:RESult? | Search Carrier Once. Executes the carrier search and returns found flag, carrier frequency and power. | None |
| SENSe:PN:POWer:INPut<Input>:LEVel:AUTO ON | Auto Level On. Turns the auto leveling of RF input 1 or 2 at pre-measurement on. | Number: 1,2 |
| SENSe:PN:POWer:INPut<Input>:LEVel:AUTO OFF | Auto Level Off. Turns the auto leveling of RF input 1 or 2 at pre-measurement off. | Number: 1,2 |
| SENSe:PN:POWer:INPut<Input>:LEVel:MAXimum <Level> | Set Max Input Level. Sets the maximum input level in dBm (-45 to 27) used to calculate the attenuation. | Number: 1,2; Number: -45,27 |
| SENSe:PN:POWer:INPut<Input>:ATTenuation:AUTO ON | Auto Attenuation On. Turns the automatic attenuator setting from the maximum input level on. | Number: 1,2 |
| SENSe:PN:POWer:INPut<Input>:ATTenuation:AUTO OFF | Auto Attenuation Off. Turns the automatic attenuator setting from the maximum input level off. | Number: 1,2 |
| SENSe:PN:POWer:INPut<Input>:ATTenuation <Attenuation> | Set Attenuation. Sets the manual attenuation in dB of RF input 1 or 2 (0 to 30 dB in 1 dB steps). | Number: 1,2; Number: 0,30 |
| SENSe:PN:POWer:INPut<Input>:CORRection:STATe ON | Power Correction On. Turns the power correction of RF input 1 or 2 on. | Number: 1,2 |
| SENSe:PN:POWer:INPut<Input>:CORRection:STATe OFF | Power Correction Off. Turns the power correction of RF input 1 or 2 off. | Number: 1,2 |
| SENSe:PN:POWer:INPut<Input>:CORRection:LOAD <FileName> | Load Power Correction. Loads power correction data of RF input 1 or 2 from a csv file (string in quotes). | Number: 1,2; Text |
| SENSe:PN:PULSe:STATe ON | Pulse Mode On. Turns pulsed phase noise measurement on. | None |
| SENSe:PN:PULSe:STATe OFF | Pulse Mode Off. Turns pulsed phase noise measurement off. | None |
| SENSe:PN:PULSe:PERiod <Period> | Set Pulse Period. Sets the pulse modulation period in s. | Number |
| SENSe:PN:PULSe:PORT<Port>:GATE:STATe ON | Pulse Gate On. Turns pulse modulation gating of the RF port on. | Number: 1,2 |
| SENSe:PN:PULSe:PORT<Port>:GATE:STATe OFF | Pulse Gate Off. Turns pulse modulation gating of the RF port off. | Number: 1,2 |
| SENSe:PN:PULSe:PORT<Port>:GATE:LEVel <Level> | Set Pulse Gate Level. Sets the pulse gate level in dBm. | Number: 1,2; Number |
| SENSe:PN:ROSCillator:SOURce:FIXed:STATe ON | Fixed Internal Reference On. Turns the always-use-internal-reference for phase noise measurements on. | None |
| SENSe:PN:ROSCillator:SOURce:FIXed:STATe OFF | Fixed Internal Reference Off. Turns the always-use-internal-reference for phase noise measurements off. | None |
| SENSe:PN:MEXTernal:STATe ON | External Mixer On. Turns the external mixer measurement on. | None |
| SENSe:PN:MEXTernal:STATe OFF | External Mixer Off. Turns the external mixer measurement off. | None |
| SENSe:PN:MEXTernal:HARMonic <Harmonic> | Set Mixer Harmonic. Sets the LO harmonic number used for external mixing. | Number |
| SENSe:PN:MEXTernal:LO<LO>:STATe ON | Mixer LO Output On. Turns the LO 1 or 2 output power on. | Number: 1,2 |
| SENSe:PN:MEXTernal:LO<LO>:STATe OFF | Mixer LO Output Off. Turns the LO 1 or 2 output power off. | Number: 1,2 |
| SENSe:PN:MEXTernal:LO<LO>:LEVel <Level> | Set Mixer LO Level. Sets the LO 1 or 2 output power level. | Number: 1,2; Number |
| SENSe:PN:MIXer:STATe ON | Frequency Translation On. Turns frequency translation (DUT count 2 residual measurement) on. | None |
| SENSe:PN:MIXer:STATe OFF | Frequency Translation Off. Turns frequency translation (DUT count 2 residual measurement) off. | None |
| SENSe:PN:MIXer:LO:FREQuency <Frequency> | Set Translation LO Frequency. Sets the external LO frequency in Hz for frequency conversion. | Number |
| SENSe:PN:MIXer:INPut:FREQuency:NUMerator <M> | Set Translation Multiplier. Sets the multiplier M of the frequency conversion equation. | Number |
| SENSe:PN:MIXer:INPut:FREQuency:DENominator <N> | Set Translation Divisor. Sets the divisor N of the frequency conversion equation. | Number |
| SENSe:PN:PDELay:AUTO:STATe ON | Path Delay Auto On. Turns automatic path delay compensation on. | None |
| SENSe:PN:PDELay:AUTO:STATe OFF | Path Delay Auto Off. Turns automatic path delay compensation off. | None |
| SENSe:PN:PDELay:VALue <Delay> | Set Path Delay. Sets the manual path delay in s. | Number |
| SENSe:PN:RESidual:DUT:COUNt <Count> | Set Residual DUT Count. Sets the DUT count (1 or 2) for residual measurement. | Number: 1,2 |
| SENSe:PN:RESidual:SUBTract:STATe ON | Subtract 3dB On. Turns the subtract 3 dB (identical DUT) correction on. | None |
| SENSe:PN:RESidual:SUBTract:STATe OFF | Subtract 3dB Off. Turns the subtract 3 dB (identical DUT) correction off. | None |
| SENSe:PN:SOURce:FREQuency <Frequency> | Set Residual Source Frequency. Sets the source 1/2 output frequency in Hz for residual measurement. | Number |
| SENSe:PN:SOURce:POWer:LEVel <Level> | Set Residual Source Power. Sets the source 1/2 output power in dBm for residual measurement. | Number |
| SENSe:PN:SOURce:RESolution <Resolution> | Set Residual Source Resolution. Selects the source power resolution (HIGH 0.1 dB / NORMal 0.5 dB). | Text: HIGH,NORMal |
| SENSe:PN:BASeband:INPut:COUPling <Coupling> | Set Baseband Coupling. Selects the DC block capacitor type for baseband measurement. | Text: ACLF,ACHF |
| SENSe:PN:BASeband:GAIN <Gain> | Set Baseband Gain. Sets the baseband path gain. | Number |
| SENSe:PN:BASeband:DISCharge | Discharge DC Block. Discharges the baseband DC block capacitor. | None |
| SENSe:PN:XGINdicator:STATe ON | XCorr Gain Indicator On. Turns the cross-correlation gain indicator on. | None |
| SENSe:PN:XGINdicator:STATe OFF | XCorr Gain Indicator Off. Turns the cross-correlation gain indicator off. | None |
| SENSe:PN:XCOunter:STATe ON | XCorr Counter On. Turns the cross-correlation counter on. | None |
| SENSe:PN:XCOunter:STATe OFF | XCorr Counter Off. Turns the cross-correlation counter off. | None |
| SENSe:PN:SPURious:SYSTem:OMISsion:STATe ON | Omit System Spurs On. Turns omission of carrier derived system spurs on. | None |
| SENSe:PN:SPURious:SYSTem:OMISsion:STATe OFF | Omit System Spurs Off. Turns omission of carrier derived system spurs off. | None |
| SENSe:SWEep:MODE <Mode> | Set Sweep Mode. Sets the number of triggers channel 1 accepts. | Text: HOLD,CONTinuous,GROups,SINGle |
| SENSe:SWEep:POINts? | Read Sweep Points. Returns the number of data points of the measurement. | None |
| SENSe:AVERage:STATe ON | Averaging On. Turns trace averaging on. | None |
| SENSe:AVERage:STATe OFF | Averaging Off. Turns trace averaging off. | None |
| SENSe:AVERage:COUNt <Count> | Set Average Count. Sets the number of measurements to average (1 to 65536). | Number: 1,65536 |
| SENSe:AVERage:CLEar | Clear Average. Clears and restarts averaging. | None |
| SENSe:ROSCillator:SOURce <Source> | Set Reference Source. Selects the internal or external reference oscillator. | Text: INTernal,EXTernal |
| SENSe:ROSCillator:SOURce:CONDition? | Read Reference Condition. Returns whether the analyzer is locked to the external reference. | None |
| Command | What it does | Arguments |
|---|---|---|
| CALCulate:MEASure<Meas>:DEFine "<Parameter>" | Define Measurement. Creates a phase noise class measurement with the given measurement number (string in quotes e.g. 'PN:Phase Noise'). | Number; Text: PN:Phase Noise,AM:Phase Noise,PN+AM:Phase Noise,L(f):Phase Noise,Sphi(f):Phase Noise,Sv(f):Phase Noise,Sy(f):Phase Noise,AVAR:Phase Noise,ADEV:Phase Noise |
| CALCulate:MEASure<Meas>:DELete | Delete Measurement. Deletes the specified measurement. | Number |
| CALCulate:MEASure:DELete:ALL | Delete All Measurements. Deletes all measurements on the analyzer. | None |
| CALCulate:MEASure<Meas>:PN:CARRier:FREQuency? | Measure Carrier Frequency. Returns the measured carrier frequency in Hz. | Number |
| CALCulate:MEASure<Meas>:PN:CARRier:LEVel? | Measure Carrier Level. Returns the measured carrier power level in dBm. | Number |
| CALCulate:MEASure<Meas>:DATA:X? | Read Trace X. Returns the x-axis (offset frequency) values of the measurement. | Number |
| CALCulate:MEASure<Meas>:DATA:FDATA? | Read Trace Data. Returns the formatted trace data of the measurement (e.g. dBc/Hz). | Number |
| CALCulate:MEASure<Meas>:DATA:FMEM? | Read Memory Data. Returns the formatted memory trace data of the measurement. | Number |
| CALCulate:MEASure<Meas>:PN:DATA:PDATa? | Read PN Data. Returns the dBc data array of the measurement. | Number |
| CALCulate:MEASure<Meas>:PN:DATA:SPData? | Read Spur Flags. Returns the spurious flag (0 or 1) for each data point. | Number |
| CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:TYPE <Type> | Set Integration Type. Sets integration range 1-4 type (off / full span / custom). | Number; Number: 1,2,3,4; Text: OFF,FULL,CUSTom |
| CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:STARt <Frequency> | Set Integration Start. Sets the start offset in Hz of a custom integration range. | Number; Number: 1,2,3,4; Number |
| CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:STOP <Frequency> | Set Integration Stop. Sets the stop offset in Hz of a custom integration range. | Number; Number: 1,2,3,4; Number |
| CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:WEIGhting <Filter> | Set Integration Weighting. Selects a weighting filter file for the integration range (string in quotes). | Number; Number: 1,2,3,4; Text |
| CALCulate:MEASure<Meas>:PN:INTegral:RANGe<Range>:DATA? <Result> | Measure Integrated Result. Returns integrated noise / residual FM / AM / PM / RMS jitter / RMS radian / RMS degree for the range. | Number; Number: 1,2,3,4; Text: IPN,RFM,RAM,RPM,RMSJ,RMSR,RMSD |
| CALCulate:MEASure<Meas>:PN:AVARiance:DEViation? <AvgTime>,<Cutoff> | Measure Allan Deviation. Returns the Allan deviation for the average time in s and cut-off frequency in Hz. | Number; Number; Number |
| CALCulate:MEASure<Meas>:PN:AVARiance:VARiance? <AvgTime>,<Cutoff> | Measure Allan Variance. Returns the Allan variance for the average time in s and cut-off frequency in Hz. | Number; Number; Number |
| CALCulate:MEASure<Meas>:PN:AVARiance:JITTer? <AvgTime>,<Cutoff> | Measure Allan Jitter. Returns the jitter for the average time in s and cut-off frequency in Hz. | Number; Number; Number |
| CALCulate:MEASure<Meas>:PN:SNOise:STATe ON | Spot Noise On. Turns spot noise calculation on. | Number |
| CALCulate:MEASure<Meas>:PN:SNOise:STATe OFF | Spot Noise Off. Turns spot noise calculation off. | Number |
| CALCulate:MEASure<Meas>:PN:SNOise:DECades:STATe ON | Spot Noise Decades On. Turns spot noise calculation at every decade offset on. | Number |
| CALCulate:MEASure<Meas>:PN:SNOise:DECades:STATe OFF | Spot Noise Decades Off. Turns spot noise calculation at every decade offset off. | Number |
| CALCulate:MEASure<Meas>:PN:SNOise:DECades:X? | Read Spot Noise Decades X. Returns the decade offset frequencies in Hz of the spot noise. | Number |
| CALCulate:MEASure<Meas>:PN:SNOise:DECades:Y? | Read Spot Noise Decades Y. Returns the spot noise values at the decade offset frequencies. | Number |
| CALCulate:MEASure<Meas>:PN:SNOise:USER<Position>:STATe ON | Spot Noise User On. Turns the user defined spot noise offset 1-6 on. | Number; Number: 1,2,3,4,5,6 |
| CALCulate:MEASure<Meas>:PN:SNOise:USER<Position>:STATe OFF | Spot Noise User Off. Turns the user defined spot noise offset 1-6 off. | Number; Number: 1,2,3,4,5,6 |
| CALCulate:MEASure<Meas>:PN:SNOise:USER<Position>:X <Frequency> | Set Spot Noise Offset. Sets the user defined spot noise offset frequency in Hz. | Number; Number: 1,2,3,4,5,6; Number |
| CALCulate:MEASure<Meas>:PN:SNOise:USER<Position>:Y? | Measure Spot Noise. Returns the spot noise value at the user defined offset frequency. | Number; Number: 1,2,3,4,5,6 |
| CALCulate:MEASure<Meas>:PN:SPURious:ANALysis:STATe ON | Spur Analysis On. Turns spurious analysis on. | Number |
| CALCulate:MEASure<Meas>:PN:SPURious:ANALysis:STATe OFF | Spur Analysis Off. Turns spurious analysis off. | Number |
| CALCulate:MEASure<Meas>:PN:SPURious:DATA? | Measure Spurs. Returns the list of detected spurs (frequency, power and jitter per spur). | Number |
| CALCulate:MEASure<Meas>:PN:SPURious:OMISsion:STATe ON | Spur Omission On. Turns omission of displayed spurs on. | Number |
| CALCulate:MEASure<Meas>:PN:SPURious:OMISsion:STATe OFF | Spur Omission Off. Turns omission of displayed spurs off. | Number |
| CALCulate:MEASure<Meas>:PN:SPURious:SENSibility <Value> | Set Spur Sensibility. Sets the spurious sensibility (*RST 3 = 3 x standard deviation). | Number; Number |
| CALCulate:MEASure<Meas>:PN:SPURious:SORT <Order> | Set Spur Sort. Selects the spurious table sort order. | Number; Text: POWer,OFFSet |
| CALCulate:MEASure<Meas>:PN:SPURious:THReshold:LEVel:MINimum <Level> | Set Spur Min Level. Sets the minimum spurious threshold level in dBc (-500 to 500). | Number; Number: -500,500 |
| CALCulate:MEASure<Meas>:PN:SPURious:THReshold:TABle:DELete | Clear Spur Threshold Table. Deletes all entries of the spurious threshold table. | Number |
| CALCulate:MEASure<Meas>:PN:SPURious:OSSPur:STATe ON | User Spur Omission On. Turns omission of user specified spurs on. | Number |
| CALCulate:MEASure<Meas>:PN:SPURious:OSSPur:STATe OFF | User Spur Omission Off. Turns omission of user specified spurs off. | Number |
| CALCulate:MEASure<Meas>:PN:SPURious:OSSPur:DELete | Clear User Spur Table. Deletes all entries of the user spur table. | Number |
| CALCulate:MEASure<Meas>:PN:QTRace ON | Quick Trace On. Turns the quick trace (intermediate trace update) function on. | Number |
| CALCulate:MEASure<Meas>:PN:QTRace OFF | Quick Trace Off. Turns the quick trace (intermediate trace update) function off. | Number |
| CALCulate:MEASure<Meas>:FORMat <Format> | Set Display Format. Sets the display format of the measurement. | Number; Text: MLOGarithmic,MLINear,PHASe,UPHase,REAL,IMAGinary,FREQuency,DFRequency |
| CALCulate:MEASure<Meas>:SMOothing:STATe ON | Smoothing On. Turns trace smoothing on. | Number |
| CALCulate:MEASure<Meas>:SMOothing:STATe OFF | Smoothing Off. Turns trace smoothing off. | Number |
| CALCulate:MEASure<Meas>:SMOothing:APERture <Percent> | Set Smoothing Aperture. Sets the smoothing aperture in percent of the data points (1 to 25). | Number; Number: 1,25 |
| CALCulate:MEASure<Meas>:MATH:MEMorize | Data To Memory. Stores the measurement trace into memory. | Number |
| CALCulate:MEASure<Meas>:MARKer<Marker>:STATe ON | Marker On. Turns the marker (1 to 15) on. | Number; Number: 1,15 |
| CALCulate:MEASure<Meas>:MARKer<Marker>:STATe OFF | Marker Off. Turns the marker (1 to 15) off. | Number; Number: 1,15 |
| CALCulate:MEASure<Meas>:MARKer:AOFF | All Markers Off. Turns all markers of the measurement off. | Number |
| CALCulate:MEASure<Meas>:MARKer<Marker>:X <Position> | Set Marker X. Moves the marker to the given x position (offset frequency in Hz). | Number; Number: 1,15; Number |
| CALCulate:MEASure<Meas>:MARKer<Marker>:X? | Read Marker X. Returns the marker x position (offset frequency in Hz). | Number; Number: 1,15 |
| CALCulate:MEASure<Meas>:MARKer<Marker>:Y? | Read Marker Y. Returns the marker y value (value and 0). | Number; Number: 1,15 |
| CALCulate:MEASure<Meas>:MARKer<Marker>:FUNCtion:EXECute <Function> | Marker Search. Executes a marker search function. | Number; Number: 1,15; Text: MAXimum,MINimum,RPEak,LPEak,NPEak,TARGet,LTARget,RTARget,PEAK,SPURious |
| CALCulate:MEASure<Meas>:MARKer<Marker>:DELTa ON | Delta Marker On. Turns delta (relative to reference marker) mode of the marker on. | Number; Number: 1,15 |
| CALCulate:MEASure<Meas>:MARKer<Marker>:DELTa OFF | Delta Marker Off. Turns delta (relative to reference marker) mode of the marker off. | Number; Number: 1,15 |
| CALCulate:MEASure<Meas>:LIMit:STATe ON | Limit Test On. Turns the limit line test on. | Number |
| CALCulate:MEASure<Meas>:LIMit:STATe OFF | Limit Test Off. Turns the limit line test off. | Number |
| CALCulate:MEASure<Meas>:LIMit:FAIL? | Read Limit Fail. Returns the limit test result: 0 = pass and 1 = fail. | Number |
| Command | What it does | Arguments |
|---|---|---|
| DISPlay:WINDow<Window>:STATe ON | Create Window. Creates the display window with the given number. | Number |
| DISPlay:WINDow<Window>:STATe OFF | Delete Window. Deletes the display window with the given number. | Number |
| DISPlay:MEASure<Meas>:FEED <Window> | Feed Measurement To Window. Displays the measurement in the given window. | Number; Number |
| DISPlay:WINDow<Window>:TABLe:SPURious:ENABle ON | Spur Table On. Turns the spurious table display on. | Number |
| DISPlay:WINDow<Window>:TABLe:SPURious:ENABle OFF | Spur Table Off. Turns the spurious table display off. | Number |
| DISPlay:WINDow<Window>:TABLe:INOise:ENABle ON | Integrated Noise Table On. Turns the integrated noise table display on. | Number |
| DISPlay:WINDow<Window>:TABLe:INOise:ENABle OFF | Integrated Noise Table Off. Turns the integrated noise table display off. | Number |
| DISPlay:WINDow<Window>:TABLe:SNOise:ENABle ON | Spot Noise Table On. Turns the spot noise table display on. | Number |
| DISPlay:WINDow<Window>:TABLe:SNOise:ENABle OFF | Spot Noise Table Off. Turns the spot noise table display off. | Number |
| DISPlay:WINDow<Window>:TRACe<Trace>:Y:AUTO | Autoscale Trace. Autoscales the specified trace in the window once. | Number; Number |
| DISPlay:WINDow<Window>:Y:AUTO | Autoscale All. Autoscales all traces in the window once. | Number |
| DISPlay:MEASure<Meas>:Y:PDIVision <Value> | Set Scale Per Division. Sets the y-axis scale per division of the measurement trace. | Number; Number |
| DISPlay:MEASure<Meas>:Y:RLEVel <Level> | Set Reference Level. Sets the y-axis reference level of the measurement trace. | Number; Number |
| DISPlay:MEASure<Meas>:Y:RPOSition <Position> | Set Reference Position. Sets the reference position (graticule from the bottom) of the measurement trace. | Number; Number |
| DISPlay:UPDate:STATe ON | Display Update On. Turns display updates on. | None |
| DISPlay:UPDate:STATe OFF | Display Update Off. Turns display updates off. | None |
| Command | What it does | Arguments |
|---|---|---|
| SOURce:DC:CATalog? | Read DC Sources. Returns the names of the configured and active DC sources. | None |
| SOURce:DC:STATe "<Name>",ON | DC Source On. Turns the specified DC source on (name string in quotes). | Text: VControl,VSupply1,VSupply2 |
| SOURce:DC:STATe <Name>,OFF | DC Source Off. Turns the specified DC source off (name string in quotes). | Text: VControl,VSupply1,VSupply2 |
| SOURce:DC:OUTPut <Name>,<State> | Set DC Output State. Sets the DC channel output state of the specified DC source. | Text: VControl,VSupply1,VSupply2; Text: ON,OFF,DEFault |
| SOURce:DC:STARt <Name>,<Value> | Set DC Start. Sets the start DC value of the source (same start and stop for a fixed voltage). | Text: VControl,VSupply1,VSupply2; Number |
| SOURce:DC:STOP <Name>,<Value> | Set DC Stop. Sets the stop DC value of the source. | Text: VControl,VSupply1,VSupply2; Number |
| SOURce:DC:CENTer <Name>,<Value> | Set DC Center. Sets the center DC value of the source (use span 0 for a fixed voltage). | Text: VControl,VSupply1,VSupply2; Number |
| SOURce:DC:SPAN <Name>,<Value> | Set DC Span. Sets the span DC value of the source. | Text: VControl,VSupply1,VSupply2; Number |
| SOURce:DC:LIMit:MAXimum <Name>,<Value> | Set DC Limit Max. Sets the maximum DC limit of the source. | Text: VControl,VSupply1,VSupply2; Number |
| SOURce:DC:LIMit:MINimum <Name>,<Value> | Set DC Limit Min. Sets the minimum DC limit of the source. | Text: VControl,VSupply1,VSupply2; Number |
| SOURce:DC:LNOise 'VControl',ON | Low Noise VControl On. Turns the low noise V Control on. | None |
| SOURce:DC:LNOise 'VControl',OFF | Low Noise VControl Off. Turns the low noise V Control off. | None |
| Command | What it does | Arguments |
|---|---|---|
| SYSTem:FPReset | Preset Blank. Performs a standard preset and then deletes all traces, measurements and windows (blank screen). | None |
| SYSTem:DC:ENABle ON | DC Outputs On. Turns all configured DC sources for all channels on. | None |
| SYSTem:DC:ENABle OFF | DC Outputs Off. Turns all configured DC sources for all channels off. | None |
| SYSTem:ERRor? | Read Error. Returns the next error in the error queue. | None |
| SYSTem:ERRor:COUNt? | Read Error Count. Returns the number of errors in the error queue. | None |
| SYSTem:PRESet | Preset. Presets the analyzer to factory defaults (same as *RST except data format). | None |
| Command | What it does | Arguments |
|---|---|---|
| MMEMory:STORe:STATe <FileName> | Save State. Saves the instrument state to a .sta file (string in quotes). | Text |
| MMEMory:STORe:CSARchive <FileName> | Save State And Cal. Saves instrument state and calibration data to a .csa file (string in quotes). | Text |
| MMEMory:LOAD:STATe <FileName> | Load State. Loads an instrument state file (string in quotes). | Text |
| MMEMory:STORe:DATA "<FileName>","CSV Formatted Data","<Scope>","Displayed",<Selector> | Save Trace CSV. Saves formatted trace data to a csv file (file and scope strings in quotes; selector = measurement number or -1). | Text; Text: Trace,Channel,Auto; Number |
| MMEMory:LOAD:PN:SPURious:THReshold <FileName> | Load Spur Threshold Table. Loads the spurious threshold table from a file (string in quotes). | Text |
| MMEMory:LOAD:PN:SPURious:OSSPur <FileName> | Load User Spur Table. Loads the user spur omission table from a file (string in quotes). | Text |
| Command | What it does | Arguments |
|---|---|---|
| INITiate:CONTinuous ON | Continuous Measurement. Sets internal (continuous) triggering. | None |
| INITiate:CONTinuous OFF | Manual Trigger Mode. Sets manual triggering (use INITiate:IMMediate to trigger). | None |
| INITiate:IMMediate | Start Measurement. Sends one trigger to start a measurement (trigger source manual); synchronize with *OPC?. | None |
| Command | What it does | Arguments |
|---|---|---|
| FORMat:DATA ASCii,0 | Set Data Format ASCII. Selects ASCII data transfer format. | None |
| FORMat:DATA REAL,64 | Set Data Format Real64. Selects 64-bit binary data transfer format. | None |
| FORMat:BORDer <Order> | Set Byte Order. Selects the byte order of binary data transfers. | Text: NORMal,SWAPped |
| Command | What it does | Arguments |
|---|---|---|
| OUTPut:STATe ON | RF Output On. Turns the RF source power on. | None |
| OUTPut:STATe OFF | RF Output Off. Turns the RF source power off. | None |
| Command | What it does | Arguments |
|---|---|---|
| TRIGger:SOURce <Source> | Set Trigger Source. Selects the sweep trigger source. | Text: IMMediate,EXTernal,MANual |
| Command | What it does | Arguments |
|---|---|---|
| ABORt | Abort Measurement. Stops all sweeps and returns the trigger system to idle. | None |
| Command | What it does | Arguments |
|---|---|---|
| HCOPy:FILE <FileName> | Save Screenshot. Saves the screen image to a .png/.bmp/.jpg file (string in quotes). | Text |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *OPT? | Read Options. Returns the analyzer option configuration. | None |
| *IDN? | Identify. Returns the instrument identification string (manufacturer, model, serial, firmware). | None |
| *RST | Reset. Resets the instrument to its default state. | None |
| *CLS | Clear Status. Clears all event registers and the error queue. | None |
| *TST? | Self Test. Returns the hardware status: 0 = no failures, otherwise the sum of failure weights. | None |
| *OPC? | Operation Complete. Returns 1 when all pending operations are complete. | None |
| *WAI | Wait To Continue. Waits until all pending operations are complete before continuing. | None |
This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the E5055A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.
Yes. The Keysight E5055A accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 204 SCPI commands for it: 35 queries that read values or state, and 169 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, SENSe:PN:ADJust:CONFigure:FREQuency:SEARch:STATe ON and SENSe:CLASs:NAME?.
SENSe:CLASs:NAME? (read class name). Returns the measurement class name of channel 1. The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight E5055A through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.