Vector Network Analyzer · SCPI command reference

Keysight 8753ET SCPI commands

The Keysight 8753ET vector network analyzer is controlled remotely with SCPI commands. TestFlow's instrument library holds 51 commands for it, 9 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 8753ET on a real bench.

Commands
51
Queries
9
Subsystems
46
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight 8753ET and writes the report from the measured results.

Get started

Control the Keysight 8753ET from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keysight 8753ET's VISA address

# Paste the address of your Keysight 8753ET (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("AUTO")  # Autoscale
print(inst.query("STAR?"))  # Read Start Freq

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight 8753ET and writes the report from the measured results.

Get started

Run a 8753ET test without writing code

Describe the test in a sentence, for example “autoscale on the 8753ET, then read start freq and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

51 commands in 46 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

S subsystem (2)

CommandWhat it doesArguments
S11Measure S11. Selects the S11 parameter for measurement.None
S21Measure S21. Selects the S21 parameter for measurement.None

STAR subsystem (2)

CommandWhat it doesArguments
STAR <Frequency>Set Start Freq. Sets the start frequency (append units such as HZ, MHZ or GHZ).Number
STAR?Read Start Freq. Returns the start frequency in Hz.None

Run control (2)

CommandWhat it doesArguments
STOP <Frequency>Set Stop Freq. Sets the stop frequency (append units such as HZ, MHZ or GHZ).Number
STOP?Read Stop Freq. Returns the stop frequency in Hz.None

POWE subsystem (2)

CommandWhat it doesArguments
POWE <Power>Set Power Level. Sets the source output power level in dBm.Number
POWE?Read Power Level. Returns the source output power level in dBm.None

MARK subsystem (2)

CommandWhat it doesArguments
MARK1Marker 1 On. Turns on marker 1 and makes it active.None
MARK1 <Frequency>Set Marker Freq. Places marker 1 at the specified frequency (append units).Number

CENT subsystem (1)

CommandWhat it doesArguments
CENT <Frequency>Set Center Freq. Sets the center frequency (append units).Number

SPAN subsystem (1)

CommandWhat it doesArguments
SPAN <Frequency>Set Span Freq. Sets the frequency span (append units).Number

CWFREQ subsystem (1)

CommandWhat it doesArguments
CWFREQ <Frequency>Set CW Freq. Sets the CW (fixed) frequency (append units).Number

POIN subsystem (1)

CommandWhat it doesArguments
POIN <Points>Set Points. Sets the number of measurement points per sweep.Number: 201,401,801,1601,51,101

IFBW subsystem (1)

CommandWhat it doesArguments
IFBW <Bandwidth>Set IF Bandwidth. Sets the IF bandwidth in Hz.Number: 3000,1000,300,100,30,10

AVEROON subsystem (1)

CommandWhat it doesArguments
AVEROONAverage On. Turns trace averaging on.None

AVEROOFF subsystem (1)

CommandWhat it doesArguments
AVEROOFFAverage Off. Turns trace averaging off.None

AVERFACT subsystem (1)

CommandWhat it doesArguments
AVERFACT <Factor>Set Average Factor. Sets the averaging factor (number of sweeps).Number

AVERREST subsystem (1)

CommandWhat it doesArguments
AVERRESTAverage Restart. Restarts the averaging measurement.None

LOGM subsystem (1)

CommandWhat it doesArguments
LOGMFormat LogMag. Sets the display format to log magnitude.None

PHAS subsystem (1)

CommandWhat it doesArguments
PHASFormat Phase. Sets the display format to phase.None

DELA subsystem (1)

CommandWhat it doesArguments
DELAFormat Delay. Sets the display format to group delay.None

SMIC subsystem (1)

CommandWhat it doesArguments
SMICFormat Smith. Sets the display format to Smith chart.None

POLA subsystem (1)

CommandWhat it doesArguments
POLAFormat Polar. Sets the display format to polar.None

LINM subsystem (1)

CommandWhat it doesArguments
LINMFormat LinMag. Sets the display format to linear magnitude.None

SWR subsystem (1)

CommandWhat it doesArguments
SWRFormat SWR. Sets the display format to SWR.None

REAL subsystem (1)

CommandWhat it doesArguments
REALFormat Real. Sets the display format to real.None

IMAG subsystem (1)

CommandWhat it doesArguments
IMAGFormat Imaginary. Sets the display format to imaginary.None

OUTPFORM subsystem (1)

CommandWhat it doesArguments
OUTPFORMRead Formatted Data. Outputs the formatted trace data array over GPIB.None

OUTPDATA subsystem (1)

CommandWhat it doesArguments
OUTPDATARead Raw Data. Outputs the corrected (raw) complex data array over GPIB.None

MARKOFF subsystem (1)

CommandWhat it doesArguments
MARKOFFMarkers Off. Turns all markers off.None

OUTPMARK subsystem (1)

CommandWhat it doesArguments
OUTPMARKRead Marker. Outputs the active marker value (and stimulus) over GPIB.None

SEAMAX subsystem (1)

CommandWhat it doesArguments
SEAMAXMarker Search Max. Moves the active marker to the trace maximum.None

SEAMIN subsystem (1)

CommandWhat it doesArguments
SEAMINMarker Search Min. Moves the active marker to the trace minimum.None

Run control (1)

CommandWhat it doesArguments
SINGSingle Sweep. Triggers a single sweep and then holds.None

CONT subsystem (1)

CommandWhat it doesArguments
CONTContinuous Sweep. Sets continuous-sweep trigger mode.None

HOLD subsystem (1)

CommandWhat it doesArguments
HOLDHold. Holds the sweep (stops triggering).None

NUMG subsystem (1)

CommandWhat it doesArguments
NUMG <Groups>Number Of Groups. Triggers the specified number of sweeps then holds.Number

AUTO subsystem (1)

CommandWhat it doesArguments
AUTOAutoscale. Autoscales the active channel trace.None

SCAL subsystem (1)

CommandWhat it doesArguments
SCAL <Per_Div>Set Scale Per Div. Sets the vertical scale per division for the active trace.Number

REFV subsystem (1)

CommandWhat it doesArguments
REFV <Value>Set Reference Value. Sets the reference value for the active trace.Number

CORRON subsystem (1)

CommandWhat it doesArguments
CORRONCorrection On. Turns measurement error correction on.None

CORROFF subsystem (1)

CommandWhat it doesArguments
CORROFFCorrection Off. Turns measurement error correction off.None

OUTPERRO subsystem (1)

CommandWhat it doesArguments
OUTPERRORead Error. Outputs the next message from the error queue.None

PRES subsystem (1)

CommandWhat it doesArguments
PRESPreset. Presets the analyzer to its default state.None

OUTPIDEN subsystem (1)

CommandWhat it doesArguments
OUTPIDENIdentify. Outputs the instrument identification string over GPIB.None

OPC subsystem (1)

CommandWhat it doesArguments
OPC?Operation Complete. Returns 1 when all pending operations are complete.None

CLES subsystem (1)

CommandWhat it doesArguments
CLESClear Status. Clears the status registers and error queue.None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the 8753ET and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the Keysight 8753ET support SCPI remote control?

Yes. The Keysight 8753ET accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 51 SCPI commands for it: 9 queries that read values or state, and 42 settings and actions.

How do I control the Keysight 8753ET from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, AUTO and STAR?.

Which command reads a value from the Keysight 8753ET?

STAR? (read start freq). Returns the start frequency in Hz. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Keysight 8753ET without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight 8753ET through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.