Tunable Laser Source · SCPI command reference

Keysight 81950A, N7711A, and N7714A SCPI commands

The Keysight 81950A, N7711A, and N7714A tunable laser source is controlled remotely with SCPI commands. TestFlow's instrument library holds 20 commands for it, 8 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 81950A on a real bench.

Commands
20
Queries
8
Subsystems
8
Models covered
3

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight 81950A, N7711A, or N7714A and writes the report from the measured results.

Get started

Control the Keysight 81950A from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keysight 81950A's VISA address

# Paste the address of your Keysight 81950A (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write(":OUTP ON")  # Enable Output
print(inst.query(":OUTP?"))  # Read Output State

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Keysight 81950A, N7711A, or N7714A and writes the report from the measured results.

Get started

Run a 81950A test without writing code

Describe the test in a sentence, for example “enable output on the 81950A, then read output state and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

20 commands in 8 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The 81950A, N7711A, and N7714A share this command set in TestFlow's library.

Source (SOURce) (6)

CommandWhat it doesArguments
:SOURce:WAVelength <val>Set Wavelength. Set laser wavelength (nm)Number
:SOURce:WAVelength?Read Wavelength. Reads current wavelengthNone
:SOURce:FREQuency <val>Set Frequency. Sets optical frequencyNumber
:SOURce:FREQuency?Read Frequency. Reads optical frequencyNone
:SOURce:POWer <val>Set Power. Sets output powerNumber
:SOURce:POWer?Read Power. Reads current powerNone

Output (OUTPut) (3)

CommandWhat it doesArguments
:OUTP ONEnable Output. Turns laser output ONNone
:OUTP OFFDisable Output. Turns laser output OFFNone
:OUTP?Read Output State. Returns ON/OFF stateNone

System (SYSTem) (3)

CommandWhat it doesArguments
:SYSTem:ERRor?Read Error. Reads next errorNone
:SYSTem:COMM:ETHernet:IPAD <addr>Set IP Address. Sets Ethernet IPText
:SYSTem:COMM:ETHernet:IPAD?Read IP Address. Reads Ethernet IPNone

Other subsystem (2)

CommandWhat it doesArguments
<SCPI>SCPI. Sends a custom SCPI command with optional parameters.Text
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

Status (STATus) (1)

CommandWhat it doesArguments
:STATus:OPERation:CONDition?Read Status. Reads instrument statusNone

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last mearurment to a variableText

IEEE 488.2 common commands (3)

CommandWhat it doesArguments
*IDN?Identification. Reads instrument IDNone
*RSTReset. Resets instrumentNone
*CLSClear Status. Clears status queuesNone

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keysight's programming documentation for the 81950A, N7711A, and N7714A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keysight's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the Keysight 81950A support SCPI remote control?

Yes. The Keysight 81950A accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 20 SCPI commands for it: 8 queries that read values or state, and 12 settings and actions.

How do I control the Keysight 81950A from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, :OUTP ON and :OUTP?.

Which command reads a value from the Keysight 81950A?

:OUTP? (read output state). Returns ON/OFF state The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Keysight 81950A without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keysight 81950A through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.