Multimeter · SCPI command reference

Keithley DAQ6510 SCPI commands

The Keithley DAQ6510 multimeter is controlled remotely with SCPI commands. TestFlow's instrument library holds 96 commands for it, 26 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the DAQ6510 on a real bench.

Commands
96
Queries
26
Subsystems
13
Models covered
1

Control the Keithley DAQ6510 from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Keithley DAQ6510's VISA address

# Paste the address of your Keithley DAQ6510 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write(":SENSe:FUNCtion \"VOLT:DC\"")  # Func DCV
print(inst.query(":MEASure:VOLTage:DC?"))  # Measure DCV

inst.close()

Run a DAQ6510 test without writing code

Describe the test in a sentence, for example “func dcv on the DAQ6510, then measure dcv and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

96 commands in 13 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

Sense (SENSe) (59)

CommandWhat it doesArguments
:SENSe:FUNCtion "VOLT:DC"Func DCV. Selects DC voltage as the active measurement function.None
:SENSe:FUNCtion "VOLT:AC"Func ACV. Selects AC voltage as the active measurement function.None
:SENSe:FUNCtion "CURR:DC"Func DCI. Selects DC current as the active measurement function.None
:SENSe:FUNCtion "CURR:AC"Func ACI. Selects AC current as the active measurement function.None
:SENSe:FUNCtion "RES"Func 2W Resistance. Selects 2-wire resistance as the active measurement function.None
:SENSe:FUNCtion "FRES"Func 4W Resistance. Selects 4-wire resistance as the active measurement function.None
:SENSe:FUNCtion "FREQ:VOLT"Func Frequency. Selects frequency (voltage input) as the active measurement function.None
:SENSe:FUNCtion "PER:VOLT"Func Period. Selects period (voltage input) as the active measurement function.None
:SENSe:FUNCtion "CAP"Func Capacitance. Selects capacitance as the active measurement function.None
:SENSe:FUNCtion "CONT"Func Continuity. Selects continuity as the active measurement function.None
:SENSe:FUNCtion "DIOD"Func Diode. Selects diode test as the active measurement function.None
:SENSe:FUNCtion "TEMP"Func Temperature. Selects temperature as the active measurement function.None
:SENSe:FUNCtion <Function>Set Function. Sets the active measurement function using its SCPI function string.Text: "VOLT:DC","VOLT:AC","CURR:DC","CURR:AC","RES","FRES","FREQ:VOLT","PER:VOLT","CAP","CONT","DIOD","TEMP"
:SENSe:FUNCtion?Read Function. Queries the currently selected measurement function string.None
:SENSe:VOLTage:DC:RANGe <Range>Set DCV Range. Sets the DC voltage measurement range in volts.Number: 0.1,1,10,100,1000
:SENSe:VOLTage:AC:RANGe <Range>Set ACV Range. Sets the AC voltage measurement range in volts.Number: 0.1,1,10,100,750
:SENSe:CURRent:DC:RANGe <Range>Set DCI Range. Sets the DC current measurement range in amps.Number: 0.00001,0.0001,0.001,0.01,0.1,1,3
:SENSe:CURRent:AC:RANGe <Range>Set ACI Range. Sets the AC current measurement range in amps.Number: 0.001,0.01,0.1,1,3
:SENSe:RESistance:RANGe <Range>Set 2W Res Range. Sets the 2-wire resistance measurement range in ohms.Number: 10,100,1000,10000,100000,1000000,10000000,100000000
:SENSe:FRESistance:RANGe <Range>Set 4W Res Range. Sets the 4-wire resistance measurement range in ohms.Number: 1,10,100,1000,10000,100000,1000000,10000000,100000000
:SENSe:CAPacitance:RANGe <Range>Set Capacitance Range. Sets the capacitance measurement range in farads.Number: 0.000000001,0.00000001,0.0000001,0.000001,0.00001
:SENSe:VOLTage:DC:RANGe:AUTO ONDCV Autorange On. Enables autorange for the DC voltage function.None
:SENSe:VOLTage:DC:RANGe:AUTO OFFDCV Autorange Off. Disables autorange for the DC voltage function.None
:SENSe:VOLTage:AC:RANGe:AUTO ONACV Autorange On. Enables autorange for the AC voltage function.None
:SENSe:VOLTage:AC:RANGe:AUTO OFFACV Autorange Off. Disables autorange for the AC voltage function.None
:SENSe:CURRent:DC:RANGe:AUTO ONDCI Autorange On. Enables autorange for the DC current function.None
:SENSe:CURRent:DC:RANGe:AUTO OFFDCI Autorange Off. Disables autorange for the DC current function.None
:SENSe:RESistance:RANGe:AUTO ONRes Autorange On. Enables autorange for the 2-wire resistance function.None
:SENSe:RESistance:RANGe:AUTO OFFRes Autorange Off. Disables autorange for the 2-wire resistance function.None
:SENSe:VOLTage:DC:NPLCycles <NPLC>Set DCV NPLC. Sets the integration time for DC voltage in number of power line cycles.Number: 0.0005,0.1,1,10,15
:SENSe:CURRent:DC:NPLCycles <NPLC>Set DCI NPLC. Sets the integration time for DC current in number of power line cycles.Number: 0.0005,0.1,1,10,15
:SENSe:RESistance:NPLCycles <NPLC>Set 2W Res NPLC. Sets the integration time for 2-wire resistance in number of power line cycles.Number: 0.0005,0.1,1,10,15
:SENSe:FRESistance:NPLCycles <NPLC>Set 4W Res NPLC. Sets the integration time for 4-wire resistance in number of power line cycles.Number: 0.0005,0.1,1,10,15
:SENSe:VOLTage:DC:INPutimpedance AUTODCV Input Impedance Auto On. Sets DC voltage input impedance to AUTO (>10 GOhm on 100 mV/1 V/10 V ranges).None
:SENSe:VOLTage:DC:INPutimpedance MOHM10DCV Input Impedance 10M. Sets the DC voltage input impedance to a fixed 10 MOhm on all ranges.None
:SENSe:VOLTage:DC:AZERo:STATe ONSet DCV AZero On. Enables autozero for the DC voltage function.None
:SENSe:VOLTage:DC:AZERo:STATe OFFSet DCV AZero Off. Disables autozero for the DC voltage function.None
:SENSe:VOLTage:AC:DETector:BANDwidth <Bandwidth>Set ACV Detector BW. Sets the AC voltage detector bandwidth in Hz (3, 30 or 300).Number: 3,30,300
:SENSe:CURRent:AC:DETector:BANDwidth <Bandwidth>Set ACI Detector BW. Sets the AC current detector bandwidth in Hz (3, 30 or 300).Number: 3,30,300
:SENSe:VOLTage:DC:AVERage:COUNt <Count>Set DCV Average Count. Sets the number of readings in the DC voltage filter average (1 to 100).Number
:SENSe:VOLTage:DC:AVERage:TCONtrol <Type>Set DCV Average TCon. Sets the DC voltage averaging filter type to repeating or moving.Text: REPeat,MOVing
:SENSe:VOLTage:DC:AVERage:STATe ONDCV Average On. Enables the averaging filter for the DC voltage function.None
:SENSe:VOLTage:DC:AVERage:STATe OFFDCV Average Off. Disables the averaging filter for the DC voltage function.None
:SENSe:VOLTage:DC:RELative:ACQuireAcquire DCV Rel. Acquires the present DC voltage reading and stores it as the relative (null) offset.None
:SENSe:VOLTage:DC:RELative <Offset>Set DCV Rel Offset. Sets the relative (null) offset value for the DC voltage function.Number
:SENSe:VOLTage:DC:RELative:STATe ONDCV Rel On. Enables the relative (null) offset for the DC voltage function.None
:SENSe:VOLTage:DC:RELative:STATe OFFDCV Rel Off. Disables the relative (null) offset for the DC voltage function.None
:SENSe:CURRent:DC:RELative:ACQuireAcquire DCI Rel. Acquires the present DC current reading and stores it as the relative (null) offset.None
:SENSe:CURRent:DC:RELative:STATe ONDCI Rel On. Enables the relative (null) offset for the DC current function.None
:SENSe:CURRent:DC:RELative:STATe OFFDCI Rel Off. Disables the relative (null) offset for the DC current function.None
:SENSe:VOLTage:AC:UNIT DBSet ACV Unit dB. Sets the AC voltage measurement unit to dB.None
:SENSe:VOLTage:AC:UNIT DBMSet ACV Unit dBm. Sets the AC voltage measurement unit to dBm.None
:SENSe:VOLTage:AC:UNIT VOLTSet ACV Unit Volt. Sets the AC voltage measurement unit back to volts.None
:SENSe:VOLTage:AC:DB:REFerence <Reference>Set ACV dB Reference. Sets the dB reference voltage for the AC voltage function (1e-7 to 1000 V).Number
:SENSe:VOLTage:AC:DBM:REFerence <Reference>Set ACV dBm Reference. Sets the dBm reference impedance for the AC voltage function (1 to 9999 ohms).Number
:SENSe:TEMPerature:TRANsducer <Type>Set Temp Transducer. Sets the temperature transducer type for the temperature function.Text: TCouple,RTD,FRTD,THERmistor
:SENSe:TEMPerature:UNIT <Unit>Set Temp Unit. Sets the units used for temperature measurements.Text: CELSius,FAHRenheit,KELVin
:SENSe:COUNt <Count>Set Sample Count. Sets the number of measurements made when a measurement is requested (1 to 1000000).Number
:SENSe:COUNt?Read Sample Count. Queries the number of measurements made per measurement request.None

Measurement (MEASure) (13)

CommandWhat it doesArguments
:MEASure?Measure. Makes a measurement using the active function and returns the reading.None
:MEASure:VOLTage:DC?Measure DCV. Sets the function to DC voltage and returns one measured reading in volts.None
:MEASure:VOLTage:AC?Measure ACV. Sets the function to AC voltage and returns one measured reading in volts.None
:MEASure:CURRent:DC?Measure DCI. Sets the function to DC current and returns one measured reading in amps.None
:MEASure:CURRent:AC?Measure ACI. Sets the function to AC current and returns one measured reading in amps.None
:MEASure:RESistance?Measure 2W Resistance. Sets the function to 2-wire resistance and returns one reading in ohms.None
:MEASure:FRESistance?Measure 4W Resistance. Sets the function to 4-wire resistance and returns one reading in ohms.None
:MEASure:FREQuency?Measure Frequency. Sets the function to frequency and returns one reading in Hz.None
:MEASure:PERiod?Measure Period. Sets the function to period and returns one reading in seconds.None
:MEASure:CAPacitance?Measure Capacitance. Sets the function to capacitance and returns one reading in farads.None
:MEASure:CONTinuity?Measure Continuity. Sets the function to continuity and returns the measured resistance in ohms.None
:MEASure:DIODe?Measure Diode. Sets the function to diode test and returns the forward voltage in volts.None
:MEASure:TEMPerature?Measure Temperature. Sets the function to temperature and returns one reading in the configured units.None

Trace (TRACe) (7)

CommandWhat it doesArguments
:TRACe:CLEarClear Buffer. Clears all readings and statistics from the default reading buffer.None
:TRACe:ACTual?Read Buffer Actual. Returns the number of readings currently stored in the default reading buffer.None
:TRACe:STATistics:AVERage?Read Buffer Stat Mean. Returns the mean (average) of the readings stored in the default reading buffer.None
:TRACe:STATistics:MINimum?Read Buffer Stat Min. Returns the minimum reading stored in the default reading buffer.None
:TRACe:STATistics:MAXimum?Read Buffer Stat Max. Returns the maximum reading stored in the default reading buffer.None
:TRACe:STATistics:STDDev?Read Buffer Stat StdDev. Returns the standard deviation of the readings stored in the default reading buffer.None
:TRACe:STATistics:CLEarClear Buffer Statistics. Clears the statistical information from the default reading buffer.None

Trigger (TRIGger) (3)

CommandWhat it doesArguments
:TRIGger:BLOCk:MDIGitize <BlockNumber>Set Trigger Source. Defines a trigger-model block that makes measurements at the specified block number.Number
:TRIGger:LOAD "SimpleLoop", <Count>Trigger Load SimpleLoop. Loads the SimpleLoop trigger model template that makes the specified number of readings.Number
:TRIGger:EXTernal:IN:EDGE <Edge>Set Trigger Ext In Edge. Sets the edge used to detect external input triggers on the rear-panel line.Text: FALLing,RISing,EITHer

Read (READ) (1)

CommandWhat it doesArguments
:READ?Read. Triggers and returns one reading from the active function using the default reading buffer.None

Fetch (FETCh) (1)

CommandWhat it doesArguments
:FETCh?Fetch. Returns the latest reading from the buffer without triggering a new measurement.None

Initiate (INITiate) (1)

CommandWhat it doesArguments
:INITiateInit Trigger Model. Starts the trigger model and begins the configured measurement sequence.None

Abort (1)

CommandWhat it doesArguments
:ABORtAbort Trigger Model. Stops all trigger-model commands and returns the instrument to idle.None

System (SYSTem) (1)

CommandWhat it doesArguments
:SYSTem:ERRor:NEXT?Read Error. Reads and removes the oldest message from the error queue (0,"No error" if empty).None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

SAVE subsystem (1)

CommandWhat it doesArguments
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

IEEE 488.2 common commands (6)

CommandWhat it doesArguments
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).None
*RSTReset. Resets the instrument to its default state.None
*CLSClear Status. Clears all event registers and the error queue.None
*TST?Self Test. Runs the self-test and returns 0 for pass or 1 for failure.None
*OPC?Operation Complete. Returns 1 when all pending operations are complete.None
*WAIWait To Continue. Waits until all pending operations are complete before continuing.None

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Keithley's programming documentation for the DAQ6510 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Keithley's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.

Frequently asked questions

Does the Keithley DAQ6510 support SCPI remote control?

Yes. The Keithley DAQ6510 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 96 SCPI commands for it: 26 queries that read values or state, and 70 settings and actions.

How do I control the Keithley DAQ6510 from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, :SENSe:FUNCtion "VOLT:DC" and :MEASure:VOLTage:DC?.

Which command reads a value from the Keithley DAQ6510?

:MEASure:VOLTage:DC? (measure dcv). Sets the function to DC voltage and returns one measured reading in volts. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Keithley DAQ6510 without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Keithley DAQ6510 through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.