In TestFlow's instrument library, the Chroma 63102A takes 31 SCPI commands, 2 of them queries, and the Siglent SDL1020X-E takes 99 SCPI commands, 42 of them queries. That is 68 more on the SDL1020X-E. Counting distinct command headers (the command path without its arguments), 0 are accepted by both, 25 exist only on the 63102A, and 99 only on the SDL1020X-E. None of the 63102A's command headers appears in the SDL1020X-E library, so a script written for one needs every command rewritten for the other.
Both libraries have commands under MEAS. Only the 63102A has CURR, VOLT, RES, CONFigure, LOAD, and MODE and 1 more. Only the SDL1020X-E has SOURce, TRIGger, SENSe, and SYSTem.
Only the SDL1020X-E library has an operation-complete command (*OPC?); the 63102A library has none. Neither library has an error-queue read such as SYSTem:ERRor?.
TestFlow's library gives the 63103A, 63105A, 63106A, 63107A, 63108A, 63110A, 63112A, 63113A, 63123A, and 6314A the same command set as the 63102A, so the 63102A column applies to them too. TestFlow's library gives the SDL1030X-E the same command set as the SDL1020X-E, so the SDL1020X-E column applies to it too.
Describe the test; TestFlow's agent uses these command sets to drive your Chroma 63102A or Siglent SDL1020X-E and writes the report from the measured results.
Counts come from TestFlow's instrument library, the command sets its AI agent uses to drive each instrument. A distinct command header is the command path without its arguments, so :OUTPut ON and :OUTPut OFF count once.
| Chroma 63102A | Siglent SDL1020X-E | |
|---|---|---|
| Commands in TestFlow's library | 31 | 99 |
| Queries (read a value or state) | 2 | 42 |
| Settings and actions | 29 | 57 |
| Subsystems | 11 | 6 |
| Distinct command headers | 25 | 99 |
| Headers the other lacks | 25 | 99 |
| Command language | SCPI | SCPI |
| Operation complete | Not in library | *OPC? |
| Error queue | Not in library | Not in library |
| Models on this command set | 63102A, 63103A, 63105A, 63106A, 63107A, 63108A, 63110A, 63112A, 63113A, 63123A, 6314A | SDL1020X-E, SDL1030X-E |
None. No command header in TestFlow's 63102A library matches one in its SDL1020X-E library, so every command needs its own version for each instrument.
25 command headers in the 63102A library have no match in the SDL1020X-E library. Below are 17 of them from the 6 largest subsystems; the full list is in the Chroma 63102A command reference.
| Command | What it does |
|---|---|
| CURR:STAT:L1 <Amp> | Static Current low. Sets the static current lower value |
| CURR:STAT:L2 <Amp> | Static Current High. Sets the static current higher value (Always set low and high) |
| CURR:STAT:RISE <Slew> | Current rise slew. Sets the current rising slew rate |
| CURR:STAT:FALL <Slew> | Current fall slew. Sets the current falling slew rate |
| Command | What it does |
|---|---|
| RES:L1 <OHM> OHM | Resistance low. Sets the lower value of dynamic resistive load |
| RES:L2 <OHM> OHM | Resistance high. Sets the higher value of dynamic resistive load (Set low and high for constant) |
| RES:RISE <Slew> | Resistance rise. Sets resistive load rise |
| RES:FALL <Slew> | Resistance fall. Sets resistive load fall |
| Command | What it does |
|---|---|
| VOLT:L1 <Volt>V | Static voltage low. Sets the lower value of CV mode |
| VOLT:L2 <Volt>V | Static voltave high. Sets the higher value of the CV mode |
| VOLTage:CURRent <Amp> | Static voltage current limit. Sets the current limit |
| VOLT:MODE FAST | Voltage fast. Sets voltage fast mode |
| Command | What it does |
|---|---|
| :CONFigure:REMote ON | Remote control. Sets control to remote |
| CONF:VOLT:LATC ON | Voltage latch ON. Enables voltage latch |
| Command | What it does |
|---|---|
| MEAS:VOLT? | Measure voltage. Measures the voltage |
| MEAS:CURR? | Measure current. Measures the current |
| Command | What it does |
|---|---|
| CHANnel <Channel> | Channel select. Selects the channel |
99 command headers in the SDL1020X-E library have no match in the 63102A library. Below are 17 of them from the 6 largest subsystems; the full list is in the Siglent SDL1020X-E command reference.
| Command | What it does |
|---|---|
| :SOURce:INPut:STATe <state> | Set Input State. Turns the load input on or off. |
| :SOURce:INPut:STATe? | Query Input State. Returns 1 if the load input is on, otherwise 0. |
| :SOURce:SHORt:STATe <state> | Set Short Circuit State. Sets the short circuit status of the load. |
| :SOURce:SHORt:STATe? | Query Short Circuit State. Returns 1 if the short circuit is on, otherwise 0. |
| Command | What it does |
|---|---|
| *IDN? | Identify. Returns the manufacturer, model, serial number and software version string. |
| *RST | Reset. Restores the instrument to its initial state. |
| *CLS | Clear Status. Clears the event registers and the error list. |
| *ESE <number> | Set Standard Event Enable. Sets the bits in the standard event status enable register. |
| Command | What it does |
|---|---|
| MEASure:VOLTage:DC? | Measure Voltage. Returns the real-time voltage measurement. |
| MEASure:CURRent:DC? | Measure Current. Returns the real-time current measurement. |
| MEASure:POWer:DC? | Measure Power. Returns the real-time power measurement. |
| MEASure:RESistance:DC? | Measure Resistance. Returns the real-time resistance measurement. |
| Command | What it does |
|---|---|
| SENSe:AVERage:COUNt <index> | Set Average Count. Sets the averaging point count for read-back current and voltage. |
| SENSe:AVERage:COUNt? | Query Average Count. Returns the averaging point count index. |
| Command | What it does |
|---|---|
| TRIGger:SOURce <source> | Set Trigger Source. Sets the trigger source of the load. |
| TRIGger:SOURce? | Query Trigger Source. Returns the trigger source. |
| Command | What it does |
|---|---|
| SYSTem:SENSe:STATe <state> | Set Sense State. Enables or disables the Sense function. |
With PyVISA and any VISA backend, this opens both instruments, identifies them, sends one setting, and reads one value from each. For the setting, the two libraries have no common command with a worked example on both, so each instrument uses its own: :CONFigure:REMote ON on the 63102A and :SOURce:LIST:STATe:ON on the SDL1020X-E. For the reading, the two libraries have no common command with a worked example on both, so each instrument uses its own: MEAS:VOLT? on the 63102A and MEASure:VOLTage:DC? on the SDL1020X-E. For the basics, see the PyVISA tutorial.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find both VISA addresses
a = rm.open_resource("PASTE_63102A_ADDRESS") # Chroma 63102A
b = rm.open_resource("PASTE_SDL1020X_E_ADDRESS") # Siglent SDL1020X-E
for inst in (a, b):
inst.timeout = 5000 # ms
print(inst.query("*IDN?"))
# No shared setting with an example on both, so each uses its own command
a.write(":CONFigure:REMote ON") # Remote control
b.write(":SOURce:LIST:STATe:ON") # Enter List Test
# No shared reading with an example on both, so each uses its own command
print(a.query("MEAS:VOLT?")) # Measure voltage
print(b.query("MEASure:VOLTage:DC?")) # Measure Voltage
a.close()
b.close()Describe the test; TestFlow's agent uses these command sets to drive your Chroma 63102A or Siglent SDL1020X-E and writes the report from the measured results.
In TestFlow's library the 63102A has 31 SCPI commands and the SDL1020X-E has 99 SCPI commands. 0 command headers are common to both, 25 are only on the 63102A (mostly CURR and RES), and 99 are only on the SDL1020X-E (mostly SOURce and IEEE 488.2 common (*)).
Not without changes. None of the 25 command headers in the 63102A library matches one of the 99 in the SDL1020X-E library: the 63102A uses commands such as CURR:STAT:L1 and the SDL1020X-E uses commands such as :SOURce:INPut:STATe. PyVISA opens both the same way, but each command needs its own version for each instrument.
The SDL1020X-E: 31 commands for the 63102A against 99 for the SDL1020X-E in TestFlow's library. Of those, 2 on the 63102A and 42 on the SDL1020X-E are queries that read a value or state, and the library splits them into 11 and 6 subsystems.
Only the SDL1020X-E library has an operation-complete command (*OPC?); the 63102A library has none. Neither library has an error-queue read such as SYSTem:ERRor?. Each command named here is listed in the tables on this page and on each instrument's command reference.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.