Dynamic Load · SCPI command comparison

Chroma 63102A vs Kikusui PLZ164WA: SCPI commands and remote control compared

In TestFlow's instrument library, the Chroma 63102A takes 31 SCPI commands, 2 of them queries, and the Kikusui PLZ164WA takes 35 SCPI commands, 13 of them queries. That is 4 more on the PLZ164WA. Counting distinct command headers (the command path without its arguments), 2 are accepted by both, 23 exist only on the 63102A, and 23 only on the PLZ164WA. 8% of the 63102A's command headers also exist on the PLZ164WA.

Both libraries have commands under MEAS. Only the 63102A has CURR, VOLT, RES, CONFigure, LOAD, and MODE and 1 more. Only the PLZ164WA has SOUR, INP, and SYST.

Neither library has an operation-complete command such as *OPC?. Only the PLZ164WA library has an error-queue read (SYSTem:ERRor?); the 63102A library has none.

TestFlow's library gives the 63103A, 63105A, 63106A, 63107A, 63108A, 63110A, 63112A, 63113A, 63123A, and 6314A the same command set as the 63102A, so the 63102A column applies to them too. TestFlow's library gives the PLZ334WA and PLZ664WA the same command set as the PLZ164WA, so the PLZ164WA column applies to them too.

63102A commands
31
PLZ164WA commands
35
Shared headers
2
Unique headers
46

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Chroma 63102A or Kikusui PLZ164WA and writes the report from the measured results.

Get started

63102A vs PLZ164WA at a glance

Counts come from TestFlow's instrument library, the command sets its AI agent uses to drive each instrument. A distinct command header is the command path without its arguments, so :OUTPut ON and :OUTPut OFF count once.

Chroma 63102AKikusui PLZ164WA
Commands in TestFlow's library3135
Queries (read a value or state)213
Settings and actions2922
Subsystems118
Distinct command headers2525
Headers the other lacks2323
Command languageSCPISCPI
Operation completeNot in libraryNot in library
Error queueNot in librarySYSTem:ERRor?
Models on this command set63102A, 63103A, 63105A, 63106A, 63107A, 63108A, 63110A, 63112A, 63113A, 63123A, 6314APLZ164WA, PLZ334WA, PLZ664WA

Commands both accept

2 command headers are in both libraries.

CommandWhat it does
MEAS:VOLT?Measure voltage. Measures the voltage
MEAS:CURR?Measure current. Measures the current

Only on the Chroma 63102A

23 command headers in the 63102A library have no match in the PLZ164WA library. Below are 16 of them from the 6 largest subsystems; the full list is in the Chroma 63102A command reference.

CURR (10 only on the 63102A)

CommandWhat it does
CURR:STAT:L1 <Amp>Static Current low. Sets the static current lower value
CURR:STAT:L2 <Amp>Static Current High. Sets the static current higher value (Always set low and high)
CURR:STAT:RISE <Slew>Current rise slew. Sets the current rising slew rate
CURR:STAT:FALL <Slew>Current fall slew. Sets the current falling slew rate

RES (4 only on the 63102A)

CommandWhat it does
RES:L1 <OHM> OHMResistance low. Sets the lower value of dynamic resistive load
RES:L2 <OHM> OHMResistance high. Sets the higher value of dynamic resistive load (Set low and high for constant)
RES:RISE <Slew>Resistance rise. Sets resistive load rise
RES:FALL <Slew>Resistance fall. Sets resistive load fall

VOLT (4 only on the 63102A)

CommandWhat it does
VOLT:L1 <Volt>VStatic voltage low. Sets the lower value of CV mode
VOLT:L2 <Volt>VStatic voltave high. Sets the higher value of the CV mode
VOLTage:CURRent <Amp>Static voltage current limit. Sets the current limit
VOLT:MODE FASTVoltage fast. Sets voltage fast mode

CONFigure (2 only on the 63102A)

CommandWhat it does
:CONFigure:REMote ONRemote control. Sets control to remote
CONF:VOLT:LATC ONVoltage latch ON. Enables voltage latch

CHANnel (1 only on the 63102A)

CommandWhat it does
CHANnel <Channel>Channel select. Selects the channel

LOAD (1 only on the 63102A)

CommandWhat it does
LOAD 1Turn On. Turns the load on

Only on the Kikusui PLZ164WA

23 command headers in the PLZ164WA library have no match in the 63102A library. Below are 11 of them from the 5 largest subsystems; the full list is in the Kikusui PLZ164WA command reference.

SOUR (16 only on the PLZ164WA)

CommandWhat it does
SOUR:FUNC:MODE <Mode>Mode. Sets the load mode constant(voltage, current, resistance or power)
SOUR:FUNC:MODE?Read Mode. Reads the mode
SOUR:CURR <Amp>Set Current. Sets the current
SOUR:CURR?Read Set Current. Reads set current

INP (3 only on the PLZ164WA)

CommandWhat it does
INP:STAT ONTurn On. Turns the load on
INP:STAT?Read State. Read the load state (On or off)
INP:PROT:CLEClear Protection. Clears all protection

IEEE 488.2 common (*) (2 only on the PLZ164WA)

CommandWhat it does
*IDN?Identify. Identify the instrument
*RSTReset. Reset the instrument

MEAS (1 only on the PLZ164WA)

CommandWhat it does
MEAS:POW?Measure power. Measure the power

SYST (1 only on the PLZ164WA)

CommandWhat it does
SYST:ERR?Read Error. Reads errors

Control both from Python

With PyVISA and any VISA backend, this opens both instruments, identifies them, sends one setting, and reads one value from each. For the setting, the two libraries have no common command with a worked example on both, so each instrument uses its own: :CONFigure:REMote ON on the 63102A and INP:STAT ON on the PLZ164WA. For the reading, both libraries hold MEAS:VOLT? (measure voltage), and the same line runs on each. For the basics, see the PyVISA tutorial.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find both VISA addresses

a = rm.open_resource("PASTE_63102A_ADDRESS")  # Chroma 63102A
b = rm.open_resource("PASTE_PLZ164WA_ADDRESS")  # Kikusui PLZ164WA
for inst in (a, b):
    inst.timeout = 5000  # ms
    print(inst.query("*IDN?"))

# No shared setting with an example on both, so each uses its own command
a.write(":CONFigure:REMote ON")  # Remote control
b.write("INP:STAT ON")  # Turn On

# The same reading on both: Measure voltage
print(a.query("MEAS:VOLT?"))
print(b.query("MEAS:VOLT?"))

a.close()
b.close()

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Chroma 63102A or Kikusui PLZ164WA and writes the report from the measured results.

Get started

Frequently asked questions

What is the difference between the Chroma 63102A and the Kikusui PLZ164WA for remote control?

In TestFlow's library the 63102A has 31 SCPI commands and the PLZ164WA has 35 SCPI commands. 2 command headers are common to both, 23 are only on the 63102A (mostly CURR and RES), and 23 are only on the PLZ164WA (mostly SOUR and INP).

Can I use the same SCPI code for the Chroma 63102A and the Kikusui PLZ164WA?

Partly. 2 command headers are accepted by both, including MEAS:VOLT? and MEAS:CURR?. 46 are not, for example CURR:STAT:L1 on the 63102A and SOUR:FUNC:MODE on the PLZ164WA, so a script that uses those needs a separate branch for each instrument.

Which has more remote commands, the 63102A or the PLZ164WA?

The PLZ164WA: 31 commands for the 63102A against 35 for the PLZ164WA in TestFlow's library. Of those, 2 on the 63102A and 13 on the PLZ164WA are queries that read a value or state, and the library splits them into 11 and 8 subsystems.

How do the 63102A and PLZ164WA signal operation complete and report errors?

Neither library has an operation-complete command such as *OPC?. Only the PLZ164WA library has an error-queue read (SYSTem:ERRor?); the 63102A library has none. Each command named here is listed in the tables on this page and on each instrument's command reference.