Power Supply · SCPI command comparison

Aim-TTi CPX400DP vs Siglent SPD3303X-E: SCPI commands and remote control compared

In TestFlow's instrument library, the Aim-TTi CPX400DP takes 33 SCPI commands, 16 of them queries, and the Siglent SPD3303X-E takes 20 SCPI commands, 9 of them queries. That is 13 more on the CPX400DP. Counting distinct command headers (the command path without its arguments), 1 are accepted by both, 27 exist only on the CPX400DP, and 15 only on the SPD3303X-E. 4% of the CPX400DP's command headers also exist on the SPD3303X-E.

Only the CPX400DP has OP1, OP2, V1, V2, V1O, and V2O and 10 more. Only the SPD3303X-E has CH, OUTPut, SYSTem, INSTrument, and MEASure.

Only the CPX400DP library has an operation-complete command (*OPC?); the SPD3303X-E library has none. Only the SPD3303X-E library has an error-queue read (SYSTem:ERRor?); the CPX400DP library has none.

TestFlow's library gives the MX100QP, MX100TP, MX103QP, MX180TP, PL303QMDP, PL303QMTP, QL355TP, QL564TP, and QPX600DP the same command set as the CPX400DP, so the CPX400DP column applies to them too.

CPX400DP commands
33
SPD3303X-E commands
20
Shared headers
1
Unique headers
42

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Aim-TTi CPX400DP or Siglent SPD3303X-E and writes the report from the measured results.

Get started

CPX400DP vs SPD3303X-E at a glance

Counts come from TestFlow's instrument library, the command sets its AI agent uses to drive each instrument. A distinct command header is the command path without its arguments, so :OUTPut ON and :OUTPut OFF count once.

Aim-TTi CPX400DPSiglent SPD3303X-E
Commands in TestFlow's library3320
Queries (read a value or state)169
Settings and actions1711
Subsystems119
Distinct command headers2816
Headers the other lacks2715
Command languageSCPISCPI
Operation complete*OPC?Not in library
Error queueNot in librarySYSTem:ERRor?
Models on this command setCPX400DP, MX100QP, MX100TP, MX103QP, MX180TP, PL303QMDP, PL303QMTP, QL355TP, QL564TP, QPX600DPSPD3303X-E

Commands both accept

1 command header is in both libraries.

CommandWhat it does
*IDN?Identify. Returns instrument identification string (manufacturer model serial firmware)

Only on the Aim-TTi CPX400DP

27 command headers in the CPX400DP library have no match in the SPD3303X-E library. Below are 12 of them from the 6 largest subsystems; the full list is in the Aim-TTi CPX400DP command reference.

I1 (2 only on the CPX400DP)

CommandWhat it does
I1 <Amp>Set Current Ch1. Sets the current limit on channel 1 in Amps (0 to 20A, resolution 1mA)
I1?Read Set Current Ch1. Returns the set current limit of channel 1 in Amps

I2 (2 only on the CPX400DP)

CommandWhat it does
I2 <Amp>Set Current Ch2. Sets the current limit on channel 2 in Amps (0 to 20A, resolution 1mA)
I2?Read Set Current Ch2. Returns the set current limit of channel 2 in Amps

OCP1 (2 only on the CPX400DP)

CommandWhat it does
OCP1 <Amp>Set OCP Ch1. Sets the over-current protection trip level on channel 1 in Amps (resolution 10mA)
OCP1?Read OCP Ch1. Returns the set OCP level of channel 1 in Amps

OCP2 (2 only on the CPX400DP)

CommandWhat it does
OCP2 <Amp>Set OCP Ch2. Sets the over-current protection trip level on channel 2 in Amps (resolution 10mA)
OCP2?Read OCP Ch2. Returns the set OCP level of channel 2 in Amps

OP1 (2 only on the CPX400DP)

CommandWhat it does
OP1 1Turn On Ch1. Turns on output channel 1
OP1?Read State Ch1. Returns the output state of channel 1 (1=ON 0=OFF)

OP2 (2 only on the CPX400DP)

CommandWhat it does
OP2 1Turn On Ch2. Turns on output channel 2
OP2?Read State Ch2. Returns the output state of channel 2 (1=ON 0=OFF)

Only on the Siglent SPD3303X-E

15 command headers in the SPD3303X-E library have no match in the CPX400DP library. Below are 15 of them from the 6 largest subsystems; the full list is in the Siglent SPD3303X-E command reference.

CH (4 only on the SPD3303X-E)

CommandWhat it does
CH<CH>:VOLTage <Volt>Set Voltage.
CH<CH>:VOLTage?Read Voltage.
CH<CH>:CURRent <Amp>Set Current.
CH<CH>:CURRent?Read Current.

SYSTem (3 only on the SPD3303X-E)

CommandWhat it does
SYSTem:ERRor?Read Error.
SYSTem:VERSion?Read Version.
SYSTem:STATus?Read Status.

IEEE 488.2 common (*) (2 only on the SPD3303X-E)

CommandWhat it does
*LOCKLock Keys.
*UNLOCKUnlock Keys.

INSTrument (2 only on the SPD3303X-E)

CommandWhat it does
INSTrument CH<CH>Select Channel.
INSTrument?Read Selected Channel.

MEASure (2 only on the SPD3303X-E)

CommandWhat it does
MEASure:VOLTage? CH<CH>Measure Voltage.
MEASure:CURRent? CH<CH>Measure Current.

OUTPut (2 only on the SPD3303X-E)

CommandWhat it does
OUTPut CH<CH>,ONOutput On.
OUTPut:TRACK <Mode>Set Track Mode.

Control both from Python

With PyVISA and any VISA backend, this opens both instruments, identifies them, sends one setting, and reads one value from each. For the setting, the two libraries have no common command with a worked example on both, so each instrument uses its own: OP1 1 on the CPX400DP and OUTPut CH1,ON on the SPD3303X-E. For the reading, the two libraries have no common command with a worked example on both, so each instrument uses its own: OP1? on the CPX400DP and MEASure:VOLTage? CH1 on the SPD3303X-E. For the basics, see the PyVISA tutorial.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find both VISA addresses

a = rm.open_resource("PASTE_CPX400DP_ADDRESS")  # Aim-TTi CPX400DP
b = rm.open_resource("PASTE_SPD3303X_E_ADDRESS")  # Siglent SPD3303X-E
for inst in (a, b):
    inst.timeout = 5000  # ms
    print(inst.query("*IDN?"))

# No shared setting with an example on both, so each uses its own command
a.write("OP1 1")  # Turn On Ch1
b.write("OUTPut CH1,ON")  # Output On

# No shared reading with an example on both, so each uses its own command
print(a.query("OP1?"))  # Read State Ch1
print(b.query("MEASure:VOLTage? CH1"))  # Measure Voltage

a.close()
b.close()

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Aim-TTi CPX400DP or Siglent SPD3303X-E and writes the report from the measured results.

Get started

Frequently asked questions

What is the difference between the Aim-TTi CPX400DP and the Siglent SPD3303X-E for remote control?

In TestFlow's library the CPX400DP has 33 SCPI commands and the SPD3303X-E has 20 SCPI commands. 1 command headers are common to both, 27 are only on the CPX400DP (mostly I1 and I2), and 15 are only on the SPD3303X-E (mostly CH and SYSTem).

Can I use the same SCPI code for the Aim-TTi CPX400DP and the Siglent SPD3303X-E?

Partly. 1 command headers are accepted by both. 42 are not, for example I1 on the CPX400DP and CH<CH>:VOLTage on the SPD3303X-E, so a script that uses those needs a separate branch for each instrument.

Which has more remote commands, the CPX400DP or the SPD3303X-E?

The CPX400DP: 33 commands for the CPX400DP against 20 for the SPD3303X-E in TestFlow's library. Of those, 16 on the CPX400DP and 9 on the SPD3303X-E are queries that read a value or state, and the library splits them into 11 and 9 subsystems.

How do the CPX400DP and SPD3303X-E signal operation complete and report errors?

Only the CPX400DP library has an operation-complete command (*OPC?); the SPD3303X-E library has none. Only the SPD3303X-E library has an error-queue read (SYSTem:ERRor?); the CPX400DP library has none. Each command named here is listed in the tables on this page and on each instrument's command reference.