The Berkeley Nucleonics Model 645 function generator is controlled remotely with SCPI commands. TestFlow's instrument library holds 66 commands for it, 8 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the Model 645 on a real bench.
Commands
66
Queries
8
Subsystems
19
Models covered
1
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Berkeley Nucleonics Model 645 and writes the report from the measured results.
Control the Berkeley Nucleonics Model 645 from Python
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the Berkeley Nucleonics Model 645's VISA address
# Paste the address of your Berkeley Nucleonics Model 645 (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write("OUTPut ON") # Turn On Output
print(inst.query("OUTPut?")) # Read Output State
inst.close()
Run this on your bench
Describe the test; TestFlow's agent uses this command set to drive your Berkeley Nucleonics Model 645 and writes the report from the measured results.
Describe the test in a sentence, for example “turn on output on the Model 645, then read output state and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
66 commands in 19 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.
Output (OUTPut) (6)
Command
What it does
Arguments
OUTPut ON
Turn On Output. Enables the output.
None
OUTPut OFF
Turn Off Output. Disables the output.
None
OUTPut?
Read Output State. Reads whether the output is on.
None
OUTPut:LOAD <Ohm>
Set Output Load. Sets the output load in ohms; use INF for high-Z.
Number
OUTPut:POLarity NORMal
Set Output Polarity Normal. Sets output polarity to normal.
None
OUTPut:POLarity INVerted
Set Output Polarity Inverted. Sets output polarity to inverted.
None
Voltage (VOLTage) (6)
Command
What it does
Arguments
VOLTage <Amp>
Set Amplitude. Sets the output amplitude (Vpp).
Number
VOLTage?
Read Amplitude. Reads the output amplitude.
None
VOLTage:OFFSet <Volt>
Set Voltage Offset. Sets the DC offset voltage.
Number
VOLTage:HIGH <Volt>
Set Voltage High. Sets the high voltage level.
Number
VOLTage:LOW <Volt>
Set Voltage Low. Sets the low voltage level.
Number
VOLTage:UNIT <Unit>
Set Voltage Unit. Sets the amplitude unit.
Text: VPP,VRMS,DBM
Function (FUNCtion) (5)
Command
What it does
Arguments
FUNCtion <Shape>
Set Function. Sets the output waveform shape.
Text: SINusoid,SQUare,RAMP,PULSe,NOISe,DC,USER
FUNCtion?
Read Function. Reads the output waveform shape.
None
FUNCtion:RAMP:SYMMetry <Pct>
Set Ramp Symmetry. Sets the ramp symmetry percentage.
Number
FUNCtion:SQUare:DCYCle <Pct>
Set Square Duty Cycle. Sets the square-wave duty cycle.
Number
FUNCtion:USER <Name>
Select Arb Waveform. Selects the active arbitrary waveform.
Text
Burst (BURSt) (5)
Command
What it does
Arguments
BURSt:STATe ON
Enable Burst. Enables burst mode.
None
BURSt:STATe OFF
Disable Burst. Disables burst mode.
None
BURSt:MODE <Mode>
Set Burst Mode. Sets the burst mode.
Text: TRIGgered,GATed
BURSt:NCYCles <N>
Set Burst Cycles. Sets the number of cycles per burst.
Number
BURSt:INTernal:PERiod <Sec>
Set Burst Period. Sets the internal burst period.
Number
Frequency (FREQuency) (4)
Command
What it does
Arguments
FREQuency <Freq>
Set Frequency. Sets the output frequency in Hz.
Number
FREQuency?
Read Frequency. Reads the output frequency.
None
FREQuency:STARt <Freq>
Set Sweep Start Freq. Sets the sweep start frequency.
Number
FREQuency:STOP <Freq>
Set Sweep Stop Freq. Sets the sweep stop frequency.
Number
Apply (APPLy) (4)
Command
What it does
Arguments
APPLy:SINusoid <Freq>,<Amp>,<Offset>
Apply Sine. Sets sine with the given frequency, amplitude and offset.
Number; Number; Number
APPLy:SQUare <Freq>,<Amp>,<Offset>
Apply Square. Sets square with the given frequency, amplitude and offset.
Number; Number; Number
APPLy:RAMP <Freq>,<Amp>,<Offset>
Apply Ramp. Sets ramp with the given frequency, amplitude and offset.
Number; Number; Number
APPLy:PULSe <Freq>,<Amp>,<Offset>
Apply Pulse. Sets pulse with the given frequency, amplitude and offset.
Number; Number; Number
Sweep (SWEep) (4)
Command
What it does
Arguments
SWEep:STATe ON
Enable Sweep. Enables the frequency sweep.
None
SWEep:STATe OFF
Disable Sweep. Disables the frequency sweep.
None
SWEep:TIME <Sec>
Set Sweep Time. Sets the sweep time in seconds.
Number
SWEep:SPACing <Type>
Set Sweep Spacing. Sets the sweep spacing.
Text: LINear,LOGarithmic
PULS subsystem (3)
Command
What it does
Arguments
PULSe:WIDTh <Sec>
Set Pulse Width. Sets the pulse width in seconds.
Number
PULSe:PERiod <Sec>
Set Pulse Period. Sets the pulse period in seconds.
Number
PULSe:TRANsition <Sec>
Set Pulse Edge. Sets the pulse edge transition time.
Number
AM subsystem (3)
Command
What it does
Arguments
AM:STATe ON
Enable AM. Enables AM modulation.
None
AM:STATe OFF
Disable AM. Disables AM modulation.
None
AM:DEPTh <Pct>
Set AM Deviation. Sets the AM depth percentage.
Number
FM subsystem (3)
Command
What it does
Arguments
FM:STATe ON
Enable FM. Enables FM modulation.
None
FM:STATe OFF
Disable FM. Disables FM modulation.
None
FM:DEViation <Freq>
Set FM Deviation. Sets the FM frequency deviation.
Number
PM subsystem (3)
Command
What it does
Arguments
PM:STATe ON
Enable PM. Enables PM modulation.
None
PM:STATe OFF
Disable PM. Disables PM modulation.
None
PM:DEViation <Deg>
Set PM Deviation. Sets the PM phase deviation.
Number
FSK subsystem (3)
Command
What it does
Arguments
FSKey:STATe ON
Enable FSK. Enables FSK modulation.
None
FSKey:STATe OFF
Disable FSK. Disables FSK modulation.
None
FSKey:FREQuency <Freq>
Set FSK Hop Freq. Sets the FSK hop frequency.
Number
PWM subsystem (2)
Command
What it does
Arguments
PWM:STATe ON
Enable PWM. Enables PWM modulation.
None
PWM:STATe OFF
Disable PWM. Disables PWM modulation.
None
Trigger (TRIGger) (2)
Command
What it does
Arguments
TRIGger:SOURce <Src>
Set Trigger Source. Sets the trigger source.
Text: IMMediate,EXTernal,BUS
TRIGger
Trigger Now. Issues an immediate trigger.
None
System (SYSTem) (1)
Command
What it does
Arguments
SYSTem:ERRor?
Read Error. Reads and clears the next error-queue entry.
None
Other subsystem (1)
Command
What it does
Arguments
<Output Variable>=<Equation>
Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)
Number; Text
WAIT subsystem (1)
Command
What it does
Arguments
wait(<ms>)
Wait. Waits for number of milliseconds
Number
SAVE subsystem (1)
Command
What it does
Arguments
save2var <Variable_name>
Save2Variable. Saves last measurement to a variable
Text
IEEE 488.2 common commands (9)
Command
What it does
Arguments
*IDN?
Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).
None
*RST
Reset. Resets the instrument to its default state.
None
*CLS
Clear Status. Clears all event registers and the error queue.
None
*TST?
Self Test. Runs the self-test and returns 0 for pass.
None
*OPC?
Operation Complete. Returns 1 when all pending operations are complete.
None
*WAI
Wait To Continue. Waits until all pending operations are complete.
None
*TRG
Trigger Bus. Generates a bus trigger.
None
*SAV <Slot>
Save State Slot. Stores the current state to a memory slot.
Number
*RCL <Slot>
Recall State Slot. Recalls a state from a memory slot.
Number
Where these commands come from
This reference is generated from TestFlow's instrument library, which is drafted from Berkeley Nucleonics's programming documentation for the Model 645 and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Berkeley Nucleonics's programming manual: check it before you rely on an edge case. Last updated July 8, 2026.
Frequently asked questions
Does the Berkeley Nucleonics Model 645 support SCPI remote control?
Yes. The Berkeley Nucleonics Model 645 accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 66 SCPI commands for it: 8 queries that read values or state, and 58 settings and actions.
How do I control the Berkeley Nucleonics Model 645 from Python?
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, OUTPut ON and OUTPut?.
Which command reads a value from the Berkeley Nucleonics Model 645?
OUTPut? (read output state). Reads whether the output is on. The full list of queries is in the command reference on this page, grouped by subsystem.
Can I automate the Berkeley Nucleonics Model 645 without writing code?
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Berkeley Nucleonics Model 645 through the TestFlow desktop app, and writes the report from the measured results.
Your next test, automated today