The TestEquity 1007H, 1016H, 101H, and 123H temperature chamber is controlled remotely with SCPI commands. TestFlow's instrument library holds 55 commands for it, 26 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 1007H on a real bench.
Describe the test; TestFlow's agent uses this command set to drive your TestEquity 1007H, 1016H, 101H, or 123H and writes the report from the measured results.
With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find the TestEquity 1007H's VISA address
# Paste the address of your TestEquity 1007H (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000 # ms
print(inst.query("*IDN?")) # identify the instrument
inst.write(":OUTPut1:STATe ON") # Turn On Chamber
print(inst.query(":SOURce:CLOop1:SPOint?")) # Read Temperature Setpoint
inst.close()Describe the test; TestFlow's agent uses this command set to drive your TestEquity 1007H, 1016H, 101H, or 123H and writes the report from the measured results.
Describe the test in a sentence, for example “turn on chamber on the 1007H, then read temperature setpoint and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.
55 commands in 8 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values. The 1007H, 1016H, 101H, and 123H share this command set in TestFlow's library.
| Command | What it does | Arguments |
|---|---|---|
| :SOURce:CLOop1:SPOint <Setpoint> | Set Temperature Setpoint. Writes the temperature user setpoint (°C or °F per UNIT setting). | Number |
| :SOURce:CLOop1:SPOint? | Read Temperature Setpoint. Reads the active temperature setpoint (°C or °F per UNIT setting). | None |
| :SOURce:CLOop1:PVALue? | Read Temperature. Reads the temperature process value (°C or °F per UNIT setting). | None |
| :SOURce:CLOop1:IDLE <Setpoint> | Set Temperature Idle Setpoint. Writes the temperature idle setpoint used when no profile is running (°C or °F per UNIT setting). | Number |
| :SOURce:CLOop1:IDLE? | Read Temperature Idle Setpoint. Reads the temperature idle setpoint (°C or °F per UNIT setting). | None |
| :SOURce:CLOop1:ERRor? | Read Temperature Input Error. Reads the temperature loop input error status (returns ERROR or NONE). | None |
| :SOURce:CLOop1:RACTion <Action> | Set Temperature Ramp Action. Sets the temperature ramp action: OFF, STArtup (at power-up), SETPoint (on setpoint change) or BOTH. | Text: OFF,STArtup,SETPoint,BOTH |
| :SOURce:CLOop1:RSCAle <Scale> | Set Temperature Ramp Scale. Sets the temperature ramp rate time base (per minute or per hour). | Text: MINutes,HOURS |
| :SOURce:CLOop1:RRATe <Rate> | Set Temperature Ramp Rate. Writes the temperature ramp rate (°C or °F per UNIT setting per ramp-scale unit). | Number |
| :SOURce:CLOop1:RRATe? | Read Temperature Ramp Rate. Reads the temperature ramp rate. | None |
| :SOURce:CLOop1:RTIMe <Time> | Set Temperature Ramp Time. Writes the temperature ramp time (kept for backward compatibility; mirrors ramp rate). | Number |
| :SOURce:CLOop1:RTIMe? | Read Temperature Ramp Time. Reads the temperature ramp time. | None |
| :SOURce:CLOop2:SPOint <Setpoint> | Set Humidity Setpoint. Writes the humidity user setpoint (%RH). | Number |
| :SOURce:CLOop2:SPOint? | Read Humidity Setpoint. Reads the active humidity setpoint (%RH). | None |
| :SOURce:CLOop2:PVALue? | Read Humidity. Reads the humidity process value (%RH). | None |
| :SOURce:CLOop2:IDLE <Setpoint> | Set Humidity Idle Setpoint. Writes the humidity idle setpoint used when no profile is running (%RH). | Number |
| :SOURce:CLOop2:IDLE? | Read Humidity Idle Setpoint. Reads the humidity idle setpoint (%RH). | None |
| :SOURce:CLOop2:ERRor? | Read Humidity Input Error. Reads the humidity loop input error status (returns ERROR or NONE). | None |
| :SOURce:CLOop2:RACTion <Action> | Set Humidity Ramp Action. Sets the humidity ramp action: OFF, STArtup (at power-up), SETPoint (on setpoint change) or BOTH. | Text: OFF,STArtup,SETPoint,BOTH |
| :SOURce:CLOop2:RSCAle <Scale> | Set Humidity Ramp Scale. Sets the humidity ramp rate time base (per minute or per hour). | Text: MINutes,HOURS |
| :SOURce:CLOop2:RRATe <Rate> | Set Humidity Ramp Rate. Writes the humidity ramp rate (%RH per ramp-scale unit). | Number |
| :SOURce:CLOop2:RRATe? | Read Humidity Ramp Rate. Reads the humidity ramp rate. | None |
| :SOURce:CLOop2:RTIMe <Time> | Set Humidity Ramp Time. Writes the humidity ramp time (kept for backward compatibility; mirrors ramp rate). | Number |
| :SOURce:CLOop2:RTIMe? | Read Humidity Ramp Time. Reads the humidity ramp time. | None |
| Command | What it does | Arguments |
|---|---|---|
| :PROGram:SELected:NUMBer <Profile> | Select Profile. Selects the profile (program) number to run (1 to 40). | Number: 1,40 |
| :PROGram:SELected:NUMBer? | Read Profile Number. Reads the selected profile number. | None |
| :PROGram:SELected:NAME? | Read Profile Name. Reads the name of the selected profile. | None |
| :PROGram:SELected:STEP <Step> | Select Profile Step. Selects the profile step to start from (1 to 50). | Number: 1,50 |
| :PROGram:SELected:STATe STArt | Start Profile. Starts the selected profile. | None |
| :PROGram:SELected:STATe PAUSe | Pause Profile. Pauses the running profile. | None |
| :PROGram:SELected:STATe RESume | Resume Profile. Resumes a paused profile. | None |
| :PROGram:SELected:STATe STOP | Stop Profile. Stops the running profile. | None |
| :PROGram:SELected:STATe? | Read Profile State. Reads the profile run state. | None |
| :PROGram:SELected:SPOint1? | Read Profile Temp Setpoint. Reads the profile temperature setpoint (loop 1). | None |
| :PROGram:SELected:SPOint2? | Read Profile Humidity Setpoint. Reads the profile humidity setpoint (loop 2). | None |
| :PROGram:SELected:STEP? | Read Profile Step. Reads the current profile step number. | None |
| :PROGram:SELected:STEP:TYPE? | Read Step Type. Reads the current profile step type. | None |
| :PROGram:SELected:STEP:TIME:ELApsed? | Read Step Time Elapsed. Reads the elapsed time of the current profile step. | None |
| :PROGram:SELected:STEP:TIME:REMain? | Read Step Time Remaining. Reads the remaining time of the current profile step. | None |
| Command | What it does | Arguments |
|---|---|---|
| :OUTPut1:STATe ON | Turn On Chamber. Turns on Event 1 (chamber power / conditioning on OEM chambers). | None |
| :OUTPut1:STATe OFF | Turn Off Chamber. Turns off Event 1 (chamber power / conditioning on OEM chambers). | None |
| :OUTPut2:STATe ON | Turn On Humidity System. Turns on Event 2 (humidity system on OEM chambers). | None |
| :OUTPut2:STATe OFF | Turn Off Humidity System. Turns off Event 2 (humidity system on OEM chambers). | None |
| :OUTPut<Output>:STATe ON | Turn On Event Output. Turns on the selected event output (1 to 8). | Number: 1-7 |
| :OUTPut<Output>:STATe OFF | Turn Off Event Output. Turns off the selected event output (1 to 8). | Number: 1-7 |
| :OUTPut<Output>:STATe? | Read Event Output State. Reads the selected event output state (ON or OFF). | Number: 1-7 |
| :OUTPut<Output>:NAME? | Read Event Output Name. Reads the name of the selected event output. | Number: 1-7 |
| Command | What it does | Arguments |
|---|---|---|
| :UNIT:TEMPerature <Unit> | Set Comm Temp Units. Sets the temperature units used for communications (C or F). | Text: C,F |
| :UNIT:TEMPerature? | Read Comm Temp Units. Reads the temperature units used for communications. | None |
| :UNIT:TEMPerature:DISPlay <Unit> | Set Display Temp Units. Sets the temperature units shown on the front panel (C or F). | Text: C,F |
| :UNIT:TEMPerature:DISPlay? | Read Display Temp Units. Reads the temperature units shown on the front panel. | None |
| Command | What it does | Arguments |
|---|---|---|
| <Output Variable>=<Equation> | Math. Variable equals an equation like (${Var_name}= ${Var_name}/2) | Number; Text |
| Command | What it does | Arguments |
|---|---|---|
| wait(<ms>) | Wait. Waits for number of milliseconds | Number |
| Command | What it does | Arguments |
|---|---|---|
| save2var <Variable_name> | Save2Variable. Saves last measurement to a variable | Text |
| Command | What it does | Arguments |
|---|---|---|
| *IDN? | Identify. Returns the F4T controller identification (Watlow Electric, model, serial, firmware). | None |
This reference is generated from TestFlow's instrument library, which is drafted from TestEquity's programming documentation for the 1007H, 1016H, 101H, and 123H and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by TestEquity's programming manual: check it before you rely on an edge case. Last updated November 17, 2026.
Yes. The TestEquity 1007H accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 55 SCPI commands for it: 26 queries that read values or state, and 29 settings and actions.
Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, :OUTPut1:STATe ON and :SOURce:CLOop1:SPOint?.
:SOURce:CLOop1:SPOint? (read temperature setpoint). Reads the active temperature setpoint (°C or °F per UNIT setting). The full list of queries is in the command reference on this page, grouped by subsystem.
Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the TestEquity 1007H through the TestFlow desktop app, and writes the report from the measured results.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.