Phase Noise Analyzer · SCPI command reference

Microchip 53100A SCPI commands

The Microchip 53100A phase noise analyzer is controlled remotely with SCPI commands. TestFlow's instrument library holds 56 commands for it, 26 of them queries that read values or state, grouped below by subsystem with their arguments. It is the same library TestFlow's AI agent uses to drive the 53100A on a real bench.

Commands
56
Queries
26
Subsystems
22
Models covered
1

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Microchip 53100A and writes the report from the measured results.

Get started

Control the Microchip 53100A from Python

With PyVISA, any VISA backend, and the instrument on USB, LAN or GPIB, this opens a session, identifies the instrument, sends one setting, and reads one value. Every command in it comes from the reference below. For a full walkthrough, see the PyVISA tutorial and the SCPI command cheat sheet.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find the Microchip 53100A's VISA address

# Paste the address of your Microchip 53100A (USB, LAN or GPIB).
inst = rm.open_resource("PASTE_VISA_ADDRESS_HERE")
inst.timeout = 5000  # ms

print(inst.query("*IDN?"))  # identify the instrument
inst.write("pause freq")  # Pause Frequency Difference
print(inst.query("show state"))  # Read State

inst.close()

Run this on your bench

Describe the test; TestFlow's agent uses this command set to drive your Microchip 53100A and writes the report from the measured results.

Get started

Run a 53100A test without writing code

Describe the test in a sentence, for example “pause frequency difference on the 53100A, then read state and log it against a limit”. TestFlow's agent builds the steps from this command library, runs them on your bench through the TestFlow desktop app, and writes the report from the measured values. See AI test automation for lab instruments for how that works and where it stops.

Get started

SCPI command reference

56 commands in 22 groups. A ? marks a query that returns a value. Placeholders in angle brackets take the listed values.

SHOW subsystem (21)

CommandWhat it doesArguments
show stateRead State. Returns the measurement status (Ready / Initializing / Collecting) and elapsed time.None
show messageRead Message. Returns the most recent error or warning message from the 53100A driver (empty = none).None
show inputsRead Inputs. Returns rounded frequency and amplitude readings for the REFERENCE and DUT input ports.None
show spectrumMeasure Spectrum. Returns the phase noise spectrum as offset frequency and dBc/Hz pairs followed by the noise floor estimate.None
show spursMeasure Spurs. Returns the list of identified coherent spurs (offset and amplitude in dBc).None
show amspectrumMeasure AM Spectrum. Returns the AM noise spectrum as offset frequency and dBc/Hz pairs.None
show amspursMeasure AM Spurs. Returns the list of identified coherent spurs in the AM noise spectrum.None
show ipnMeasure Integrated Phase Noise. Returns the integrated phase noise chart in radians (with and without spurs) for lower limits 0.01 Hz to 10 kHz.None
show adevMeasure ADEV. Returns the current Allan deviation chart for the current tau0.None
show phasediffMeasure Phase Difference. Returns the phase difference strip chart (default last 600 s).None
show freqdiffMeasure Frequency Difference. Returns the fractional frequency difference strip chart (default last 600 s).None
show fcounterMeasure Frequency Counter. Returns the reference frequency and frequency count chart at averaging times up to 1000 s.None
show phaserateRead Phase Rate. Returns the current phase data rate.None
show tau0Read Tau0. Returns the current tau0 value.None
show referencefreqRead Reference Freq. Returns the reference frequency used for counter and phase/frequency difference calculations.None
show inputfreqRead Input Freq. Returns the nominal DUT input frequency used for phase difference calculations.None
show titleRead Title. Returns the current title (default shows the instrument serial number).None
show dateRead Date. Returns the current date and time of the TSERVE host.None
show dateformatRead Date Format. Returns the date and time formats.None
show timeformatRead Time Format. Returns the date and time formats.None
show versionRead Version. Returns software and hardware version information.None

SET subsystem (9)

CommandWhat it doesArguments
set phaserate <Rate>Set Phase Rate. Sets the phase data rate in readings per second for the next measurement (1, 10, 100 or 1000; default 100).Number: 1,10,100,1000
set tau0 <Tau0>Set Tau0. Sets the tau0 period in s for show adev (0.001, 0.01, 0.1 or 1); recomputes phaserate.Number: 0.001,0.01,0.1,1
set referencefreq autoSet Reference Freq Auto. Estimates the REFERENCE input frequency automatically (rounded to 100 kHz) for the next measurement.None
set referencefreq <Frequency>Set Reference Freq. Sets the nominal reference frequency in MHz explicitly (takes effect at the next measurement).Number
set inputfreq autoSet Input Freq Auto. Estimates the nominal DUT input frequency automatically (rounded to 100 kHz).None
set inputfreq <Frequency>Set Input Freq. Sets the precise nominal DUT input frequency in MHz used for phase difference calculations.Number
set title <Title>Set Title. Sets the title returned by show title (enclose titles containing spaces in double quotes).Text
set dateformat <Format>Set Date Format. Sets the show date format (1 = dd mon yyyy; 2 = mm/dd/yyyy; 3 = dd/mm/yyyy).Number: 1,2,3
set timeformat <Format>Set Time Format. Sets the time format (24 = 24 hour; 12 = 12 hour).Number: 24,12

PAUSE subsystem (2)

CommandWhat it doesArguments
pause phasediffPause Phase Difference. Inhibits updates to the phase difference chart.None
pause freqPause Frequency Difference. Inhibits updates to the frequency difference chart.None

RESUME subsystem (2)

CommandWhat it doesArguments
resume phasediffResume Phase Difference. Resumes updates to the phase difference chart.None
resume freqResume Frequency Difference. Resumes updates to the frequency difference chart.None

REMOVELINEAR subsystem (2)

CommandWhat it doesArguments
removelinear onRemove Linear On. Turns removal of the measured linear trend from show phasediff data on.None
removelinear offRemove Linear Off. Turns removal of the measured linear trend from show phasediff data off.None

SAVE subsystem (2)

CommandWhat it doesArguments
save <FileName>Save TIM File. Saves the current phase record and phase noise trace data to the given .TIM file.Text
save2var <Variable_name>Save2Variable. Saves last measurement to a variableText

PROMPT subsystem (2)

CommandWhat it doesArguments
prompt onPrompt On. Turns the Telnet command prompt on.None
prompt offPrompt Off. Turns the Telnet command prompt off.None

HELP subsystem (2)

CommandWhat it doesArguments
helpHelp. Returns a list of commands with brief usage information.None
help <Command>Help Command. Returns the help text for the given command.Text

START subsystem (1)

CommandWhat it doesArguments
startStart Measurement. Initiates a new phase noise / stability measurement; streaming data is sent on the data port (default 1298).None

Run control (1)

CommandWhat it doesArguments
stopStop Measurement. Terminates the measurement in progress.None

MEASURELINEAR subsystem (1)

CommandWhat it doesArguments
measurelinearMeasure Linear Trend. Computes the slope and intercept of the current phase difference chart.None

OPEN subsystem (1)

CommandWhat it doesArguments
open <FileName>Open Stream File. Opens (overwrites) an ASCII file to which streaming phase difference or frequency readings are written.Text

CLOSE subsystem (1)

CommandWhat it doesArguments
closeClose Stream File. Closes the stream output data file most recently opened.None

SELFTEST subsystem (1)

CommandWhat it doesArguments
selftestSelf Test. Runs the 53100A built-in self test (only when no measurement is running) and returns a 32-bit hex fault code (0x00000000 = no faults).None

HISTORY subsystem (1)

CommandWhat it doesArguments
historyCommand History. Returns a list of previously entered commands.None

BEEP subsystem (1)

CommandWhat it doesArguments
beepBeep. Plays a sound on the TSERVE host PC.None

RESET subsystem (1)

CommandWhat it doesArguments
resetReset. Stops any measurement and restores the server default properties (same as resettofactorydefaults).None

RESETTOFACTORYDEFAULTS subsystem (1)

CommandWhat it doesArguments
resettofactorydefaultsReset To Factory Defaults. Stops any measurement and restores the server default properties (same as reset).None

SHUTDOWN subsystem (1)

CommandWhat it doesArguments
shutdownShutdown Server. Shuts down the TSERVE server process and disconnects all clients.None

QUIT subsystem (1)

CommandWhat it doesArguments
quitQuit. Closes the Telnet connection; TSERVE and any measurement keep running (exit and logout are aliases).None

Other subsystem (1)

CommandWhat it doesArguments
<Output Variable>=<Equation>Math. Variable equals an equation like (${Var_name}= ${Var_name}/2)Number; Text

WAIT subsystem (1)

CommandWhat it doesArguments
wait(<ms>)Wait. Waits for number of millisecondsNumber

Where these commands come from

This reference is generated from TestFlow's instrument library, which is drafted from Microchip's programming documentation for the 53100A and checked when the agent runs it. Ranges, timing, and option-dependent commands are defined by Microchip's programming manual: check it before you rely on an edge case. Last updated September 26, 2026.

Frequently asked questions

Does the Microchip 53100A support SCPI remote control?

Yes. The Microchip 53100A accepts SCPI commands over its remote interfaces. TestFlow's instrument library holds 56 SCPI commands for it: 26 queries that read values or state, and 30 settings and actions.

How do I control the Microchip 53100A from Python?

Install PyVISA and a VISA backend, find the instrument's address with ResourceManager().list_resources(), open it, then send commands with write() and read values with query(). The example on this page sends *IDN?, pause freq and show state.

Which command reads a value from the Microchip 53100A?

show state (read state). Returns the measurement status (Ready / Initializing / Collecting) and elapsed time. The full list of queries is in the command reference on this page, grouped by subsystem.

Can I automate the Microchip 53100A without writing code?

Yes. TestFlow is an AI agent for electronics testing: you describe the test, and it builds the automation from this command library, runs it on the Microchip 53100A through the TestFlow desktop app, and writes the report from the measured results.

Your next test, automated today

Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.