In TestFlow's instrument library, the Rigol DG1022Z takes 201 SCPI commands, 49 of them queries, and the Tektronix AFG1022 takes 67 SCPI commands, 10 of them queries. That is 134 more on the DG1022Z. Counting distinct command headers (the command path without its arguments), 38 are accepted by both, 142 exist only on the DG1022Z, and 18 only on the AFG1022. 21% of the DG1022Z's command headers also exist on the AFG1022.
Both libraries have commands under SOURce, OUTPut, and SYSTem. Only the DG1022Z has COUNter, DISPlay, COUPling, MMEMory, ROSCillator, and HCOPy and 1 more.
Only the DG1022Z library has an operation-complete command (*OPC?); the AFG1022 library has none. Both libraries read errors with SYSTem:ERRor?.
TestFlow's library gives the DG1032Z the same command set as the DG1022Z, so the DG1022Z column applies to it too. TestFlow's library gives the AFG1062 the same command set as the AFG1022, so the AFG1022 column applies to it too.
Describe the test; TestFlow's agent uses these command sets to drive your Rigol DG1022Z or Tektronix AFG1022 and writes the report from the measured results.
Counts come from TestFlow's instrument library, the command sets its AI agent uses to drive each instrument. A distinct command header is the command path without its arguments, so :OUTPut ON and :OUTPut OFF count once.
| Rigol DG1022Z | Tektronix AFG1022 | |
|---|---|---|
| Commands in TestFlow's library | 201 | 67 |
| Queries (read a value or state) | 49 | 10 |
| Settings and actions | 152 | 57 |
| Subsystems | 14 | 7 |
| Distinct command headers | 180 | 56 |
| Headers the other lacks | 142 | 18 |
| Command language | SCPI | SCPI |
| Operation complete | *OPC? | Not in library |
| Error queue | SYSTem:ERRor? | SYSTem:ERRor? |
| Models on this command set | DG1022Z, DG1032Z | AFG1022, AFG1062 |
38 command headers are in both libraries. Here are 12, taken across subsystems. Where the AFG1022 library writes the command differently, its form is shown under the DG1022Z's.
| Command | What it does |
|---|---|
| :SOURce1:AM:STATe ONAFG1022: SOURce<Channel>:AM:STATe ON | Enable AM. Enables amplitude modulation on channel 1. |
| :OUTPut1:STATe ONAFG1022: OUTPut<Channel>:STATe ON | Turn On Output. Enables the output of channel 1. |
| :SYSTem:ERRor? | Read Error. Reads and removes the next error from the error queue. |
| *SAV <Location>AFG1022: *SAV <Slot> | Save State. Stores the current instrument state or arbitrary waveform to an internal location. |
| :SOURce1:AM:SOURce <Source>AFG1022: SOURce<Channel>:AM:SOURce <Src> | Set AM Source. Sets the AM modulation source to internal or external. |
| :OUTPut1:STATe?AFG1022: OUTPut<Channel>:STATe? | Read Output State. Reads whether channel 1 output is ON or OFF. |
| *RCL <Location>AFG1022: *RCL <Slot> | Recall State. Recalls a state or arbitrary waveform from an internal location. |
| :SOURce1:AM:INTernal:FREQuency <Freq>AFG1022: SOURce<Channel>:AM:INTernal:FREQuency <Freq> | Set AM Mod Frequency. Sets the internal AM modulating frequency in Hz. |
| :OUTPut1:IMPedance <Ohm>AFG1022: OUTPut<Channel>:IMPedance <Ohm> | Set Output Impedance. Sets the output load impedance in ohms (1 to 10k) or INFinity for HighZ. |
| *IDN? | Identify. Returns the instrument identification string (manufacturer model serial firmware). |
| :SOURce1:AM:INTernal:FUNCtion <Name>AFG1022: SOURce<Channel>:AM:INTernal:FUNCtion <Shape> | Set AM Mod Waveform. Sets the internal AM modulating waveform. |
| :OUTPut1:IMPedance?AFG1022: OUTPut<Channel>:IMPedance? | Read Output Impedance. Reads the current output load impedance setting. |
142 command headers in the DG1022Z library have no match in the AFG1022 library. Below are 24 of them from the 6 largest subsystems; the full list is in the Rigol DG1022Z command reference.
| Command | What it does |
|---|---|
| :SOURce1:APPLy:SINusoid <Freq>,<Amp>,<Offset> | Apply Sine. Configures channel 1 to output a sine waveform with the given frequency amplitude and offset. |
| :SOURce1:APPLy:SQUare <Freq>,<Amp>,<Offset> | Apply Square. Configures channel 1 to output a square waveform with the given frequency amplitude and offset. |
| :SOURce1:APPLy:RAMP <Freq>,<Amp>,<Offset> | Apply Ramp. Configures channel 1 to output a ramp waveform with the given frequency amplitude and offset. |
| :SOURce1:APPLy:PULSe <Freq>,<Amp>,<Offset> | Apply Pulse. Configures channel 1 to output a pulse waveform with the given frequency amplitude and offset. |
| Command | What it does |
|---|---|
| :OUTPut1:POLarity NORMal | Set Output Polarity Normal. Sets the output polarity of channel 1 to normal. |
| :OUTPut1:POLarity? | Read Output Polarity. Reads the current output polarity of channel 1. |
| :OUTPut1:MODE <Mode> | Set Output Mode. Sets the output mode of channel 1 to normal or gated. |
| :OUTPut1:MODE? | Read Output Mode. Reads the current output mode of channel 1. |
| Command | What it does |
|---|---|
| :COUNter:STATe ON | Enable Counter. Enables the frequency counter function. |
| :COUNter:STATe? | Read Counter State. Reads the running status of the frequency counter. |
| :COUNter:MEASure? | Read Counter Measurement. Reads the frequency period duty cycle and pulse width measured by the counter. |
| :COUNter:COUPling <Mode> | Set Counter Coupling. Sets the input coupling of the frequency counter. |
| Command | What it does |
|---|---|
| :DISPlay:BRIGhtness <Level> | Set Display Brightness. Sets the screen brightness level (1 to 100). |
| :DISPlay:BRIGhtness? | Read Display Brightness. Reads the current screen brightness. |
| :DISPlay:MODE <Mode> | Set Display Mode. Sets the display mode to dual parameter dual graph or single channel. |
| :DISPlay:TEXT <Str> | Set Display Text. Displays a text string on the instrument screen. |
| Command | What it does |
|---|---|
| *OPC? | Operation Complete. Returns 1 when all pending operations are complete. |
| *OPC | Set Operation Complete Bit. Sets the operation complete bit after all previous commands are executed. |
| *ESE <Value> | Set Event Status Enable. Sets the standard event status enable register. |
| *ESR? | Read Event Status. Reads and clears the standard event status register. |
| Command | What it does |
|---|---|
| :MMEMory:STORe:STATe <File> | Store State To File. Stores the current instrument state to a file on external memory. |
| :MMEMory:LOAD:STATe <File> | Load State From File. Loads an instrument state file from external memory. |
| :MMEMory:CATalog? | List Directory. Returns the files and folders in the current external memory directory. |
| :MMEMory:CDIRectory <Path> | Set Directory. Sets the current working directory on external memory. |
18 command headers in the AFG1022 library have no match in the DG1022Z library. Below are 4 of them from the 1 largest subsystem; the full list is in the Tektronix AFG1022 command reference.
| Command | What it does |
|---|---|
| SOURce<Channel>:FUNCtion:SHAPe <Shape> | Set Function. Sets the output waveform shape for the selected channel. |
| SOURce<Channel>:FUNCtion:SHAPe? | Read Function. Reads the output waveform shape for the selected channel. |
| SOURce<Channel>:FREQuency:FIXed <Freq> | Set Frequency. Sets the output frequency in Hz for the selected channel. |
| SOURce<Channel>:FREQuency:FIXed? | Read Frequency. Reads the output frequency for the selected channel. |
With PyVISA and any VISA backend, this opens both instruments, identifies them, sends one setting, and reads one value from each. For the setting, both libraries hold :OUTPut1:STATe (turn on output), and the same line runs on each. For the reading, both libraries hold :OUTPut1:STATe? (read output state), and the same line runs on each. For the basics, see the PyVISA tutorial.
import pyvisa
rm = pyvisa.ResourceManager()
print(rm.list_resources()) # find both VISA addresses
a = rm.open_resource("PASTE_DG1022Z_ADDRESS") # Rigol DG1022Z
b = rm.open_resource("PASTE_AFG1022_ADDRESS") # Tektronix AFG1022
for inst in (a, b):
inst.timeout = 5000 # ms
print(inst.query("*IDN?"))
# The same setting on both: Turn On Output
a.write(":OUTPut1:STATe ON")
b.write("OUTPut1:STATe ON")
# The same reading on both: Read Output State
print(a.query(":OUTPut1:STATe?"))
print(b.query("OUTPut1:STATe?"))
a.close()
b.close()Describe the test; TestFlow's agent uses these command sets to drive your Rigol DG1022Z or Tektronix AFG1022 and writes the report from the measured results.
In TestFlow's library the DG1022Z has 201 SCPI commands and the AFG1022 has 67 SCPI commands. 38 command headers are common to both, 142 are only on the DG1022Z (mostly SOURce and OUTPut), and 18 are only on the AFG1022 (mostly SOURce).
Partly. 38 command headers are accepted by both, including :SOURce1:AM:STATe, :OUTPut1:STATe, and :SYSTem:ERRor?. 160 are not, for example :SOURce1:APPLy:SINusoid on the DG1022Z and SOURce<Channel>:FUNCtion:SHAPe on the AFG1022, so a script that uses those needs a separate branch for each instrument.
The DG1022Z: 201 commands for the DG1022Z against 67 for the AFG1022 in TestFlow's library. Of those, 49 on the DG1022Z and 10 on the AFG1022 are queries that read a value or state, and the library splits them into 14 and 7 subsystems.
Only the DG1022Z library has an operation-complete command (*OPC?); the AFG1022 library has none. Both libraries read errors with SYSTem:ERRor?. Each command named here is listed in the tables on this page and on each instrument's command reference.
Describe the test. TestFlow builds the workflow, runs it on your instruments, and writes the report. See it on a bench like yours.