Vector Signal Generator · SCPI command comparison

Anritsu MG3710A vs Keysight E8267D: SCPI commands and remote control compared

In TestFlow's instrument library, the Anritsu MG3710A takes 176 SCPI commands, 53 of them queries, and the Keysight E8267D takes 130 SCPI commands, 39 of them queries. That is 46 more on the MG3710A. Counting distinct command headers (the command path without its arguments), 63 are accepted by both, 81 exist only on the MG3710A, and 47 only on the E8267D. 44% of the MG3710A's command headers also exist on the E8267D.

Both libraries have commands under SOURce, OUTPut, MMEMory, and SYSTem. Only the MG3710A has UNIT.

Both libraries synchronise with *OPC?. Both libraries read errors with SYSTem:ERRor?.

TestFlow's library gives the MG3710E the same command set as the MG3710A, so the MG3710A column applies to it too.

MG3710A commands
176
E8267D commands
130
Shared headers
63
Unique headers
128

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Anritsu MG3710A or Keysight E8267D and writes the report from the measured results.

Get started

MG3710A vs E8267D at a glance

Counts come from TestFlow's instrument library, the command sets its AI agent uses to drive each instrument. A distinct command header is the command path without its arguments, so :OUTPut ON and :OUTPut OFF count once.

Anritsu MG3710AKeysight E8267D
Commands in TestFlow's library176130
Queries (read a value or state)5339
Settings and actions12391
Subsystems98
Distinct command headers144110
Headers the other lacks8147
Command languageSCPISCPI
Operation complete*OPC?*OPC?
Error queueSYSTem:ERRor?SYSTem:ERRor?
Models on this command setMG3710A, MG3710EE8267D

Commands both accept

63 command headers are in both libraries. Here are 12, taken across subsystems.

CommandWhat it does
:SOURce:FREQuency:MODE <Mode>Set Frequency Mode. Sets the frequency entry mode (fixed CW value or list sweep).
:OUTPut:STATe ONTurn On RF Output. Enables the RF output of the selected signal generator.
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).
:SOURce:FREQuency:MODE?Read Frequency Mode. Reads the current frequency entry mode.
:OUTPut:STATe?Read RF Output State. Reads whether the RF output is currently ON or OFF.
*RSTReset. Resets the instrument to its default state.
:SOURce:FREQuency:CENTer <Freq>Set Frequency Center. Sets the center frequency of the sweep range in Hz.
:OUTPut:MODulation:STATe ONTurn On Modulation. Enables modulation of the RF output signal.
*CLSClear Status. Clears all event registers and the error queue.
:SOURce:FREQuency:CENTer?Read Frequency Center. Reads the sweep center frequency in Hz.
:OUTPut:MODulation:STATe?Read Modulation State. Reads whether RF modulation is enabled.
*TST?Self Test. Runs the self-test and returns 0 for pass or 1 for failure.

Only on the Anritsu MG3710A

81 command headers in the MG3710A library have no match in the E8267D library. Below are 16 of them from the 5 largest subsystems; the full list is in the Anritsu MG3710A command reference.

SOURce (64 only on the MG3710A)

CommandWhat it does
:SOURce:FREQuency:CW <Freq>Set Frequency. Sets the CW/fixed RF output frequency in Hz (100 kHz to 6 GHz depending on installed options).
:SOURce:FREQuency:CW?Read Frequency. Reads the current CW/fixed RF output frequency in Hz.
:SOURce:FREQuency:OFFSet?Read Frequency Offset. Reads the current frequency offset in Hz.
:SOURce:FREQuency:MULTiplier:STATe ONTurn On Frequency Multiplier. Applies the configured frequency multiplier to the displayed frequency.

MMEMory (7 only on the MG3710A)

CommandWhat it does
:MMEMory:STORe:STATe <File>Save State. Stores the current instrument state to the named file.
:MMEMory:LOAD:STATe <File>Load State. Recalls an instrument state from the named file.
:MMEMory:LOAD:LIST <File>Load List. Loads a frequency/level list from the named file.
:MMEMory:LOAD:WFM:ALLLoad All Waveforms. Loads all waveform patterns from storage into waveform memory.

SYSTem (6 only on the MG3710A)

CommandWhat it does
:SYSTem:ERRor:NEXT?Read Error. Reads and removes the next entry from the error queue.
:SYSTem:ERRor:CODE:NEXT?Read Error Code. Reads the numeric code of the next error queue entry.
:SYSTem:BEEPer ONTurn On Beeper. Enables the system beeper.
:SYSTem:LANGuage <Mode>Set Language Mode. Selects the remote command language mode used by the instrument.

OUTPut (2 only on the MG3710A)

CommandWhat it does
:OUTPut:PROTection:COUNt?Read Output Protection Count. Reads how many times the reverse power protection has tripped.
:OUTPut:PROTection:ERRor?Read Output Protection Error. Reads the reverse power protection error status.

UNIT (2 only on the MG3710A)

CommandWhat it does
:UNIT:POWer <Unit>Set Power Unit. Sets the unit used for RF level entry and display.
:UNIT:POWer?Read Power Unit. Reads the unit currently used for RF level display.

Only on the Keysight E8267D

47 command headers in the E8267D library have no match in the MG3710A library. Below are 11 of them from the 5 largest subsystems; the full list is in the Keysight E8267D command reference.

SOURce (40 only on the E8267D)

CommandWhat it does
:SOURce:FREQuency:FIXed <Freq>Set Frequency. Sets the CW/fixed RF output frequency in Hz (100 kHz to 44 GHz depending on installed options).
:SOURce:FREQuency:FIXed?Read Frequency. Reads the current CW/fixed RF output frequency in Hz.
:SOURce:FREQuency:SPAN?Read Frequency Span. Reads the sweep frequency span in Hz.
:SOURce:FREQuency:MANual <Freq>Set Frequency Manual. Sets the frequency output when the sweep is in manual mode.

IEEE 488.2 common (*) (2 only on the E8267D)

CommandWhat it does
*SAV <Register>,<Sequence>Save State Slot. Saves the current instrument state to an internal register.
*RCL <Register>,<Sequence>Recall State Slot. Recalls the instrument state from an internal register.

MMEMory (2 only on the E8267D)

CommandWhat it does
:MMEMory:DELete <File>Delete File. Deletes the named file from instrument memory.
:MMEMory:CATalog?List Files. Lists the files in the current instrument directory.

SYSTem (2 only on the E8267D)

CommandWhat it does
:SYSTem:ERRor:NEXt?Read Error. Reads and removes the next entry from the error queue.
:SYSTem:PRESetSystem Preset. Returns the instrument to its preset state.

OUTPut (1 only on the E8267D)

CommandWhat it does
:OUTPut:PROTection:STATe ONTurn On Output Protection. Enables the reverse power protection on the RF output.

Control both from Python

With PyVISA and any VISA backend, this opens both instruments, identifies them, sends one setting, and reads one value from each. For the setting, both libraries hold :OUTPut:STATe (turn on rf output), and the same line runs on each. For the reading, both libraries hold :OUTPut:STATe? (read rf output state), and the same line runs on each. For the basics, see the PyVISA tutorial.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find both VISA addresses

a = rm.open_resource("PASTE_MG3710A_ADDRESS")  # Anritsu MG3710A
b = rm.open_resource("PASTE_E8267D_ADDRESS")  # Keysight E8267D
for inst in (a, b):
    inst.timeout = 5000  # ms
    print(inst.query("*IDN?"))

# The same setting on both: Turn On RF Output
a.write(":OUTPut:STATe ON")
b.write(":OUTPut:STATe ON")

# The same reading on both: Read RF Output State
print(a.query(":OUTPut:STATe?"))
print(b.query(":OUTPut:STATe?"))

a.close()
b.close()

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Anritsu MG3710A or Keysight E8267D and writes the report from the measured results.

Get started

Frequently asked questions

What is the difference between the Anritsu MG3710A and the Keysight E8267D for remote control?

In TestFlow's library the MG3710A has 176 SCPI commands and the E8267D has 130 SCPI commands. 63 command headers are common to both, 81 are only on the MG3710A (mostly SOURce and MMEMory), and 47 are only on the E8267D (mostly SOURce and IEEE 488.2 common (*)).

Can I use the same SCPI code for the Anritsu MG3710A and the Keysight E8267D?

Partly. 63 command headers are accepted by both, including :SOURce:FREQuency:MODE, :OUTPut:STATe, and :SOURce:FREQuency:MODE?. 128 are not, for example :SOURce:FREQuency:CW on the MG3710A and :SOURce:FREQuency:FIXed on the E8267D, so a script that uses those needs a separate branch for each instrument.

Which has more remote commands, the MG3710A or the E8267D?

The MG3710A: 176 commands for the MG3710A against 130 for the E8267D in TestFlow's library. Of those, 53 on the MG3710A and 39 on the E8267D are queries that read a value or state, and the library splits them into 9 and 8 subsystems.

How do the MG3710A and E8267D signal operation complete and report errors?

Both libraries synchronise with *OPC?. Both libraries read errors with SYSTem:ERRor?. Each command named here is listed in the tables on this page and on each instrument's command reference.