Vector Signal Generator · SCPI command comparison

Keysight N5172B vs Rohde & Schwarz SMBV100B: SCPI commands and remote control compared

In TestFlow's instrument library, the Keysight N5172B takes 193 SCPI commands, 64 of them queries, and the Rohde & Schwarz SMBV100B takes 124 SCPI commands, 53 of them queries. That is 69 more on the N5172B. Counting distinct command headers (the command path without its arguments), 42 are accepted by both, 126 exist only on the N5172B, and 68 only on the SMBV100B. 25% of the N5172B's command headers also exist on the SMBV100B.

Both libraries have commands under SOURce, OUTPut, and SYSTem. Only the N5172B has MMEMory. Only the SMBV100B has TRIGger.

Both libraries synchronise with *OPC?. Both libraries read errors with SYSTem:ERRor?.

TestFlow's library gives the N5182B the same command set as the N5172B, so the N5172B column applies to it too.

N5172B commands
193
SMBV100B commands
124
Shared headers
42
Unique headers
194

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Keysight N5172B or Rohde & Schwarz SMBV100B and writes the report from the measured results.

Get started

N5172B vs SMBV100B at a glance

Counts come from TestFlow's instrument library, the command sets its AI agent uses to drive each instrument. A distinct command header is the command path without its arguments, so :OUTPut ON and :OUTPut OFF count once.

Keysight N5172BRohde & Schwarz SMBV100B
Commands in TestFlow's library193124
Queries (read a value or state)6453
Settings and actions12971
Subsystems88
Distinct command headers168110
Headers the other lacks12668
Command languageSCPISCPI
Operation complete*OPC?*OPC?
Error queueSYSTem:ERRor?SYSTem:ERRor?
Models on this command setN5172B, N5182BSMBV100B

Commands both accept

42 command headers are in both libraries. Here are 12, taken across subsystems.

CommandWhat it does
:SOURce:FREQuency:MODE <Mode>Set Frequency Mode. Sets the frequency entry mode (fixed CW value or list/step sweep).
:OUTPut:STATe ONTurn On RF Output. Enables the RF output signal.
*IDN?Identify. Returns the instrument identification string (manufacturer, model, serial, firmware).
:SOURce:FREQuency:MODE?Read Frequency Mode. Reads the current frequency entry mode.
:OUTPut:STATe?Read RF Output State. Reads whether the RF output is currently ON or OFF.
*RSTReset. Resets the instrument to its default state.
:SOURce:FREQuency:CENTer <Freq>Set Frequency Center. Sets the center frequency of the frequency sweep range in Hz.
*CLSClear Status. Clears all event registers and the error queue.
:SOURce:FREQuency:CENTer?Read Frequency Center. Reads the sweep center frequency in Hz.
*TST?Self Test. Runs the self-test and returns 0 for pass or 1 for failure.
:SOURce:FREQuency:SPAN <Freq>Set Frequency Span. Sets the span of the frequency sweep range in Hz.
*OPC?Operation Complete. Returns 1 when all pending operations are complete.

Only on the Keysight N5172B

126 command headers in the N5172B library have no match in the SMBV100B library. Below are 17 of them from the 5 largest subsystems; the full list is in the Keysight N5172B command reference.

SOURce (112 only on the N5172B)

CommandWhat it does
:SOURce:FREQuency:FIXed <Freq>Set Frequency. Sets the CW/fixed RF output frequency in Hz (9 kHz to 6 GHz).
:SOURce:FREQuency:FIXed?Read Frequency. Reads the current CW/fixed RF output frequency in Hz.
:SOURce:POWer:LEVel:IMMediate:OFFSet <Offset>Set Power Offset. Sets a power offset in dB added to the displayed RF level to account for external gain or loss.
:SOURce:POWer:LEVel:IMMediate:OFFSet?Read Power Offset. Reads the current RF power offset in dB.

OUTPut (5 only on the N5172B)

CommandWhat it does
:OUTPut:MODulation:STATe ONTurn On Modulation. Enables the modulation of the RF output signal (front-panel MOD ON/OFF).
:OUTPut:MODulation:STATe?Read Modulation State. Reads whether RF modulation is enabled.
:OUTPut:BLANking:STATe ONTurn On Output Blanking. Enables RF output blanking so the RF is muted while the synthesizer settles.
:OUTPut:BLANking:STATe?Read Output Blanking State. Reads the RF output blanking state.

SYSTem (4 only on the N5172B)

CommandWhat it does
:SYSTem:ERRor:NEXt?Read Error. Reads and removes the next entry from the error queue.
:SYSTem:PRESetSystem Preset. Returns the instrument to its preset state.
:SYSTem:PRESet:TYPE <Type>Set Preset Type. Selects whether preset restores the factory (normal) or a user-defined state.
:SYSTem:CAPability?Read Capability. Returns the instrument capability string.

IEEE 488.2 common (*) (3 only on the N5172B)

CommandWhat it does
*SAV <Reg>,<Seq>Save State Slot. Saves the current instrument state to an internal register.
*RCL <Reg>,<Seq>Recall State Slot. Recalls the instrument state from an internal register.
*OPT?Read Options. Returns the list of options installed in the instrument.

MMEMory (2 only on the N5172B)

CommandWhat it does
:MMEMory:DELete <File>Delete File. Deletes the named file from instrument memory.
:MMEMory:CATalog?List Files. Lists the files in the current instrument directory.

Only on the Rohde & Schwarz SMBV100B

68 command headers in the SMBV100B library have no match in the N5172B library. Below are 12 of them from the 4 largest subsystems; the full list is in the Rohde & Schwarz SMBV100B command reference.

SOURce (60 only on the SMBV100B)

CommandWhat it does
SOURce:FREQuency:CW <Freq>Set Frequency. Sets the frequency of the RF output signal in CW/fixed mode in Hz (8 kHz to 6 GHz depending on installed options).
SOURce:FREQuency:CW?Read Frequency. Reads the current CW/fixed RF output frequency in Hz.
SOURce:ROSCillator:EXTernal:FREQuency <Freq>Set External Reference Frequency. Selects the expected frequency of the external reference input signal (10 MHz, 5 MHz, 13 MHz, or a variable/user-defined value).
SOURce:ROSCillator:EXTernal:FREQuency?Read External Reference Frequency. Reads the configured external reference input frequency setting.

OUTPut (4 only on the SMBV100B)

CommandWhat it does
OUTPut:STATe:PON <Pon>Set Power On Output State. Defines the state of the RF output signal when the instrument is switched on.
OUTPut:STATe:PON?Read Power On Output State. Reads the configured RF output state used at power-on.
OUTPut:AMODe FIXedSet Attenuator Mode Manual. Sets the RF output attenuator to manual mode so the attenuation value can be fixed.
OUTPut:AMODe?Read Attenuator Mode. Reads the current RF output attenuator mode.

SYSTem (2 only on the SMBV100B)

CommandWhat it does
SYSTem:ERRor:ALL?Read Error Queue. Queries and clears all unread entries from the error/event queue; returns 0,'No error' if the queue is empty.
SYSTem:ERRor:COUNt?Read Error Count. Reads the number of entries currently in the error/event queue.

TRIGger (2 only on the SMBV100B)

CommandWhat it does
TRIGger:FSWeep:SOURce <Source>Set Frequency Sweep Trigger Source. Selects the trigger source for the RF frequency sweep (free-running, single/bus, external, or external-auto).
TRIGger:FSWeep:SOURce?Read Frequency Sweep Trigger Source. Reads the current RF frequency sweep trigger source.

Control both from Python

With PyVISA and any VISA backend, this opens both instruments, identifies them, sends one setting, and reads one value from each. For the setting, both libraries hold :OUTPut:STATe (turn on rf output), and the same line runs on each. For the reading, both libraries hold :OUTPut:STATe? (read rf output state), and the same line runs on each. For the basics, see the PyVISA tutorial.

import pyvisa

rm = pyvisa.ResourceManager()
print(rm.list_resources())  # find both VISA addresses

a = rm.open_resource("PASTE_N5172B_ADDRESS")  # Keysight N5172B
b = rm.open_resource("PASTE_SMBV100B_ADDRESS")  # Rohde & Schwarz SMBV100B
for inst in (a, b):
    inst.timeout = 5000  # ms
    print(inst.query("*IDN?"))

# The same setting on both: Turn On RF Output
a.write(":OUTPut:STATe ON")
b.write("OUTPut:STATe ON")

# The same reading on both: Read RF Output State
print(a.query(":OUTPut:STATe?"))
print(b.query("OUTPut:STATe?"))

a.close()
b.close()

Run this on your bench

Describe the test; TestFlow's agent uses these command sets to drive your Keysight N5172B or Rohde & Schwarz SMBV100B and writes the report from the measured results.

Get started

Frequently asked questions

What is the difference between the Keysight N5172B and the Rohde & Schwarz SMBV100B for remote control?

In TestFlow's library the N5172B has 193 SCPI commands and the SMBV100B has 124 SCPI commands. 42 command headers are common to both, 126 are only on the N5172B (mostly SOURce and OUTPut), and 68 are only on the SMBV100B (mostly SOURce and OUTPut).

Can I use the same SCPI code for the Keysight N5172B and the Rohde & Schwarz SMBV100B?

Partly. 42 command headers are accepted by both, including :SOURce:FREQuency:MODE, :OUTPut:STATe, and :SOURce:FREQuency:MODE?. 194 are not, for example :SOURce:FREQuency:FIXed on the N5172B and SOURce:FREQuency:CW on the SMBV100B, so a script that uses those needs a separate branch for each instrument.

Which has more remote commands, the N5172B or the SMBV100B?

The N5172B: 193 commands for the N5172B against 124 for the SMBV100B in TestFlow's library. Of those, 64 on the N5172B and 53 on the SMBV100B are queries that read a value or state, and the library splits them into 8 and 8 subsystems.

How do the N5172B and SMBV100B signal operation complete and report errors?

Both libraries synchronise with *OPC?. Both libraries read errors with SYSTem:ERRor?. Each command named here is listed in the tables on this page and on each instrument's command reference.